TERRA UNIVERSAL 1788-26 IsoDry RH Control System Instruction Manual
- June 12, 2024
- TERRA UNIVERSAL
Table of Contents
Supplemental Manual
Document No. 1788-26
IsoDry™ RH Control System
1788-26 IsoDry RH Control System
Proprietary Notice
This manual pertains to proprietary devices manufactured by Terra Universal,
Inc. Neither this document nor any portion of it may be reproduced in any way
without prior written permission from Terra Universal.
Safety Notice
A thorough familiarity with all operating guidelines is essential to safe
operation of the product. Failure to observe safety precautions could result
in poor performance, damage to the system or other property, or serious bodily
injury or death.
The following symbols are intended to call your attention to two levels of
hazard involved in operation:
The information presented here is subject to change without notice.
Terra Universal makes no warranties applying to information contained in this
manual or its suitability for any implied or inferred purpose. Terra Universal
shall not be held liable for any errors this manual contains or for any
damages that result from its use.
CAUTION
Cautions are used when failure to observe instructions could result in
significant damage to equipment.
WARNING
Warnings are used when failure to observe instructions or precautions could
result in injury or death.
Introduction
This manual is intended to be a supplement to Doc. #180040: Desiccators and RH
Controllers that outlines the setup and operating instructions that are
specific to the IsoDry™ System and its components.
The defining feature of this system is the IsoDry™ Gas Dilution Module. The
Adjust-A-Shelf cabinet, Dual Purge, and NitroWatch are slightly modified to
integrate with the Gas Dilution Module, but are for the most part identical to
the standard versions described in Doc. #1800-40. The IsoDry™ RH Control
System is only compatible with Adjust-A-Shelf desiccators that carry the
IsoDry™ moniker.
IsoDry™ System Compatibility
Single Column IsoDry™ Adjust-A-Shelf™
Double Column IsoDry™ Adjust-A-Shelf™
This manual focuses on specific setup and operating instructions for the
IsoDry™ RH Control System.
For installation instructions and general specifications for the
Adjust-A-Shelf™ Cabinet, Dual Purge, or Nitro Watch, refer to Doc. #1800-40:
Desiccators and RH Controllers.
All of Terra’s IsoDry™ Desiccator Cabinets employ a similar nitrogen-flow
design:
- High-purity nitrogen gas is fed into a plenum at the rear of the desiccator (single-column design) or directly into one of the chambers (double-column design) where it mixes with the air inside the cabinet. When the Relative Humidity (RH) climbs above set-point level, a high-flow purge is activated.
- The IsoDry™ Gas Dilution Module continuously circulates this nitrogen-rich air mixture throughout the desiccator, further blending the nitrogen gas with the air and diluting any moisture present inside the cabinet.
- The influx of nitrogen gas pressurizes the desiccator, and air is released through the automatic relief/bleed (RB) valves, allowing the concentration of nitrogen gas in the desiccator to rise until the desired RH level is reached. At this time, the nitrogen purge reverts to a low flow.
The IsoDry™ System is made up of four main components:
- The IsoDry™ Desiccator Cabinet
- The IsoDry™ Dual Purge
- The IsoDry™ Nitro Watch
- The IsoDry™ Gas Dilution Module
IsoDry™ Dual Purge System
The Dual Purge System controls the nitrogen gas input. The system provides
two levels of nitrogen purging to protect your stored materials: Low-level
purge and high-level purge.
- Low-level purge maintains the minimum pressurization necessary to block out moisture and contaminants while the desiccator doors are closed.
- The high-level purge activates when the RH level rises above the set-point and is designed to minimize the impact of introducing outside air to the stored materials. An adjustable time delay maintains the high-level purge for a set time after the set-point is reached, followed by a return to low-level purging.
The IsoDry™ Dual Purge System is designed for enclosures that feature several
access doors. It relies on the IsoDry™ Nitro Watch to activate the high- and
low-level purges.
Each unit features a system line pressure gauge and regulator, internal
positive pressure gauge, a 2-amp fuse (located inside the housing), a
flowmeter, and all necessary gas line connections.
In addition, each unit features an alarm that flashes if the inlet line
pressure falls below 30 psi. This alarm alerts the user of a problem with the
supply line.
WARNING
Automatic RB Valves are required in nitrogen-purged desiccators to prevent
dangerous pressure buildup.
IsoDry™ Nitro Watch
The Nitro Watch offers set-point control of RH levels via automatic control
of the Dual Purge System and IsoDry™ Gas Dilution Module.
The NitroWatch senses and displays the relative humidity level (from ambient
to 0%) inside a desiccator within 2% RH.
