KYORITSU 2608A AC Clamp Meter with DC Volt Range Instruction Manual
- June 13, 2024
- KYORITSU
Table of Contents
2608A AC Clamp Meter with DC Volt Range
Product Information
The KEW SNAP 2608A Transformer Jaw is an AC clamp meter with a
DC volt range. It is designed to incorporate many safety features
for reliable and safe operation. The instrument is equipped with a
protective fingerguard, a range switch, and a barrier for
electrical shock protection. It also has a zero adjuster, a scale
plate, and a jaw trigger for easy usage. The transformer jaw has V
terminal, COM terminal, and OHM terminal for convenient
connections. The product is manufactured by KYORITSU ELECTRICAL
INSTRUMENTS WORKS, LTD.
Specifications
-
Measuring Range and Accuracy:
-
AC current: 6/15/60/150/300A
-
AC voltage: 150/300/600V
-
DC voltage: 60V x1, 1k (25 mid-scale) x10, 10k (250
mid-scale) -
Resistance: -20-+150 (with no temperature compensation)
-
-
Overload Protection:
- Maximum overload: 60A AC for 10 sec, 300A AC for 10 sec, 360A
AC for 10 sec, 600V AC for 10 sec, 720V AC for 10 sec, 230V AC for
10 sec, 230V AC (protection by fuse)
- Maximum overload: 60A AC for 10 sec, 300A AC for 10 sec, 360A
-
Location for Use: Indoor use, Altitude up to 2000m
-
Storage Temperature and Humidity: -10-+50, relative humidity up
to 75% without condensation -
Operating Temperature and Humidity: 0-+40, relative humidity up
to 90% without condensation -
Conductor Size: Safety Standard Withstand Voltage
-
Dimensions: (Not provided in the manual)
-
Weight: (Not provided in the manual)
-
Power Source: (Not provided in the manual)
-
Accessories (included): (Not provided in the manual)
-
Accessories (optional): (Not provided in the manual)
Product Usage Instructions
1. Before using the KEW SNAP 2608A Transformer Jaw, make sure to
read and understand the safety warnings and instructions in the
user manual.
2. Connect the V terminal, COM terminal, and OHM terminal of the
transformer jaw to the appropriate electrical connections.
3. Ensure that the protective fingerguard is in place to prevent
electrical shock.
4. Use the range switch to select the desired measurement range
for AC current and voltage.
5. Adjust the zero value using the zero adjuster as needed.
6. The scale plate shows the measured values for easy
reading.
7. Use the jaw trigger to open and close the transformer jaw for
clamping around conductors.
8. The instrument is equipped with overload protection to
prevent damage from excessive current or voltage.
9. When using the instrument, ensure that it is within the
specified temperature and humidity range.
10. Follow proper storage conditions, including temperature and
humidity requirements.
11. The conductor size and safety standards should be considered
when using the instrument.
12. The dimensions, weight, power source, and included
accessories are not provided in the manual. Please refer to the
product packaging or contact the manufacturer for this
information.
INSTTRRUUCCTTIIOONNMMAANNUUAALL
TRANSFORMER JAW
V TERMINAL COM TERMINAL OHM TERMINAL
BARRIER
JAW TRIGGER RANGE SWITCH
POINTER LOCK BUTTON ZERO ADJUSTER
OHM ZERO ADJUSTER
SCALE PLATE
AC CLAMP METER WITH DC VOLT RANWGRISET STRAP
KEW SNAP 6 Easy to use and dependable, the instrument is especially
designed to incorporate as many safety features as possible.
KEAWC CSLNAAMPPM2E6T0E8R A
WITH DC VOLT RANGE
Easy to use and dependable, the instrument is especially designed to
incorporate as many safety features as possible.
KEW SNAP 6
KEW SNAP 2608A TRANSFORMER JAW
BARRIER
V TERMINAL
COM TERMINAL
OHCMaTpERMINAL
JAW TRIGGER
Protective fingerguard
RANGE SWITCH
POINTER LOCK BUTTON ZERO ADJUSTERÝ
PIZtErRoiOtseAcDatJUiOvpSeTHaEMfrRitngperroguv ai drdi n(gBaprrrioert)ec t Si oCAnLEaPgLAaTiEn s t electrical shock and ensuring the minimum
required air and creepage distances.
Cap
WRIST STRAP
Uncapped condition for CAT II environment
Capped condition for CAT III/ IV environments
The Cap shuld be firmly attached to the probes.
