AMATECH F-TKS Total Knee Stabilizer Instruction Manual
- June 5, 2024
- AMATECH
Table of Contents
AMATECH F-TKS Total Knee Stabilizer Instruction Manual
IMPORTANT NOTICES
Prior to using this or any other type of medical apparatus with a patient, it is recommended that you read the Instructions for Use and familiarize yourself with the product.
-
Read and understand all warnings in this manual and on the device itself prior to use with a patient.
-
The symbol is intended to alert the user to important procedures or safety instructions regarding the use of this device.
The symbol on the labels is intended to show when the IFU should be referenced for use. -
The techniques detailed in this manual are only manufacturer’s suggestions. The final responsibility for patient care with respect to this device remains with the attending physician.
-
Device function should be checked prior to each usage.
-
This device should only be operated by trained personnel.
-
All modifications, upgrades, or repairs must be performed by an authorized specialist.
-
Keep this manual available for future reference.
-
Any serious incident that has occurred in relation to the device should be reported to the manufacturer and the competent authority listed in this document.
General Information
AMATECH Corporation is a subsidiary of Allen Medical Systems, Inc. which is a subsidiary of Hill-Rom, Inc. (NYSE: HRC), a leading worldwide manufacturer and provider of medical technologies and related services for the health care industry. As an industry leader in patient positioning, our passion is improving patient outcomes and caregiver safety, while enhancing our customers’ efficiency. Our inspiration comes from providing innovative solutions to address our customers’ most pressing needs. We immerse ourselves in our customers’ world, to better address these needs and the daily challenges of their environment. Whether developing a solution to address patient positioning challenges or creating a system to offer safe and effective surgical site access for the surgical team, we are committed to providing products of exceptional value and quality.
Copyright Notice:
Revision
© 2019 Allen Medical Systems Inc. ALL RIGHTS RESERVED.
No part of this text shall be reproduced or transmitted in any form or by any means, electronic or mechanical, including photocopying, recording, or by any information or retrieval system without written permission from Allen Medical Systems, Inc. (Allen Medical).
The information in this manual is confidential and may not be disclosed to third parties without the prior written consent of Allen Medical.
Trademarks:
Trademark information can be found in www.allenmedical.com/pages/termsconditions.
Product may be covered by one or more patents. Please consult listing at www .hill-rom.com/patents for any patent(s).
Contact Details:
For complains or ordering information, please contact your supplier and refer to the catalog. All modifications, upgrades, or repairs must be performed by an authorized specialist.
Safety Considerations:
Safety hazard symbol notice:
DO NOT USE IF PRODUCT SHOWS VISIBLE DAMAGE AND MATERIAL DEGRADATION.
Equipment misuse notice:
Do not use the product if package is damaged or unintentionally opened before use. All modifications, upgrades, or repairs must be performed by an authorized specialist.
Notice to users and/or patients:
Any serious incident that has occurred in relation to the device should be reported using the contact details provided in section 1.3 of this Instruction for Use and to the competent authority of the Member State in which the user and/or patient is established.
Note: Refer to the surgical table manufacturer’s user guide for instructions on use. Always refer to the surgical table manufacturer’s weight limits.
NEVER EXCEED THE WEIGHT CAPACITY OF THE OPERATING ROOM TABLE
Safe disposal:
Customers should adhere to all federal, state, regional, and/or local laws and regulations as it pertains to the safe disposal of medical devices and accessories.
If in doubt, the user of the device shall first contact your supplier for guidance on safe disposal protocols.
Operating the system:
Applicable Symbols:
Symbol used | Description | Reference |
---|---|---|
Indicates the device is a medical device | MDR 2017/745 | |
**** | Indicates the medical device manufacturer | EN ISO 15223-1 |
**** | Indicates the manufacturer’s serial number. The device serial number is |
encoded as 1YYWWSSSSSSS.
- YY indicates the year of manufacture. i.e. 1 18 WWSSSSSSS where 18 represents the year 2018.
- WW indicates the number of the manufacturing week per a standard shop calendar. (Leading zeros included.)
