Pyxis SP-350 Handheld PTSA Meter User Manual

June 5, 2024
Pyxis

Pyxis SP-350 Handheld PTSA Meter User Manual
Pyxis SP-350 Handheld PTSA Meter

Introduction

The Pyxis SP-350 is a uniquely designed handheld multimeter that measures PTSA. It is a cuvette-less device.

Less than 5 mL water sample is needed to fill the sample cell for proper measurement.

Main Features

The SP-350 includes the following features:

  • Measurement of the concentration of fluorescent tracer PTSA of a water sample.
  • Pre-calibrated for measuring PTSA (pyrenetetrasulfonic acid) in the range of 0 to 300 ppb.
  • The fluorescence PTSA measurement is automatically compensated for sample color and turbidity interference up to 60 NTU without prefiltration.
  • PTSA fluorescence can be calibrated using a single standard through a user-friendly, menu-driven procedure.
  • Large color graphic screen that can be read in direct sunlight.

Specifications

Table 1. SP-350 Specifications

Item Specification *****
P/N 50206
PTSA Range 0–300 ppb
PTSA Precision ± 1% or ± 1 ppb
Temperature Range 40–106 °F (4–41 °C)
Display 320 × 240 TFT-LCD, visible under direct sunlight
Power Supply 4 AA alkaline batteries
Typical Battery Life 10,000 readings
Dimension (L × W × H) 6.30 × 2.91 × 1.30 inch (160 × 74 × 33 mm)
Weight 0.68 lbs (310 g)
Enclosure Rating IP67

With Pyxis’s continuous improvement policy, these specifications are subject to change without notice.
Batteries excluded

Unpacking Instrument

Remove the instrument and accessories from the shipping container and inspect each item for any damage that may have occurred during shipment. Verify that all items listed on the packing slip are included. If any items are missing or damaged, please contact Pyxis Customer Service at service@pyxislab.com.

Standard Accessories
Optional Accessories

The following optional accessories can be ordered from Pyxis Customer Service (order@pyxis-lab.com) or Pyxis E-Store at <https ://pyxis-lab.com/shop/>.
Table 2.

Accessory Name P/N
Pyxis Carrying Case for SP-350 50725
Pyxis 100 ppb PTSA Calibration Standard — 500 mL 21001
Pyxis 200 ppb PTSA Calibration Standard — 500 mL 21000
Pyxis 300 ppb PTSA Calibration Standard — 500 mL 21003
Pyxis Handheld Cleaning Kit SER-02

Installation

Battery Installation

The SP-350 is powered by four alkaline batteries. Typical battery life lasts for 10,000 measurements or 10 months. When the battery capacity is critically low, the SP-350 will display a ”LOW BATTERY” warning for 5 seconds and then automatically turn off.

**NOTE*** Do not use rechargeable nickel cadmium (NiCad) or lithium batteries.

Replace the batteries to resume operation of the SP-350 after the battery warning. The SP 350 will automatically turn on in the measurement mode after new batteries are installation.

The SP-350 battery compartment, shown in Figure 1, is on the back side of the instrument. Batteries are held in place by a cover secured with two Phillips- head screws.

Figure 1. The SP-350 battery compartment
The SP-350 battery compartment

Install batteries using the following steps:

  1. Remove the battery compartment cover by loosening the two screws.
  2. Remove old batteries and dispose of them properly.
  3. Following the positive and negative terminal signs in the compartment bottom, snap four new AA alkaline batteries firmly into the battery holder.
  4. Replace the battery compartment cover and ensure that the sealing O-ring is lying flat on the battery holder.
  5. Fasten the two screws.
    NOTE*
    Failure to properly seat the O-ring may result in water damage to the SP-350.

Instrument Overview

Control Keys

The SP-350 has three control keys, as shown in Figure 2. The left ( ), right ( ), and ok ( ) keys are used to launch actions indicated on the LCD display directly above the keys. The labels above the keys indicate the function associated with each key and functions can be changed in different operation modes.
Figure 2.
Control Keys

Main Module On/Of

To turn on the SP-350: Press momentarily and release.
To turn off the SP-350: Press and hold for about three seconds. Release when the LCD display turns off. The SP-350 turns itself off after 60 seconds without user interaction detected. This is done to conserve battery life.