KYORITSU ELECTRICAL INSTRUMENTS WORKS, LTD
3. Specifications
Measuring Range and Accuracy
Ranges
Accuracy
AC current 6/15/60/150/300A
±3% of full scale
AC voltage 150/300/600V
±3% of full scale
DC voltage Resistance Temperature
60V x1 1k (25 mid-scale) x10 10k (250 mid-scale) -20-+150 ((wCiathn noopt tuiosnea.)l Model 7060)
±3% of full scale ±2% of scale length ±5(0-+100) ±10 (other ranges)
Overload Protection
Range AC6/15A AC60/150A AC300A AC150/300V AC600V DC60V ×1/×10
Maximum overload 60A AC for 10 sec 300A AC for 10 sec 360A AC for 10 sec 600V AC for 10 sec 720V AC for 10 sec 230V AC for 10 sec 230V AC, protection by fuse
Location for use :Indoor use, Altitude up to
2000m
Storage Temperature
and Humidity
:-10-+50,relative humidity up
to 75% without condensation
Operating Temperature and Humidity
:0-+40, relative humidity up to 90% without condensation
Conductor Size Safety Standard Withstand Voltage Dimensions Weight PowerSource
Accessories (included)
(optional)
:Approx. 33mm diameter max. :IEC61010-1CAT.IIIIII330000VV CAT.III660000VV IEC61010–20-03311 IEC61010-2-032 :347700VVACACforf5orsec1ondms inute between electrical circuit and housing cases or metal parts of jaws :193(L)×78(W)×39(D)mm :Approx.275g(battery included) :R6P(DC1.5V) battery or equivalent :Instruction manual Carrying case Model 9052 Test leads Model 7066 A R6P battery(included within the instrument) Two 0.5A/62500VV fuses(spare fuse included) :Mooddeell80048,(8M0u0l8t(iM-Turaltni-)T ran) Model7060(temperature probe)
1. Safety Warnings
This instrument has been designed and tested according to IEC Publication
61010: Safety Requirements for Electronic Measuring Apparatus. This
instruction manual contains warnings and safety rules which must be observed
by the user to ensure safe operation of the instrument and retain it in safe
condition. Therefore, read through these operating instructions when using the
instrument.
WARNING ReRadeathdrtohurgohuganhdaunndduenrdstearnsdtaindstructions
conintasitnreudctiinontshicsomntaaninueadl winhethnisusminagnutahle
insbtreufomreenst.tartingusingtheinstrument. Saveandkeepthemanualhandyto
Saveenanbdlekqeueipckthreemfearneunaclehwanhdeynetoveernable
quincekcreesfsearernyc. e whenever necessary. T heinstrumentistobeusedonlyin
Beitssurientteonduesedtahpepliincsattriuomnse.nt only in its inteU
nnddeedrstapnpdliacnadtiofonlslowanadlltthoe follow
mesaasuferteymiennsttrpurocctieodnusrecsondteasicnreibdedininthtehe
mamnuaanlu. al. F ailuretofollowtheinstructions
Bemsuaryectaouusnedienrjsutrany,diannsdtrfuomlloewntall safety
insdtraumctaiognesacnodn/toarindeadminagtheetmoaenquuaipl.ment
undertest.Kyoritsuisbynomeans
Failliuarbeletofofrolalnoywdtahmeaagbeorveesuinltsitnrgucftrionms
matyhecainussteruimnjeunrty,inincostnrturmadeincttiodnamtoage
andt/hoisrdcamutaiogneatroyeqnuotipem. ent under test.
The symbol indicated on the instrument means that the user must refer to
related parts in the manual for safe operation of the instrument. Be sure to
carefully read the instructions following each symbol in this manual.
DANGERis reserved for conditions and actions that are likely to cause serious
or fatal injury. WARNING is reserved for conditions and actions that can cause
serious or fatal injury. CAUTIONis reserved for conditions and actions that
can cause injury or instrument damage
FFoolllloowwiinnggsymsybmolsboalrseusaerdeonutsheedinsotrnumtehnet
ainnsdtriunmtheentinasntrducintiotnhemainnustarlu.ction manual.
AAtttteennttiioonnsshhoouulldd bbee ppaaiiddttooeeaacchhssyymmbboollttoo e#
ennsmndroRITcIcIbsRrTcIrunnneaoaunoaneehaedurtihrdedrmtrfudnfieiiereisfeeieniscsbitgsfcecetarrcofoaeyoasayargocsarstrrtporttoptcoeyeretoyceyoetoeuesanuemcfe.mo,snserrtdusertdwbdhtcsrhubtsdtshisoaoheohcehieantoaaalnnnitoinegnaciltffoinapsinedenthsrtlstuieiistttshiusnonsuitiyhrlmynttsnhsaecsrlsir..staatjswtutmtiomuhtuiirsorrtocnherncruauiikartteunnosayircsemiarinnstmnokl.oktdprotse.inniepuwerernotmdwnsulrndcnmjhiieumhutntctniseaianeemarsewoswnnrxbywitnnutnuehertlthiierouatnemttmteihtrlmcohtnrMhhaeeagmsieetecnndeoanhdnaauasositmautoncsashtsnuuuardulsaeeusarbuvlbarcrnymianoemlmlslnumeueollammpo.tagatmebsauelgoaeoonog.tnsepaslrrornettt..
vInodltiacgaetescoArrCes(pAolntedrinnagtintogCthuerraepnptl)i.cable
MIndeiacsautreesmDeCnt(cDaitreegcotrCy,uwrrheincth).is marked
nInedxitctaotethsiAs sCymanbdolD. C.
