MEAN WELL MPM-15 Series 15W High Reliable Green Medical Encapsulated Type
Product Information
The 15W High Reliable Green Medical Encapsulated Type MPM-15 series is a compact-sized power supply unit. It is designed to meet the safety standards of EN 60601-1, IEC 60601-1, BS EN/EN60335-1, ANSI/AAMI ES60601-1, BS EN/EN60601-1, IEC60601-1, and TPTC004. This power supply unit is specifically approved for medical use and provides a reliable and efficient power solution.
Key Features:
- Compact size: The MPM-15 series is designed in a compact size, making it suitable for applications with limited space.
- Medical safety approved: This power supply unit complies with the safety standards of ANSI/AAMI ES60601-1 and IEC/BS EN/EN60601-1, ensuring safe usage in medical environments.
- Suitable for BF application: The MPM-15 series can be used in BF (Body Floating) applications with appropriate system consideration.
- No load power consumption: The power supply unit has low power consumption when no load is connected, resulting in energy efficiency.
Product Usage Instructions
To use the MPM-15 series power supply unit, please follow the instructions below:
- Ensure that the power supply unit is compatible with your specific medical device or application.
- Verify that the input voltage and frequency of the power supply match the requirements of your system.
- Connect the AC input to a suitable power source using the provided power cord.
- Connect the output terminals of the power supply unit to your medical device, ensuring a secure and proper connection.
- Before turning on the power supply, double-check all connections to avoid any potential hazards.
- Once all connections are verified, switch on the power supply unit using the designated power switch.
- Monitor the power supply unit during operation to ensure it is functioning properly.
- In case of any abnormality or malfunction, immediately disconnect the power supply and consult the user manual for troubleshooting steps or contact customer support for assistance.
It is essential to follow the user manual guidelines and safety precautions when using the MPM-15 series power supply unit to ensure safe and reliable operation.
Features
-
2.06″×1.07″ compact size
-
Medical safety approved (2 x MOPP) according to ANSI/AAMI ES60601 -1 and IEC/BS EN/EN60601-1
-
Suitable for BF application with appropriate system consideration
-
No load power consumption<0. 1W
-
Extremely low leakage current
-
Wide operating temp. range – 40 ~ +85°C
-
EMI class B for class II configuration
-
Protections:
Short circuit/Overload/Over voltage/Over temperature -
No minimum load required
-
3 years warranty
Applications
- Portable medical device
- Mobile clinical workstation
- Medical computer monitor
- Medical examination instrument
GTIN CODE
MW Search: https://www.meanwell.com/serviceGTIN.aspx.
Description
MPM-15 is a 15W high-density and small size (52.427.224mm) AC/DC module type medical power supply series offered in pin type. It features the operation for 80~264VAG, a low no-load power consumption of less than 0. 1W, high efficiency of up to 87%, Class II (no FG) double insulation, outstanding dissipation, and high lifespan thanks to the interior potting, 2G anti-vibration, high EMC performance, 4KVAC isolation, etc. The design observes IEC/BS EN/EN60601-1 and ANSI/AAMI ES60601-1version three with 2xMOPP level and ultra-low leakage current (<80 A). It is very suitable for BF (patient contact) type medical devices or relevant equipment.
Model Encoding
SPECIFICATION
MODEL| MPM-15-3.3| MPM-15-5| MPM-15-12| MPM-15-15|
MPM-15-24
---|---|---|---|---|---
OUTPUT
| DC VOLTAGE| 3.3V| 5V| 12V| 15V| 24V
RATED CURRENT| 3.5A| 3A| 1.25A| 1A| 0.63A
CURRENT RANGE Note.2| 0 ~ 3.5A| 0 ~ 3A| 0 ~ 1.25A| 0 ~ 1A| 0 ~ 0.63A
PEAK CURRENT| 3.85A| 3.3A| 1.38A| 1.1A| 0.69A
RATED POWER| 11.6W| 15W| 15W| 15W| 15.1W
PEAK LOAD(10sec.) **Note.3| 12.7W| 16.5W| 16.6W| 16.5W| 16.6W
RIPPLE & NOISE (max.) **Note.4| 150mVp-p| 150mVp-p| 150mVp-p|
180mVp-p| 180mVp-p
VOLTAGE TOLERANCE Note.5| ±2.0%| ±2.0%| ±2.0%| ±2.0%| ±2.0%
LINE REGULATION| ±0.5%| ±0.5%| ±0.3%| ±0.3%| ±0.3%
LOAD REGULATION| ±0.5%| ±0.5%| ±0.5%| ±0.5%| ±0.5%
SETUP, RISE TIME| 1500ms, 30ms/230VAC 1500ms, 30ms/115VAC at
full load
HOLD UP TIME (Typ.)| 40ms/230VAC 10ms/115VAC at full load
INPUT
| VOLTAGE RANGE Note.6| 80 ~ 264VAC 113V~370VDC
FREQUENCY RANGE| 47 ~ 440Hz
EFFICIENCY (Typ.)| 83.5%| 85.5%| 86.5%| 87%| 86.5%
AC CURRENT (Typ.)| 0.6A/115VAC 0.3A/230VAC
INRUSH CURRENT (Typ.)| COLD START 20A/115VAC
45A/230VAC
LEAKAGE CURRENT (max.) Note.7| Touch current <80μA/264VAC
PROTECTION
| ****
OVERLOAD
| 110% ~ 150% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is
removed
OVER VOLTAGE
| 3.8 ~ 5V| 5.8~ 6.8V| 13.8 ~ 16.2V| 17.3 ~ 20.3V| 27.6 ~ 32.4V
Protection type : Shut off o/p voltage, clamping by zener diode
OVER TEMPERATURE| Protection type : Shut down o/p voltage, recovers
automatically after temperature goes down
ENVIRONMENT
| WORKING TEMP.| -40 ~ +85℃ (Refer to “Derating Curve”)
WORKING HUMIDITY| 20 ~ 90% RH non-condensing
STORAGE TEMP., HUMIDITY| -40 ~ +85℃, 10 ~ 95% RH non-condensing
TEMP. COEFFICIENT| ±0.03%/℃ (0 ~ 60℃)
SOLDERING TEMPERATURE| Wave soldering: 265℃,5s (max.); Manual soldering:
390℃,3s (max.)
