UEi C163 Test Instruments User Guide

June 4, 2024
UEi

UEi C163 Test Instruments User Guide
UEi C163 Test Instruments

OVERVIEW

Product Overview
Product Overview

Screen Menu

Press ”Power” Button for 2 seconds. The analyzer will perform a 60 second purge. This should be done in fresh air.
Rotate Selector Dial to MENU to set up or customize your settings.
Selected parameter are centered, highlighted with arrow icons on either side.

  • Use and buttons to scroll menu options,
  • Press to select and edit a parameter.
  • Press and buttons to change contents of a field.
  • Press button to scroll fields.
  • Press to enter content and move to next field.
STATUS SCREEN

Screen Instruction

Rotate Selector Dial to STATUS to view parameters. Selected Fuel

  • Ti = Ambient Temperature
  • ATM = Atmospheric Pressure (measured in mbar)
  • CAL = Days remaining until calibration

Press and buttons to scroll Status Bar options.

Time, Pump Status, Battery Status
Fuel Type
Date

Change Fuel Type in STATUS SCREEN

  • Press and buttons to the currently selected fuel.
  • Press and Hold button until arrows indicate fuel edit mode is activated.
  • The and to scroll fuel types.
  • Press to select desired fuel.

BEFORE TESTING

CHECKLIST

  • Clean particle filter
  • Water trap and probe line are free of condensate
  • All hose and thermocouple connections are properly secured
  • Water trap is fitted correctly
  • Power on and zero
  • Flue gas probe is sampling ambient FRESH air

Remove the rubber boot and red drain plug. Allow the water to drain out. Re- insert the rubber plug and replace boot cover.

Remove protective rubber boot. Remove water trap from analyzer. Remove particle filter from water trap and replace with new one. Reattach trap and boot.
Battery Installation

COMBUSTION ANALYSIS

  • Insert flue probe into combustion input. Insert K-type temperature probe into input T1.
    COMBUSTION ANALYSIS

  • Insert flue probe into stack. Adjust cone so the probe tip is approximately at the center of the stack.
    COMBUSTION ANALYSIS

  • Rotate function rotary dial to O2/Eff for Test Screen 1 parameters.
    COMBUSTION ANALYSIS

NOTE: If there is no hole, check with service and installation instructions of appliance for best testing method or check with your local authority on best practice.

  • O2/Eff displays CO2, O2, Efficiency, Flue temperature, Inlet temperature.
    COMBUSTION ANALYSIS

  • Rotate the function rotary dial to CO/CO2 for Test Screen 2 parameters.
    COMBUSTION ANALYSIS

  • CO/CO2 displays CO (ppm), CO2, Losses calculated, Excess air %. Make any adjustments needed for proper Combustion and wait for analyzer to display change in readings. (Repeat as necessary).
    COMBUSTION ANALYSIS

CHANGE FUEL TYPE

  • Rotate the dial to either STATUS, AUX, O2/Eff, CO/CO2 or Temp setting.
    Changing Fuel Type

  • Press button until the Fuel Type displays on the Status Bar. Press and Hold until two arrow icons display.
    Changing Fuel Type

  • Press and buttons to scroll fuel selections. Press to select fuel.
    Changing Fuel Type

AUXILIARY SCREEN
  • When performing a combustion Test, Rotate function rotary dial to AUX to view the Auxiliary Test Screen’s custom parameters.
    AUXILIARY SCREEN

  • AUX displays 5 lines of editable parameters. To edit line values Press button until Edit? displays on the Status Bar. Press and Hold until the Status Bar reads LINE 1.
    AUXILIARY SCREEN

  • Press and buttons to scroll parameter selections. Press to select a parameter and advance to the next Line. Continue until cursor is back to Status Bar to exit.
    AUXILIARY SCREEN

The AUX screen allows you to assign any of the following parameters on any display line. O2, CO, Display Fuel Type, CAL, XAIR, LOSS, Efgc, ATM, Ti, ∆T, T2, T1, CO2, COa, or COn. Pressure(C163 and !64 only). NOX and No (C164 only).

