The KEW SNAP 2608A Transformer Jaw is an AC clamp meter with a
DC volt range. It is designed to incorporate many safety features
for reliable and safe operation. The instrument is equipped with a
protective fingerguard, a range switch, and a barrier for
electrical shock protection. It also has a zero adjuster, a scale
plate, and a jaw trigger for easy usage. The transformer jaw has V
terminal, COM terminal, and OHM terminal for convenient
connections. The product is manufactured by KYORITSU ELECTRICAL
INSTRUMENTS WORKS, LTD.

Specifications

  • Measuring Range and Accuracy:

    • AC current: 6/15/60/150/300A

    • AC voltage: 150/300/600V

    • DC voltage: 60V x1, 1k (25 mid-scale) x10, 10k (250
      mid-scale)

    • Resistance: -20-+150 (with no temperature compensation)

  • Overload Protection:

    • Maximum overload: 60A AC for 10 sec, 300A AC for 10 sec, 360A
      AC for 10 sec, 600V AC for 10 sec, 720V AC for 10 sec, 230V AC for
      10 sec, 230V AC (protection by fuse)
  • Location for Use: Indoor use, Altitude up to 2000m

  • Storage Temperature and Humidity: -10-+50, relative humidity up
    to 75% without condensation

  • Operating Temperature and Humidity: 0-+40, relative humidity up
    to 90% without condensation

  • Conductor Size: Safety Standard Withstand Voltage

  • Dimensions: (Not provided in the manual)

  • Weight: (Not provided in the manual)

  • Power Source: (Not provided in the manual)

  • Accessories (included): (Not provided in the manual)

  • Accessories (optional): (Not provided in the manual)

Product Usage Instructions

1. Before using the KEW SNAP 2608A Transformer Jaw, make sure to
read and understand the safety warnings and instructions in the
user manual.

2. Connect the V terminal, COM terminal, and OHM terminal of the
transformer jaw to the appropriate electrical connections.

3. Ensure that the protective fingerguard is in place to prevent
electrical shock.

4. Use the range switch to select the desired measurement range
for AC current and voltage.

5. Adjust the zero value using the zero adjuster as needed.

6. The scale plate shows the measured values for easy
reading.

7. Use the jaw trigger to open and close the transformer jaw for
clamping around conductors.

8. The instrument is equipped with overload protection to
prevent damage from excessive current or voltage.

9. When using the instrument, ensure that it is within the
specified temperature and humidity range.

10. Follow proper storage conditions, including temperature and
humidity requirements.

11. The conductor size and safety standards should be considered
when using the instrument.

12. The dimensions, weight, power source, and included
accessories are not provided in the manual. Please refer to the
product packaging or contact the manufacturer for this
information.

INSTTRRUUCCTTIIOONNMMAANNUUAALL

TRANSFORMER JAW

V TERMINAL COM TERMINAL OHM TERMINAL

BARRIER
JAW TRIGGER RANGE SWITCH
POINTER LOCK BUTTON ZERO ADJUSTER

OHM ZERO ADJUSTER

SCALE PLATE

AC CLAMP METER WITH DC VOLT RANWGRISET STRAP

KEW SNAP 6 Easy to use and dependable, the instrument is especially
designed to incorporate as many safety features as possible.
KEAWC CSLNAAMPPM2E6T0E8R A
WITH DC VOLT RANGE
Easy to use and dependable, the instrument is especially designed to incorporate as many safety features as possible.
KEW SNAP 6

KEW SNAP 2608A TRANSFORMER JAW

BARRIER

V TERMINAL
COM TERMINAL
OHCMaTpERMINAL

JAW TRIGGER
Protective fingerguard
RANGE SWITCH
POINTER LOCK BUTTON ZERO ADJUSTERÝ

PIZtErRoiOtseAcDatJUiOvpSeTHaEMfrRitngperroguv ai drdi n(gBaprrrioert)ec t Si oCAnLEaPgLAaTiEn s t electrical shock and ensuring the minimum

required air and creepage distances.

Cap

WRIST STRAP

Uncapped condition for CAT II environment

Capped condition for CAT III/ IV environments

The Cap shuld be firmly attached to the probes.

