Pyxis ST-500 Series Inline PRSA Sensors User Manual
- June 6, 2024
- Pyxis
Table of Contents
**Pyxis ST-500 Series Inline PRSA Sensors User Manual
**
Introduction
The Pyxis ST-500 Series inline fluorimeter sensor measures the concentration of fluorescence tracer PTSA (pyre netters sulfonic acid) in water. The standard ST-001 Tee Assembly provided with each ST-500/ST-500RO sensor, has two 3/4” female NPT ports and can be placed to an existing 3/4” sample water line. Pyxis Lab also offers 2” and 3” Tee formats for larger flow installations. The 4–20mA current output of the ST-500 Series sensor can be connected to any controller that accepts an isolated or non-isolated 4–20mA input. The ST-500 Series sensor is a smart device. In addition to measuring fluorescence, the ST-500 Series sensor has an extra photo-electric components that monitors the color and turbidity of the sample water. This extra feature allows automatic color and turbidity compensation to eliminate interference commonly associated with real-world waters.
The Pyxis ST-500 Series sensor has a short fluidic channel and can be easily cleaned. The fluidic and optical arrangement of the ST-500 Series sensor is designed to overcome shortcomings associated with other fluorimeters that have a distal sensor surface or a long, narrow fluidic cell. Traditional inline fluorimeters are susceptible to color, turbidity interference, and fouling, making them very difficult to properly clean.
The Pyxis ST-500 Series sensor uses a narrow wavelength band gallium phosphide photodiode and high temperature-tolerant and humidity-resistant optical filters. This combination greatly enhances the robustness of the sensor. It can be operated under a wide range of ambient conditions without the need of humidity and temperature regulation. The performance of the ST-500 Series sensor can be stable and consistent for a long period of time.
Specifications
Table 1. ST-500 Series Specifications
- With Pyxis’s continuous improvement policy, these specifications are subject to change without notice.
- See Figure 4 for ST-500SS dimensions.
Unpacking Instrument
Remove the instrument and accessories from the shipping container and inspect each item for any damage that may have occurred during shipping. Verify that all accessory items are included. If any item is missing or damaged, please contact Pyxis Lab Customer Service at service@pyxis-lab.com.
Standard Accessories
Tee Assembly 3/4” NPT (1x Tee, O-ring, and Nut) P/N: ST-001
NOTE ST-001 is not included for ST-500SS
- 7-Pin Female Adapter/Flying Leads Cable (2 ft) P/N: MA-1100
- User Manual available online at https://pyxis-lab.com/support/
Optional Accessories
The following optional accessories can be ordered from Pyxis Customer Service
(order@pyxis-lab.com) or
Pyxis E-Store at https://pyxis-lab.com/shop/.
Figure 1
Installation
ST-500 and ST-500RO Piping
The provided ST-001 Tee Assembly can be connected to a pipe system through the 3/4” female ports, either socket or NPT threaded. To properly install the ST-500/ST-500RO sensor into the ST-001 Tee Assembly, follow the steps below:
- Insert the provided O-ring into the O-ring groove on the tee.
- Insert the ST-500/ST-500RO sensor into the tee.
- Tighten the tee nut onto the tee to form a water-tight, compression seal.
Figure 3. Dimension of the ST-500/ST-500RO and the ST-001 Tee Assembly (mm)
ST-500SS Piping
The ST-500SS sensor has 3/4” female NPT threaded ports on the sensor itself
and therefore does not require a custom tee assembly. It is recommended that
two 3/4” NPT to 1/4” tubing adapters are used to connect the sensor to the
sampling system. Sample water entering the sensor must be cooled down to below
120 °F (49 °C). The sensor can be held by a 1.75-inch pipe clamp or mounted to
a panel with four 1/4-28 bolts. See Figure 4 for ST-500SS dimensions.
Figure 4. Dimension of the ST-500SS (inch)
Wiring
If the power ground terminal and the negative 4–20mA terminal in the controller are internally connected (non-isolated 4–20mA input), it is unnecessary to connect the 4–20mA negative wire (green) to the 4–20mA negative terminal in the controller. If a separate DC power supply other than that from the controller is used, make sure that the output from the power supply is rated for 22–26 VDC @ 65mA.
- NOTE The negative 24V power terminal (power ground) and the negative 4–20mA terminal on the ST-500 Series sensor are internally connected.
Follow the wiring table below to connect the ST-500 Series sensor to a
controller
Table 2.
Wire Color | Designation |
---|---|
Red | 24V + |
Black | 24V Power ground |
White | 4–20mA + |
Green ***** | 4–20mA – |
Blue | RS-485 A |
Yellow | RS-485 B |
Clear | Shield, earth ground |
- Internally connected to the power ground.