It operates in tandem with the Dual Purge System and the Automatic RB Valves
to precisely and automatically control the flow of nitrogen into the
desiccator and maintain a preset humidity level, no matter how operating
conditions may change. It thus compensates for factors you may not have time
to worry about, such as how often parts are accessed, the size and condition
of the desiccator, variations in ambient humidity, and long periods without
operator supervision.
The NitroWatch consists of two basic components: a humidity sensor and an
indicator/controller unit.
The Humidity Sensor, which senses the humidity level, is mounted inside the
controlled environment. It employs a fast-response capacitive probe whose
capacitance is proportional to the humidity level. The sensor is capable of
measuring humidity over the entire humidity range, from 0 to 100% RH, with an
accuracy of ± 2% RH at a temperature of 68°F. The voltage output of the sensor
is directly proportional to the humidity.
A 3-prong connector allows easy connection of the sensor to the Control Unit.
The Control Unit provides the indicators and controls necessary to set the
desired humidity level and to measure the actual humidity level inside the
desiccator.
IsoDry™ Gas Dilution Module
The Gas Dilution Module, unique to IsoDry™ Desiccator Systems, rapidly
blends high-purity nitrogen with the ambient air and circulates the nitrogen-
rich mixture throughout the desiccator, essentially diluting the amount of
moisture present in the air. By expediting this process, the gas dilution
module provides three key benefits: it quickly lowers humidity levels after
moisture enters the cabinet, minimizes the moisture exposure time of sensitive
materials, and provides humidity uniformity throughout the cabinet. The speed
of the fans can be adjusted using the Speed Dial on the back of the Dual
Purge.
Installation
CAUTION
- Do not position the desiccator components so that the power supply and nitrogen source connections are inaccessible. The power supply serves as the main disconnect for the system.
- Never use alcohol or other cleaning agents on acrylic surfaces.
To prevent dangerously low oxygen levels and risk of asphyxiation, nitrogen-
purged systems should only be installed in a well-ventilated area.
Before installation and operation, carefully unpack the desiccator and
accessories and check for signs of damage or missing parts. Wipe down with a
particle-free cloth.
Installation requires a supply of clean, dry air or nitrogen (less than 5
ppm(v) H2O) and 1/4″ polyethylene tubing. The Dual Purge universal power
supply operates on 110/120VAC, 60Hz or 220/240VAC, 50Hz with no required
switch settings. All grounded cabinets feature grounding terminals in the
upper-right corner of the cabinet. For standard grounding, simply connect the
grounding terminal to your in-house ground.
IsoDry™ Dual Purge™ and IsoDry™ Nitro Watch™ Set-Up
The Nitro Watch draws its power from the Dual Purge System and will not
operate unless the Dual Purge System is plugged in and turned ON.
- Plug one end of the low-voltage phone cable into the “Purge Control” jack on the back of the Dual Purge and plug the other end into the “Purge Control” jack on the back of the NitroWatch.
- Plug the 5-pin cable into the “RH sensor” jack on the back of the Nitro Watch and connect it to the desiccator’s RH sensor.
- Use the 12V power cable to connect the IsoDry™ Gas Dilution Module to the “IsoDry Power” jack on the back of the Dual Purge.
- Insert the provided cord plug into the “Door Switch” jack. Leave the “R/H Output” jack empty.
- Use ¼” tubing to connect the nitrogen gas supply to the “Gas In” port on the back of the Dual Purge.
- Use ¼” tubing to connect the “Gas Out” port on the back of the Dual Purge to the nitrogen inlet port on the desiccator.
- Plug the 120VAC power cable into the Dual Purge and an appropriate grounded outlet.
CAUTION
Installation of at least one Terra Automatic RB Valve (Cat. #1600-60B) is
required to protect against permanent damage to seals and doors.
IsoDry™ System Setup (Rear View)
Operation
Dual Purge System Operation
-
Zero Calibration: This routine is run before shipping to ensure that the pressure sensor that activates the Low Pressure Alarm is properly calibrated. You need to run the routine only if the Low Pressure Alarm is activated even though the Dual Purge is connected to a nitrogen supply with at least 30 psi of pressure.
To perform the zero calibration routine, disconnect the incoming gas supply line (Any pressure on the internal pressure transducers will be subtracted from the reading during normal operation, resulting in incorrect low-pressure alarms).