TInhdisicinasttersumAeCnt(sAatltisefriensatthinegmaCrukirnrgernetq)u. irement
dInefdinicedatienstDheCW(EDEirEecDtirCeuctrirveen(t2)0.02/96/EC).
TInhdisicasytemsboAlCinadnicdatDesC .separate collection for
electricalandelectronicequipment.
DDAANNGGEERR
atvilnehbaaNNDTtvilaDTssepoenbhhadterrooaeosvorepopaavdturvevtronrrnnoeneepmtretu6sosororroasfm6f0rtetteoooneh0a0mnmesrarrdnean0VenetmmtatucaVndttxemtckeeeAkscuexepimremteesArraCpml.otcoayljCpj..fpomouyamsatOsftci.ifwwetiolacetOoataftonauahnltumottsuai.ssi.henmpupsmuemedmrsresrwarmeseaaewakrspaiamkbmrsepirabeeltseeaeerleeem,nrekdmsn,kdgtttieetenhgiet.aasnhasaogesoIisgeissngs,fgun,usuenwnurerewasseeessaehqe,hddem,muicifccicfouineneophuihrfnmofrnommccttctttcteueuehahatitnisisinnoennoettt,,
shuonrdtetrh etesctirchuasit euxnpdoesredtecsotn. dIuf cetqivueippmaretnst,
unhdoewrevteers,texhtarsaperxepcoasuetidoncsohnoduuldctbievetapkeanrttso,
homwineivmeirz,eetxhterapopsrseibcialiuttyioonfsshhoorutilndg.be taken toN
mevinerimaitztemthpetptoosussiebitlihteyoinfssthruomrteintg.ifits
sNuerfvaecreaottreymoputrthoaundseisthweeitn. strument if its
suDrofacneootreyxocueredhatnhdeismwaexti.mum allowable
iDnpoutnooftaenxycemeedastuhreemmenatxirmanugme. allowable
inNpeuvteorfoapneynmtheeasbuartetemryenctormapnagretm. entcover Otdc
caTiOthanhaobiTtNaTwreuNNTtmntehvanhbeshhrtehieeneaeeoeennaeenvrergvvsdnrrnnsodeweefew.idmwtisorirnenmrrrefniimrhodoussasdssamtterrtmetleoatkrrrram,rn,yeepliysuuspsunceanraeaemmpmmgrctotpftfnrjloeeaeooniapeimdejcnwttoankdlnonymatyetimuwdihseictttanstisasiaifaesfioogitusnokuukstopnri,innernedmtorsboettceccsooornnawamocemtmt,sestissarioaobeboteos,eesrsenesnnceanenakurtuisoasssss,,y.oruiunlutcuusearsndeshcerrimnecienqonqheetddoodwtujiemumceuoatminiaoeooednprosipestdnnrpnsynp.nn.psk..ltaedletytbbym,rdrdiirrtituaiffmioionnmwyowkakaaeneiinienienbtntnttshdnyhsynestt.
Vinesrtirfuymperonpterdaompeargaetioonr osneraiokunsowpnerssoounrcael
binejfuorryem uasyeboercatuaskeindg. action as a result
indicationoftheinstrument. Keep your fingers and hands behind the
barrierduringmeasWurAeRmNenINt.G
taemirnhebxaifNiDmNDsopnneopsnteoosdasroeorovctdtiiuvsrerrfrrrniniumeearmfeucooimocidrpnamtcreakataetasatmneelriinittititodnertnnetttrieensy-
mtoustccothtmontaapmaaratatsltl# olopKssleitrleplbtuoteyanaaKostrstc-onhnrW
oaututchyytrdmhtebba.oimietaoAslsesrsikitxttnaabinhuiiireRtpsk.ttrnsuasuouoaestacNusstrtttrnteeceauiruyIdooyrnumrNpnrapoaymmmaa.ceaulGeeerymnkrnraattnotyestasbesdtu.ludn.aiorropsosrertaRrudRmrscmrmeriittaeebmeastr.nstaultumuauiterkrtt,kcrioenyebetninrufuasattafrutnth,ntohnaohcdiyeryfehlre
DDoo nnootttrtryy ttoo rreeppllaaccee tthhee bbaatttteerriieessiifftthhee
susurfrafcaeceofotfhtehiensintrsutrmuemnetnitsiwsewt.et. RReemmoovvee tthhee
tteessttleadsffrroommtthheeininssttrruummeenntt
bebfeofroereopoepennininggththeebboottttoommccaseffoor bbaatttteerryyoorr
fufsuesererpelpalcaecmemeenntt..
