microlife IR310 Ear Thermometer User Guide

June 2, 2024
microlife

IR310 Ear Thermometer
User Guide

IR310 Ear Thermometer

Microlife IR 310

Guarantee Card
Name of Purchaser ……………..
Serial Number ……………..
Date of Purchase ……………..
Specialist Dealer ……………..

Microlife Ear Thermometer IR 310

1. Measuring sensor
2. Display
3. M-button (memory)
4. Probe cover ejector
5. Battery compartment cover
6. START/IO button
7. All segments displayed
8. Memory
9. Ready for measuring
10. Correct position indication
11. Measurement complete
12. Incorrect location indication
13. Probe cover not detected
14. Error function display
15. Low battery indicator| 16. Changing between Celsius and Fahrenheit
17. Recall mode
18. Recall the last 30 readings
19. Measured temperature too high
20. Measured temperature too low
21. Ambient temperature too high
22. Ambient temperature too low
23. Blank display
24. Flat battery
25. Date/Time
26. Beeper function setting
27. Replacing the battery
28. Storage holder
29. Probe cover
30. How to reload a new probe cover
---|---

Read the instructions carefully before using this device.
Type BF applied part

This Microlife thermometer is a high quality product incorporating the latest technology and tested in accordance with international standards. With its unique technology, this device can provide a stable, heat-interference-free reading with each measurement. The device performs a self-test every time it is switched on to always guarantee the specified accuracy of any measurement. The Microlife Ear Thermometer is intended for the periodic measurement and monitoring of human body temperature. It is intended for use on people of all ages.
This thermometer has been clinically tested and proven to be safe and accurate when used in accordance to the operating instruction manual.
Please read through these instructions carefully in order for you to understand all functions and safety information.

The Advantages of this Thermometer

Measurement in 1 second
The innovative infrared technology allows measurement of ear temperature in only 1 second.
Correct position indication
ACCUsens guidance system confirms the right position in the ear with «good» displayed on the LCD display and a beep.
Probe cover
This thermometer is user-friendly and more hygienic with the usage of a probe cover.
Probe LED
This thermometer includes a probe LED light which enables the user to find the correct ear position in the dark.
Accurate and reliable
The unique probe assembly construction incorporates an advanced infrared sensor, ensuring that each measurement is accurate and reliable.
Gentle and easy to use

  • The ergonomic design enables simple and easy use of the thermometer.
  • This thermometer can even be used on a sleeping child without causing any interruption.
  • This thermometer is quick, therefore child-friendly.

Multiple readings recall
Users will be able to recall the last 30 readings with a record of both time and date when entering the recall mode, enabling efficient tracking of temperature variations.
Safe and hygienic

  • No risk of broken glass or mercury ingestion.
  • Completely safe for use on children.
  • Using a new probe cover each time, ensures this thermometer is completely hygienic for use by the whole family.

Fever alarm
10 short beeps and a red LCD backlight alert the patient that he/she may have a temperature equal to or higher than 37.5 °C.

Important Safety Instructions

  • Follow instructions for use. This document provides important product operation and safety information regarding this device. Please read this document thoroughly before using the device and keep for future reference.

  • This device may only be used for the purposes described in these instructions. The manufacturer cannot be held liable for damage caused by incorrect application.

  • Never immerse this device in water or other liquids. For cleaning please follow the instructions in the « Cleaning and disinfecting» section.

  • Do not use this device if you think it is damaged or notice anything unusual.

  • Never open this device.

  • Earwax in the ear canal may cause a lower temperature reading. Therefore it is important to ensure the subject’s ear canal is clean.

  • Only use this thermometer with a new Microlife branded and undamaged probe cover to prevent cross-infection.

  • If the measurement result is not consistent with the patient’s finding or unusually low, repeat the measurement every 15 minutes or double check the result by another core body temperature measurement.

  • This device comprises sensitive components and must be treated with caution. Observe the storage and operating conditions described in the «Technical Specifications»
    section.

  • Ensure that children do not use this device unsupervised; some parts are small enough to be swallowed.

  • Do not use this device close to strong electromagnetic fields such as mobile telephones or radio installations. Keep a minimum distance of 3.3 m from such devices when using this device.

  • Protect it from:
    • extreme temperatures
    • impact and dropping
    • contamination and dust
    • direct sunlight
    • heat and cold

  • If the device is not going to be used for a prolonged period the battery should be removed.