KYORITSU ELECTRICAL INSTRUMENTS WORKS, LTD

3. Specifications

Measuring Range and Accuracy

Ranges

Accuracy

AC current 6/15/60/150/300A

±3% of full scale

AC voltage 150/300/600V

±3% of full scale

DC voltage Resistance Temperature

60V x1 1k (25 mid-scale) x10 10k (250 mid-scale) -20-+150 ((wCiathn noopt tuiosnea.)l Model 7060)

±3% of full scale ±2% of scale length ±5(0-+100) ±10 (other ranges)

Overload Protection

Range AC6/15A AC60/150A AC300A AC150/300V AC600V DC60V ×1/×10

Maximum overload 60A AC for 10 sec 300A AC for 10 sec 360A AC for 10 sec 600V AC for 10 sec 720V AC for 10 sec 230V AC for 10 sec 230V AC, protection by fuse

Location for use :Indoor use, Altitude up to

2000m

Storage Temperature

and Humidity

:-10-+50,relative humidity up

to 75% without condensation

Operating Temperature and Humidity

:0-+40, relative humidity up to 90% without condensation

Conductor Size Safety Standard Withstand Voltage Dimensions Weight PowerSource Accessories (included)
(optional)

:Approx. 33mm diameter max. :IEC61010-1CAT.IIIIII330000VV CAT.III660000VV IEC61010–20-03311 IEC61010-2-032 :347700VVACACforf5orsec1ondms inute between electrical circuit and housing cases or metal parts of jaws :193(L)×78(W)×39(D)mm :Approx.275g(battery included) :R6P(DC1.5V) battery or equivalent :Instruction manual Carrying case Model 9052 Test leads Model 7066 A R6P battery(included within the instrument) Two 0.5A/62500VV fuses(spare fuse included) :Mooddeell80048,(8M0u0l8t(iM-Turaltni-)T ran) Model7060(temperature probe)

1. Safety Warnings
This instrument has been designed and tested according to IEC Publication 61010: Safety Requirements for Electronic Measuring Apparatus. This instruction manual contains warnings and safety rules which must be observed by the user to ensure safe operation of the instrument and retain it in safe condition. Therefore, read through these operating instructions when using the instrument.
WARNING ReRadeathdrtohurgohuganhdaunndduenrdstearnsdtaindstructions conintasitnreudctiinontshicsomntaaninueadl winhethnisusminagnutahle insbtreufomreenst.tartingusingtheinstrument. Saveandkeepthemanualhandyto Saveenanbdlekqeueipckthreemfearneunaclehwanhdeynetoveernable quincekcreesfsearernyc. e whenever necessary. T heinstrumentistobeusedonlyin Beitssurientteonduesedtahpepliincsattriuomnse.nt only in its inteU nnddeedrstapnpdliacnadtiofonlslowanadlltthoe follow mesaasuferteymiennsttrpurocctieodnusrecsondteasicnreibdedininthtehe mamnuaanlu. al. F ailuretofollowtheinstructions Bemsuaryectaouusnedienrjsutrany,diannsdtrfuomlloewntall safety insdtraumctaiognesacnodn/toarindeadminagtheetmoaenquuaipl.ment
undertest.Kyoritsuisbynomeans Failliuarbeletofofrolalnoywdtahmeaagbeorveesuinltsitnrgucftrionms matyhecainussteruimnjeunrty,inincostnrturmadeincttiodnamtoage andt/hoisrdcamutaiogneatroyeqnuotipem. ent under test.
The symbol indicated on the instrument means that the user must refer to related parts in the manual for safe operation of the instrument. Be sure to carefully read the instructions following each symbol in this manual.
DANGERis reserved for conditions and actions that are likely to cause serious or fatal injury. WARNING is reserved for conditions and actions that can cause serious or fatal injury. CAUTIONis reserved for conditions and actions that can cause injury or instrument damage

FFoolllloowwiinnggsymsybmolsboalrseusaerdeonutsheedinsotrnumtehnet ainnsdtriunmtheentinasntrducintiotnhemainnustarlu.ction manual. AAtttteennttiioonnsshhoouulldd bbee ppaaiiddttooeeaacchhssyymmbboollttoo e# ennsmndroRITcIcIbsRrTcIrunnneaoaunoaneehaedurtihrdedrmtrfudnfieiiereisfeeieniscsbitgsfcecetarrcofoaeyoasayargocsarstrrtporttoptcoeyeretoyceyoetoeuesanuemcfe.mo,snserrtdusertdwbdhtcsrhubtsdtshisoaoheohcehieantoaaalnnnitoinegnaciltffoinapsinedenthsrtlstuieiistttshiusnonsuitiyhrlmynttsnhsaecsrlsir..staatjswtutmtiomuhtuiirsorrtocnherncruauiikartteunnosayircsemiarinnstmnokl.oktdprotse.inniepuwerernotmdwnsulrndcnmjhiieumhutntctniseaianeemarsewoswnnrxbywitnnutnuehertlthiierouatnemttmteihtrlmcohtnrMhhaeeagmsieetecnndeoanhdnaauasositmautoncsashtsnuuuardulsaeeusarbuvlbarcrnymianoemlmlslnumeueollammpo.tagatmebsauelgoaeoonog.tnsepaslrrornettt..
vInodltiacgaetescoArrCes(pAolntedrinnagtintogCthuerraepnptl)i.cable MIndeiacsautreesmDeCnt(cDaitreegcotrCy,uwrrheincth).is marked nInedxitctaotethsiAs sCymanbdolD. C. TInhdisicinasttersumAeCnt(sAatltisefriensatthinegmaCrukirnrgernetq)u. irement dInefdinicedatienstDheCW(EDEirEecDtirCeuctrirveen(t2)0.02/96/EC). TInhdisicasytemsboAlCinadnicdatDesC .separate collection for electricalandelectronicequipment.

