UEi DL579 Digital HVAC Clamp Meter Instruction Manual
- June 10, 2024
- UEi
Table of Contents
DL579
TEST iNSTRUMENTS™
Digital HVAC Clamp Meter w/ CATIII 600V
INSTRUCTION
MANUAL
Functions
- True RMS
- 750V AC/1000V DC
- 600A AC
- AC/DC microamps: 2000µA
- Capacitance: 6000µF
- Frequency: 99.99kHz
- Duty cycle
- Diode test
- Audible continuity
- NCV
- LRA Inrush
- Temperature range: -328˚ to 2462˚F
- Resistance: 60MΩ
- Low Z (Low Impedance)
Features
- Dual display
- Auto/Manual ranging
- Back light
- Low battery indicator
- Data Hold
- Auto power off
- Test lead storage
- Auto calibration
- Built-in Magnet w/hanging strap (Optional)
- Visible high-voltage alert
- Input jack locks
- Min/Max
- Battery compartment latches
General Specifications
- Operating Temperature: 32˚ to 122˚F (0˚ to 50˚C)
- Storage Temperature: -4˚ to 140˚F (-20˚ to 60˚C)
- Operating Humidity: <80%
- Pollution Degree: 2
- Display: 3 5/6 digits 6,000 count
- Back light: Yes
- Refresh Rate: 3/sec
- Over-range: “OL” is displayed
- Apo: Auto power off after 30 minutes of use.
- Dimensions: 8.7” x 2.52” x1.41”
- Item Weight: 0.65 lb
- CAT Rating: CATIV 300V, CATIII 600V
- Certifications: cETLus UL 61010-1: 2012, IP42
- Battery Type: (AAA) 2
- Test leads: Test lead ATL55 w/alligator clips
Important Safety Warnings
WARNING
Read entire Safety Notes section regarding potential hazards and proper
instructions before using this meter. In this manual the word “WARNING” is
used to indicate conditions or actions that may pose physical hazards to the
user. The word “CAUTION” is used to indicate conditions or actions that may
damage this instrument.
WARNING
To ensure safe operation and service of the tester, follow these
instructions.
Failure to observe these warnings can result in severe injury or death.
WARNING
- Before each use, verify meter operation by measuring a known voltage or current.
- Never use the meter on a circuit with voltages that exceed the category based rating of this meter.
- Do not use this meter during electrical storms or in wet weather.
- Do not use the meter or test leads if they appear damaged.
- Ensure meter leads are fully seated and keep fingers away from the metal probe contact when making measurements. Always grip the leads behind the finger guards molded into the probe. For information on test lead shields instructions on page 19.
- Do not open the meter to replace batteries while the probes are connected.
- Use caution when working with voltages above 60 DC or 25 AC RMS. Such voltages pose shock hazards.
- To avoid false readings that can lead to electrical shock, replace batteries if a low battery indicator appears.
- Unless measuring voltage or current, shut off and lockout power before measuring resistance or capacitance.
- Always adhere to national and local safety codes. Use proper personal protective equipment (PPE) to prevent shock and arc blast injury where hazardous live conductors are exposed.
- Always turn off power to a circuit or assembly under test before cutting, unsoldering or breaking the current path. Even small amounts of current can be dangerous.
- Always disconnect the live test lead before disconnecting the common test lead from the circuit.
- In the event of electrical shock, ALWAYS bring the victim to the emergency room for evaluation, regardless of victim’s apparent recovery. Electrical shock can cause unstable heart rhythms that may need medical attention.
- If any of the following occur during testing, turn off the power source to the circuit being tested: arcing, flame, smoke, extreme heat, smell of burning materials or discoloration or melting of components.
WARNING
Higher voltages and currents require greater awareness of physical safety
hazards. Before connecting the test leads; turn off power to the circuit under
test, set meter to the desired function and range; connect the test leads to
the meter first, then connect to the circuit under test. Reapply power. If an
erroneous reading is observed, disconnect power immediately and recheck all
settings and connections.
WARNING
This meter is designed to provide HVAC/R technicians with the capabilities
they need to diagnose and repair HVAC/R system. Observe all recommended safety
procedures that include proper lockout utilization and use of personal
protective equipment that includes safety glasses, gloves and flame resistant
clothing.