WARNING: The measurement results given by this device is not a diagnosis. It is not replacing the need for the consultation of a physician, especially if not matching the patient’s symptoms. Do not rely on the  measurement result only, always consider other potentially occurring symptoms and the patient’s feedback. Calling a doctor or an ambulance is advised if needed.

How this Thermometer measures Temperature

This thermometer measures infrared energy radiated from the eardrum and the surrounding tissue. This energy is collected through the lens and converted to a temperature  value. The measured reading obtained directly from the eardrum (Tympanic Membrane) ensures the most accurate ear temperature. Measurements taken from the surrounding tissue of the ear canal generate lower readings and may result in misdiagnosis of fever.
To avoid an inaccurate measurement

  1. Fit a new Microlife branded and undamaged probe cover 29 on the measuring sensor 1.
  2. Switch on the thermometer by pressing the START/IO button 6.
  3. After one beep is heard (and the temperature scale icon is flashing), straighten the ear canal by gently pulling the middle of the ear back and up.
  4. Place the probe 1 firmly into the ear canal. «Good» will be displayed with a short beep to confirm the device has detected the correct position. Press the START/IO  utton 6 and keep the probe in the ear until the thermometer beeps to signal the completion of the measurement.

Control Displays and Symbols

  • All segments displayed 7: Press the START/IO button 6 to turn on the unit; all segments will be shown for 1 second.
  • Ready for measurement 9 : When the unit is ready for measurement, the «°C» or «°F» icon will keep flashing.
  • The probe LED light is activated and will keep flashing.
  • Correct position indication 10 : The probe LED light will stop flashing (stays illuminated) and «good» will be displayed on the LCD, when the measuring sensor detects an appropriate position.
  • Measurement complete 11: The reading will be shown on the display 2 with the «°C» or «°F» icon; the unit is ready for the next measurement, when the «°C» or «°F» icon is flashing again.
  • Low battery indicator 15: When the unit is turned on, the icon will keep flashing to remind the user to replace the battery.

Setting Date, Time and Beeper Functions

Setting the date and time

  1. After the new battery is fitted, the year number flashes in the display 25. You can set the year by pressing the M-button 3.To confirm and then set the month, press the START/IO button 6.
  2. Press the M-button 3 to set the month. Press the START/IO 6 to confirm and then set the day.
  3. Follow the previously mentioned instructions to set the day, hours and minutes.
  4. Once you have set the minutes, wait for 10 seconds until the device automatically switches to ready for measurement 9, the date and time are set and the time is displayed.
    Change current date and time: Press and hold the Mbutton 3 for approx. 10 seconds until the year number starts to flash 25. Now you can enter the new values as described above.

Setting the beeper

  1. Press and hold the START/IO button 6 for 5 seconds to set the beeper 26.
  2. Press the M-button 3 to either turn the beeper on or off. The beeper is activated when the beeper icon 26 is shown without a cross.
    When the beeper setting has been chosen, press the START/IO button 6 to enter the «ready for measuring» mode; otherwise the device automatically switches to ready for measuring after 10 seconds 9.

Directions for Use

Important: Before each measurement, fit a new undamaged probe cover 29 on the measuring sensor 1. Failure to do so will result in incorrect temperature measurements. How to correctly fit a new probe cover 29 is pictured at the beginning of this instruction 30.

  1. Press the START/IO button 6. The display 2 is activated to show all segments for 1 second.

  2. When the «°C» or «°F» icon is flashing, a beep sound is heard and the thermometer is ready for measuring 9.

  3. The probe LED light is activated and will keep flashing.

  4. Straighten the ear canal by pulling the ear up and back to give a clear view of the eardrum.
    • For children under 1 year; pull the ear straight back.
    • For children of 1 year to adult; pull the ear up and back.
    Also refer to the short instructions at the front.

  5. While gently pulling the ear, insert the probe snugly into the ear canal.

  6. The probe LED light will stop flashing (stays illuminated) and «good» will be displayed on the LCD, when the measuring sensor detects an appropriate position.

  7. Immediately press the START/IO button 6. Release the button and wait for the beep sound. This indicates the end of the measurement.

  8. Remove the thermometer from the ear canal. The display shows the measured temperature 11.

  9. Press and release the cover ejector 4 to remove the used probe cover before starting a new measurement. Fit a new undamaged probe 29 cover on the measuring sensor 1.

  10. For the next measurement wait until the «°C»/«°F» icon is flashing and follow steps 5 – 7 above.

  11. Press and hold the START/IO button 6 for 3 seconds to turnoff the device; otherwise the device will automatically switch ofafter approx. 60 seconds.

