iFUGE D08
PRODUCT USER MANUAL https://www.praxisdienst.com/en/Lab+Equipment/Lab+Devices/
Centrifuges/iFuge+D08+Doctor+Centrifuge.html
INTRODUCTION
This manual provides important safety information for this centrifuge. lt should be kept near the centrifuge for quick & easy reference. This centrifuge is equipped with maintenance free brushless DC motor drive and a safety feature of imbalance detection. The centrifuge can be used for multiple application diagnostics & will work with varity of blood collection tu bes.
INTENDED USE
This is a bench top centrifuge which is designed to separate, sediment, spin
down aqueous solutions & solvent suspensions of differing densities in
compatible sample containers.
NOTE: Before using the centrifuge, please read this user manual
carefully. This usermanual is intended to assist with the operation and care
of the unit & is not a document which aids in repair. For repair please
contact the supplier.
SYMBOL
Warning
**** Waste electrical
FEATURES
- Maximum volume of4x50ml
- lmbalance detector with auto cutoff for safe operation
- Brushless DC motor for maintenance free performance
- Lid lock safety-the lid opens automatically on run completion
- Digital display with speed and time indicator
- RPM setting from 500-4500 RPM with a count of 100
- Long press changeover from RPM to RCF functionality
- Digital countdown Timer
- Reduction adaptors for tubes with volume less than 15 ml
- Emergency lid release du ring power failure
- Automatie internal diagnosis & error display
- Silent operation programing option
TECHNICAL SPECIFICATION
Option to choose 8 tube rotor that comes with Autoclavable and Tube holders.
Motor Type | Brushless Motor |
---|---|
Maximum Volume | 4 x 50 ml |
Speed Setting | 500 to 45 00 RPM |
Maximum RCF | 2109 x g |
Speed Accuracy | ± 100 RPM |
Speed Steps | ± 100 RPM |
Run time | 1 to 99 mins & infinite mode |
Noise Level | ≤62 dB (A) |
Ambient Temperature | 5-40ºc |
Permissible Relative Humidity | ≤80% |
Acceleration Time | 35 ± 5 seconds |
Deceleration Time | 45 ± 5 seconds |
Size (L x W x H) | 330 x 265 x 161 mm |
Weight | 4 Kg (with roter) |
Altitude | Use upto an altitude of 2000 m above MSL |
Pollution Degree | 2 |
Environment | For indoor use only |
Power Adapter Details | Input: 100-240 VAC, 50/60 Hz |
Output: 24V ****3A
Power Consumption| 40W
STANDARD ACCESSORIES
- Power adaptor
- 8 – slots roter (Pre-installed)
- T- Allen Key
- 15 ml Reduction adaptors (set of4)
- Product user manual & warranty card
- Reduction adaptor and spacer (set of 8)
INSTALLATION
Gently remove the upper packaging and take out the centrifuge by holding it
from the bottom. When the centrifuge is used for the first time, ensure that
all the packaging and accessories are removed from the roter chamber. The
roter is pre-installed. Please make sure that rotor is firmly tightened.
Accessories and user manual are kept inside the box with the centrifuge.
Please keep all the packaging in safe storage for at least 2 years for
warranty purpose.
PRECAUTIONS
- Make sure to install the centrifuge on flat stable surfaces, devoid of vibration.
- Not to lean on or insert outside weight on centrifuge to be strictly followed.
- Keep the installation area uncrowded, specially with potentially hazardous materials.
- A single technician / lab personal should only be allowed near the centrifuge while its running.
The following points are important forproper installation:
| Must be used on Fiat & Non – Slippery Surface
---|---
| Must have Clearance of 20 cm from all 4 sides
7.1 MOUNTING
Place the centrifuge on a flat and leveled surface, ensure that the four feet
of this centrifuge stand on the surface firmly. Avoid installing on a slippery
surface or surface prone to vibration.
- Ideal ambient temperature is 20° C ± S° C, avoid placing the centrifuge in direct sunlight.
- Keep clearance of at least 30-35 cm on both sides and at least 20 cm behind it to guarantee cooling efficiency.
- Keep away from heat or water to avoid sample temperature issues or centrifuge failures.
- Do not place the centrifuge in any area where operating the unit maybe difficult.
7.2 CONNECTING POWER ADAPTOR
-
Connect one slde of power adaptor to rear side of centrifuge and other to supply as shown in the figure below.
-
Rotate clockwise the adaptor nut to tighten the adaptor with the centrifuge.