DDAANNGGEERR

atvilnehbaaNNDTtvilaDTssepoenbhhadterrooaeosvorepopaavdturvevtronrrnnoeneepmtretu6sosororroasfm6f0rtetteoooneh0a0mnmesrarrdnean0VenetmmtatucaVndttxemtckeeeAkscuexepimremteesArraCpml.otcoayljCpj..fpomouyamsatOsftci.ifwwetiolacetOoataftonauahnltumottsuai.ssi.henmpupsmuemedmrsresrwarmeseaaewakrspaiamkbmrsepirabeeltseeaeerleeem,nrekdmsn,kdgtttieetenhgiet.aasnhasaogesoIisgeissngs,fgun,usuenwnurerewasseeessaehqe,hddem,muicifccicfouineneophuihrfnmofrnommccttctttcteueuehahatitnisisinnoennoettt,, shuonrdtetrh etesctirchuasit euxnpdoesredtecsotn. dIuf cetqivueippmaretnst, unhdoewrevteers,texhtarsaperxepcoasuetidoncsohnoduuldctbievetapkeanrttso, homwineivmeirz,eetxhterapopsrseibcialiuttyioonfsshhoorutilndg.be taken toN mevinerimaitztemthpetptoosussiebitlihteyoinfssthruomrteintg.ifits
sNuerfvaecreaottreymoputrthoaundseisthweeitn. strument if its suDrofacneootreyxocueredhatnhdeismwaexti.mum allowable
iDnpoutnooftaenxycemeedastuhreemmenatxirmanugme. allowable inNpeuvteorfoapneynmtheeasbuartetemryenctormapnagretm. entcover Otdc caTiOthanhaobiTtNaTwreuNNTtmntehvanhbeshhrtehieeneaeeoeennaeenvrergvvsdnrrnnsodeweefew.idmwtisorirnenmrrrefniimrhodoussasdssamtterrtmetleoatkrrrram,rn,yeepliysuuspsunceanraeaemmpmmgrctotpftfnrjloeeaeooniapeimdejcnwttoankdlnonymatyetimuwdihseictttanstisasiaifaesfioogitusnokuukstopnri,innernedmtorsboettceccsooornnawamocemtmt,sestissarioaobeboteos,eesrsenesnnceanenakurtuisoasssss,,y.oruiunlutcuusearsndeshcerrimnecienqonqheetddoodwtujiemumceuoatminiaoeooednprosipestdnnrpnsynp.nn.psk..ltaedletytbbym,rdrdiirrtituaiffmioionnmwyowkakaaeneiinienienbtntnttshdnyhsynestt. Vinesrtirfuymperonpterdaompeargaetioonr osneraiokunsowpnerssoounrcael binejfuorryem uasyeboercatuaskeindg. action as a result
indicationoftheinstrument. Keep your fingers and hands behind the barrierduringmeasWurAeRmNenINt.G
taemirnhebxaifNiDmNDsopnneopsnteoosdasroeorovctdtiiuvsrerrfrrrniniumeearmfeucooimocidrpnamtcreakataetasatmneelriinittititodnertnnetttrieensy- mtoustccothtmontaapmaaratatsltl# olopKssleitrleplbtuoteyanaaKostrstc-onhnrW oaututchyytrdmhtebba.oimietaoAslsesrsikitxttnaabinhuiiireRtpsk.ttrnsuasuouoaestacNusstrtttrnteeceauiruyIdooyrnumrNpnrapoaymmmaa.ceaulGeeerymnkrnraattnotyestasbesdtu.ludn.aiorropsosrertaRrudRmrscmrmeriittaeebmeastr.nstaultumuauiterkrtt,kcrioenyebetninrufuasattafrutnth,ntohnaohcdiyeryfehlre DDoo nnootttrtryy ttoo rreeppllaaccee tthhee bbaatttteerriieessiifftthhee susurfrafcaeceofotfhtehiensintrsutrmuemnetnitsiwsewt.et. RReemmoovvee tthhee tteessttleadsffrroommtthheeininssttrruummeenntt bebfeofroereopoepennininggththeebboottttoommccaseffoor bbaatttteerryyoorr fufsuesererpelpalcaecmemeenntt..