**NOTE*
This auto-time off setting may be customized by the user as desired through the
uPyxis®** Mobile or Desktop App.

Measurement

PTSA Measurement

When powered on, the SP-350 will default to the PTSA measurement mode. The water sample can be transferred to the sample cell using a pipette or filled directly from a faucet, sample bottle, or sample valve.

**NOTE***
Special care should be taken when pouring the sample into the cell to avoid air bubble entrainment, which can interfere with reading accuracy.

Before beginning a measurement, use the sample water to rinse the sample cell at least three times. Allow 5–10 seconds for the SP-350 to stabilize. The values will be displayed in white with a blue background if a stable value is reached (Figure 2). For a sample containing 100 ppb PTSA, the measured PTSA should be stabilized within the range of 98–102 ppb.

High Color and Turbidity Warning

The SP-350 has extra channels to measure sample turbidity and color to automatically compensate sample color and turbidity interference. If sample turbidity and color values determined are too high, a PTSA measurement warning will be displayed. In such a case, the user should filter the sample for PTSA measurement.

Calibration

PTSA Calibration (Two-Point with Zero)
  1. Rinse the sample cell three times with DI water. Fill the sample cell with DI water.
    **NOTE***
    In emergency, “non-PTSA” water, such as city water, may be used, but recalibrate using DI water for the zero step as soon as it is available.

  2. Power on the SP-350 by pressing . Allow 5–10 seconds for the SP-350 to stabilize.

  3. The unit is actively reading and displaying PTSA. The values will be very low if DI water is used; PTSA value should be near zero. A low non-zero value (e.g. 0.2 or 0.4, etc.) is not problematic.

  4. Press P-Cal ( or ) to launch the PTSA CALIBRATION screen (Figure 3).

  5. Press Zero ( ) to start the zero (blank) calibration.

  6. If the calibration succeeds, a checkmark ( ) and instructions for the slope calibration will appear (Figure 4).

  7. Rinse the sample cell three times with the desired PTSA standard. Fill the sample cell with the desired PTSA standard.

  8. Press Cycle ( ) to cycle between the PTSA standards 100, 200, and 300 ppb (it repeats). Ensure the value selected matches the desired PTSA standard in the sample cell.

  9. Press Slope ( ) to start the slope calibration.

  10. If the calibration succeeds, a checkmark ( ), a ”Calibration Success” message will appear (Figure 5). Otherwise, a warning message is displayed.

  11. Calibration is now complete. Press Exit ( ) to return to measurement mode.
    **NOTE***
    If Exit is pressed before the second checkmark appears, the calibration will not be completed and must be re-done.
    Calibration

Device Information and Diagnosis

The DEVICE INFORMATION screen is launched when System ( ) is pressed in the measurement mode.

This screen contains the device serial number, software version, and hardware version (Figure 6). The battery life as a percentage and the MAC addresses for main module also shown.
Device Information and Diagnosis

Press Diagnosis( ) to launch the SYSTEM DIAGNOSIS screen where raw measurement data are displayed (Figure 7). The information has no use for normal operation, but instead is used for device troubleshooting.
Device Information and Diagnosis

Provide an image of both the DEVICE INFORMATION screen and the SYSTEM DIAGNOSIS screen when you contact Pyxis (service@pyxis- lab.com) for troubleshooting your device or call +1 866-203-8397.

Sample Cell Cleanliness Check

The SP-350 is designed to provide reliable and accurate measurement on PTSA. Heavy fouling will prevent the light from reaching the sensor, resulting in inaccurate readings. It is suggested that the SP-350 be checked for fouling and cleaned on a monthly basis. Heavily contaminated waters may require more frequent cleanings. Cleaner water sources with less contamination may not require cleaning for several months. The SP-350 is designed to carry out a Cleanliness Check as described below:

  1. Power on the SP-350 by pressing .

  2. Press System ( ) to launch the DEVICE INFORMATION screen.

  3. Press Diagnosis ( ) to launch the SYSTEM DIAGNOSIS screen.

  4. Allow 5–10 seconds for the message in the top-right corner of the display change from Starting BTLE… to BTLE Started

  5. Press Cleanliness ( ). An instruction prompt appears to ask the user to put DI water into the sample cell (Figure 8).

  6. Pour DI water into the sample cell.

  7. Press Confirm ( , , or ). The instruction prompt will disappear and the SP-350 displays a countdown toward the bottom of the display.

