Terra Universal CleanBoothTM Laminar Flow Station User Manual
- June 1, 2024
- TERRA UNIVERSAL
Table of Contents
Clean Booth TM Laminar Flow Station
User Manual Terra Universal Product Manual Document No. 1800-73
CleanBooth™ Laminar Flow Station
CleanBoothTM Laminar Flow Station
General Information
The laminar flow station is ideal for kitting operations, assembly, screen printers and other free-standing or benchtop processing equipment with the ability to effectively block the inrush of contaminants. The backbone of the system is one or more filter/fan units (FFUs), each with industry-leading low-noise impeller blowers. Horizontal flow versions have the FFUs mounted on the side of the frame. The catered frame positions the FFU behind benchtop or large process equipment for repair operations wherever you need a temporary horizontal laminar flow of filtered air. Following the instructions and recommended maintenance will help insure a long and efficient service life from the unit.
1.1 Part Numbers Covered by this Manual
Part Numbers| Configuration| Width| Depth|
Height| Voltage| Airflow Direction
---|---|---|---|---|---|---
1870-01A| Knockdown| 48.25”| 26.3”| 70.1”| 120 V| Vertical
1870-02A| Knockdown| 72.5”| 26.3”| 70.1”| 120 V| Vertical
1870-03B| Knockdown| 49.25”| 49.25”| 84.1”| 120 V| Vertical
1870-03B-220| Knockdown| 49.25”| 49.25”| 84.1”| 220 V| Vertical
1870-04B| Knockdown| 96.25”| 48.25”| 84.2”| 120 V| Vertical
1870-04B-220| Knockdown| 96.25”| 48.25”| 84.2”| 220 V| Vertical
1870-05A| Knockdown| 72.5”| 26.3”| 70.1”| 120 V| Vertical
1870-05A-220| Knockdown| 72.5”| 26.3”| 70.1”| 220 V| Vertical
1870-06B| Knockdown| 49.25”| 50.25”| 80”| 120 V| Vertical
1870-06B-220| Knockdown| 49.25”| 50.25”| 80”| 220V| Vertical
1870-07B| Knockdown| 96.25”| 48.25”| 84.2”| 120 V| Vertical
1870-07B-220| Knockdown| 96.25”| 48.25”| 84.2”| 220 V| Vertical
1870-08B| Knockdown| 96.25”| 48.25”| 84.2”| 120 V| Vertical
1870-08B-220| Knockdown| 96.25”| 48.25”| 84.2”| 220 V| Vertical
1870-09B| Knockdown| 96.25”| 48.25”| 84.2”| 120 V| Vertical
1870-09B-220| Knockdown| 96.25”| 48.25”| 84.2”| 220 V| Vertical
1870-20| Knockdown| 48.25”| 29.5”| 71”| 120 V| Horizontal
1870-21| Knockdown| 100.5”| 29.5”| 71”| 120 V| Horizontal
1870-22| Knockdown| 48.25”| 29.5”| 71”| 120 V| Horizontal
1870-22-220| Knockdown| 48.25”| 29.5”| 71”| 220 V| Horizontal
1870-23| Knockdown| 72.5”| 29.5”| 71”| 120 V| Horizontal
1870-23-220| Knockdown| 72.5”| 29.5”| 71”| 220 V| Horizontal
1870-25| knockdown| 102”| 29.5”| 71”| 120 V| Horizontal
1870-25-220| Knockdown| 102”| 29.5”| 71”| 220 V| Horizontal
Proprietary Notice| Safety Notice
---|---
This manual pertains to proprietary devices manufactured by Terra Universal,
Inc. Neither this document nor any portion of it may be reproduced in any way
without prior written permission from Terra Universal.| A thorough familiarity
with all operating guidelines is essential to safe operation of the product.
Failure to observe safety precautions could result in poor performance, damage
to the system or other property, or serious bodily injury or death. The
following symbols are intended to call your attention to two levels of hazard
involved in operation.