dSatmopaguesdinagndthteheteisntnelCeraAmdUeiTtfaIOlthNoerocuotleorr
jacket jacket
is is
expMosaekde. sure that the range switch is set to an
Daoppnrootprtiuatrenp tohseitiofnunbcetfioornemsealkeicntgormesawsiutrcehm wenitth.
pluAgglwedaysinm taeksetslueraedtso icnosnenret cthtedpltuog othf eeacihrcteusitt unledaedrtfeusllty. into the appropriate terminal on the
instrument. When the
inst#rumCeAntUwTilIlOnNot
be
in
use
for
a
lretoheliWlMaAxneonDUemptlasnarghpwsioteodkgnrereamveosnnupyfipptnomaeeusrserttgruleuhtrilmideemanyoriteamtaxhotdeeem.kpdieipnnenotppobetthsoso.osfrateafscuartDtittlttrtuttoeuihheoimimtrtoremehhientneeeeoesiaons,nab,pt.toirsneppnuprtafdwslrrlsonadueaoeirgcedrlmcpeeelawteerebmtitnseafnirihaotawratttelktgesliti.iiietponnvnbtcnlegeetuhthsssrgmttmiefioonoooserirdrrafnausaissaregcosguelealeleterecvcoetaehtnfmaannsooffteiuttrttnesenhnaes.grtnear,t.
removingthebatteries. DMoeansoutreemxpeonstec athteeg ionrsietrsu(Omveenrt-
vtoolttahgee dciarteecgtorsiuens),
eTxotremenestuermepersaatfueresoporerdaetwiofnall.of measuring
Tinhsitsruinmsternutms,enItE iCsn6’1t0 d1u0ste &stawbalitsehre psrosoaffeedt.y
Ksteaenpdaawrdaysffroormvadruisotuasnedlewcattreicr.al environments,
MiUicmHennlasssaeieetttgkcreraheutaugsrmemuiosdcrereru-
aaneninrmltzmtue.eepmDbndetecvonobfliatoeornrostrrcoheenetamCdmtuameosnAegaevdcnTkoeaatidrsnbtiteegrehwttagseeocsio.rtuihgrvtrCieeeregsnsAestrtnoTetrflacoemtsoraeoe,rdrlcarvsalmeseenunsfaotrprdnsmeooi.mnmnecgednanltttlathhe.rteedoy
Meenaesrugrye,mseonat cmateeagsourrieinsg(Oivnesrt-
rvuomltaegnetcdaetesgigonrieeds)for TCoAenTsure seanfveiorpoenrmateionntsof
mcaenasuerinndguirnestrugmreeanttesr,
IEmCo6m1e0n1t0areysetanbelrigshyetshsaanfeotnyestdaensdiganrdesdffoorrvCaAriTous
ealnCedActTcra.icllae:ldeSamnenvceoiarnAosdunCarmreyemeneletelnesc,cttctrcraiiaccttaeaelglgocoriorriiuzceeutsldie.tstascoothntrnooeCucAgtehTdIVtoa,
O(None, Other) traCnirscfourimtsewr ohricshimairlaerndoetvidceir.ectly
CCAATT.II :
cpPooreePcniwomqlrneeniueacrmncritcpyetraoecmirdrcetdyelaee.tdlnoecotlttreaouicncctottalhrAnleeiCtnccmiaebrelccyalutceieanicitrdtssprciptuocoooawiftlwaseoneerourqftAculseoCiuptrpmdbp.yelnya.t
CCAATT.III: PrPimraimryaerlyecetlreiccatlrciciraclucitisrcouf
itthseoefqtuhipement
cppoaaedtnnnhqieenseullte,ritdcopiatibmesnoutdudrettiinlbfoedetutniesrct.deopioecnartnnlnsyepeflcar,totnoaemendldtdhtthofeiereoeeducddtiitessllyrtetrrstiitsbbof.ruuottthiimooenn
CAT.IV: ThTehecircciurcitufirtofmrotmhethseervsiecrevidcreopdrtopthe
setrovictheeensetrravniccee, eanndtrtaontchee, apnowdetromtheeter
anpdowperrimmaeryter oavnedr-cpurrirmenatry protection
deovviceer(cduisrtrriebnuttiopnroptaencetl)i.on device
(distributionInpcoamningewli)re.
Interior wiring
CAT.
CAT.
TransformerCAT.
SocketO:
DevicCeATw.hich not directly
is
connected to the
mains power supply
2. Features
Tear-drop-shaped jaws for ease of use in
2cr.owFdeeadtcuabrelesareas and other tight places Safety design throughout.
