Safety precautions
This instruction manual contains warnings and safety rules which must be
observed by the user to ensure the safe operation of the instrument and retain
it in a safe condition. Therefore, read through these operating instructions
before using the instrument.
1-1. DANGER The connector for the thermocouple is not insulated. Do not
make measurements using a thermocouple when there is a risk of electrical
shock hazard.
1-2. WARNING KEW5515 is one of the portable laser applied instruments
which are regulated by the consumer product safety law.
- Do not look directly into the laser.
- Do not point the laser directly at persons or animals.
- When measuring a reflective object such as mirrors, take extra caution so as not to look at diffuse reflections.
- Keep out of reach of child or person who doesn’t have the Product handling knowledge.
- Use the Product only as specified. Do not disassemble or modify the Product.
1-3. CAUTION Please follow the precautions below for use to keep the instrument’s functions properly and to enable accurate measurement.
- Do not contact this product with a measuring object. This is a contactless thermometer. Contact with a hot section may cause unrepairable damages or inaccurate readings.
- Do not damage the measuring window (plastic lens). Do not let a foreign object penetrate the measuring window, or drop a hard object on the measuring window.
- This instrument applies a precise optical system.
- Do not give an excessive shock to the instrument. Do not bring the instrument closer to an electrostatic object.
- The emissivity of this instrument can be selected. When the emissivity setting is different from that of a measuring object, an inaccurate reading will occur.
- The sudden changes in the ambient temperature will cause inaccurate readings. Wait a while to let the temperature of the instrument stabilize, and then measure.
- Remove the battery from the instrument, if the instrument is not in use for a long period or is put in storage.
1-4. Environmental precautions
- Do not use or store the instrument in a location where the instrument is exposed to direct sunlight, dust, lampblack, and corrosive gas, or where the temperature and/ or humidity is high. Otherwise, the measuring window gets soiled or deteriorated, which causes inaccurate readings.
- This instrument is not water-proof. Do not use in water or liquid nor make measurements or store it where it may get wet.
- Keep the instrument away from an object which radiates a strong electromagnetic wave.
1-5. Maintenance
Enclosure
Clean it with a soft cloth. To remove heavy stains, wipe it with a cloth wrung
out with diluted neutral detergent.
Cautions: Do not use chemical agents such as thinner, benzine or alcohol
since these may cause a crack or fog on the surface of the measuring window,
and consequently it may change the infrared radiation transmittance of the
plastic lens.
1-6. Relation of distance and measuring diameter
The relation between the distance and measuring diameter is as follows.
However, the ratio between distance and measuring diameter varies depending on
the distance. The more the distance increases, the larger area is needed.
Features
This is a contactless handheld infrared thermometer. Temperature measurement
with a K-type thermocouple is also possible.
Alarm function
The upper and lower temperature limits can be set. The red blinking backlight
indicates that the measured value is below or over the pre-set limits.
Auto-power-off function
To save battery life, the power is shut off automatically if no key is pressed
for 6 seconds.
Specifications
Measuring range | -32 to 535°C |
---|---|
Accuracy | ±3.0°C:-32 to -20 °C, ±2.0°C: -20 to +100°C, ±2%:100 to 535°C |
Infrared spectral band | 5 to 14 pm |
Repeatability | Within ±1°C |
Resolution | 0.1°C |
Response | 500ms |
Emissivity | Variable between 0.10 and 1.00 (by 0.01 steps): Before shipment: |
0.95
Measuring diameter| 1000mm/078mm (Distance/ Measuring dia.: 12:1)
Battery| One 9 V Dry cell battery (006P)
Operating temperature & humidity| 0 to 50°C/ 10 to 90% RH
Auto power off| If no key is pressed for 6 seconds, the power is shut off
automatically.
---|---
LCD display| LCD with backlight (blinks in red when alarm function is
activated)
Collimation| Laser beam (630 to 670nm 1mW or less) specifies the center.
Display| Max/ Min/ Average value display
Dual display| Simultaneous display: Measured value and either of max,
min, average or thermocouple value.
Thermocouple| K-type
Measuring range of thermocouple| -199 to 1372ºC
Accuracy of thermocouple| ±(1.5%+1ºC): -40 to 1372ºC
Dimension| 180 × 130 × 40mm
Weight| Approx. 195g (excluding battery)
Accessories| Carrying case, One 9 V Dry cell battery (006P), Instruction
manual
Approved standards| EN61326, IEC61000-4-2, IEC61000-4-3, IEC61000-4-8,
IEC60825-1, CE Mark
Environmental standards| **** EU RoHS Directive compliant
Name of each part
4-1. Body
LASER RADIATION DOES NOT STARE INTO BEAM CLASS 2 LASER PRODUCT
4-2. Display part (Buttons) ****
-
Battery mark ………………… Appears to indicate low battery.
-
Laser mark ………………….. Appears when powering on the instrument.