Terra Universal makes no warranties applying to information contained in this
manual or its suitability for any implied or inferred purpose. Terra Universal
shall not be held liable for any errors this manual contains or for any
damages that result from its use.| | Cautions are used when failure to
observe instructions could result in significant damage to equipment.
| Warnings are used when failure to observe instructions or precautions
could result in injury or death.
The information presented here is subject to change without notice.
Critical Operation Conditions
- Because of their panel design, soft wall cleanrooms cannot control pressure, humidity or temperature levels inside the enclosure, making them unsuitable for many requirements, including many ISO 5 applications. The flexible curtains can also be more difficult to clean than rigid hard wall panels.
- Please refer to our other cleanrooms if pressure, humidity and/or temperature control or frequent cleaning, sterilization is required.
Installation
Step 1:
a. Part number 1870-01A: Connect both corner frames pieces with middle frame
and bottom frame rods.
b. Part numbers 1870-02A & 1870-05A: Connect indicated parts shown.
c. Part numbers 1870-03B & 1870-06B: Connect indicated parts shown.
d. Part numbers 1870-05B, 1870-07B, 1870-08B & 1870-09B: Connect indicated parts shown.
Step 3:
a. 1870-03B / 1870-03B-220 – Slide the right frame into the left frame.
b. 1870-04B / 1870-04B-220 – Attach the left and right frame to the back frame.
a. Attach the top frame to the base assembly.
Step 4 (Optional): Apply gasketing along the ceiling grid perimeter.
Step 5: Once the frame structure is assembled, the LED light panel wiring must be installed through the openings provided and secured with sheet metal screws.
Step 6: Place the fan filter units onto the ceiling grid.
Step 7: Raise the outer cover of the Grip-Track™ mount. Seat the curtain bead inside the track, and snap the cover down to secure the curtain.
3.1 System Operation
- Plug the power cord into an appropriate grounded receptacle (115VAC, 50/60Hz or 230VAC, 50/60Hz).
- FFUs and light are controlled by means of a single ON/OFF switch, located on the right-rear portion of the work area.
- The FFUs incorporate a 3-position impeller blower, which is normally shipped in the “High” position. Under normal operating conditions, this setting should not be changed. The blower speed control is located on the upper-right side of each unit.
3.2 Specifications
Terra Part #| Voltage| Airflow Direction|
Inside Width| Inside Depth| Inside Height
---|---|---|---|---|---
1870-01A| 120 V| Vertical| 48.25”| 26.3”| 70.1”
1870-02A| 120 V| Vertical| 72.5”| 26.3″| 70.1”
1870-03B| 120 V| Vertical| 49.25”| 49.25″| 84.1″
1870-03B-220| 220 V| Vertical| 49.25″| 49.25″| 84.1″
1870-04B| 120 V| Vertical| 96.25”| 48.25″| 84.2”
1870-04B-220| 220 V| Vertical| 96.25”| 48.25″| 84.2″
1870-05A| 120 V| Vertical| 72.5”| 26.3″| 70.1″
1870-05A-220| 220 V| Vertical| 72.5”| 26.3″| 70.1″
1870-06B| 120 V| Vertical| 49.25”| 50.25″| 80″
1870-06B-220| 220 V| Vertical| 49.25″| 50.25″| 80″
1870-07B| 120 V| Vertical| 96.25”| 48.25″| 84.2″
1870-07B-220| 220 V| Vertical| 96.25”| 48.25”| 84.2”
1870-08B| 120 V| Vertical| 96.25”| 48.25”| 84.2”
1870-08B-220| 220 V| Vertical| 96.25”| 48.25”| 84.2”
1870-09B| 120 V| Vertical| 96.25”| 48.25”| 84.2”
1870-09B-220| 220 V| Vertical| 96.25”| 48.25”| 84.2”
1870-20| 120 V| Horizontal| 48.25”| 29.5”| 71”
1870-21| 120 V| Horizontal| 100.5”| 29.5”| 71”
1870-22| 120 V| Horizontal| 48.25”| 29.5”| 71”
1870-22-220| 220 V| Horizontal| 48.25”| 29.5”| 71”
1870-23| 120 V| Horizontal| 72.5”| 29.5”| 71”
1870-23-220| 220 V| Horizontal| 72.5”| 29.5”| 71”
1870-25| 120 V| Horizontal| 102”| 29.5”| 71”
1870-25-220| 220 V| Horizontal| 102”| 29.5”| 71″
3.2.1 Strip Curtain
- Static-dissipative vinyl helps prevent surface charges and particle attraction.
