KEITHLEY 7710 Multiplexer Module Instructions
- June 12, 2024
- KEITHLEY
Table of Contents
- Introduction
- Con nections
- Module configuration
- Typical connections
- Connection log
- Installation
- Remove a switching module
- Operating instructions
- Operating considerations
- Calibration
- Temprerature calibration
- Factory service
- Safety precautions
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
7710 Multiplexer Module
Instructions Model 7710
Multiplexer Module
Instructions for use with DAQ6510
Keithley Instruments
28775 Aurora Road
Cleveland, Ohio 44139
1-800-833-9200
tek.com/keithley
Introduction
The 7710 20-channel Solid-state Differential Multiplexer with Automatic Cold
Junction Compensation (CJC) module offers 20 channels of 2-pole or 10 channels
of 4-pole relay input that can be configured as two independent banks of
multiplexers. The relays are solid state, providing long life and low
maintenance. It is ideal for long-term data logging applications and for
demanding high-speed applications.
Figure 1: 7710 20-Channel Differential Multiplexer Module Item
shipped may vary from model pictured here.
The 7710 includes the following features:
- Fast-actuating, long-life solid-state relays
- DC and AC voltage measurement
- Two-wire or four-wire resistance measurements (automatically pairs relays for four-wire measurements)
- Temperature applications (RTD, thermistor, thermocouple)
- Built-in cold junction reference for thermocouple temperature
- Screw terminal connections
NOTE
The 7710 can be used with the DAQ6510 Data Acquisition and Multimeter System.
If you are using this switching module with the 2700, 2701, or 2750, please
see Model 7710 Multiplexer
Card User’s Guide, Keithley Instruments PA-847.
Con nections
Screw terminals on the switching module are provided for connection to device
under test (DUT) and external circuitry. The 7710 uses quick-disconnect
terminal blocks. You can make connections to a terminal block when it is
disconnected from the module. These terminal blocks are rated for 25 connects
and disconnects.
WARNING
Connection and wiring procedures in this document are intended for use by
qualified personnel only, as described by the types of product users in the
Safety precautions (on page 25). Do not perform these procedures unless
qualified to do so. Failure to recognize and observe normal safety precautions
could result in personal injury or death.
The following information describes how to make connections to the switching
module and define the channel designations. A connection log is provided that
you can use to record your connections.
Wiring procedure
Use the following procedure to make connections to the 7710 module. Make all
connections using the correct wire size (up to 20 AWG). For maximum system
performance, all measurement cables should be less than three meters. Add
supplementary insulation around the harness for voltages above 42 VPEAK.
WARNING
All wiring must be rated for the maximum voltage in the system. For example,
if 1000 V is applied to the front terminals of the instrument, the switching
module wiring must be rated for 1000 V. Failure to recognize and observe
normal safety precautions could result in personal injury or death.
Equipment needed:
- Flat-blade screwdriver
- Needle-nose pliers
- Cable ties
To wire the 7710 module:
-
Make sure all power is discharged from the 7710 module.
-
Using a screwdriver, turn the access screw to unlock and open the cover, as shown in the following figure.
Figure 2: Screw terminal access -
If needed, remove the appropriate quick-disconnect terminal block from the module.
a. Place a flat-head screwdriver under the connector and gently push up to loosen it, as shown in the following figure.
b. Use needle-nose pliers to pull the connector straight up.
CAUTION
Do not rock the connector from side to side. Damage to the pins could result.
Figure 3: Proper procedure to remove terminal blocks -
Using a small flat-blade screwdriver, loosen the terminal screws and install the wires as needed. The following figure shows the connections, including the connections to source and sense.
Figure 4: Screw terminal channel designations -
Plug the terminal block into the module.
-
Route wire along the wire path and secure with cable ties as shown. The following figure shows connections to channels 1 and 2.
Figure 5: Wire dressing -
Fill in a copy of the connection log. See Connection log (on page 8).
-
Close the screw terminal access cover.
-
Using a screwdriver, press in the access screw and turn to lock the cover.
Module configuration
The following figure shows a simplified schematic of the 7710 module. As
shown, the 7710 has channels that are grouped into two banks of 10 channels
(20 channels total). Backplane isolation is provided for each bank. Each bank
includes separate cold junction reference points. The first bank contains
channels 1 to 10, while the second bank contains channels 11 to 20. Each
channel of the 20-channel multiplexer module is wired with separate inputs for
HI/LO providing fully isolated inputs.