Buzzer mark ………………… Appears when the alarm function (P. 8) is enabled. -
Thermocouple TC ………… Lights up while measuring with a K-type thermocouple
Emissivity [EMS] …………. Lights up while setting emissivity. -
Upper limit alarm [HAL] …. Lights up while setting an upper limit for alarm.
Lower limit alarm [LAL] ….. Lights up while setting a lower limit for alarm. -
Cursor button (Down) ……. Alters number.
-
Buzzer button ………………. Turns on/ off the buzzer sound.
-
SCAN …………………………. Blinks during measurement.
HOLD …………………………. Lights up while the display retains its last measurement (P. 8). -
Unit …………………………….. Indicates measurement unit ºC (P. 9).
-
Main display ………………… Shows measured values.
-
MAX/ MIN/ AVE ……………. Lights up when using each function (P. 7).
-
Sub display ………………….. Shows the selected measurement mode and values in parameter setting mode (P. 7).
-
Cursor button (Up)………… Alters number.
-
Function button…………….. Switches between measurement and parameter setting modes (P. 7).
Getting started
5-1. A measurement model and parameter setting mode
Pull the trigger on the instrument to take a measurement. Press the F button
(Function button) to use other functions. The main display shows measured
values and the sub-display shows values obtained with each function. The
following flow chart shows the sequence and corresponding action.
Measurement mode
-
SCAN Shows measured values.
-
MAX Shows the maximum value taken during one measurement on the sub-display.
-
MINShows the minimum value taken during one measurement on the sub-display.
-
AVGAverages the values taken in one measurement and shows it on the sub-display.
-
TC Shows measured values using K-type thermocouples.
-
EMS Shows emissivity. (P. 10) Use the cursor buttons to alter emissivity.
-
HAL Shows the upper limit for alarm. (P. 8) Use the cursor buttons to alter the limit value. The red backlight blinks when a measured value exceeds this limit.
-
LALShows the lower limit for alarm. (P. 8) Use the cursor buttons to alter the limit value. The red backlight blinks when a measured value falls behind this limit.
-
Auto power-off function activates in 6 seconds. When the display becomes blank due to this power-off function, pull the trigger.
5-2. Other functions
-
K-type thermocouple measurement
Plug a K-type thermocouple into the connector for the thermocouple, and press the function button to select the TC function. Measured values using a thermocouple are displayed on the sub-display. -
Alarm function
The red backlight blinks when a measured value exceeds or falls behind the values which are set on HAL and LAL functions. On both functions, it is possible to enable or disable the buzzer sound. -
HOLD function
The measured value is held and displayed for 6 seconds after releasing the trigger. -
Low battery warning
A battery mark appears when the battery voltage becomes low.
Battery replacement
When the battery is low, the battery mark **** appears on the display. Replace the battery with a new one. When nothing is displayed on the display if attempted, it may also be battery exhaustion. Replace the battery with a new one. Open the battery cover as the below figure shows, insert a new battery with correct polarity, and then close the cover.
Setting emissivity (reference)
The quantity of infrared emitted radiation from objects depends on the objects’ material, surface conditions, measuring temperatures, and so on. The table below shows the emissivity of some objects. Please note that these values are just for reference. Please also refer to the estimated emissivity using the separately sold black body tape MODEL8085.
Material
| Emissivity| Material|
Emissivity
---|---|---|---
Water, Ice| 0.98| Cloth, Fabric (colored)| 0.95
Soil| 0.92 to 0.96| Leather, Fur| 0.96
Concrete (wet)| 0.96 to 0.98| Human skin| 0.99
Concrete (dry)| 0.91 to 0.95| Vegetable, Fruit| 0.98
Ceramics| 0.85 to 0.95| Dough| 0.98
Stone, Asbestos| 0.92| Meat| 0.98
Plastics| 0.90 to 0.95| Copper oxide| 0.50 to 0.60
Rubber (black)| 0.95| Ferric oxide| 0.70 to 0.80
Wood| 0.98| Painted surface| 0.80
Paper| 0.92| Tile| 0.80
Estimated emissivity using the black body tape
If a measuring object accepts sticking of adhesive tape, stick a piece of
the black body tape (emissivity: 0.94) to the object, and measure the
temperature after setting the emissivity to 0.94.
- Black body is a substance that absorbs all incident radiation: no transmission and no reflection and would have an emissivity value of 1.
- Infrared thermometers can easily measure temperature from a distance with safe. They can be used for electrical or equipment maintenance, building control and process monitoring, and other applications.
Kyoritsu reserves the right to change specifications or designs described in this manual without notice and without obligations.
KYORITSU ELECTRICAL INSTRUMENTS WORKS, LTD.
2-5-20, Nakane, Meguro-ku, Tokyo, 152-0031
Japan Phone: +81-3-3723-0131
Fax: +81-3-3723-0152
Factory: Ehime,Japan www.kew-ltd.co.jp
92-1954D
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>