Ensure the power switch is OFF while connecting the power adaptor. -
The sticker on the body & jack are indicating the position of the rib on the jack. While fitting the adaptor to the unit-piease match the sticker position as shown in the figure below & gently push it in.
-
lf matched correctly the plug will go in easily without much force (do not try to force the plug in if it is not matched properly). Once the plug is inserted into position, turn the loose upper ring clockwise to tighten & to secure the fitment.
STANDARD PARTS USTING
- View window
- Operation and setting buttons & Display
- Emergency lid release
- Power switch
- Adapter port
Bottom View
REDUCTION ADAPTORS
This centrifuge features centrifugation for wide range of volumes from 5ml to 50ml with the combination of its adaptors. The accessories included are:
Rotor Options
| Standard Blood Collection Tubes
Tube Capacity| 50 ml| 15 ml| 10 ml| 7 ml| 5 ml| 2 ml
Dimension (mm)| Ø29 x 115| Ø17 x 120| Ø16 x 100| Ø13 x 100| Ø12 x 75| Ø13 x 55
Rotor Radius (mm)| 88.| 93| 89| 87.| 73| 64
Speed (RPM)| 4500| 4500| 4500| 4500| 4500| 4500
Max. RCF| 1985 x g| 2109 x g| 2019 x g| 1966xg| 1656 x g| 1452 x g
Total no. of tube| 4| 8| 8| 8| 8| 8
Adaptor Required | No| No| Yes| Yes| Yes| Yes
DO8SP1 | No| No| No| No| Yes| Yes
DO8SP2 | No| No| No| Yes| No| No
DO8SP3 | No| No| No| No| No| Yes
ROTOR INSTALLATION
The rotor of the centrifuge is pre-installed. In case of any spillage, remove
the rotor as shown in the above figure to clean the rotor and its chamber
thoroughly.
10.1 REMOVING ROTOR
- To remove rotor, first remove the Tube Holders.
- Using T-Allen key, loosen the rotor nut by rotating rotor nut eo unter clockwise and take the out rotorvertically.
10.2 INSTALLING THE ROTOR
- Make sure the rotor shaft and the rotor mounting hole are clean. Load the rotor vertically into the motor shaft.
- Put the T-Allen key inthe rotor nut& turn clockwise to tighten the rotor.
- After properly fastening the rotor, place the Tube Holders into the rotor.
NOTE: Care should be taken to open lid of centrifuge via the function button only, do not use external pressure to pul/ it open. In case of lid getting jammed p/ease use emergency lid release.
BALANCING THE ROTOR
- Always balance the rotor before centrifugation Symmetrical loading of tubes in the rotor is necessary. Otherwise vibration or imbalance can occur which can cause serious damage to the centrifuge. The following are process of symmetrical loading of centrifuge tu bes inside the rotor.
- The samples in the tubes should be of equal volume. To achieve a balanced condition of the 8 reduction adaptor are mandatory to place inside the rotor whether 2, 3, 4 or6 / 8 tubes are being used.
- lf the tu bes are not loaded symmetrically then the imbalance detector will cut off the running centrifuge for safety reasons. This will stop the centrifuge and Err 55 will be seen indicating the tubes are not loaded symmetrically. To resume operation, load the tu bes symmetrically & restart the centrifuge.
- Below are examples of incorrect method of loading tu bes into the centrifuge rotor.
USER INTERFACE & DISPLAY
Item | Name | Function |
---|---|---|
1 | RPM | Press this button to select RPM value. Then Press “+/-” to set RPM |
value.
2| RCF| Long press “RPM/RCF” to toggle between RPM/RCF value.
Select the RCF value to be set for centrifugation.
3| START| Single press “START” button centrifuge will start as per set RPM and
TIME. START button works only if lid is closed.
4| STOP/OPEN| Single press “STOP/OPEN” button will stop centrifuge.
Centrifuge lid will open automatically after end of centrifugation.
5| TIME| Press this button to select Time value. Then Press “+/-” to set Time
value.
6| P1| Single press “P1” button to set program mode.
7| P2| Single press “P2” button to set program mode.
8| +1-| “+” is used to increase and “-” is used to decrease values.
OPERATION
After properly installing the centrifuge, switch ON the centrifuge from the rear side by pressing the power switch. The lid status will be seen in the display. lf the lid is open, the display will show “Lid” and if it is closed, the set RPM will be displayed. Please ensure that the Tube Holders are insidethe rotor.