dSatmopaguesdinagndthteheteisntnelCeraAmdUeiTtfaIOlthNoerocuotleorr

jacket jacket

is is

expMosaekde. sure that the range switch is set to an

Daoppnrootprtiuatrenp tohseitiofnunbcetfioornemsealkeicntgormesawsiutrcehm wenitth.

pluAgglwedaysinm taeksetslueraedtso icnosnenret cthtedpltuog othf eeacihrcteusitt unledaedrtfeusllty. into the appropriate terminal on the

instrument. When the

inst#rumCeAntUwTilIlOnNot

be

in

use

for

a

lretoheliWlMaAxneonDUemptlasnarghpwsioteodkgnrereamveosnnupyfipptnomaeeusrserttgruleuhtrilmideemanyoriteamtaxhotdeeem.kpdieipnnenotppobetthsoso.osfrateafscuartDtittlttrtuttoeuihheoimimtrtoremehhientneeeeoesiaons,nab,pt.toirsneppnuprtafdwslrrlsonadueaoeirgcedrlmcpeeelawteerebmtitnseafnirihaotawratttelktgesliti.iiietponnvnbtcnlegeetuhthsssrgmttmiefioonoooserirdrrafnausaissaregcosguelealeleterecvcoetaehtnfmaannsooffteiuttrttnesenhnaes.grtnear,t.

removingthebatteries. DMoeansoutreemxpeonstec athteeg ionrsietrsu(Omveenrt- vtoolttahgee dciarteecgtorsiuens),
eTxotremenestuermepersaatfueresoporerdaetwiofnall.of measuring

Tinhsitsruinmsternutms,enItE iCsn6’1t0 d1u0ste &stawbalitsehre psrosoaffeedt.y Ksteaenpdaawrdaysffroormvadruisotuasnedlewcattreicr.al environments,
MiUicmHennlasssaeieetttgkcreraheutaugsrmemuiosdcrereru- aaneninrmltzmtue.eepmDbndetecvonobfliatoeornrostrrcoheenetamCdmtuameosnAegaevdcnTkoeaatidrsnbtiteegrehwttagseeocsio.rtuihgrvtrCieeeregsnsAestrtnoTetrflacoemtsoraeoe,rdrlcarvsalmeseenunsfaotrprdnsmeooi.mnmnecgednanltttlathhe.rteedoy

Meenaesrugrye,mseonat cmateeagsourrieinsg(Oivnesrt- rvuomltaegnetcdaetesgigonrieeds)for TCoAenTsure seanfveiorpoenrmateionntsof mcaenasuerinndguirnestrugmreeanttesr,
IEmCo6m1e0n1t0areysetanbelrigshyetshsaanfeotnyestdaensdiganrdesdffoorrvCaAriTous

ealnCedActTcra.icllae:ldeSamnenvceoiarnAosdunCarmreyemeneletelnesc,cttctrcraiiaccttaeaelglgocoriorriiuzceeutsldie.tstascoothntrnooeCucAgtehTdIVtoa, O(None, Other) traCnirscfourimtsewr ohricshimairlaerndoetvidceir.ectly CCAATT.II : cpPooreePcniwomqlrneeniueacrmncritcpyetraoecmirdrcetdyelaee.tdlnoecotlttreaouicncctottalhrAnleeiCtnccmiaebrelccyalutceieanicitrdtssprciptuocoooawiftlwaseoneerourqftAculseoCiuptrpmdbp.yelnya.t CCAATT.III: PrPimraimryaerlyecetlreiccatlrciciraclucitisrcouf itthseoefqtuhipement
cppoaaedtnnnhqieenseullte,ritdcopiatibmesnoutdudrettiinlbfoedetutniesrct.deopioecnartnnlnsyepeflcar,totnoaemendldtdhtthofeiereoeeducddtiitessllyrtetrrstiitsbbof.ruuottthiimooenn CAT.IV: ThTehecircciurcitufirtofmrotmhethseervsiecrevidcreopdrtopthe
setrovictheeensetrravniccee, eanndtrtaontchee, apnowdetromtheeter anpdowperrimmaeryter oavnedr-cpurrirmenatry protection deovviceer(cduisrtrriebnuttiopnroptaencetl)i.on device

(distributionInpcoamningewli)re.

Interior wiring

CAT.

CAT.

TransformerCAT.