  8. Once the Cleanliness Check is completed a Clean message (Figure 9) or Sample cell fouled message (Figure 10) will appear towards the bottom of the display.

  9. Cleanliness check is now complete. Press Exit ( ) to return to measurement mode.

Bluetooth Connection to Devices

The SP-350 uses a built-in Bluetooth Low Energy Connection (BTLE) to connect wirelessly to a smart phone via the uPyxis® Mobile App or to a computer via the included Bluetooth Adapter (P/N: MA-NEB) and the uPyxis® Desktop App. To allow the SP-350 to connect via Bluetooth with other devices follow the steps below:

  1. Power on the SP-350 by pressing .
  2. Press System ( ) to launch the DEVICE INFORMATION screen.
  3. Press Diagnosis ( ) to launch the SYSTEM DIAGNOSIS screen.
  4. Allow 5–10 seconds for the message in the top-right corner of the display change from Starting BTLE… to BTLE Started (Figure 7).
  5. Choose to connect via one of two options:
  6. The uPyxis® Mobile App (see the Use with uPyxis® Mobile App section), or
  7. The uPyxis® Desktop App (see the Use with uPyxis® Desktop App section).

Calibrate an ST-500 Series Sensor with the SP-350 via Bluetooth

The SP-350 can be used to verify the result of an inline Pyxis ST-500 Series sensor by measuring the sample water taken from the inline sensor sample line. The SP-350 can then be used to calibrate the inline sensors over the Bluetooth connection. To calibrate an inline sensor, follow the steps below:

  1. Power on the SP-350 by pressing .

  2. Press System ( ) to launch the DEVICE INFORMATION screen.

  3. Press Comm ( ) to launch the COMMUNICATION screen (Figure 11).
    Sample Cell Cleanliness Check

  4. Press Scan ( ) to begin scanning for Bluetooth devices.

  5. Discoverable devices will begin to populate on the display with their name and MAC-Address (Figure 12).
    Sample Cell Cleanliness Check

  6. If more than one device appears in the Device list, press » ( ) to cycle through the devices.

  7. If no devices or the incorrect device appear in the Device list, press Scan ( ) to re scan for discoverable devices.

  8. Press Connect ( ) to begin pairing to the selected sensor.

  9. When the connection is established, the SP-350 displays the latest PTSA measurement from the connected sensor (Figure 13).
    Sample Cell Cleanliness Check

  10. Fill the main module sample cell with the same sample water that the sensor is measuring.

  11. Press Read ( ) to see PTSA measurement from the SP-350along with the sensor measurement (Figure 14).
    Sample Cell Cleanliness Check

  12. Press Calib ( ) to begin sensor PTSA calibration.

  13. The SP-350 will take the sensor PTSA measurement three times to verify the calibration (Figure 15).
    Sample Cell Cleanliness Check
    **NOTE***
    It takes about one minute for the sensor to approach the calibrated reading and the three verifying readings may not be exactly the same as the value measured by the SP-350Ṗress Read ( ) again to take more readings from the sensor, if necessary.

  14. If the calibration is successful, a ”Verify calibration – OK” message will appear on the top of the display (Figure 16).
    Sample Cell Cleanliness Check

  15. Calibration is now complete. Long press Calib ( ) to return to measurement mode.

Factory Reset

Use the following steps to restore all device parameters to factory default:

  1. Power on the SP-350 by pressing .

  2. Press System ( ) to launch the DEVICE INFORMATION screen.

  3. Press Diagnosis ( or ) to launch the SYSTEM DIAGNOSIS screen.

  4. Allow 5–10 seconds for the message in the top-right corner of the display change from Starting BTLE… to BTLE Started .

  5. Press Help ( ) to launch the HELP screen (Figure 17).
    Factory Reset

  6. Press Factory Reset ( or ). The display updates as shown in Figure 18 appear and the user can choose one of three options:
    Factory Reset
    (a) Press to start the factory reset, or
    (b) Press Cancel ( ) to return to the HELP screen, or
    (c) Press Exit ( ) to abandon the factory reset entirely.