Designed to CAT.III
T3e0a0rV-d/rCopA-sTh.IaIpe6d0j0aVwsafnodr epasoellouftiuosneindcergorwedeed2
csapbeleciafireedasbayndthoethienrtetirgnhattipolnaacel
ssSaafefetytysdteasnidganrd tIhEroCu6g1h0o1u0t. Designed to CAT III 300V/CAT
II 6D0C0V avnodltpagolelutiroanndgeegreeesp2escpieaclliyfiedubsyeftuhle for
iDafncPleiChpgtaoeotehivruncwtontrkoleaeetirrtrnaifoghgoslonuearacarperkdlppaa-
solntsfayoweygf-
afeeerrotetreueyraasrsdesdueptmiplaenofpnceogcildrraaygiatlnefilerooyadndnrsuciIysemEysmCerlufyee6suarel1ldiggf0Pieonh1onrgt0icnco.yihtrneuercsdkleiomincklgy
hUasrdes-tos-rheraodudloecdatitorannssUfosrems sehrrojauwdsedto further
tirmanpsrfoovremsearfejtayws to further improve safety
OOpptitoionnaalltetmempepreartauturereprporboebfeorfoter mtepmerpaeturarteure
mmeeaassuurreemmeenntt
4. Preparation for Measurement
41Releasing Pointer Lock Slide the pointer lock button to the right position
to unlock the pointer.
42Meter Zero Adjustment Set the pointer at the center of the
“0”mark on the scale by rotating the zero adjuster with a screw driver.
POINTER LOCK BUTTON
POINTER
ZERO ADJUSTER
43Checking Battery Voltage a. Set the range switch to the “×1”
position. b. Insert the red test lead into the OHM
terminal and the black test lead to the COM terminal. c. With the test leads
shorted, try to set the pointer over the “0” mark at the right end of the
resistance scale, using the OHM zero adjuster. d. When the adjustment brings
the pointer over the “0” mark, proceed to measurement. If not, replace the
battery. (See section 7 for battery and fuse replacement.)
NOTE The battery is needed only in resistance
measurement. AC/DC voltage and AC current measurement can be made without the
battery. If the instrument does not operate properly even after battery
replacement, check the fuse and test leads.
5. Measurement
51AC Current Measurement
WARNING Do not make measurement on a circuit
above 600V AC. This may cause shock hazard or damage to the instrument or
equipment under test. Transformer jaw tips are designed not to short the
circuit under test. If equipment under test has exposed conductive parts,
however, extra precaution should be taken to minimize the possibility of
shorting. Do not make measurement with the bottom case removed. Do not make
current measurement with the test leads connected to the instrument. Do not
measure current exceeding the limit for the overload protection feature. Keep
your fingers and hands behind the barrier during measurement.
CAUTION When the order of the current under
test is not known, make measurement first on the highest 300A range, then
switch to the appropriate range.
a. Set the range switch to the “AC 300A” position.
b. Press the trigger to open the transformer jaws and clamp onto one conductor
only. It is recommended that the conductor be placed at the center of the
closed jaws.
CORRECT
WRONG
c. Take the reading on the 300A current scale.
d. Set the range switch to the appropriate position based on the reading.
e. Take the reading on the appropriate scale.
Range 6 15 60 150 300
Scale used 60 150 60 150 300
Multiply reading by ×0.1 ×0.1 ×1 ×1 ×1
NOTE During current measurement, keep the
transformer jaws fully closed. Otherwise, accurate measurement cannot be made.
The maximum conductor size is 33mm in diameter. When measuring a larger
current, the transformer jaws may buzz. This is not a fault and does not
affect the accuracy at all.
5-2 AC Voltage Measurement
WARNING Do not make measurement on a circuit
above 600V AC. This may cause shock hazard or damage to the instrument or
equipment under test. Do not make measurement with the bottom case removed. Do
not apply to the instrument voltage exceeding the limit for the overload
protection feature. Keep your fingers and hands behind the barrier during
measurement.
a. Set range switch to an AC voltage position. When the order of the voltage
under test is not known, set the range switch to the highest 600V range.
b. Insert the red test lead to the V terminal and the black test lead to the
COM terminal.
c. Connect the tip of the test leads to the circuit under test.
RED TEST LEAD RED TEST LEAD
BLACK TEST LEAD
BLACK
dTE.STTaLkEAeD the reading on the scale for the selected range.
de. ATaftkeer thceomrepaledtiinngg onthethemsecasleurfeomr etnhte,
sremlecotveed rthanegete.st leads from the circuit
e. Aunftdeerr tecsotm. pleting the measurement, remove the test leads from the
circuit
Raunngdeer test. Scale used Multiply reading by
RACan1g5e0
1S5c0ale used M×u1ltiply reading by
AACC310500
310500
××11
AACC630000
63000
××110
5AC6300DC Vol6t0age Measur×em10ent
53DC VoltagWe AMReNasINuGrement Daboovneot60mVaDkeCm.WeAaRsuNreINmGent on a
circuit DDoo nnoottmmakakeemmeaesausruermeemnetnotnwaitchirctuhiet
baboottvoem60cVasDe Cre.moved. KDeoepnoytoumr
afiknegemrseaansdurheamndesntbewhiitnhd tthhee
bbaorttroiemr dcuasreinrgemmeoavseudr.ement. Keep your fingers and hands
behind the a. Sbaertrierradnugrieng msweaitscuhremteont. the DC
60V position. ab. SInestert rtahnegreed tsewstitlecahd totothe tVheterm”iDnaCl
6an0Vd”pthoesitibolna.ck test lead to the COM b. tIenrsmerint athl.e red test
lead to the V terminal c. aCnodnntehcet bthlaeckretdesteslet aldeatdo ttihpe
toCOthMe
pteorsmitinveal.side of the circuit under test and c.