SocketO:

DevicCeATw.hich not directly

is

connected to the

mains power supply

2. Features

Tear-drop-shaped jaws for ease of use in
2cr.owFdeeadtcuabrelesareas and other tight places Safety design throughout. Designed to CAT.III
T3e0a0rV-d/rCopA-sTh.IaIpe6d0j0aVwsafnodr epasoellouftiuosneindcergorwedeed2 csapbeleciafireedasbayndthoethienrtetirgnhattipolnaacel ssSaafefetytysdteasnidganrd tIhEroCu6g1h0o1u0t. Designed to CAT III 300V/CAT II 6D0C0V avnodltpagolelutiroanndgeegreeesp2escpieaclliyfiedubsyeftuhle for iDafncPleiChpgtaoeotehivruncwtontrkoleaeetirrtrnaifoghgoslonuearacarperkdlppaa- solntsfayoweygf- afeeerrotetreueyraasrsdesdueptmiplaenofpnceogcildrraaygiatlnefilerooyadndnrsuciIysemEysmCerlufyee6suarel1ldiggf0Pieonh1onrgt0icnco.yihtrneuercsdkleiomincklgy hUasrdes-tos-rheraodudloecdatitorannssUfosrems sehrrojauwdsedto further tirmanpsrfoovremsearfejtayws to further improve safety OOpptitoionnaalltetmempepreartauturereprporboebfeorfoter mtepmerpaeturarteure mmeeaassuurreemmeenntt

4. Preparation for Measurement
41Releasing Pointer Lock Slide the pointer lock button to the right position to unlock the pointer.
42Meter Zero Adjustment Set the pointer at the center of the
“0”mark on the scale by rotating the zero adjuster with a screw driver.

POINTER LOCK BUTTON
POINTER

ZERO ADJUSTER

43Checking Battery Voltage a. Set the range switch to the “×1”
position. b. Insert the red test lead into the OHM
terminal and the black test lead to the COM terminal. c. With the test leads shorted, try to set the pointer over the “0” mark at the right end of the resistance scale, using the OHM zero adjuster. d. When the adjustment brings the pointer over the “0” mark, proceed to measurement. If not, replace the battery. (See section 7 for battery and fuse replacement.)
NOTE The battery is needed only in resistance
measurement. AC/DC voltage and AC current measurement can be made without the battery. If the instrument does not operate properly even after battery replacement, check the fuse and test leads.

5. Measurement
51AC Current Measurement
WARNING Do not make measurement on a circuit
above 600V AC. This may cause shock hazard or damage to the instrument or equipment under test. Transformer jaw tips are designed not to short the circuit under test. If equipment under test has exposed conductive parts, however, extra precaution should be taken to minimize the possibility of shorting. Do not make measurement with the bottom case removed. Do not make current measurement with the test leads connected to the instrument. Do not measure current exceeding the limit for the overload protection feature. Keep your fingers and hands behind the barrier during measurement.
CAUTION When the order of the current under
test is not known, make measurement first on the highest 300A range, then switch to the appropriate range.
a. Set the range switch to the “AC 300A” position.
b. Press the trigger to open the transformer jaws and clamp onto one conductor only. It is recommended that the conductor be placed at the center of the closed jaws.

CORRECT

WRONG

c. Take the reading on the 300A current scale.
d. Set the range switch to the appropriate position based on the reading.
e. Take the reading on the appropriate scale.

Range 6 15 60 150 300

Scale used 60 150 60 150 300

Multiply reading by ×0.1 ×0.1 ×1 ×1 ×1

NOTE During current measurement, keep the
transformer jaws fully closed. Otherwise, accurate measurement cannot be made. The maximum conductor size is 33mm in diameter. When measuring a larger current, the transformer jaws may buzz. This is not a fault and does not affect the accuracy at all.
5-2 AC Voltage Measurement
WARNING Do not make measurement on a circuit
above 600V AC. This may cause shock hazard or damage to the instrument or equipment under test. Do not make measurement with the bottom case removed. Do not apply to the instrument voltage exceeding the limit for the overload protection feature. Keep your fingers and hands behind the barrier during measurement.

a. Set range switch to an AC voltage position. When the order of the voltage under test is not known, set the range switch to the highest 600V range.
b. Insert the red test lead to the V terminal and the black test lead to the COM terminal.
c. Connect the tip of the test leads to the circuit under test.

RED TEST LEAD RED TEST LEAD

BLACK TEST LEAD

BLACK
dTE.STTaLkEAeD the reading on the scale for the selected range.
de. ATaftkeer thceomrepaledtiinngg onthethemsecasleurfeomr etnhte, sremlecotveed rthanegete.st leads from the circuit
e. Aunftdeerr tecsotm. pleting the measurement, remove the test leads from the circuit
Raunngdeer test. Scale used Multiply reading by