NOTE:

  • Patients and thermometer should stay in similar room condition for at least 30 minutes.

  • To ensure accurate readings, wait at least 30 sec. after 3-5continuous measurements.

  • It is essential that a new undamaged probe cover 29 is used for each measurement. Therefore, this device remindsthe user to take off the used probe cover when turning off the device. The «probe cover» icon 13 is displayed and the probeLED light will flash for 3 seconds. For cleaning, follow the instructions in the «Cleaning and disinfecting» section.

  • The device detects if there is no probe cover placed on the measuring sensor when a measurement is performed; a probecover icon 13 and a red LCD backlight will alert the patient accordingly.

  • After cleaning the measuring sensor 1with alcohol, wait 5minutes before taking the next measurement, in order to allow the thermometer to reach its operating reference
    temperature.

  • 10 short beeps and a red LCD backlight alert the patient that heshe may have a temperature equal to or higher than 37.5 °C.

  • For an infant, it is best to have the child lying flat with his/her head sideways so the ear is facing upwards. For an older childor adult, it is best to stand behind and slightly to the side of thepatient.

  • Always take the temperature in the same ear, since the temperature readings may be different from ear to ear.

  • In the following situations it is recommended that three temperatures in the same ear be taken and the highest one taken as the reading:
    – New born infants in the first 100 days.
    – Children under three years of age with a compromised immune system and for whom the presence or absence of fever is critical.
    – When the user is learning how to use the thermometer for the first time until he/she has familiarized himself/herself with the device and obtains consistent readings.
    – If the measurement is surprisingly low.

  • Don’t take a measurement while or immediately after nursing a baby.

  • Don’t use the thermometer in high humidity environments.

  • Patients should not drink, eat or exercise before/while taking the measurement.

  • Doctors recommend rectal measurement for newborn infants within the first 6 months, as all other measuring methods might lead to ambiguous results.

  • Readings from different measuring sites should not be compared as the normal body temperature varies by measuring site and time of day, being highest in the evening and lowest about one hour before waking up.

  • Normal body temperature ranges:
    – Axillar: 34.7 – 37.3 °C / 94.5 – 99.1 °F
    – Oral: 35.5 – 37.5 °C / 95.9 – 99.5 °F
    – Rectal: 36.6 – 38.0 °C / 97.9 – 100.4 °F
    – Microlife IR 310: 35.4 – 37.4 °C / 95.7 – 99.3 °F

Changing between Celsius and Fahrenheit

This thermometer can display temperature readings in either Fahrenheit or Celsius 16.

  • To switch the display between °C and °F, press and hold the M- button 3 for 5 seconds when device is power off; «°C»/«°F» icon is shown on the display.
  • Press the M-button 3 to change the measurement scale between °C and °F .
  • When the measurement scale has been chosen, press the START/IO button 6 to enter the «ready for measuring» mode; otherwise the device automatically switches to ready for measuring after 5 seconds 9.

How to recall 30 readings in Memory Mode

This thermometer can recall the last 30 readings with a record of both time and date.

  • Recall mode 30: Press the M-button 3 to enter recall mode when the power is off. The memory icon «M» will show.
  • Reading 1 – the last reading 18: Press and release the Mbutton 3 to recall the last reading. Number «1» and a flashing «M» are displayed.

Pressing and releasing the M-button 3 after the last 30 readings have been recalled will resume the above sequence from reading 1.

Error Messages

  • Measured temperature too high 19: Displays «H» when measured temperature is higher than 43 °C / 109.4 °F.
  • Measured temperature too low 20: Displays «L» when measured temperature is lower than 32 °C / 89.6 °F.
  •   Ambient temperature too high 21: Displays «AH» when ambient temperature is higher than 40.0 °C / 104.0 °F.
  • Ambient temperature too low 22: Display «AL» when ambient temperature is lower than 10.0 °C / 50.0 °F.
  • Incorrect location indication 12: Display «ERP» when the probe is not correctly inserted in the ear canal. Please insert the probe as described in this manual.
  • No probe cover on the measuring sensor 13: Displays probe cover icon 13 with a red LCD backlight to remind the user to put on a new undamaged probe cover before each measurement.
  • Error function display 14: The system has a malfunction.
  • Blank display 23: Check if the battery has been loaded correctly. Also check polarity (<+> and <->) of the battery.
  • Flat battery indication 24: If only the icon is shown on the display, the battery should be replaced immediately.