tChoennbelacctkthteestreledadtetsipt lteoadthetipnetgoattihve
spiodsei.tive side of the circuit under test and
d.Tthaekebtlhaeckretaedsitngleoand thtiep 6t0oVtshcealen.egative e. Asifdte.r
completing the measurement, d.Treamkeovtheetrheeadtiensgt olenatdhse
6fr0oVmsctahle. circuit e. Aunftdeerr tecsotm. pleting the measurement,
5r4emoRveesitshteantceestMleeaadssurfermomentthe circuit
under test.
54ResistanceWMAReaNsINurGement bDoottonmotcamseakreeWmmoAevReaNsdu.INreGment
with the MDoakneost umreaktheatmtehaesruereismneont vowlittahgethine
bthoettcoimrcucaitsoerreeqmuoipvmede.nt under test.
MKeaekpeysouurrefinthgaetrsthaenrdehiasnndos bveohltiangdethine
tbhaerrcieirrcduuitroinrgeqmueipasmuernemt uenndt.er test. Keep your fingers
and hands behind the
barrier during measurement.
72Fuse Replacement
7cTi2hrceuFiitnuisssterpuRrmoetepenlcattecsdermbeeysinsatta0n.5cAe
/m25e0aVsufruisneg.
WThheenintshtreuminesntrtusmreenstisdtaonecsenmoteaospuerriantge
cpirrocpueitrliys pirnotercetseidstbayncae 0.m5Ae/a62s05u0rVeVmfuesnet.,
Wchheecnk tthhee infusstreumanednt rdeopelascenoitt oipnertahtee
pforlolopweirnlyg sitnepsr,ewshisetnannceecesmsaerays.urement, a.
Rcheemcokvethethfeuseteasntd lreeapdlasce friot min tthhee
finolsltorwuminegnstt.eps, when necessary. ab.. RReemmoovvee thethsecretwesotn
thleeabdasck fsriodme of tthhee ibnostttroummceanste. to open the instrument.
bc.. RReepmlaocvee tthheesbcrloewwnonfutshee bwaicthk stihdee osfptahree
bfuostetoimnstcaalsleedtobeonpeeanththteheinbstartutemrye.nt. cd..
RSecrpelawcethtehebobtlotowmn cfuassee wbaitchk thoentosptahree
finussteruinmsteanllte. d beneath the battery. d. Screw the bottom case back
onto the instrument.
CAUTION
Make from
tshueretetromCirAneUamlTsoIvOweNhtehne
test leads resistance
mMeaakseurseumreenttoisroevmeor.veIfththee tteesstt lleeaaddss
farroemleftthiensetertremdintoalsthewhteernmirneaslsis, ttahnecier
minaedasvuerretemnetnst hisorotivnegr. caIfntheexhteasutstleathdes
abraetteleryft. inserted to the terminals, their
inadvertent shorting can exhaust the
a. Sbeattttehrey.range switch to the ×1 or ×
10 position. ba.. ISnestetrhtetrhaengreedswtitecsht tloeathde t”o×1the”
OorH”M×
1te0rm”inpaol saitniodn.the black test lead to the b. CInOseMrttetrhmeinraeld.
test lead to the OHM c. Wteritmhintahl
eantdesttheleabdlasckshteosrttelde,adsetto tthhee
CpoOinMtetreormveirnathl.e 0 mark at the right end c. Wof itthhe
threesistteasntcelesacdasle,shuosrintegd,theseOt HthMe
pzeorinotaedrjouvseterrt.he “0” mark at the right end of the resistance scale,
using the OHM zeroBaLdAjCusKteTrE.ST LEAD
BLACK TEST LEAD
RED TEST LEAD RED TEST LEAD
d. Connect the tip of the test leads to the circuit under test.
ed.. TCaoknenethcte trheeadtiinpgoof nthtehetersetsilsetaadnscetosctahlee
acnirdcumituultnipdleyritteasst.follows.