RACan1g5e0

1S5c0ale used M×u1ltiply reading by

AACC310500

310500

××11

AACC630000

63000

××110

5AC6300DC Vol6t0age Measur×em10ent

53DC VoltagWe AMReNasINuGrement Daboovneot60mVaDkeCm.WeAaRsuNreINmGent on a circuit DDoo nnoottmmakakeemmeaesausruermeemnetnotnwaitchirctuhiet
baboottvoem60cVasDe Cre.moved. KDeoepnoytoumr afiknegemrseaansdurheamndesntbewhiitnhd tthhee
bbaorttroiemr dcuasreinrgemmeoavseudr.ement. Keep your fingers and hands behind the a. Sbaertrierradnugrieng msweaitscuhremteont. the DC
60V position. ab. SInestert rtahnegreed tsewstitlecahd totothe tVheterm”iDnaCl
6an0Vd”pthoesitibolna.ck test lead to the COM b. tIenrsmerint athl.e red test lead to the V terminal c. aCnodnntehcet bthlaeckretdesteslet aldeatdo ttihpe toCOthMe
pteorsmitinveal.side of the circuit under test and c. tChoennbelacctkthteestreledadtetsipt lteoadthetipnetgoattihve
spiodsei.tive side of the circuit under test and d.Tthaekebtlhaeckretaedsitngleoand thtiep 6t0oVtshcealen.egative e. Asifdte.r completing the measurement, d.Treamkeovtheetrheeadtiensgt olenatdhse 6fr0oVmsctahle. circuit e. Aunftdeerr tecsotm. pleting the measurement, 5r4emoRveesitshteantceestMleeaadssurfermomentthe circuit
under test.
54ResistanceWMAReaNsINurGement bDoottonmotcamseakreeWmmoAevReaNsdu.INreGment with the MDoakneost umreaktheatmtehaesruereismneont vowlittahgethine
bthoettcoimrcucaitsoerreeqmuoipvmede.nt under test. MKeaekpeysouurrefinthgaetrsthaenrdehiasnndos bveohltiangdethine
tbhaerrcieirrcduuitroinrgeqmueipasmuernemt uenndt.er test. Keep your fingers and hands behind the
barrier during measurement.

72Fuse Replacement 7cTi2hrceuFiitnuisssterpuRrmoetepenlcattecsdermbeeysinsatta0n.5cAe /m25e0aVsufruisneg. WThheenintshtreuminesntrtusmreenstisdtaonecsenmoteaospuerriantge
cpirrocpueitrliys pirnotercetseidstbayncae 0.m5Ae/a62s05u0rVeVmfuesnet., Wchheecnk tthhee infusstreumanednt rdeopelascenoitt oipnertahtee pforlolopweirnlyg sitnepsr,ewshisetnannceecesmsaerays.urement, a. Rcheemcokvethethfeuseteasntd lreeapdlasce friot min tthhee finolsltorwuminegnstt.eps, when necessary. ab.. RReemmoovvee thethsecretwesotn thleeabdasck fsriodme of tthhee ibnostttroummceanste. to open the instrument. bc.. RReepmlaocvee tthheesbcrloewwnonfutshee bwaicthk stihdee osfptahree bfuostetoimnstcaalsleedtobeonpeeanththteheinbstartutemrye.nt. cd.. RSecrpelawcethtehebobtlotowmn cfuassee wbaitchk thoentosptahree finussteruinmsteanllte. d beneath the battery. d. Screw the bottom case back onto the instrument.

CAUTION

Make from

tshueretetromCirAneUamlTsoIvOweNhtehne

test leads resistance

mMeaakseurseumreenttoisroevmeor.veIfththee tteesstt lleeaaddss

farroemleftthiensetertremdintoalsthewhteernmirneaslsis, ttahnecier

minaedasvuerretemnetnst hisorotivnegr. caIfntheexhteasutstleathdes

abraetteleryft. inserted to the terminals, their

inadvertent shorting can exhaust the

a. Sbeattttehrey.range switch to the ×1 or ×

10 position. ba.. ISnestetrhtetrhaengreedswtitecsht tloeathde t”o×1the” OorH”M×
1te0rm”inpaol saitniodn.the black test lead to the b. CInOseMrttetrhmeinraeld. test lead to the OHM c. Wteritmhintahl eantdesttheleabdlasckshteosrttelde,adsetto tthhee
CpoOinMtetreormveirnathl.e 0 mark at the right end c. Wof itthhe threesistteasntcelesacdasle,shuosrintegd,theseOt HthMe
pzeorinotaedrjouvseterrt.he “0” mark at the right end of the resistance scale, using the OHM zeroBaLdAjCusKteTrE.ST LEAD

BLACK TEST LEAD

RED TEST LEAD RED TEST LEAD

d. Connect the tip of the test leads to the circuit under test.
ed.. TCaoknenethcte trheeadtiinpgoof nthtehetersetsilsetaadnscetosctahlee acnirdcumituultnipdleyritteasst.follows.
e. Take the reading on the resistance scale RaanngdemultipSlycaitleasusfoedllowMsu.ltiply reading by

R×a1nge

Scale used M×u1ltiply reading by

××110

××110

×10

×10

8. Optional Accessaries

M8.oOdepl 8t0io04naanldA8c00c8e(sMsualtrii-Tersans)