- Tests performed per ASTM D257 show surface resistivity of 1.8 x 107 ohms/sq.
- Heavy-gauge materials hang straight, without the need for sewn hems or sash weights.
- Static-Dissipative PVC includes built-in static inhibitors.
- Terra offers panels made from static-dissipative polyurethane that provides permanent, humidity- independent ESD protection from sensitive manufacturing conditions.
- This material contains no plasticizer, making it virtually non-gassing.
3.2.2 Fan Filter Unit
WHISPERFLOW® FFU PERFORMANCE DATA
Nominal Unit Size (ft.)| Filter| Sound (dB) @90FPM| AIR
FLOW (CFM)
HIGH| MED| LOW
2’ x 4’| HEPA| 49| 800| 720| 590
2’ x 4’| ULPA| 49| 660| 640| 580
WHISPERFLOW® FFU ELECTRICAL DATA
Nominal Unit Size (ft.)| Filter| Motor HP| Voltage| Full Load Amps| Frequency
(Hz)| Power Consumption (Watts)
HIGH| MED| LOW
2’ x 4’| HEPA| 1/4| 120| 3.8| 50/60| 420| 380| 360
2’ x 4’| HEPA| 1/4| 208-230/240| 1.8/1.6| 50/60| 380| 310| 240
2’ x 4’| ULPA| 1/4| 120| 3.8| 50/60| 440| 425| 415
Maintenance
4.1 Filter Replacement
Air velocity should be monitored to ensure effective filtration using an
appropriate air-speed monitor. Under typical conditions (8 hours of operation
per day), filters should be replaced every year. As a HEPA filter clogs, you
will notice increased blower noise and vibration, along with decreased air
speed. The filter assembly is separate from the fan module, which stays in
place in the ceiling grid.
Topside replaceable FFUs require the units to be removed from the hood for
filter replacement.
- Remove the unit from the ceiling grid.
- Place the unit on a flat work surface and remove the 10 sheet metal screws that hold the HEPA/ULPA filter to the case.
- Replace with a new HEPA/ULPA filter carefully handling edges only.
- Secure filter to the case with the same screws used previously.
- Place unit back onto the ceiling grid.
4.2 CleanBooth™ Frame Cleaning and Care
A thorough cleaning of coated components is required to conserve the façade
decorative appearance and to reduce the corrosion strain. Powder coated
surfaces proper maintenance and regular servicing are prerequisites for claims
related to any guarantee and require regular cleaning at least once per year.
When a coated component is soiled during transportation, storage or assembly,
cleaning the component must take place immediately with clear cold or
lukewarm water.
- Buildings must be cleaned more often when they are located in severe polluted environments such as; a region with increased salt contamination and/or chemical exhausts, a direct area of influence or within the vicinity of an industrial or chemical enterprise, the immediate vicinity of a sea coast or within a defined chemical/radioactive precipitation zone.
- A neutral or weak alkaline detergent can be used against severe soiling.
- Use only clean water with slight additives of neutral washing agents (pH 5-8) with the aid of non-abrasive soft cloths, rags or industrial cotton. Avoid strong rubbing.
- Use white spirit free of aromatic compounds or isopropyl alcohol (IPA) to remove greasy, oily or sooty substances. Can also be applied for adhesives, silicone cartouche, adhesive tapes or other residues.
- Use unidirectional (not circular) motion to remove surface contaminants from the edges of all surfaces.
