Wiring Guides

ST-710 & ST-711

  • Depending on the controller, you may need to add a jumper from 24V- (Power Ground) to 4–20mA-

ST-712

Using uPyxis®

Place your sensor into the necessary Pyxis Lab® pH or OPR solution, then on uPyxis® select your desired Calibration Method & enter the specific concentration. Press ‘Calibrate’ to finish!pyxis-ST-71X-Series-Inline-pH-
Sensor-fig-3

Under ‘Diagnosis’ you can send diagnostic information straight to our Customer Service department. Under ‘Device Info’ you can customize Device Name, Product Name & Modbus Address!pyxis-ST-71X-Series-Inline-pH-Sensor-
fig-4