Multi-Trans extend existing current Mmeoadseulr88in00g004r8aa(nnMgdeu8lu0t0ip- 8Tto(rMa3nu0)l0ti0-TAraasnsw)ell as the Mmauxlitmi-Turmancsondeuxctteonr dsize.existing current measuring range up to 3000A as well as the maximum conducWtoAr RsiNzeIN. G

Daboovneot60m0aVkAeCmW. eAaRsuNrIeNmGent on a circuit DThoentorat nmsfaokremmerejaaswusreamreenmtaodne aofcmirceutaitl
aabnodvteh6e0ir0VtipAsCa.re not insulated. Never Ttohuechtratnhsefoermxpeorsjeadwsmaeretaml padaretsofumnedtearl
atensdt wthitehirjatwipstipasr.e not insulated. Never tDooucnhotthme aekxepomseedasmureetmalenpat rwtsituhndtheer
tbeostttowmithcajaswe rteipms.oved. DDoo nnoottmmakaekememaseuarseumreemntewntithwtihthe ttehset
bleoatdtosmcocnanseecrteemd toovethde. instrument. Do not make measurement with the test

leads connectedCAtoUtTheIOinNstrument.

When test is

the not

okrndoCewrAnUo, fTmItOahkeNe

current under measurement

fWirhstenonthteheorhdiegrheosft t3h0e0Acurrraenngte,unthdeenr

tsewsittcihs tnoottheknaopwprno,pmriaatkeeramnegaes.urement

first on the highest 300A range, then

a. Ssewt tihtcehratongteheswaiptcphrotportihaetedreasnirgeed. position.

b. Clamp Kew Snap 2608A onto the pick-up coil a. Soef tMthueltir- aTnrgane.switch to the desired position. bc.. CCllaammpp KMewulStin- Tapra2n608oAntoontothtehe bpuicsk-b-uapr cooirl
ocof nMduulctti-oTrruann.der test. cd.. CTalakme ptheMruelatid- TinrganonoKnteow tShneapb2u6s0-b8Aar anodr
cmounldtiupclytoirt buynd1e0r. test. d. Take the reading on Kew Snap 2608A and

multiply Models
MMooddeells8004

it Mbayx. 1co0n.ductor size M(mamx.incdoinadmuecteor) s(mizme6in0diameter)

Measuring range A1Mra0Cen00ag0seAuring

Input to output ratio Input to o1u0t:p1ut ratio

Model80048 6100

AC0 31000A

10:1

MMooddeel8l0708060(D10is0continA3u0Ce00d0A)

10:1

Model 7060 is an temperature probe with a Mmeoadseulr7in0g60range from 20 to 150. Model 7060 is an temperature probe with a

measuring range fWroAmRNI2N0G to 150.

Do not with a

vmoaltkaegemmeaWosruAerRetmhNaIeNnnGt3o0nVaAmCetoarl

part 60V

DDoC.not make measurement on a metal part

with a voltage more than 30V AC or 60V

DC.

6. Using Pointer Lock T6h. eUpsoiningterPlooicnkteferatLuoreckcan be used for
measurement in dimly light or hard-to-read Tlohceatipoonisn.ter lock feature can be used for measurement in dimly light or hard-to-read lao.cMataikoensA. C current, AC voltage, DC voltage
or resistance measurement as described a. MinaskeectAioCn c5u. rrent, AC voltage, DC voltage
or resistance measurement as described in section 5.
POINTER LOCK BUTTON PBOUTINTTOENRLLOOCCKK
RLOECLEKASE RELEASE
b. Slide the pointer lock button to the left position.
bc.. TSlaidkee thtehepoirneteardilnogck abwutatyon ftroomthe tlhefet pcoonsidtiuocnt.or or the circuit under test.
cd.. TTaokereletahsee trheeadpinogintearwlaoyck,frsolimde tthhee cbounttdouncttoorthoer rthigehct.ircuit under test.
d. To release the pointer lock, slide the button to the right.

7. Battery and Fuse Replacement 7. Battery andWFAuRsNeINRGeplacement
sTuoreavtooidremeleocvteWritcAhResNhtoeINsctkGlehaadzsarfdro, mmathkee iTnostrauvomidenet lebcetrfiocreshtorcykinhgaztaordr, emplaakcee
sbuartteertioesr.emove the test leads from the iMnsatkreumsuenret tboefsocrreewtrtyhinegbotottomrepclaacsee
bbaatctkeroienst.o the instrument after battery oMrafkuseesruerpelamtoemscernetw. the bottom case bDaocknootnintostathllea ibnastttreurmy eonrtfuasfetetrhabtadttoeerys
onrotfuhsaeveretphleamspeemciefinetd. rating. Do not install a battery or fuse that does
not have the specified rating.
SCREW
SCREW