TEMPERATURE TEST
  • Insert K-type temperature probes into input T1 and T2.
    Testing Temperature

  • Rotate function rotary dial to TEMP for Temperature Test Screen.
    Testing Temperature

  • TEMP displays T1, T2 and Differential Temperature. (Pressure C163 C164 only).
    Testing Temperature

PRINTING & LOGGING

  • To Print: Align printer (sold separately) above Analyzer.
    Painting logging

  • During testing, Press to start printing.
    Press again to abort.
    Painting logging

  • To Log: During testing, Press and Hold to log a report.
    To view these Logs, see View Saved Logs section.
    Painting logging

VIEW SAVED LOGS
  • Rotate the dial to MENU.
    Press button until the LOGS is selected between two arrow icons. Press to select.
    Saved Logos

  • Screen displays all saved LOGS. With VIEW selected in the Status Bar, Press to select.
    Saved Logos

  • Press and buttons to scroll Test Report categories. Press to select a category.
    Saved Logos

  • Line 1 displays the LOG number on screen.
    To view another LOG, Press and  buttons to NEXT.
    Press to advance to next log.
    Repeat until you reach desired LOG.
    Saved Logs

  • To print the LOG, Press and buttons until PRINT is selected between two arrow icons. Press to select to start printing.
    Saved Logos

  • To EXIT, Press and buttons until EXIT is selected between two arrow icons. Press to EXIT screen.
    Repeat until return to MENU screen.
    Saved Logos

POST TESTING

Remove the probe from the flue and allow analyzer to purge with fresh air until readings return to zero. O2 to 20.9% and CO below 20 ppm
CAUTION: probe tip will be HOT.
POST TESTING

POWERING OFF
When you power off the Analyzer, there is a 10 second purge.

CAUTION
Make sure you do not exceed the analyzer’s operating specifications.

  • Do not exceed the flue probes maximum temperature (1112˚F)
  • Do not exceed the analyzer’s internal temperature range 112°F (50°C)
  • Do not put the analyzer on a hot surface
  • Do not exceed the water trap’s level
  • Do not let the particle filter become dirty and blocked

**BATTERIES

**

The analyzer is fitted with 3 (AA) size rechargeable batteries. Observe correct polarity when replacing.
Always check operation immediately after battery replacement.
Do not leave batteries in unit for extended periods of non-use.

WARNING
Do not recharge with Alkaline batteries fitted. Alkaline batteries are not rechargeable. Attempting to recharge alkaline batteries may result in product damage and create a fire risk.
Time and Date will need to be reset after changing batteries. Always dispose batteries using approved disposal methods that protect the environment.

Download the full Owners Manual at UEiTEST.COM or SCAN THIS CODE

WARRANTY

The C162/C163/C164 combustion analyzer is warranted for one year including sensors. If within the warranty period your instrument should become inoperative from such defects, the unit will be repaired or replaced at UEi’s option. This warranty covers normal use and does not cover damage which occurs in shipment or failure which results from alteration, tampering, accident, misuse, abuse, neglect or improper maintenance (calibration). Batteries and consequential damage resulting from failed batteries are not covered by warranty. Any implied warranties, including but not limited to implied warranties of merchantability and fitness for a particular purpose, are limited to the express warranty. UEi shall not be liable for loss of use of the instrument or other incidental or consequential damages, expenses, or economic loss, or for any claim or claims for such damage, expenses or economic loss. A purchase receipt or other proof of original purchase date will be required before warranty repairs will be rendered. Instruments out of warranty will be repaired (when repairable) for a service charge. Warranty only covers hardware and does not extend to software applications or accessories. Contact UEi for specific warranty and service information.

UEi Test Instruments
1-800-547-5740
info@ueitest.com

This warranty gives you specific legal rights. You may also have other rights which vary from state to state.

ANNUAL RECERTIFICATION SERVICES

When you:
Request Recertification or Service Online Within 1 Year of Purchase or Service.
Symbol

UEi will:

6-Year Warranty: All UEi combustion analyzers have a standard 1-year warranty. Each recertification includes a 30-point inspection and extends the warranty for 1 more year for up to 6 years from the date of warranty.

****Contractors who book recertification of a C160 series analyzer atwww.ueitest.com/service within 12 months from either the date of purchase or the date of the last recertification will receive reduced service pricing that lowers the cost of ownership and 2 additional benefits:

24-hour Service: Analyzers received for standard recertification and service through UEi Service+ are shipped back on the next business day.

Free Shipping: UEi Service+ offers free shipping both to and from our service center. When customers book their recertification, they receive a prepaid UPS Ground shipping label.

Get full details at www.ueitest.com/service

Register Online
Registering you analyzer online is quick and easy. Just log in or setup an account, it only takes a couple of minutes. Once logged in you can register you analyzer by providing some product information and uploading a proof-of- purchase.

When it’s time to request recertifcation, just log into your account, select the analyzer, select the service and place your order.

Support

Copyright © 2021 Kane USA Inc. All Rights Reserved. 201916B 0921