13.1 SPEED SETTING
After closing the centrifuge lid, press “RPM/RCF” button to select speed
setting in RPM/RCF mode. Speed display will blink for 5 seconds. Now press(+)
button to increase the speed value and press (-) button to decrease the speed
value. Minimum and Maximum RPM of the centrifuge is 500 RPM to 4500 RPM
respectively. Blinking of RPM value will stop and the RPM value will be stored
automatically if no button is pressed after around 1 seconds ofthe adjustment.
The speed can also be changed while the centrifuge is under operation. Press
the “RPM/RCF” button & use setting (+) and (-) to change speed. Changing the
speed between the ongoing centrifugation will run the centrifuge at updated
speed for the rest of time as indicates by the timer. The speed can be
selected in RCF value, too. Long Press “RPM/RCF” button to change the RCF
setting. Press the (+) & (-) button to increase or decrease the value.
Blinking of RCF value will stop and the RCF value will be stored automatically
if no button is pressed after around 1 seconds of the adjustment. The value
will blink 5 time to indicate acceptance.
13.2 TIME SETTING
Operating time can be selected from 1 min to 99 min or “continuous”. Press
“TIME” button to change the time setting. Press the (+) and (-) button to
increase or decrease the value. and after 5 time blink of the set value. The
centrifuge timer set for run between lmin to 99 minutes or operated in
infinite time mode or roll over mode. Infinite timer will be indicated by
](:]( in display. The timer in the centrifugeis countdown timer in the display
will be in “Min/Sec” mode. Time will showed on after 5 sec. Time indicate on
Time LED light.
The input will be accepted if we leave the setting for around 1 seconds. Te
value will blink 5 time to indicate acceptance.
13.3 START & STOP OPERATION
Press “START” button to start operation and press “STOP” button to stop the
ongoing operation. When the centrifuge is running LED light will be ON.
Pressing the “STOP/OPEN” button will stop the operation. Once operation is
over centrifuge lid will open automatically. lf lid is not open press
“STOP/OPEN” button to open the centrifuge lid.
NOTE: In case of power failure once the rotor has stopped, pul/ out the
emergency lid release as shown in the figure. This will open the centrifuge
lid.
13.4 PROGRAM MODE
Single press “P1” button to enter into program mode. and Press “RPM/RCF”
Button to change in RPM/RCF in program mode. and select to (+) and (-) button
to change RPM/RCF value and press “TIME” button to change in time in program
mode. Values for any parameter gets save after 5 blink. Once program is set,
press “Start” button to start program.
Single press “P1” button to enter into only one program. and to use second
program press “P2” button. On first time usage, all program will have zero (0)
values and once the “Pl” button is pressed display will appear indicating Pl.
To enter into normal mode long press “P1/P2” button.
SAFETY PRECAUTION
Read all safety & usage information provided in this manual carefully before usingthe device.
- Never usethe centrifuge in any manner not specified in the manual.
- Never move the centrifuge while it’s operating.
- The rotor must always be securely fastened. lf the centrifuge makes an unusual noise du ring the operation, the rotor fit needs to be checked.
- The rotor must be loaded symmetrically.
- Do not use the centrifuge or rotor, if it is not installed correctly or it is damaged.
- Do not use the centrifuge or rotor if it shows any sign of external mechanical or electrical damages.
- The centrifuge must be used only for the specified application. lt must not be operated in hazardous or flammable environment & no hazardous or fla m ma ble content should be centrifuged.
- Do not fill tu bes while they are in the rotor. Liquid spillage may harm the device.
- lf liquid is spilled on the rotor or rotor chamber, the centrifuge must be cleaned immediately.
MAINTENANCE & CLEANING
Before cleaning the centrifuge, turn it off device and disconnect the power
cord.
OUTSIDE OFTHE DEVICE
- Clean the outside of the centrifuge with a soft and dry cloth.
- Do not use aggressive chemicals such as alcohol, benzene, acetone or phenol.
- lf the device is contaminated, use a 70% IPA solution to decontaminate it.
- Be careful to not scratch the surface of equipment when cleaning it.
- Do not use a metal sponge.
- The brushless motor in the centrifuge requires no routine maintenance. Any required service should be performed by authorized, qualified personnel only. Repairs performed by unauthorized personnel may void the warranty. Protective bonding continuity would be checked by authorised service person only.
INTERIOR OF THE DEVICE (CHAMBER)
- lf the chamber is not dry, wipe away any moisture from the chamber with a dry cloth.
- Clean the chamber and motor shaft at least once a week with a mild cleaning fluid.