- SSSSSSS is a sequential unique number.
| EN ISO 15223-1
| Indicates the medical device Global Trade Item Number| 21 CFR 830 MDR
2017/745
| Indicates the date when the medical device was manufactured| EN ISO
15223-1
| Indicates the manufacturer’s lot code using the Julian Date yyddd,
where yy indicates the last two digits of the year and ddd indicates
the day of the year. i.e. April 4th, 2019 would be represented as 19094.| EN
ISO 15223-1
| Indicates the manufacturer’s catalogue number| EN ISO 15223-1
| Indicates the need for the user to consult the instructions for use for
important cautionary information such as warnings and precautions| EN ISO
15223-1
| Indicates the device do not contain natural rubber or dry natural
rubber latex| EN ISO 15223-1
| Indicates the authorized representative in the European Community|
EN ISO 15223-1
**| Indicates the Medical Device complies to REGULATION (EU) 2017/745|
MDR 2017/745
| Indicates a Warning| IEC 60601-1
**| Intended to show when the IFU should be referenced for use| EN
ISO 15223-1
Intended User and Patient Population:
Intended User: Surgeons, Nurses, Doctors, Physicians and OR healthcare professionals involved in the device intended procedure. Not intended for Lay persons.
Intended Populations:
This device is intended to be used with patients that do not exceed the weight
in the safe working load field specified in the product specification section
4.2
Compliance with medical device regulations:
This product is a non-invasive, Class I Medical Device. This system is CE marked according to Annex VIII, Rule 1, of the Medical Device Regulations (REGULATION (EU) 2017/745)
EMC considerations:
This is not an electromechanical device. Therefore, EMC Declarations are not applicable.
EC authorized representative:
HILL-ROM SAS
B.P. 14 – Z.I. DU TALHOUET
56330 PLUVIGNER
FRANCE
TEL: +33 (0)2 97 50 92 12
Manufacturing Information:
ALLEN MEDICAL SYSTEMS, INC.
100 DISCOVERY WAY
ACTON, MA 01720 USA USA
800-433-5774 (NORTH AMERICA)
978-266-4200 (INTERNATIONAL)
System
System components Identification:
Product Code and Description:
F-TKS – TOTAL KNEE STABILIZER
List of Accessories and Consumable Components Table:
The following list are accessories and components that may be used with this device.
Name of Accessory | Product number |
---|---|
Total Knee Stabilizer Pad | O-TKSP |
Accessory Cart | F-30010 (US) F-30010E (EU) F-30010UK (UK) |
Name of Consumable | Product number |
--- | --- |
Replacement Disposable Pad | O-TKSP |
Note: Consult the corresponding IFU for the products mentioned in the above table.
CAUTION: Reuse of disposables will lead to patient cross contamination and/or device failure.
Indication for use:
The Total Knee Stabilizer is designed to position and support the patient’s foot in a variety of surgical procedures including, but not limited to total knee surgery. These devices are intended to be used by healthcare professionals within the Operating Room setting.
Intended use:
The Total Knee Stabilizer is used in a variety of surgical procedures including, but not limited to total knee surgery. These devices are capable of being used with a broad patient population as determined appropriate by the caregiver or institution.
Equipment Setup and Use:
Prior to use:
- a. Inspect the product looking for any visible damage or sharp edges that could be caused by a drop or impact during storage.
- b. Make sure product has been properly cleaned and disinfected and wiped dry prior to each use.
Setup:
- a. Position table rail clamp to desired location for Total Knee Stabilizer.
- b. Attach the Total Knee Stabilizer to table using the rail clamp and tighten.
Device controls and indicators:
Controls and indicators of this device are described within the Setup instructions.
Storage, Handling and Removal Instructions:
Storage and Handling:
The product should be stored in a clean and safe environment to prevent
product damage.
Removal Instruction:
Remove by loosening the rail clamp.
Troubleshooting Guide:
This device does not have a troubleshooting guide. For technical support user of the device shall first contact his or her supplier.