Press and hold SET and UP buttons while turning on system power. The display will indicate “CAL” and then automatically set the zero value. -
Initial Programming: Turn ON the Dual Purge System while depressing SET (the left button on the front control panel). Each time you release and then press and hold the SET button, you will advance through the following control functions. Use UP/DOWN to change default settings (shown in parentheses):
Prg 1 Purge Delay (minutes): Number of minutes high purge remains active after door is closed (default: 0)
Prg 2 Purge Delay (seconds): Number of seconds that the high purge remains active after door is closed (default: 5). If operating the Dual Purge without a NitroWatch, you should increase this setting to 60.
door After switching to high purge, the number of seconds to delay before turning on the beeper alarm.
beeP State of the BEEPER enable (default: ON)
Glo Activates GLOVE BOX control mode–Press UP to turn ON and proceed to glove box pressure setting, DOWN to turn OFF (default: OFF). In GLOVE BOX mode, this setting lets you adjust the pressure (milliinches WG) at which the high purge is activated
(default: 0.2). See “zero calibration” above.
DonE Settings are complete; press SET once more to begin operation.
Made a mistake? To reset the system, turn the system OFF while in setup mode and restart while holding the SET button. -
Operation Display Functions: After completing initial programming, press the specified button to view these operating conditions:
High Purge Bypass: Press and hold SET
In some applications (especially using fine powders in a glove box), you may wish to deactivate the high purge function, which could create turbulence. To do so, press DOWN while holding SET. To activate high purge, hold UP while holding SET.
Incoming Line Pressure: Press and hold UP (displays line pressure in PSI)
Turn the pressure regulator (the round knob on the right side of the Dual Purge control panel) until the pressure gauge reads 30 psi (for bench-top desiccators) or 40 psi (for larger standing desiccators). Note that regulator knob is reverse-threaded: rotate clockwise to open the valve and increase pressure, or counterclockwise to close the valve and reduce pressure. For safe operation, pressure should be externally regulated below 70 psi.
Internal Pressure: Press and hold DOWN (displays internal cabinet pressure in milli-inches of WG)
On systems with very large plastic access doors, a continuous purge can cause temporary door deflection and gas leakage. This condition does not indicate a permanent leak and generally does not damage the cabinet. When pressure is removed, doors return to their natural position and re-establish an effective seal. This deflection does, however, result in nitrogen waste and indicates the need to reduce the flowmeter setting.
Purge Timer: Press and hold UP and DOWN simultaneously to review the number of seconds the high-flow purge is active after set-point is reached. To change this value, see “Initial Programming” above.
“Hi” Indicator indicates that the Dual Purge System is in high-purge mode (you will see at least one bar to the right of this display, which indicates an internal positive pressure).
Flashing “Lo Pres” indicates the absence of incoming line pressure. Check the line for proper connection, and check the nitrogen source to ensure incoming flow. -
Flowmeter Operation
Flowmeter Operation with Dual Purge System:
The Dual Purge System includes a flowmeter, which provides a continuous low- level purge to maintain a constant positive pressure inside the cabinet. The flowmeter also ensures uninterrupted nitrogen flow if power or system electronics ever fail.
If continuous purging is desired, open the flowmeter to 5 – 10 SCFH (depending on chamber size). This is recommended if you require a very low humidity set point (less than 10%RH) or if doors are frequently opened. Continuous purging will tend to drive the %RH level very low.
If continuous purging is not required, close the flowmeter. The Dual Purge System will initiate a gas purge only when the humidity level climbs above the set point.
Flowmeter Operation without Dual Purge System:
If Dual Purge System is not ordered, flowmeter is included (only applies to ISO Dry Desiccator cabinets).
Set the flowmeter to between 5 and 20 SCFH (depending on desiccator size and moisture requirements). Connect ¼” tubing from an externally regulated gas source (20-40 psi, depending on desiccator size) to flow meter inlet. Push tubing in for connection; to release, depress connector ring and pull tubing out. If flowmeter replacement is needed, order flowmeter kit (1600-32-ISO). -
Dilution Fan Adjustment: The dilution fan speed control, located on the rear of the Dual Purge panel, allows fan speed adjustment. Turn the knob clockwise to speed up the fan (maximizing dilution efficiency), or counterclockwise to slow down the fan
(reducing turbulence inside the cabinet).
Nitro Watch™ Operation
-
Initial Programming: Turn ON the Dual Purge System while holding down the SET button on the Nitro Watch control panel (the left button on the front control panel).
Each time you release and then press and hold the SET button, you will advance through the following control functions. Use the UP/DOWN buttons to change default settings (shown in parentheses):
Door Number of seconds to delay after the door is opened before turning on the beeper alarm.
rH Relative Humidity High Purge Delay Alarm: No. of seconds the RH level must remain above the set point to activate the alarm (default: 60)
beeP Enables/Disables Beeper Alarm (default: ON)
hU HUMEX operation selector – when in the OFF mode, the system operates as a Nitro Watch (maintaining below-ambient humidity). In the ON mode, the system operates with Terra’s Humex2 (maintaining above-ambient humidity). Default: OFF.