Connecting via Bluetooth
A Bluetooth adapter (P/N: MA-WB) can be used to connect a ST-500 Series sensor
to a smart phone with the uPyxis® Mobile App or a computer with a
Bluetooth/USB Adapter (P/N: MA-NEB) and the uPyxis® Desktop App. The power
should be sourced from a 24 VDC power terminal of a controller. If a
controller is not available, please purchase a Pyxis PowerPack-1 (P/N: MA-
BLE-1) or PowerPack-4 (P/N: MA-BLE-4) auxiliary power supply with Bluetooth,
or an alternative 24 V power supply that can directly connect to the ST-500
Series sensor with proper cable connectors from Pyxis.
Figure 5. Bluetooth connection to ST-500 Series sensor
Connecting via USB
A USB-RS485 adapter (P/N: MA-485) can be used to connect a ST-500 Series
sensor to a computer with the
uPyxis® Desktop App.
- NOTE Using non-Pyxis USB-RS485 adapters may result in permanent damage of the ST500 Series sensor communication hardware.
Figure 6. USB connection to ST-500 Series sensor
Setup and Calibration with uPyxis® Mobile App
Download uPyxis® Mobile App
Download uPyxis® Mobile App from Apple App Store or Google
Play.
Figure 7. uPyxis® Mobile App installation
Connecting to uPyxis® Mobile App
Connect the ST-500 Series sensor to a mobile smart phone according to the following steps:
- Open uPyxis® Mobile App.
- On uPyxis® Mobile App, pull down to refresh the list of available Pyxis devices.
- If the connection is successful, the ST-500 Series and its Serial Number (SN) will be displayed (Figure 8).
- Press on the ST-500 Series sensor image.
Figure 8.
Calibration Screen and Reading
When connected, the uPyxis® Mobile App will default to the Calibration screen. From the Calibration screen, you can perform calibrations by pressing on Zero Calibration, Slope Calibration, and 4-20mA Span.
Follow the screen instructions for each calibration step.
Figure 9.
Diagnosis Screen
From the Diagnosis screen, you can check the diagnosis condition as well as Export & Upload. This feature may be used for technical support when communicating with service@pyxis-lab.com.
To preform a Cleanliness Check, first select the Diagnosis Condition which defines the fluid type that the ST-500 Series sensor in currently measuring, then press Cleanliness Check. If the sensor is clean, a Clean message will be shown. If the sensor is severely fouled, a Dirty message will be shown. In this case, follow the procedure in the Methods to Cleaning the ST-500 Series Sensor section of this manual.
Figure 10.
Device Info Screen
From the Device Info screen. You can name the Device or Product.
Figure 11.
Setup and Calibration with uPyxis® Desktop App
Install uPyxis® Desktop App
Download the latest version of uPyxis® Desktop software package from: <https ://pyxis-lab.com/upyxis/> this setup package will download and install the Microsoft.Net Framework 4.5 (if not previously installed on the PC), the USB driver for the USB-Bluetooth adapter (MA-NEB), the USB-RS485 adapter (MA-485), and the main uPyxis® Desktop application. Double click the uPyxis.Setup.exe file to install.
Figure 12. uPyxis® Desktop App installation
Click Install to start the installation process. Follow the screen instructions to complete the USB driver and uPyxis® installation.
Connecting to uPyxis® Desktop App
Connect the ST-500 Series sensor to a Windows computer using either a Bluetooth/USB adapter (P/N: MANEB) or a USB-RS485 adapter (P/N: MA-485) according to the following steps:
- Plug the Bluetooth/USB adapter or USB-RS485 adapter into a USB port in the computer.
- Launch uPyxis® Desktop App.
- On uPyxis® Desktop App, click Device → Connect via USB-Bluetooth or Connect via USB-RS485 (Figure 13).
- If the connection is successful, the ST-500 Series and its Serial Number (SN) will be displayed in the left pane of the uPyxis® window.
**NOTE***
After the sensor and Bluetooth is powered up, it may take up to 10 seconds for the adapter to establish the wireless signal for communication.
Figure 13.
Information Screen
Once connected to the device, a picture of the device will appear on the top- left corner of the window and the uPyxis® Desktop App will default to the Information screen. On the Information screen you can set the information description for Device Name, Product Name, and Modbus Address, then click Apply Settings to save.
Figure 14.
Calibration Screen
To calibrate the device, click on Calibration. On the Calibration screen there are three calibration buttons, Zero Calibration, Slope Calibration, and 4-20mA Span. The screen also displays the reading of the device.
The reading refresh rate is every 4 seconds. Follow the screen instructions for each calibration step.
Figure 15.
Diagnosis Screen
After the device has been calibrated and installation has been completed, to check diagnosis, click on Diagnosis. When in the Diagnosis screen you can view the Diagnosis Condition of the device. This feature may be used for technical support when communicating with service@pyxis-lab.com.
To preform a Cleanliness Check, first select the Diagnosis Condition which defines the fluid type that the ST-500 Series sensor in currently measuring, then click Cleanliness Check. If the sensor is clean, a green Clean message will be shown. If the sensor is severely fouled, a red Dirty message will be shown. In this case, follow the procedure in the Methods to Cleaning the ST-500 Series Sensor section of this manual.