- Do not use solvents or similar diluents containing ester, ketones, alcohol, aromatics, ethylene glycol or halogenated hydrocarbon.
- Do not use scratching or abrasive agents.
- Do not use strong acids, alkaline detergents or nitrifiers.
- Do not use detergents of unknown composition.
- Do not use stream-jet devices.
- Maximum exposure period of detergents must not exceed one hour. When necessary, the entire cleaning process can be repeated at least after 24 hours.
- Rinse with clean cold water immediately after every cleaning process.
- Detergents must not be used at temperatures higher than 77 °F (25 °C).
- Façade components surface temperature must not exceed 77 °F (25 °C).
- Joint sealants and other aids such as glazing aids, lubricant agents, drilling and cutting lubricants which come into contact with coated surfaces, must be pH-neutral and free of paint-damaging substance. They must be first subjected to a suitability test.
4.3 Vinyl Panel Maintenance and Care
-
For critical cleanliness applications cleaning should be done at the beginning and end of each work shift.
-
Follow your specific cleaning protocols as they vary greatly across industries and applications.
-
Static-dissipative characteristics may degrade over time and require curtain replacement after two years (depending on temperature, humidity, UV exposure and other environment conditions).
-
Static dissipative testing is recommended in static-sensitive applications.
-
Anti-static vinyl’s have additive materials that continuously provide a naturally occurring hydro-scope surface which is essential for the continuation of the anti-static process. Over time and with exposure the surface will develop a viscous feel. To remove the viscous coating clean periodically.
-
To clean standard PVC and Anti-Static Strips / Soft Wall curtains use mild dish soap and water solution with any biodegradable cleaner.
– Wash the front and back of the strips and or curtains and rinse with clear water then wipe dry with a clean, lint-free soft cloth. -
To clean anti-static vinyl use 50/50 solution of isopropyl alcohol and Di water.
– Wash the front and back of the strips or curtains using a soft non-abrasive lint-free cloth.
Note: Excessive cleaning using harsh or abrasive cleaning compounds such as ammonia, 100% isopropyl alcohol, liquids with chlorinates, antifungals, highly active sterile cleaning solutions and dry powder additives are not recommended and will increase the aging process.
Replacement Orders
Provide sales associate with pertinent information, such as serial number, model number and date, for replacement parts or a new knock-down CleanBooth™ Laminar Flow Station. The label for the CleanBooth™ is on the exterior left side of the frame and the label for the Fan Filter Unit is on the lower part of the FFU assembly.
Order Number __
Serial Number __
Unit Model Number_
Date ___
Replacement Parts
Curtain Panel Kits – Includes Grip-Track™, wall panels and strip curtains
Part #| Material| Width| Depth| Height
1330-09| Anti-Static PVC| 48”| 48”| 73”
1330-01A| Anti-Static PVC| 48”| 24”| 70”
1330-06A| Anti-Static PVC| 72”| 24”| 64”
1330-11| Anti-Static PVC| 96”| 49”| 73”
1330-10| Clear Polyurethane| 48”| 48”| 73”
1330-07| Clear Polyurethane| 72”| 24”| 64”
1330-12| Clear Polyurethane| 96”| 49”| 73”
1330-14| Clear Polyurethane| 48”| 24”| 70”
1330-00A| Vinyl| 48”| 24”| 70”
1330-05A| Vinyl| 72”| 24”| 64”
1330-08| Vinyl| 48”| 48”| 73”
1330-13| Vinyl| 96”| 49”| 73”
Strip Curtains
Part #| Material| Size
1330-15| Anti-Static PVC| 53″ Width
1330-16| Anti-Static PVC| 77″ Width
1330-17| Anti-Static PVC| 100″ Width
Filters
Part #| Size| Type| Material
6601-25| 2’ x 4’| HEPA| Aluminum
6601-28| 2’ x 4’| ULPA| Aluminum
6601-28-S| 2’ x 4’| ULPA| 304 Stainless Steel
Fan Filter Units
6601-24-H| Whisper Flow 120 V
6601-24-H-220| Whisper Flow 220 V
Warranty
Products Manufactured by Terra: Terra Universal, Inc., warrants products
that it manufactures to be free from defects for a period of 12 months for
parts and 90 days for labor, commencing from the date of shipment. Terra’s
sole responsibility is to repair or replace, at its option, any part of the
product that proves defective or malfunctioning during this time limit. In
some cases, components incorporated in Terra Universal products are covered by
additional warranties from component manufacturers; obtain specific
information from Terra sales representatives. This warranty is void if the
equipment is abused or modified by the customer, is operated outside Terra’s
operating instructions or specifications, or is used in any application other
than that for which it is specified. This warranty does not include routine
maintenance or service procedures, breakage of quartz baths after 60 days,
shipping damage, nor damage from misuse, intentional or unintentional abuse,
neglect, natural disasters, or acts of God.