  7. After a successful factory reset, the message “Factory reset done.” will appear on the display.

  8. Press Exit ( ) to return to measurement mode.

Use with uPyxis® Mobile App

Download uPyxis® Mobile App

Download uPyxis® Mobile App from Apple App Store or Google Play
Download uPyxis® Mobile App
  Download uPyxis® Mobile
App
Figure 19. uPyxis® Mobile App installation

Connecting to uPyxis® Mobile App

Connect the SP-350 sensor to a mobile smart phone according to the following steps:

  1. Follow the steps in the Bluetooth Connection to Devices section to make the SP-350 discoverable.
  2. Open uPyxis® Mobile App.
  3. On uPyxis® Mobile App, pull down to refresh the list of available Pyxis devices.
  4. If the connection is successful, the SP-350 and its Serial Number (SN) will be displayed (Figure 20).
  5. Press on the SP-350 image.
    Connecting to uPyxis® Mobile App
Setting Screen

When connected, the uPyxis® Mobile App will default to the Setting screen. From the Setting screen, the user can set the Power off time and Screen off time in seconds.
Setting Screen

System Screen

From the System screen, users can change the Device Name, find the Serial Number, Hardware Version, and Firmware Version, as well as update the firmware of the SP-350 by pressing Check Update. If a firmware update is available, press Get Firmware. Once the new firmware is downloaded, press Upgrade Firmware.

**NOTE***
The firmware update process takes some time and will require the SP-350 to stay within range (approximately 10 ft without obstructions) for the entire duration of the update.

Once the update is complete, the SP-350 will reboot which will disconnect the SP-350 from the uPyxis® Mobile App.
System Screen

Use with uPyxis® Desktop App

[uPyxis® Desktop App installation]

Install uPyxis® Desktop App

Download the latest version of uPyxis® Desktop software package from: <https ://pyxis-lab.com/upyxis/> this setup package will download and install the Microsoft.Net Framework 4.5 (if not previously installed on the PC), the USB driver for the USB-Bluetooth adapter (MA-NEB), the USB-RS485 adapter (MA-485), and the main uPyxis® Desktop application. Double click the uPyxis.Setup.exe file to install.

Figure 23. uPyxis® Desktop App installation
Install uPyxis® Desktop App

Click Install to start the installation process. Follow the screen instructions to complete the USB driver and uPyxis installation.

Connecting to uPyxis® Desktop App

Connect the SP-350 to a Windows computer using a Bluetooth/USB adapter (P/N: MA NEB) according to the following steps:

  1. Follow the steps in the Bluetooth Connection to Devices section to make the SP-350 discoverable.
  2. Plug the Bluetooth/USB adapter into a USB port in the computer.
  3. Launch uPyxis® Desktop App.
  4. On uPyxis® Desktop App, click Device → Connect via USB-Bluetooth (Figure 24).
  5. If the connection is successful, the SP-350 and its Serial Number (SN) will be displayed in the left pane of the uPyxis® window.
    Connecting to uPyxis® Desktop App
System Screen

Once connected to the device, a picture of the device will appear on the top- left corner of the window and the uPyxis® Desktop App will default to the System screen. From the System screen, users can upgrade the firmware by selecting an appropriate firmware file (contact service@pyxis- lab.com for these firmware files) and clicking Upgrade Firmware.

**NOTE***
The firmware update process takes some time and will require the SP-350 to stay within range (approximately 10 ft without obstructions) for the entire duration of the update.

Once the update is complete, the SP-350 will reboot which will disconnect the SP-350 from the uPyxis® Mobile App.
System Screen

Datalog Screen

From the Datalog screen, the user can view, delete, and export the internal log files of the SP-350 by clicking Read Datalog List and selecting the desired datalog (these are separated by month). The SP-350 will then populate any relevant log event from the selected datalog which can be viewed in more detail by clicking Read Datalog, deleted by clicking Delete, or exported by clicking Export as .CSV File.
Datalog Screen

Setting Screen

From the Setting screen, the user can set the Power off time and Screen off time in seconds.
Setting Screen

Device Maintenance and Precaution

Maintenance Best Practices and Quick Tips

For greatly increased working life and ease of use of the SP-350 follow the list of maintenance best practices and quick tips below:

  • Rinse the sample cell at least three times with the desired solution before turning on.