Symbols
| AC (Alternating current)| **| DC (Direct current)
---|---|---|---
| Negative| | AC/DC Voltage or Current
AT| Auto-ranging| OL| Overload: Range Exceeded
Apo| Auto power off Active| NCV| Non-Contact Voltage
| Low Battery| HOLD| Hold/Capture Value
MIN| Minimum measured
value displayed| MAX| Maximum measured value displayed
%| Duty Cycle| Hz| Hertz/Frequency
V| Voltage| INRUSH| Inrush
A| Amperage| Ω| Ohms/Resistance
| Diode| | Capacitance
nF| Nanofarad| | Microfarad
µA| Microamps| | Continuity
| Degrees Fahrenheit| | Degrees Celsius
M| Mega (x10 6 or 1,000,000)| M| Milli (x10 -3 or 0.001)
k| Kilo (x10 3 or 1,000)| | Micro (x10 -6 or 0.000001)
| Warning or Caution| | Ground
| Dangerous Levels| | Double Insulation
(Protection to Class II)
| Safe for disconnect from live conductors| ——–| No reading detected
KΩ| Kilo Ohms| EF**| Electric Field
Category Definitions
Measurement Category| Short-Circuit (typical) kAa| Location in
the building installation
---|---|---
II| < 10| Circuits connected to mains socket outlets and similar points in the
MAINS installation
III| < 50| Mains distributions parts of the building
IV| > 50| Source of the mains installation in the building
Overview
A. Clamp: Measure inductive AC current. Opens to 1.25” (32.0mm).
B. Conductor Alignment Marks: Use to aid the visual alignment of a conductor
when measuring inductive amperage. Greatest accuracy is achieved when the
conductor inside the clamp is centered at the intersection of these marks.
C. Category Max Indicator: Maximum CAT Rating for clamp jaw.
D. NCV Alert Light: Indicates voltage when in NCV (Non Contact Voltage) mode
and High Voltage alert.
F. Clamp Lever: Opens and closes current clamp jaw.
NOTE: The clamp uses a high-tension spring to close the jaw. Do not allow
fingers or objects to become pinched in the base as the jaws close.
G. Rotary Selector Dial: Set Rotary Selector Dial desired function
H. Range Button:
- Press to set manual range desired
I. SELECT Button:
-
Press AC or DC on Low Z setting; DCmV or DCV on Voltage setting; to activate ohms, Continuity, Diode, Capacitance on Ohms/Continuity/
Diode/Capacitance setting; AC or DC on μA setting; °C or °F temperature setting
J. Category Max Indicator: Maximum CAT Rating for input jacks.
K. Test Lead Input Jacks: Multifunction and Positive input jacks. -
Multifunction input port used for measuring: AC or DC volts, resistance, continuity, diode, capacitance, AC or DC μA.
L. Wire Separation Tab/ NCV sensor: Use to isolate an individual wire from a bundle for testing. NCV sensor detects live voltage.
M. Test Lead Holder
N. Hand Guide: Used as a point of reference for the operator’s safety.
O. Hold/Back light Button: -
Press to hold the reading on the display. Press again to return to live reading.
-
Press and hold to turn on Back light.
-
Backlight turn off after 60 seconds.
-
Press hold during power up to disable the Auto power off function.
P. Min/Max Button:
-
Press to enter MIN/MAX mode.
-
Step through maximum(MAX), minimum(MIN) and present values.
-
Press and hold return to Live readings
Q. Display: -
High contrast dual display with backlit.
-
AC Amps reading will always display on upper display.
R. Inrush/Hz Duty Button: -
Press to enter LRA Inrush mode (See page 18 for details) (must be in AC Amps mode first).
-
Press again to return to live readings.
-
Press and hold to select Hz/Duty in AC volt or AC µA.
S. K-Type Temperature Probe Inputs
T. Input Jack Lock: Switch to use Temperature or Test lead inputs
U. Built-in versatile magnet to use as a mount or as a strap (optional)
V. Battery Cover: Easy access for replacing batteries without breaking calibration seal.
W. Battery Compartment Latches
X. Serial Number
Low Z (Low Impedance)
- Rotate Selector Dial to Low Z
- Default = Auto Selection (DC or AC)
- Press SELECT x1 =
- Press SELECT x2 =
Features:
AC Amps <600A Jaw Single Conductor Only
- Rotary selector dial = any position
- Reading show on upper display
- AC Amps can be measured in any position of the rotary selector dial.
- Center wire in guides for best accuracy.
- Opposing currents cancel each other (use line-splitter when necessary).
- Keep hands below guard when measuring high current levels.
- Do not attempt to measure more than 600A AC.
Features:
AC Amps Measurement – Jaw input
Range | Resolution | Accuracy | Overload Protection |
---|---|---|---|
60.00A | 0.01A | ±2.0% + 8dgts | 600V RMS |
600.0A | 0.1A |
45Hz to 400Hz True RMS
Minimum Current for Clamp Measurement: 0.3A
Non-Contact Voltage
NCV Sensor in the tip.