Device Maintenance:
Make sure that all labels are installed and can be read. Replace labels as necessary by using a plastic scraper to remove the label. Use an alcohol wipe to remove any adhesive residue.
Contact Allen Medical Systems, Inc. if you need to repair or replace the device contact us using the information from the contact details section (1.3).
Safety Precautions and General Information:
General Safety Warnings and Cautions:
WARNING:
- a. Do not use if product shows visible damage.
- b. Prior to using this device, please read the instructions for equipment set up and use. Familiarize yourself with the product before application on a patient.
- c. To prevent patient and/or user injury and/or equipment damage, examine the device and surgical-table side rails for potential damage or wear prior to use. Do not use the device if damage is visible, if parts are missing or if it does not function as expected.
- d. To prevent patient and/or user injury and/or equipment damage, verify the device attaching clamps completely touch the table-side rails and are firmly in place. Test the locking mechanism to ensure no movement when elevated or pushed.
CAUTION:
Do not exceed safe working load shown in the product specification table.
Product Specifications:
Mechanical Specifications | Description |
---|---|
Product Dimensions | 13″ (33 cm) long mounting bar |
10″ long by 2″ diameter (25.5 cm x 5 cm) foot support
Flat ¼” x 1″ (0.6 cm x 2.5 cm) cross section
Material| PVC Type I, Stainless Steel, Buna-N/PVC
Rubber Foam
Safe Working Load on the device| 500 lbs (226 kg)
Overall Weight of Complete Device| 3 lbs (1.4 kg)
Storage Specifications| Description
Storage temperature| -29ᵒ C to +60ᵒ C
Storage Relative humidity range| 15% to 85%
Operating temperature| This device is intended to be used in a controlled
Operating Room environment.
Operating Relative humidity range
Electrical Specifications| Description
---|---
Not Applicable.| Not Applicable.
Software Specifications| Description
Not Applicable.| Not Applicable.
Compatibility Specifications| Description
The Total Knee Stabilizer is compatible with:| Clip-On™ Blade Clamp
F-40040 (US), F-40041 (EU),
F-40042 (UK), F-40043 (Japan)
Rail Clamp
F-RC2, (US), F-RC2XUK (EU/UK), F-RCS (Swiss) F-RC29X32MM (Japan), F-RC2XD (Denyer)
Note: Consult the corresponding IFU for the products mentioned in the above table.
Sterilization Instruction:
This device is not intended to be sterilized. Equipment damage may occur.
Cleaning and Disinfection Instruction:
WARNING:
- Do not use bleach or products that contain bleach to clean the device. Injury or equipment damage can occur.
- After each use, clean the device with alcohol-based wipes.
- Do not put the device into water. Equipment damage can occur.
- Use a cloth and a quaternary ammonium disinfecting/cleaning solution to clean and disinfect the device.
- Read and follow the manufacturer’s recommendation for low-level disinfection.
- Read and follow the cleaning product’s instructions. Use caution in areas where liquid can get into the mechanism.
- Wipe the device with a clean, dry cloth.
- Make sure that the device is dry before you store it or use it again.
CAUTION: DO NOT IMMERSE PADS IN ANY LIQUID
CAUTION: DO NOT USE BLEACH OR PHENOLICS ON PADS
List of Applicable Standards:
Sl. no | Standards | Description |
---|---|---|
1. | EN 62366-1 | Medical devices – Part 1: Application of usability engineering |
to medical devices
2.| EN ISO 14971| Medical devices- Application of risk management to medical
devices.
3.| EN 1041| Information supplied by the manufacturer of medical devices
| ****
EN ISO 15223-1
| Medical devices – Symbols to be used with medical device labels, labelling
and information to be supplied – Part 1: General requirements
5.| EN ISO 10993-1| Biological evaluation of medical devices –
Part 1: Evaluation and testing within a risk management process
6.| IEC 60601-2-46| Medical electrical equipment – Part 2-46:
Particular requirements for the basic safety and essential performance of
operating tables
7.| ISTA| International Safe Transit Association standards for package testing
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>