Products Manufactured by Others: Terra Universal, Inc., warrants that, to
the best of its ability, Terra’s representations of products that are
manufactured by others reflect the manufacturer’s representations, subject to
change without notice. Sole warranty for these products is the original
manufacturer’s warranty that is passed forward to the purchaser and
constitutes the customer’s sole remedy for these products. Detailed warranties
for distributed products are available through Terra sales representatives.
Freight Shortage or Damage: Upon receipt of any equipment from Terra
Universal, Inc., customer shall immediately unpack and inspect for damage or
shortage. The customer shall not accept a damaged package or a short shipment
until the carrier makes a “damage or shortage” notation on both the carrier’s
and customer’s copy of the freight bill or delivery receipt. Service title
passes when the shipment is loaded, so customer is responsible for filing and
collecting a freight claim. Any replacement products must be ordered and paid
for separately. For Terra’s “Policy and Procedures for Returning Goods,” see
Terra’s Internet site:
www.TerraUniversal.com.
Generally, customers can improve the chance of collecting on a freight claim
by following these procedures: 1) formally requesting that the carrier inspect
the shipment immediately upon suspecting damage or shortage to verify
condition; 2) notifying the carrier upon discovery of concealed damage and
requesting an inspection within 15 days of receipt, both in person or phone
and following up via mail; 3) keeping the shipment as intact as possible,
including retaining original packaging materials and keeping the product as
close to the original receiving location as possible; 4) holding salvage for
disposition by the carrier.
All Claims: Terra Universal expressly disclaims all other warranties,
expressed or implied or implied by statute, including the warranties of
merchantability or fitness for intended use. Terra Universal is not
responsible for consequential or incidental damages arising out of the
purchase or use of the products supplied by Terra Universal. Terra Universal
is not liable for damage to facilities, other equipment, products, property or
personnel of others, or of their agents, suppliers, or affiliated parties,
which is caused or alleged to have been caused by products supplied by Terra
Universal. In any event or series of events, Terra Universal’s total liability
for any and all damages whatsoever is limited to the lesser of the actual
damages or the original invoice cost of the items alleged to have caused the
damage. The customer’s sole and exclusive remedy for any cause of action
whatsoever is repair or replacement of the non-conforming products or refund
of the actual purchase price, at the sole option of Terra Universal. All
claims must be made in writing within 90 days of the date the product was
shipped. Any claims not made within this time limit shall be deemed waived by
the customer. Terra Universal is not responsible for any additional costs of
repair caused by poor packaging or in-shipment damage during return.
Warranty Returns: All warranty returns must be authorized in advance by
Terra Universal and approved under an RMA. Unless approved in advance for good
reason, all returns must be in original condition, including all manuals, and
must be packaged in original packaging materials. All returned goods are to be
shipped to Terra Universal, freight prepaid at customer’s expense. See Terra’s
“Policy and Procedure for Returned Goods.”
Thank you for ordering from Terra Universal!
Terra Universal, Inc.
• TerraUniversal.com
• 800 S. Raymond Ave.
• Fullerton, CA 92831
• TEL: 714-578-6000
• FAX: 714-578-6020