- Rotate Rotary Selector Dial to NCV position move the tip of the clamp meter near voltage source.
- Non-Contact Voltage Detection is used to detect power with sensor located in the tip of the clamp head, indicates positive response with both an Audible and Visual alert.
- Do not use Non-contact voltage detector to determine if there is current on the wire. Detection operation could be affected by socket design, insulation thickness, type or other factors.
- Voltage indicator light may also light when voltage (>AC/DC 30V) is present on the meter’s input jack or from an external interference such as motors, flashlights, etc.
Features:
On Voltage
Approx. 24V AC
AC/DC Microamps: <2000µA
WARNING
- Do not attempt to measure more than 2000µA.
Features:
DC Microamps Measurement -Test lead input
Range | Resolution | Accuracy | Overload Protection |
---|---|---|---|
600.0uA | 0.1uA | ±1.2% + 3dgts | 600V RMS |
2000uA | 1uA |
AC Microamps Measurement -Test lead input
Range | Resolution | Accuracy | Overload Protection |
---|---|---|---|
600.0uA | 0.1uA | ±2.0% + 5dgts | 600V RMS |
2000uA | 1uA | ±1.5% + 5dgts |
45Hz to 400Hz True RMS
Temperature °F/°C
- Disconnect test lead probes from voltage source and meter.
- Move Input Jack Locks to “TEMP” setting.
- Use K-Type thermocouple temperature probes only.
- Stated accuracy does not account for thermocouple accuracy.
Features:
Range | Resolution | Accuracy | Overload Protection |
---|---|---|---|
-328° to 999°F | 0.1°F | ±(1.0% + 3.6°F) | 30V RMS |
(-200° to 999°C) | (0.1°C) | ±(1.0% + 2.0°C) | |
1000° to 2462°F | 1°F | ±(1.0% + 3°F) | |
(1000°C to 1350°C) | (1°C) | ±(1.0% + 2°C) |
Continuity
-
Buzzer sounds at less than < 40Ω.
WARNING -
Do not measure resistance on a live circuit.
Features:
Open circuit voltage < 0.64V | Overload Protection |
---|---|
Threshold Approx : < 40Ω | 600V RMS |
Resistance: < 60MΩ
WARNING
- Do not measure resistance on a live circuit.
Range | Resolution | Accuracy | Overload Protection |
---|---|---|---|
600.0Ω | 0.1Ω | ±1.0% + 4dgts | 600V RMS |
6.000kΩ | 0.001kΩ | ||
60.00kΩ | 0.01kΩ | ||
600.0kΩ | 0.1kΩ | ||
6.000MΩ | 0.001MΩ | ||
60.00MΩ | 0.01MΩ | ±2.0% + 4dgts |
Capacitance
Range | Resolution | Accuracy | Overload Protection |
---|---|---|---|
60.00nF | 0.01nF | 3.5% + 6dgts | 600V RMS |
600.0nF | 0.1nF | ||
6.000uF | 0.001uF | ||
60.00uF | 0.01uF | ||
600.0uF | 0.1uF | ||
6000uF | 1uF |
WARNING To avoid damaging the meter or equipment under test, safely
discharge Capacitors before measuring capacitance. Large value capacitors
should be discharged through an appropriate resistance load. Use the DC
Voltage function to confirm the capacitor discharge.
Diode
GOOD DIODE BAD DIODE
- Forward voltage drop if forward biased.
- “O.L.” if reverse biased.
Features: ****
Diode Test
Range| Open Circuit Voltage| Test Current(Typical)|
Overload Protection
---|---|---|---
6.0V| < 3.0V DC| 0.25mA| 600V RMS
Voltage 750V AC
WARNING
- Use CATIII rated test leads or higher.
- Do not attempt to measure more than 750V AC.
- Keep hands below line when measuring high current levels.
WARNING
- High Voltage indicator will display and audible alert will sound over 600V AC
- High Voltage indicator will display (without audible alert) over 30V AC
Features:
Range | Resolution | Accuracy | Overload Protection |
---|---|---|---|
600.0mV | 0.1mV | ±2.0% + 5dgts | 750V RMS |
6.000V | 0.001V | ||
60.00V | 0.01V | ||
600.0V | 0.1V | ||
750V | 1V |
True RMS: 45Hz to 400Hz
Voltage 1000V DC WARNING
- Use CATIII rated test leads or higher.
- Do not attempt to measure more than 1000V DC.
