RAVEN SID-20200 Portable Sludge Interface Detector Owner’s Manual
- June 13, 2024
- Raven
Table of Contents
RAVEN SID-20200 Portable Sludge Interface Detector
Product Usage Instructions
- Press the POWER button to activate the display.
- Lower the probe until it just contacts the tank bottom.
- Place the cable on the COUNTER ROLLER.
- Press the ZERO button to reset the display value.
- Raise the probe until the INTERFACE lamp illuminates.
- Note the PROBE TRAVEL distance.
USING THE SID-20200
- Press POWER button to activate display Lower probe to just contact tank bottom Place cable on COUNTER ROLLER
- Press ZERO button to reset display value Raise probe until INTERFACE lamp illuminates Note PROBE TRAVEL distance
IMPORTANT
- Store the probe on the case handle hook, while carrying the unit.
- Probe stored on hook prolongs the life of the cable and probe.
- Wrap cable loosely on cable rack to prolong cable.
The NEW SID-20200 replaces the SID-10200 with upgrades that include Side mounted counter roller that measures probe travel
- Digital display that is visible at night
- Switchable Inch or millimeter display units
- Auto sleep preserves batteries
- Easy clean faceplate with tactile buttons
- Light weight and ultra-smple to use
TOOL FAMILIARITY
- POWER button activates unit
- ZERO button resets travel value
- SENS button toggles probe sensitivity
- HORN button enables horn
- PROBE TRAVEL updates when cable passes across roller
- INTERFACE lamp illuminates when probe sensor obstructed
- HORN sounds when enabled and probe sensor obstructed
- SENS display indicates current sensitivity setting
- LOW BATT displays when battery change required
- The displayed SENS value indicates the probe sensitivity setting.
- The SENS button cycles LOW, MED and HIGH sensitivity settings.
- LOW sensitivity ignores relatively small solids, but detects large solids.
- HIGH sensitivity detects relatively small and large solids.
- When solids detected, the INTERFACE lamp illuminates and the horn sounds.
SLUDGE INTERFACE DETECTOR SID-20200 SPECIFICATIONS
Product Name | Sludge Interface Detector SID-20200 |
---|---|
Handheld Unit Dimensions | 6″W x 12″L x 4.5″H (152mmW x 305mmL x 114mmH) |
Handheld Unit Weight | 1.5 lbs (680 g) |
Handheld Unit Color | Yellow |
Handheld Unit Faceplate | 1/8″ (3.175mm) Aluminum |
Handheld Unit Moisture Protection | IP66 Weatherproof |
Handheld Unit Battery Type | 3 x AAA Cells (Standard Alkaline) |
Handheld Unit Battery Life | 3 Months (Typical) |
Handheld Unit Power | On/Off |
Handheld Unit Horn | On/Off |
Handheld Unit Sensitivity | Low, Medium, High |
Handheld Unit Visual Indicator | Red LED |
Handheld Unit Audible Indicator | Horn |
Handheld Unit Battery Indicator | Low Battery Display |
Handheld Unit Calibration | Factory Sealed (No Calibration Recal Required) |
Probe Model | RP-20203-N (N = Length) |
Probe Moisture Protection | IP68 |
Probe Material | Stainless Steel |
Probe Technology | IRED Emitter/Detector |
Probe Sensor Gap | 0.78″ (20mm) |
Probe Lenses | Polycarbonate Clear |
Probe Range Min | 100-2000 mg/l (varies) |
Probe Range Max | 100-10000 mg/l (varies) |
Probe Dimensions | 2.5″W x 3.5″L x 0.75″H (64mmW x 89mmL x 19mmH) |
Probe Weight | 1 lb (450 g) |
Probe Color | Silver |
Cable Moisture Protection | IP68 |
Cable Temperature | Length: 20 to 50ft (6.1m to 15m) |
QUICK CHECK
- Press POWER button.
- Press HORN to button to enable.
- Place your finger over a probe sensor len.
- The INTERFACE lamp illuminates and horn sounds.
- The horn can be switched-off without affecting unit operation.
BATTERY REPLACEMENT
- Batteries are located inside the case head.
- Remove the case faceplate four screws.
- Remove the faceplate from the case and leave wire attached.
- Slide the black battery pack out of the case holder.
- Remove battery pack cover.
- Replace all three AAA batteries with new alkaline batteries.
- Slide the black battery pack into the case holder.
- Install case faceplate with original screws snug tight.
SENSOR ADJUSTMENT
- SENS “LOW” – Detects large solids only
- SENS “MED” – Detects intermediate and larger solids
- SENS “HIGH” – Detects all solids
FINE-TUNING SLUDGE DETECTION
- When you’ve established a baseline sensitivity for a tank, you can fine-tune by increasing and decreasing the sensitivity above your baseline.
- At sensitivities higher than your baseline, you can detect the ragged edge on top of the sludge blanket.
- Sludge blankets rarely have a well defined interface with the supernatant. Experience with the SENSOR ADJ will allow the keen operator to measure the ragged edge of the sludge blanket. It is not uncommon in a
- dynamic environment such as a secondary clarifier to identify several feet of gradual density change on top of the sludge blanket.
CABLE REPLACEMENT
- The cable and probe are one assembly and can be replaced in the field.
- Contact Raven for replacement parts and probe assemblies.
DESIGNED AND MANUFACTURED IN USA BY
- Raven Environmental Products, Inc.
- St. Louis, MO 63122 USA
- 800-545-6953
- ravenep.com