SAFE GUARDS FOR THE OPERATOR
- The operator has to wear standard PPE du ring the cleaning process.
- Ensure proper seal proof gloves are warn while cleaning the interior surface of centrifuge.
TROUBLESHOOTING
This centrifuge has a seif- diagnostic function. lf a problem occurs, an error/warning code will be displayed on the display screen and the operator can determine the malfunction with the warning code below.
ERROR | PROBLEM | SOLUTION |
---|---|---|
No display | No main power connection | Check power & properly plugin power |
adaptor at both ends
Power failure| Check the main fuse
Improper connection| Connect adaptor properly
Error 1| Latch jammed or any limit switch of latch got damaged| Contact
service representative
Error 52| Rotor stuck or incorrect operating voltage| Turn off the centrifuge,
check rotor fitment or apply correct 230 VAC ± 10VAC operating voltage.
Restart unit
Error 55| Rotor not loaded symmetrically| Load rotor symmetrically & restart
centrifuge. restart unit
Centrifuge lid cannot be opened| Rotor is still spinning| Wait for the rotor
to come to a stop
Power Failure| Emergency lid release after rotor stop
Centrifuge shakes during acceleration & exceptional
running noise| Rotor not loaded symmetrically| Load rotor symmetrically &
restart operation
Either a broken tube, damage to the rotor or motor is cause for run noise|
Replace broken tube. For damaged rotor/motor contact service representative
Rotor damaged| Remove & change rotor
Display Error| Loose connection of display| Contact service representative
Power tripping| Cable not fit properly| Remove cable and connect properly
IMPORTANT NOTE:
- If the centrifuge is working slow or hot due to overdraw current, restart the centrifuge and check lt again.
- Do not use liquid with density higher than l.2g/ml forfull load operation.
WARRANTY STATEMENT
This product is warranted to be free from defects in material and workmanship
for a period of two (2) years from date of purchase. Your product will be duly
repaired upon prompt notification in compliance with the following conditions:
This warranty is valid only if the product is used for its intended purpose
and within the guidelines specified in this instruction manual. This warranty
does not cover damage caused by accident, neglect, misuse, improper service,
natural forces or other causes not arising from defects in original material
or workmanship. This warranty does not cover any incidental or consequential
damages, commercial loss or any other damages from the use of this product.
The warranty is invalidated by any non-factory modification, which will
immediately terminate all liabilities on us for the products or damages caused
by its use. The buyer and its customer shall be responsible for the product or
use of products as weil as any supervision required for safety. lf requested
the products must be returned to the distributor in weil packed and insured
manner and all shipping charges must be paid.
Some states do not allow limitation on the length of implied warranties or the
exclusion or limitation of incidental or consequential damages. This warranty
gives you specific legal rights. This warranty is given expressly in lieu of
all other warranties, expressed or implied.
The purchaser agrees that there is no warranty of merchantability or of
fitness for any intended purpose and that there are no other remedies or
warranties, expressed or implied, which extend beyond the description on the
face of the agreement. This warranty is only applicable to the original
purchaser.
Products received without proper authorization will not be entertained. All
items returned for service should be sent postage prepaid in the original
packaging or other suitable carton, padded to avoid damage. We will not be
responsible for damage incurred by im proper packaging. All items returned for
service should be set postage prepaid in the original packaging or other
suitable carton, added to avoid damage. This warranty is valid only if the
warranty is registered with the supplier within 30 days from the date of
purchase.
For you reference, make a note of serial number, date of purchase and supplier
here.
Serial No.:…………………
Purchase Date:………….
Supplier:…………………..
PRODUCT DISPOSAL
In case the product is to be disposed of, the relevant legal regulations are
to be observed.
Information on the disposal of electrical and electronic devices in the
European Community
The disposal of electrical devices is regulated within the European Community
by national regulations based on EU Directive 2012/19/EU on waste electrical
and electronic equipment (WEEE). According to these regulations, any devices
supplied after 13.06.05 in the business to business sphere, to which this
product is assigned, may no langer be disposed off in municipal or domestic
waste. They are marked with the following symbol to indicate this.
As disposal regulations within the EU may vary from country to country, please
contact your supplier if necessary.
https://www.praxisdienst.com/en/Lab+Equipment/Lab+Devices/
Centrifuges/iFuge+D08+Doctor+Centrifuge.html
Accumax Lab Devices Private Limited
Plot No. 15, 16 & 32 GIDC Electronic Park SEZ, Seetor 26,
Gandhinagar, Gujarat, lndia – 382026.
Website:www.neuation.com