e. Take the reading on the resistance scale
RaanngdemultipSlycaitleasusfoedllowMsu.ltiply reading by
R×a1nge
Scale used M×u1ltiply reading by
××110
××110
×10
×10
8. Optional Accessaries
M8.oOdepl 8t0io04naanldA8c00c8e(sMsualtrii-Tersans)
Multi-Trans extend existing current Mmeoadseulr88in00g004r8aa(nnMgdeu8lu0t0ip- 8Tto(rMa3nu0)l0ti0-TAraasnsw)ell as the Mmauxlitmi-Turmancsondeuxctteonr dsize.existing current measuring range up to 3000A as well as the maximum conducWtoAr RsiNzeIN. G
Daboovneot60m0aVkAeCmW. eAaRsuNrIeNmGent on a circuit DThoentorat
nmsfaokremmerejaaswusreamreenmtaodne aofcmirceutaitl
aabnodvteh6e0ir0VtipAsCa.re not insulated. Never
Ttohuechtratnhsefoermxpeorsjeadwsmaeretaml padaretsofumnedtearl
atensdt wthitehirjatwipstipasr.e not insulated. Never tDooucnhotthme
aekxepomseedasmureetmalenpat rwtsituhndtheer
tbeostttowmithcajaswe rteipms.oved. DDoo
nnoottmmakaekememaseuarseumreemntewntithwtihthe ttehset
bleoatdtosmcocnanseecrteemd toovethde. instrument. Do not make measurement
with the test
leads connectedCAtoUtTheIOinNstrument.
When test is
the not
okrndoCewrAnUo, fTmItOahkeNe
current under measurement
fWirhstenonthteheorhdiegrheosft t3h0e0Acurrraenngte,unthdeenr
tsewsittcihs tnoottheknaopwprno,pmriaatkeeramnegaes.urement
first on the highest 300A range, then
a. Ssewt tihtcehratongteheswaiptcphrotportihaetedreasnirgeed. position.
b. Clamp Kew Snap 2608A onto the pick-up coil a. Soef tMthueltir-
aTnrgane.switch to the desired position. bc.. CCllaammpp KMewulStin-
Tapra2n608oAntoontothtehe bpuicsk-b-uapr cooirl
ocof nMduulctti-oTrruann.der test. cd.. CTalakme ptheMruelatid-
TinrganonoKnteow tShneapb2u6s0-b8Aar anodr
cmounldtiupclytoirt buynd1e0r. test. d. Take the reading on Kew Snap 2608A and
multiply Models
MMooddeells8004
it Mbayx. 1co0n.ductor size M(mamx.incdoinadmuecteor) s(mizme6in0diameter)
Measuring range A1Mra0Cen00ag0seAuring
Input to output ratio Input to o1u0t:p1ut ratio
Model80048 6100
AC0 31000A
10:1
MMooddeel8l0708060(D10is0continA3u0Ce00d0A)
10:1
Model 7060 is an temperature probe with a Mmeoadseulr7in0g60range from 20 to 150. Model 7060 is an temperature probe with a
measuring range fWroAmRNI2N0G to 150.
Do not with a
vmoaltkaegemmeaWosruAerRetmhNaIeNnnGt3o0nVaAmCetoarl
part 60V
DDoC.not make measurement on a metal part
with a voltage more than 30V AC or 60V
DC.
6. Using Pointer Lock T6h. eUpsoiningterPlooicnkteferatLuoreckcan be used for
measurement in dimly light or hard-to-read Tlohceatipoonisn.ter lock feature
can be used for measurement in dimly light or hard-to-read lao.cMataikoensA. C
current, AC voltage, DC voltage
or resistance measurement as described a. MinaskeectAioCn c5u. rrent, AC
voltage, DC voltage
or resistance measurement as described in section 5.
POINTER LOCK BUTTON PBOUTINTTOENRLLOOCCKK
RLOECLEKASE RELEASE
b. Slide the pointer lock button to the left position.
bc.. TSlaidkee thtehepoirneteardilnogck abwutatyon ftroomthe tlhefet
pcoonsidtiuocnt.or or the circuit under test.
cd.. TTaokereletahsee trheeadpinogintearwlaoyck,frsolimde tthhee
cbounttdouncttoorthoer rthigehct.ircuit under test.
d. To release the pointer lock, slide the button to the right.
7. Battery and Fuse Replacement 7. Battery andWFAuRsNeINRGeplacement
sTuoreavtooidremeleocvteWritcAhResNhtoeINsctkGlehaadzsarfdro, mmathkee
iTnostrauvomidenet lebcetrfiocreshtorcykinhgaztaordr, emplaakcee
sbuartteertioesr.emove the test leads from the iMnsatkreumsuenret
tboefsocrreewtrtyhinegbotottomrepclaacsee
bbaatctkeroienst.o the instrument after battery
oMrafkuseesruerpelamtoemscernetw. the bottom case bDaocknootnintostathllea
ibnastttreurmy eonrtfuasfetetrhabtadttoeerys
onrotfuhsaeveretphleamspeemciefinetd. rating. Do not install a battery or fuse
that does
not have the specified rating.