  • After a successful calibration, the unit does not automatically return to the measurement mode. If Exit is held down too long the unit will power down rather than returning to the measurement mode.

  • After returning to measurement mode after calibration, rinse several times with the first sample. The unit will continue to read the sample values without any further key presses if it has not powered off.
    If there are no key presses for 20 seconds the screen will darken, and after another 20 seconds without key activity will power down. The press of any key while the screen is dark will reset the timer and the screen will re-light. (This press will not step along the calibration process; the next press needed will still be required in the sequence).

  • Always rinse the unit with clean water after use and dry by clean tissue or paper towel.

Methods to Cleaning the SP-350

A light deposit on quartz glass inside the conductivity cell can be cleaned by a Q-tip. Aged heavy deposition, especially iron oxide deposited, can be removed using a cleaning solution that is capable of removing iron, such as the Pyxis Handheld Device Cleaning Solution Kit (P/N: SER-02) available from Pyxis online E-Store
https://pyxis-lab.com/product/handheld-device-cleaning-kit/.
Figure 28. Handheld Device Cleaning Solution Kit

To clean the SP-350 pour cleaning solution into the sample cell for 10 minutes. Rinse the sample cell with distilled water and use the Cleanliness Check (see the Sample Cell Cleanliness Check section) to confirm that the SP-350 is clean. Repeat the process as needed until the Cleanliness Check shows Clean .

Storage

Do not expose the SP-350 to an extreme high or low temperature condition such as leaving the SP-350 inside an unattended automobile.

Contact Us

Pyxis Lab, Inc
1729 Majestic Dr. Suite 5
Lafayette, CO 80026 USA
www.pyxis-lab.com
Phone: +1 866-203-8397
Email: service@pyxis-lab.com

Warranty Information

Confidentiality

The information contained in this manual may be confidential and proprietary and is the property of Pyxis Lab, Inc. Information disclosed herein shall not be used to manufacture, construct, or otherwise reproduce the goods described. Information disclosed herein shall not be disclosed to others or made public in any manner without the express written consent of Pyxis Lab, Inc.

Standard Limited Warranty
Pyxis Lab warrants its products for defects in materials and workmanship. Pyxis Lab will, at its option, repair or replace instrument components that prove to be defective with new or remanufactured components (i.e., equivalent to new). The warranty set forth is exclusive and no other warranty, whether written or oral, is expressed or implied.

Warranty Term
The Pyxis warranty term is thirteen (13) months ex-works. In no event shall the standard limited warranty coverage extend beyond thirteen (13) months from original shipment date.

Warranty Service
Damaged or dysfunctional instruments may be returned to Pyxis for repair or replacement.

In some instances, replacement instruments may be available for short duration loan or lease.

Pyxis warrants that any labor services provided shall conform to the reasonable standards of technical competency and performance effective at the time of delivery. All service interventions are to be reviewed and authorized as correct and complete at the completion of the service by a customer representative, or designate. Pyxis warrants these services for 30 days after the authorization and will correct any qualifying deficiency in labor provided that the labor service deficiency is exactly related to the originating event. No other remedy, other than the provision of labor services, may be applicable.

Repair components (parts and materials), but not consumables, provided during a repair, or purchased individually, are warranted for 90 days ex-works for materials and workmanship. In no event will the incorporation of a warranted repair component into an instrument extend the whole instrument’s warranty beyond its original term.

Warranty Shipping
A Repair Authorization (RA) Number must be obtained from Pyxis Technical Support before any product can be returned to the factory. Pyxis will pay freight charges to ship replacement or repaired products to the customer. The customer shall pay freight charges for returning products to Pyxis. Any product returned to the factory without an RA number will be returned to the customer. To receive an RMA you can generate a request on our website at https://pyxis-lab.com/request-tech-support/.

Pyxis Technical Support
Contact Pyxis Technical Support at +1 866-203-8397, service@pyxis- lab.com, or by filling out a request for support at https://pyxis-lab.com/request-tech-support/.