Pyxis ST-730 Series Inline Turbidity Sensors User Guide
- June 4, 2024
- Pyxis
Table of Contents
ST-730 Series Inline Turbidity Sensors
**User Guide
**
Additional Accessories
- Pyxis Lab® Sensor Cleaning Kit
- MA-WB Bluetooth® Adapter
- Turbidity Calibration Solutions
- ST-001 Inline Sensor Tee
User Manual Download
http://qr.w69b.com/g/pZoMTx4NG
Purchase online at: www.pyxis-lab.com/shop
Wiring Guide & uPyxis® Download
You will need to download the uPyxis® app & connect your MA-WB Bluetooth® Adapter if you want to wirelessly monitor or calibrate your ST-730 Series sensor.
|
---|---
http://qr.w69b.com/g/oVf1QrkSk| http://qr.w69b.com/g/thD3nKWSA
uPyxis® Procedures
Place your ST-730 Series sensor into the desired Pyxis Lab® Turbidity
Calibration solution, then on the uPyxis® choose a calibration type. Enter the
NTU value & select Slope Calibration’. A success message will appear & your ST-730 Series sensor is now calibrated. Under the
Diagnosis’ tab, you can conduct a Cleanliness Check’, uPyxis® will then notify you if the sensor is clean or if it is fouled. If your sensor reads
Dirty’, you will need the Pyxis Lab® Sensor Cleaning Kit.