- Keep hands below line when measuring high current levels.
WARNING
- High Voltage indicator will display and audible alert will sound over 600V DC
- High Voltage indicator will display (without audible alert) over 30V DC
Features:
DC Volts
Range | Resolution | Accuracy | Overload Protection |
---|---|---|---|
600.0mV | 0.1mV | ±0.5% + 4dgts | **** 1000V RMS |
6.000V | 0.001V | ||
60.00V | 0.01V | ||
600.0V | 0.1V | ||
1000V | 1V | ±0.8% + 10dgts |
Frequency (Hz)/Duty Cycle (%) Use CAT III rated leads or higher.
Rotate Rotary selector Dial to position select AC voltage, press and hold
the button for Frequency and Duty Cycle modes.
WARNING
Do not attempt to measure more than 750V AC.
Features:
Frequency Measurement – Test lead input
Range | Resolution | Accuracy | Overload Protection |
---|---|---|---|
99.99Hz | 0.01Hz | 0.1% + 4dgts | 1000V RMS |
999.9Hz | 0.1Hz | ||
9.999kHz | 0.001kHz | ||
99.99kHz | 0.01kHz |
Minimum Frequency: 0.5Hz, DC V offset should be zero
Sensitivity: > 10% of each AC volt range except 4V (>20%) range only
Duty Cycle – Test lead input
Range | Accuracy | Overload Protection |
---|---|---|
1.0 to 99.0% | ±(0.2% per kHz + 0.1% + 5 dgts) | 1000V RMS |
LRA Inrush
The UEi LRA Inrush is programmed to properly capture the starting current forInrush current can be measured in any position of the rotary selector Dial. compressor motors.
- Select the range capable of capturing the maximum value.
- Press the INRUSH button – INRUSH will now be shown on the screen.
- Activate the compressor and read value on the display.
- Press the INRUSH button to return to live readings.
Test Lead Notes
Cat IV and CAT II Measurement Locations
- Ensure the test lead shield is pressed firmly in place. Failure to use the CAT IV shield increases arc-flash risk.
CAT II Measurement Locations
- CAT IV shields may be removed for CAT II locations. This will allow testing on recessed conductors such as standard wall outlets. Take care not to lose the shields.
WARNING: Test Lead category protections apply only to test leads and
should not be confused with the meter’s specific CAT rating. Observe the
maximum category protection indicated on the meter the test leads are plugged
into.
CAUTION: If the test leads need to be replaced, you must use a new one
which should meet EN 61010-031 standard, rated CATIII 1000V or better.
Battery Replacement
- Rotate Battery Compartment Latches to open position
- Remove battery cover
- Replace the old batteries with 2 new (AAA) batteries
- Replace the battery cover
- Rotate Battery Compartment Latches to lock
Warranty
The DL579 is warranted to be free from defects in materials and workmanship
for a period of 1 year from the date of purchase. If within the warranty
period your instrument should become inoperative from such defects, the unit
will be repaired or replaced at UEi’s option. This warranty covers normal use
and does not cover damage which occurs in shipment or failure which results
from alteration, tampering, accident, misuse, abuse, neglect or improper
maintenance. Batteries and consequential damage resulting from failed
batteries are not covered by warranty. Any implied warranties, including but
not limited to implied warranties of merchantability and fitness for a
particular purpose, are limited to the express warranty. UEi shall not be
liable for loss of use of the instrument or other incidental or consequential
damages, expenses, or economic loss, or for any
claim or claims for such damage, expenses or economic loss. Warranty only
covers hardware and does not extend to software applications. A purchase
receipt or other proof of original purchase date will be required before
warranty repairs will be rendered. Instruments out of warranty will be
repaired (when repairable) for a service charge.
For more information on warranty and service,
contact:www.ueitest.com
Email: info@ueitest.com
1-800-547-5740
This warranty gives you specific legal rights. You may also have other rights,
which vary from state to state.
Disposal
CAUTION: This symbol indicates that equipment and its accessories shall be subject to separate collection and correct disposal.
Cleaning
Periodically clean your meter’s case using a damp cloth. DO NOT use abrasive, flammable liquids, cleaning solvents, or strong detergents as they may damage the finish, impair safety, or affect the reliability of the structural components.
Storage
Remove the batteries when instrument is not in use for a prolonged period of
time.
Do not expose to high temperatures or humidity. After a period of storage in
extreme conditions exceeding the limits mentioned in the General
Specifications section, allow
the instrument to return to normal operating conditions before using it.
Copyright © 2021 Kane USA Inc. All Rights Reserved. 202153 1021 RoHS Complint