Figure 16.
Outputs
4–20mA Output Setup
The 4–20mA output of the ST-500 and ST-500SS sensor is scaled as:
- PTSA:
- 4 mA = 0 ppb
- 20 mA = 200 ppb
The 4–20mA output of the ST-500RO sensor is scaled as:
- PTSA:
- 4 mA = 0 ppb
- 20 mA = 40 ppb
Adjusting 4–20mA Span
Users may adjust the output scale using 4–20mA Span to change the PTSA value corresponding to the 20 mA output via uPyxis®. For the uPyxis® Mobile App, press 4-20mA Span found on the Calibration and Reading Screen, shown in Figure 17. For the uPyxis® Desktop App, click 4-20mA Span found on the Calibration Screen, shown in Figure 18.
Figure 17.
Figure 18.
Communication using Modbus RTU
The ST-500 Series sensor is configured as a Modbus slave device. In addition to the ppb PTSA value, many operational parameters, including warning and error messages, are available via a Modbus RTU connection. Contact Pyxis Lab Customer Service (service@pyxis-lab.com) for more information.
Sensor Maintenance and Precaution
The ST-500 Series sensor is designed to provide reliable and continuous PTSA readings even when installed in moderately contaminated industrial cooling waters. Although the optics are compensated for the effects of moderate fouling, heavy fouling will prevent the light from reaching the sensor, resulting in low readings and the potential for product overfeed if the ST-500 Series sensor is used as part of an automated control system. When used to control product dosing, it is suggested that the automation system be configured to provide backup to limit potential product overfeed, for example by limiting pump size or duration, or by alarming if the pumping rate exceeds a desired maximum limit.
The ST-500 Series sensor is designed to be easily removed, inspected, and
cleaned if required. It is suggested that the ST-500 Series sensor be checked
for fouling and cleaned/calibrated on a monthly basis.
Heavily contaminated waters may require more frequent cleanings. Cleaner water
sources with less contamination may not require cleaning for several months.
The need to clean the ST-500 Series sensor can be determined by the Cleanliness Check using either the uPyxis® Mobile App (see the Mobile Diagnosis Screen section) or the uPyxis® Desktop App (see the Desktop Diagnosis Screen section).
Methods to Cleaning the ST-500 Series Sensor
Any equipment in contact with industrial cooling systems is subject to many potential foulants and contaminants. Our inline sensor cleaning solutions below have been shown to remove most common foulants and contaminants. A small, soft bristle brush, Q-Tips cotton swab, or soft cloth may be used to safely clean the sensor housing and the quartz optical sensor channel. These components and more come with a Pyxis Lab Inline Probe Cleaning Solution Kit (P/N: SER-01) which can be purchased at our online E-Store https://pyxislab.com/product/st-series-probe-cleaning-kit/
Figure 19. Inline Probe Cleaning Solution Kit
To clean the ST-500 Series sensor, soak the lower half of the sensor in 100 mL inline sensor cleaning solution for 30 minutes. Rinse the ST-500 Series sensor with distilled water and then check for the flashing blue light inside the ST-500 Series sensor quartz tube. If the surface is not entirely clean, continue to soak the ST-500 Series sensor for an additional 30 minutes. Use the small, soft bristle brush and Q-Tips cotton swabs as necessary to remove any remaining contaminants in the ST-500 Series sensor quartz tube.
Storage
Avoid long term storage at temperature over 140 °F. In an outdoor installation, properly shield the ST-500 Series sensor from direct sunlight and precipitation.
Troubleshooting
If the ST-500 Series sensor output signal is not stable and fluctuates significantly, make an additional ground connection — connect the clear (shield, earth ground) wire to a conductor that contacts the sample water electrically such as a metal pipe adjacent to the ST-500 Series tee.
Carry out routine calibration verification against a qualified PTSA standard. After properly cleaning the ST500 Series sensor, carry out the zero point calibration with distilled water and slope calibration using the qualified PTSA standard. Pyxis Lab PTSA Standards can be purchased at our online E-Store https://pyxislab.com/product/ptsa-standards/
Figure 20. PTSA 100 Standard
Contact Us
Pyxis Lab, Inc
1729 Majestic Dr. Suite 5
Lafayette, CO 80026 USA
www.pyxis-lab.com
Phone: +1 866-203-8397
Email: service@pyxis-lab.com
ST-500 Series User Manual service@pyxis-lab.com | +1 866-203-8397
References
- lab.com - This website is for sale! - Experimental developer products Resources and Information.
- 📧order@pyxis-lab.com
- 📧service@pyxis-lab.com
- Your request has been blocked. This could be due to several reasons.
- Water Professionals Deserve Better Tools | Pyxis Lab®
- pyxis-lab.com/upyxis/
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>