SCREW
SCREW
SPARE FUSE
FUSE
R6P BATSTPEARRYE FUSE
FUSE
R6P BATTERY
71Battery Replacement
7a.Rin1esmtruoBmvaeettnetrt.yheReptelasctemleeandts from the ab.. RReemmoovvee
thethsecretwesotn thleeabdasck fsriodme of tthhee
ibnostttroummceanste. to open the instrument. bc.. RReepmlaocvee
ththeescbreawtteorny thweitbhacak snideewofRt6hPe
bboatttteormy coarseeqtouiovpaelenntth. e iTnhsterunmewentb. attery c.
Rmeupsltacebethienstbaalltetedryinwitthhe aornieewntaRti6oPn
binadtticearytedorinesqidueivtahleenint.strTumheennte. w battery d. mScursetw
btehe inbsottatlolemd cianse thbeackoroiennttoattiohne
iinndstircuamteednitn.side the instrument. d. Screw the bottom case back onto
the
instrument.
a. Set the range switch to the TEMP (×10) position.
ab..SInetstehret rtahnegersewditcthestto tlheead”TEtoMtPh(e×1O0HM)”
tpeorsmitiionna.l and the black test lead to the
b. CInOseMrttetrhmeinraeld. test lead to the OHM c. Wteritmhintahl
eantdesttheleabdlasckshteosrttelde,adsetto tthhee
CpoOinMtetreormveirnathl.e 0 mark at the right end c. Wof itthhe
threesistteasntcelesacdasle,shuosrintegd,theseOt HthMe
zpeorinotaedrjouvseterrt.he “0” mark at the right end d.
oRfemthoeveretshiesttaenscteleasdcaslefr,omusitnhge ttehremOinHalMs. e.
zInesroeratdMjuosdteerl.7060 s red plug to the OHM d. tReermmoinvael
tahnedtetshtelebaldasckfropmlugthteottehremiCnOalMs. e. tIenrsmeritnMal.odel
7060’s red plug to the OHM f. Tteorumchinathl eantidp
tohfethbelatcekmppleurgatutorethperoCbOe Mto
ttehrempianratlu. nder test. fg..TWoauicthutnhteiltipthoef
trheeadteinmgpebreactuormeeprsotbaeblteo,
tthheenpatartkuentdheerrteeasdt.ing on the temperature g. sWcaaliet. until the
reading become stable,
then take the reading on the temperature scale.
1234567812345678…………….
LLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLL
MODEL2608
MODEL26081 1
9.After-sales Service
99.A1fteSre-nsdainlgefsorSReerpvaicr e
Make sure to provide description of the 9failu1re aSnedndyionugr
fnoarmRee,paadirdress and telephone nuMmabkeer. sPuarcek
ttohepirnosvtriduemednetsscerciputrieolny soof tthhaet fitailwuirlel nanodt
byeoudranmaamgee,dadddurreinsgs atrnadnsteploerpthatoionne nanudmfboerrw.
arPdacitktothyeouinrsdtrisutmriebnuttosre.curely so that i9tw2ill
nPoetriobedicdaCmaaligberdatidounring transportation
anIdnfoorrwdearrdtoitmtoaiynotuarindmisteraibsuurtoerm. ent accuracy,
9iti2srPeceorimodmicenCdaeldibrathtiaotn you forward the insIntruomrdeenr
ttotomaKinytaoirnitsmueaRsuerpeamirenCt eanctceurrafcoyr,
ictaliibsrarteiocnomonmceenedveerdy ytehaart. you forward the iNnOstTruEm: ent
to Kyoritsu Repair Center for caClibarliabtrioantioonncisecehvaerrgyeyde.ar.
NOTE:
Calibration is charged.
KYORITSU reserves the rights to change
specifications or designs described in this
KmYaOnuRaIlTwSitUhouret sneortviceesanthdewirthigohuttsobtoligacthioannsg.e
specifications or designs described in this
manual wKithYoOutRnIoTticSeUaEndLwEiCthTouRtIoCbAligLations.
INSTRUMENTS KWYOORRKIST,SLUTEDL. ECTRICAL NINo.5S-20T,NRakUanMe 2-cEhoNmTe, MSeguro-ku,
TWokOyo,R15K2-0S03,1LJaTpaDn .
NPhoo.5n-e20:,8N1a-3k-a3n7e232–0c1h3o1me, Meguro-ku, TFaoxky:
o8,11-35-23-702033-101J5a2pan PUhRoLn:eh:t8tp1:-/3/-3w7w23w-0.k1e3w1
-Itd.co.jp FEa-mx a: i8l 1: -i3n-f3o7@23k-e0w15-I2td.co.jp
UFaRcLto:rhietstp::/U/wwawjiwm.ake&wE-Ihtdim.coe.jp
E-mail : info@kew-Itd.co.jp
04-10
Factories : Uwajima & Ehime
92-1396C
LLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLL
80-41-810
9922–11339966CE
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>