Add ADJUST parameter: Use this function to compensate for measured discrepancies between the Nitro Watch %RH readout and that of an independent %RH calibrator. Press the UP/DOWN keys to adjust the scale up or down. (Default: 0)
This adjustable offset value will wrap around if the maximum or minimum is exceeded. For example, if you try to set the adjust value above the maximum scale offset of 49.5, the value will wrap around to –50. Also, a side effect of using an offset value other than zero is that the R/H range will be reduced. For example, if the offset value is –10, then the maxi um R/H value that can occur is 90 because the input value (100) will be added to the adjust value before the system uses it.
Alr ALARM time selector, which allows you to specify the alarm scale determined above in either minutes – LONG – or seconds – SHrt. (default: SHrt)
Press SET again to advance to Normal Run Mode.
Made a mistake? To reset the system, turn the Nitro Watch OFF while in setup mode and restart while holding the SET button. -
Adjust Humidity Set Point: After completing initial programming, press the SET button to view the current % RH set point (default setting: 15%). Use the UP/DOWN keys to change this set point. Wait for a couple seconds without pressing any buttons to save the new set-point.
-
Beeper Silencing: Press any front panel button to silence the beeper during an alarm condition.
The Nitro Watch will now activate high-flow purging whenever the %RH level inside the desiccator exceeds the specified set point. Your desiccator is ready for operation.
CAUTION
Installation of at least one Terra Automatic RB Valve (Cat. #1600-60B) is required to protect against permanent damage to seals and doors
For optimal efficiency of gas delivery and reduced %RH recovery times, Terra recommends factory installation of one RB valve in each desiccator chamber.
If the Dual Purge System remains at high-level purge, or if it frequently
fluctuates between high and low-level purge, you need to increase your
flowmeter setting. Increase the flowmeter setting until the humidity level
falls a few percent below your set point. At this flowmeter setting, the
system will be able to maintain the desired humidity while on the low-level
purge—and save nitrogen.
Because the humidity sensor is exposed to moisture during shipping, the system
must generally operate for a couple of days in a dry nitrogen environment
before the sensor dries out and delivers completely accurate readings.
Desiccator Service and Maintenance
CAUTION
- Do not clean acrylic with alcohol or other strong cleaning agents.
- Do not expose static-dissipative PVC to extreme heat or direct sunlight.
- A Terra stainless steel shield is required on the bottom of each static-dissipative PVC desiccator chamber to prevent scratching.
Refer to the Doc. #1800-40: Desiccators and RH Controllers for information regarding the care and service of your desiccator.
Troubleshooting
Terra Universal’s desiccators are designed to provide years of reliable,
efficient service. If you should experience any problems that arise during
operation of your desiccator with the Dual Purge System and Nitro Watch, refer
to the appropriate troubleshooting procedure below. If the problem persists,
or if you encounter any problems not described below, call Terra Universal for
additional assistance.
WARNING
Do not attempt to disassemble any of the modules. Contact Terra for
assistance.
PROBLEM: System won’t turn on.
POSSIBLE SOLUTIONS:
- Make sure that the power cord of the Dual Purge System is plugged into an appropriate outlet and that the two telephone cables to the NitroWatch and desiccator are properly connected.
- Make sure that the power switch of the Dual Purge System is in the ON position.
- Check the fuse of the Dual Purge System. The 2 amp fuse, mounted on the circuit board controller, is accessible once the stainless steel housing cover is removed.
PROBLEM: System stays in high purge at all times.
POSSIBLE SOLUTIONS:
- Check door seals of the desiccator to make sure that there are no leaks.
- If you require a very low humidity level, you may need to increase the flowmeter setting on the Dual Purge System. You should increase the flowmeter setting until the system maintains a humidity level a few percent below the humidity set point without switching to the high-level purge. Remember, though, that the Dual Purge System should be able to maintain the humidity level you require with a low-level purge as long as access doors remain closed.
- If you are operating a large desiccator and require a low humidity level, your system may need to operate on high purge much of the time to compensate for the hygroscopic characteristics of the acrylic (or static-dissipative PVC) desiccator walls. These materials absorb moisture from outside of the desiccator and pass it inside. The higher the difference between the external and internal humidity levels, the more nitrogen you will need in order to remove this moisture. Yet if you operate on high-flow purge most of the time, you consume large amounts of nitrogen and run the risk of pressure build-ups.