SPARE FUSE

FUSE

R6P BATSTPEARRYE FUSE

FUSE

R6P BATTERY

71Battery Replacement

7a.Rin1esmtruoBmvaeettnetrt.yheReptelasctemleeandts from the ab.. RReemmoovvee thethsecretwesotn thleeabdasck fsriodme of tthhee
ibnostttroummceanste. to open the instrument. bc.. RReepmlaocvee ththeescbreawtteorny thweitbhacak snideewofRt6hPe
bboatttteormy coarseeqtouiovpaelenntth. e iTnhsterunmewentb. attery c. Rmeupsltacebethienstbaalltetedryinwitthhe aornieewntaRti6oPn
binadtticearytedorinesqidueivtahleenint.strTumheennte. w battery d. mScursetw btehe inbsottatlolemd cianse thbeackoroiennttoattiohne
iinndstircuamteednitn.side the instrument. d. Screw the bottom case back onto the

instrument.

a. Set the range switch to the TEMP (×10) position.
ab..SInetstehret rtahnegersewditcthestto tlheead”TEtoMtPh(e×1O0HM)” tpeorsmitiionna.l and the black test lead to the
b. CInOseMrttetrhmeinraeld. test lead to the OHM c. Wteritmhintahl eantdesttheleabdlasckshteosrttelde,adsetto tthhee
CpoOinMtetreormveirnathl.e 0 mark at the right end c. Wof itthhe threesistteasntcelesacdasle,shuosrintegd,theseOt HthMe
zpeorinotaedrjouvseterrt.he “0” mark at the right end d. oRfemthoeveretshiesttaenscteleasdcaslefr,omusitnhge ttehremOinHalMs. e. zInesroeratdMjuosdteerl.7060 s red plug to the OHM d. tReermmoinvael tahnedtetshtelebaldasckfropmlugthteottehremiCnOalMs. e. tIenrsmeritnMal.odel 7060’s red plug to the OHM f. Tteorumchinathl eantidp tohfethbelatcekmppleurgatutorethperoCbOe Mto
ttehrempianratlu. nder test. fg..TWoauicthutnhteiltipthoef trheeadteinmgpebreactuormeeprsotbaeblteo,
tthheenpatartkuentdheerrteeasdt.ing on the temperature g. sWcaaliet. until the reading become stable,
then take the reading on the temperature scale.

1234567812345678…………….

LLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLL

MODEL2608
MODEL26081 1

9.After-sales Service
99.A1fteSre-nsdainlgefsorSReerpvaicr e
Make sure to provide description of the 9failu1re aSnedndyionugr fnoarmRee,paadirdress and telephone nuMmabkeer. sPuarcek ttohepirnosvtriduemednetsscerciputrieolny soof tthhaet fitailwuirlel nanodt byeoudranmaamgee,dadddurreinsgs atrnadnsteploerpthatoionne nanudmfboerrw. arPdacitktothyeouinrsdtrisutmriebnuttosre.curely so that i9tw2ill nPoetriobedicdaCmaaligberdatidounring transportation anIdnfoorrwdearrdtoitmtoaiynotuarindmisteraibsuurtoerm. ent accuracy, 9iti2srPeceorimodmicenCdaeldibrathtiaotn you forward the insIntruomrdeenr ttotomaKinytaoirnitsmueaRsuerpeamirenCt eanctceurrafcoyr, ictaliibsrarteiocnomonmceenedveerdy ytehaart. you forward the iNnOstTruEm: ent to Kyoritsu Repair Center for caClibarliabtrioantioonncisecehvaerrgyeyde.ar. NOTE:
Calibration is charged.

KYORITSU reserves the rights to change

specifications or designs described in this

KmYaOnuRaIlTwSitUhouret sneortviceesanthdewirthigohuttsobtoligacthioannsg.e specifications or designs described in this
manual wKithYoOutRnIoTticSeUaEndLwEiCthTouRtIoCbAligLations.

INSTRUMENTS KWYOORRKIST,SLUTEDL. ECTRICAL NINo.5S-20T,NRakUanMe 2-cEhoNmTe, MSeguro-ku,

TWokOyo,R15K2-0S03,1LJaTpaDn .
NPhoo.5n-e20:,8N1a-3k-a3n7e232–0c1h3o1me, Meguro-ku, TFaoxky: o8,11-35-23-702033-101J5a2pan PUhRoLn:eh:t8tp1:-/3/-3w7w23w-0.k1e3w1 -Itd.co.jp FEa-mx a: i8l 1: -i3n-f3o7@23k-e0w15-I2td.co.jp UFaRcLto:rhietstp::/U/wwawjiwm.ake&wE-Ihtdim.coe.jp

E-mail : info@kew-Itd.co.jp

04-10

Factories : Uwajima & Ehime

92-1396C

LLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLLL

80-41-810

9922–11339966CE