VIBRATION| 10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X,
Y, Z axes
OPERATING ALTITUDE Note.8| 5000 meters
SAFETY & EMC
(Note.9)
| SAFETY STANDARDS| IEC60601-1, TUV BS EN/EN60601-1, IEC60335-1, Dekra BS EN/EN60335-1, EAC TP TC 004, UL ANSI/AAMI ES60601-1
(3.1 version), CAN/CSA-C22 3rd Edition approved
ISOLATION LEVEL| Primary-Secondary: 2xMOPP
WITHSTAND VOLTAGE| I/P-O/P:4KVAC
ISOLATION RESISTANCE| I/P-O/P:100M Ohms / 500VDC / 25℃/ 70% RH
EMC EMISSION
| Parameter| Standard| Test Level / Note
Conducted emission| BS EN/EN55011 (CISPR11)| Class B
Radiated emission| BS EN/EN55011 (CISPR11)| Class B
Harmonic current| BS EN/EN61000-3-2| Class A
Voltage flicker| BS EN/EN61000-3-3| —–
EMC IMMUNITY
| BS EN/EN55035, BS EN/EN60601-1-2
Parameter| Standard| Test Level / Note
ESD| BS EN/EN61000-4-2| Level 4, 15KV air ; Level 4, 8KV contact
RF field susceptibility
| ****
BS EN/EN61000-4-3
| Level 3, 10V/m( 80MHz~2.7GHz )
Table 9, 9~28V/m( 385MHz~5.78GHz )
EFT bursts| BS EN/EN61000-4-4| Level 3, 2KV
Surge susceptibility| BS EN/EN61000-4-5| Level 3, 1KV/Line-Line
Conducted susceptibility| BS EN/EN61000-4-6| Level 3, 10V
Magnetic field immunity| BS EN/EN61000-4-8| Level 4, 30A/m
Voltage dip, interruption| ****
BS EN/EN61000-4-11
| 100% dip 1 periods, 30% dip 25 periods,
100% interruptions 250 periods
OTHERS
| MTBF| 7319.8K hrs min. Telcordia SR-332 (Bellcore) ; 1210.0K hrs
min. MIL-HDBK-217F (25℃)
DIMENSION| 52.427.224mm (LWH) or 2.06″1.07″0.94″ inch
PACKING| 0.056Kg; 240pcs/14.4Kg/0.94CUFT
NOTE|
- All parameters NOT specially mentioned are measured at 230VAC input, rated load, and 25℃ of ambient temperature.
- No minimum load is required.
- 33% Duty cycle maximum within every 30 seconds. The average output power should not exceed the rated power.
- Ripple & noise are measured at 20MHz of bandwidth by using a 12″ twisted pair-wire terminated with a 0.1μf & 47μf parallel capacitor.
- Tolerance: includes set up tolerance, line regulation, and load regulation.
- Derating may be needed under low input voltages. Please check the derating curve for more details.
- Touch current was measured from primary input to DC output.
- The ambient temperature derating of 3.5℃/1000m with fanless models and of 5℃/1000m with fan models for operating altitudes higher than 2000m(6500ft).
- The power supply is considered a component that will be installed into the final equipment. The final equipment must be re-confirmed that it still meets EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”
(as available on http://www.meanwell.com)
※ Product Liability Disclaimer: For detailed information, please refer to https://www.meanwell.com/serviceDisclaimer.aspx
Block Diagram
Derating Curve
Output Derating VS Input Voltage
Mechanical Specification
Installation Manual
Please refer to: http://www.meanwell.com/manual.html.