PROBLEM: The system constantly switches between high and low purge.
POSSIBLE SOLUTIONS:
- Make sure that all access doors are fully closed. Check door seals for leaks.
- You should increase the flowmeter setting until the Dual Purge System maintains a humidity level a few percent below the set point without switching to the high-level purge.
PROBLEM: The desiccator is leaking excessively.
POSSIBLE SOLUTIONS:
- Check the condition of the gasket on the desiccator. If it is cracked or peeling, call Terra for a replacement.
- Check door alignment. Leaks may develop if doors are out of alignment.
PROBLEM: Line pressure gauge does not display any pressure, or displays the “Low Pressure” alarm at all times.
POSSIBLE SOLUTIONS:
- Supply gas is down, or for some reason is not reading the Dual Purge system. Check your supply line for kinks or blockage.
- If your gas system utilizes a filter, it may be clogged and need to be replaced.
- Check the line pressure regulator to make sure it is not closed
- Run the “Zero Calibration” routine described in Section 3.2
PROBLEM: The Nitro Watch delivers an obviously incorrect humidity reading.
POSSIBLE SOLUTIONS:
- Check the Nitro Watch sensor connection on the rear panel of the control module. If the connection to the sensor is good, and the unit still fails to deliver an accurate reading, contact Terra Universal.
PROBLEM: The Nitro Watch delivers a low-pressure alarm.
POSSIBLE SOLUTIONS:
-
Turn off the Dual Purge system using the back switch.
-
Disconnect the N2 gas line.
-
Open the flowmeter and regulator.
-
Simultaneously press and hold the “Set” and “Up” buttons while turning on the Dual Purge system. The display will show “CAL”. Press “Set” to proceed to the next setting.
-
Set under the “CAL” menu,
a. “Prg1” to 0 minutes
b. “Prg2” to 5 seconds
c. “Door” to 60
d. “Beep” to ON
e. “Glov” to OFF -
Re-connect the N2 gas line, set the flowmeter to 7 SCFH, and set the regulator to 30 psi.
PROBLEM: The Nitro Watch displays the “door” alarm.
POSSIBLE SOLUTIONS:
- Make sure that a cord plug is present in the Door Switch input on the rear panel of the Nitro Watch control module.
Specifications
Refer to the original order form for the exact specifications/configuration of
your desiccator.
WARNING
Internal chamber pressure must never exceed 0.5 psi when the desiccator is
purged with nitrogen.
IsoDry™ Dual Purge System
Operating Temperature: | 50°F to 140°F |
---|---|
Overall Dimensions: | 12″W x 11 ¼”D x 7″H |
Weight: | 16 lbs. |
Power Requirement: | 120/220 VAC, 50/60 Hz, 0.5 Amp |
Universal power supply requires no switch settings.
Degree of Protection:| IPX0
Max Gas Inlet Pressure:| 50 psi
IsoDry™ Nitro Watch
Operating Temperature: | 50°F to 140°F |
---|---|
Dimensions: | 11″W x 6 ½”D x 3″H |
Power Requirements: | 12 VDC (Supplied by Dual Purge) |
Sensor Dimensions: | 11/2″ x ¾” x 4″ |
Case Material: | Stainless steel |
Display: | 3 ½ digit LED display |
Electrical Connections: | Screw terminals |
Output: | 0-5 V |
Measuring Range: | 0-100% RH |
Accuracy (at 20° C): | ± 2% RH |
Display Resolution: | ± 0.1% RH |
Temp. Dependence: | ± 0.04% RH/°C |
Sensor Calibration: | None required, but scale offset routine allows display |
compensation. Capacitive sensor pick up should be tested and replaced as necessary (about every 5 years under normal use)
IsoDry™ Gas Dilution Module
Operating Temperature: | 50°F to 140°F |
---|---|
Dimensions: | 13 ½”W x 7”H x 2”D |
Power Requirement: | 12 V (Supplied by Dual Purge) |
Housing Material: | Stainless steel |
For a complete list of specifications for the Adjust-A-Shelf™ desiccator cabinet and the modules listed above, refer to Doc. #1800-40: Desiccators and RH Controllers.
Warranty
For more information about our warranty system, please visit our company’s warranty website.
Thank you for ordering from
Terra Universal!
© Copyright 2023 Terra Universal Inc. All rights reserved.
Revised June 2023
Terra Universal, Inc.
TerraUniversal.com
800 S. Raymond Ave.
Fullerton, CA 92831
TEL: 714-578-6000
FAX: 714-578-6020
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>