Cleaning and disinfecting

Use an alcohol swab or cotton tissue moistened with alcohol (70% Isopropyl) to clean the thermometer casing and the measuring sensor. Ensure that no liquid enters the interior of the device. Never use abrasive cleaning agents, thinners or benzene for cleaning and never immerse the device in water or other cleaning liquids. Take care not to scratch the surface of the sensor lens and the display.

Battery Replacement

This device is supplied with 2 new, long-life 1.5V, size AAA batteries. Batteries need replacing when this icon 24 is the only symbol shown on the display.
Remove the battery cover 27 by sliding it in the direction shown.
Replace the batteries – ensure correct polarity as shown by the symbols in the compartment.
Batteries and electronic devices must be disposed of in accordance with the locally applicable regulations, not with domestic waste.

Guarantee

This device is covered by a 5 year guarantee from the date of purchase. During this guarantee period, at our discretion, Microlife will repair or replace the defective product free of charge.
Opening or altering the device invalidates the guarantee.

The following items are excluded from the guarantee:

  • Transport costs and risks of transport.
  • Damage caused by incorrect application or non-compliance with the instructions for use.
  • Damage caused by leaking batteries.
  • Damage caused by accident or misuse.
  • Packaging/storage material and instructions for use.
  • Regular checks and maintenance (calibration).
  • Accessories and wearing parts: Battery.

Should guarantee service be required, please contact the dealer from where the product was purchased, or your local Microlife service. You may contact your local Microlife service through our website: www.microlife.com/support
Compensation is limited to the value of the product. The guarantee will be granted if the complete product is returned with the original invoice. Repair or replacement within guarantee does not prolong or renew the guarantee period. The legal claims and rights of consumers are not limited by this guarantee.

Technical Specifications

Type: Adjusted mode Ear Thermometer IR 310
Mode type: Adjust mode
Measuring site: Ear
Reference body site: Sublingual
Measurement range: Body mode: 32.0 – 43.0 °C / 89.6 – 109.4 °F
Resolution: 0.1 °C / °F
Measurement accu Body mode:
racy (Laboratory): ±0.2 °C, 35.0 ~ 42.0 °C / ±0.4 °F, 95.0 ~ 107.6 °F

±0.3 °C, 32.0 ~ 34.9 °C and 42.1 ~ 43.0 °C /
±0.5 °F, 89.6 ~ 94.8 °F and 107.8 ~ 109.4 °F
Clinical results:| Repeatability: 0.19 °C
Bias: 0.03 °C
Limits of agreement: 1.33 °C
Display:| Liquid Crystal Display, 4 digits plus special icons
Acoustic:| The unit is turned ON and ready for the measurement: 1 short beep.
Complete the measurement: 1 long beep.
System error or malfunction: 3 short beeps.
Fever alarm: 10 short beeps.
Memory:| 30 readings recall in the memory mode with a record of both time and date.
Backlight:| The display light will be GREEN for 1 second, when the unit is turned ON.
The display light will be GREEN for 5 seconds, when a measurement is completed with a reading less than 37.5 °C / 99.5 °F.
The display light will be RED for 5 seconds, when a measurement is completed with a reading equal to or higher than 37.5 °C / 99.5 °F.
Operating condi:| 10 – 40 °C / 50.0 – 104 °F
tions:| 15 – 95 % relative maximum humidity
Storage conditions:| -25 – +55 °C / -13 – +131 °F
15 – 95 % relative maximum humidity
Automatic Switch:| Approx. 1 minute after last measurement
off:| has been taken.
Battery:| 2 x 1.5 V alkaline batteries; size AAA
Dimensions:| approx. 800 measurements (using new batteries)
Battery lifetime:| 151 x 46 x 60.5 mm
Weight:| 60 g (with battery), 58 g (w/o battery)
IP Class:| IP22
Reference to| ISO 80601-2-56; IEC 60601-1; IEC
standards:| 60601-1-2 (EMC); IEC 60601-1-11
Expected service life:| 5 years or 12000 measurements

This device complies with the requirements of the Medical Device Directive 93/42/EEC.
Technical alterations reserved.
According to the Medical Product User Act a biennial technical inspection is recommended for professional users. Please observe the applicable disposal regulations.
14.www.microlife.com
Detailed user information about our thermometers and blood pressure monitors as well as services can be found at www.microlife.com.

Microlife AG
Espenstrasse 139
9443 Widnau
Switzerland
www.microlife.com
Microlife UAB
P. Lukšio g. 32
08222 Vilnius
Lithuania
IB IR 310 N-V10 4822
Revision Date: 2022-11-09