Connections to DMM functions are provided through the module backplane
connector.
Channels 21, 22, and 23 are configured automatically by the instrument when
using system channel operation.
When using system channel operation for 4-wire measurements (including 4-wire
ohms, RTD temperature, Ratio, and Channel Average), the channels are paired as
follows:
CH1 and CH11 | CH6 and CH16 |
---|---|
CH2 and CH12 | CH7 and CH17 |
CH3 and CH13 | CH8 and CH18 |
CH4 and CH14 | CH9 and CH19 |
CH5 and CH15 | CH10 and CH20 |
NOTE
Channels 21 to 23 in this schematic refer to designations used for control and
not actual available channels. For more information, refer to the instrument
reference manual.
Figure 6: 7710 simplified schematic
Typical connections
The following examples show typical wiring connections for the following types of measurements:
- Thermocouple
- Two-wire resistance and thermistor
- Four-wire resistance and RTD
- DC or AC voltage
Connection log
You can use the following table to record your connection information.
Connection log for the 7710
Channel | Color | Description |
---|---|---|
Card Source | H | |
L | ||
Card Sense | H | |
L | ||
CH1 | H | |
L | ||
CH2 | H | |
L | ||
CH3 | H | |
L | ||
CH4 | H | |
L | ||
CH5 | H | |
L | ||
CH6 | H | |
L | ||
CH7 | H | |
L | ||
CH8 | H | |
L | ||
CH9 | H | |
L | ||
CH10 | H | |
L | ||
CH11 | H | |
L | ||
CH12 | H | |
L | ||
CH13 | H | |
L | ||
CH14 | H | |
L | ||
CH15 | H | |
L | ||
CH16 | H | |
L | ||
CH17 | H | |
L | ||
CH18 | H | |
L | ||
CH19 | H | |
L | ||
CH2O | H | |
L |
Installation
Before operating an instrument with a switching module, verify that the
switching module is properly installed and the mounting screws are tightly
fastened. If the mounting screws are not properly connected, an electrical
shock hazard may be present.
If you are installing two switching modules, it is easier to install a
switching module into slot 2 first, then install the second switching module
into slot 1.
NOTE
If you have a Keithley Instruments Model 2700, 2701, or 2750 instrument, you
can use your existing switching module in the DAQ6510. Follow the instructions
in your original equipment documentation to remove the module from the
instrument, then use the following instructions to install it in the DAQ6510.
You do not need to remove wiring to the module.
NOTE
For inexperienced users, it is recommended that you do not connect a device
under test (DUT) and external circuitry to the switching module. This allows
you to exercise close and open operations without the dangers associated with
live test circuits. You can also set up pseudocards to experiment with
switching. Refer to “Pseudocards” in the Model DAQ6510 Data Acquisition and
Multimeter System Reference Manual for information on setting up pseudocards.
WARNING
To prevent electric shock that could result in injury or death, never handle a
switching module that has power applied to it. Before installing or removing a
switching module, make sure the instrument is turned off and disconnected from
line power. If the switching module is connected to a DUT, make sure power is
removed from all external circuitry.
WARNING
Slot covers must be installed on unused slots to prevent personal contact with
high-voltage circuits. Failure to recognize and observe standard safety
precautions could result in personal injury or death due to electric shock.
CAUTION
Before installing or removing a switching module, make sure the DAQ6510 power
is turned off and disconnected from line power. Failure to comply may result
in incorrect operation and loss of data in the memory.
Required equipment:
- Medium flat-blade screwdriver
- Medium Phillips screwdriver
To install a switching module into the DAQ6510:
- Turn off the DAQ6510.
- Disconnect the power cord from the power source.
- Disconnect the power cord and any other cables that are connected to the rear panel.
- Position the DAQ6510 so you are facing the rear panel.
- Use the screwdriver to remove the slot cover screws and the cover plate. Retain the plate and screws for future use.
- With the top cover of the switching module facing up, slide the switching module into the slot.
- Press the switching module in firmly to make sure the switching module connector is connected to the DAQ6510 connector.
- Use the screwdriver to tighten the two mounting screws to secure the switching module to the mainframe. Do not overtighten.
- Reconnect the power cord and any other cables.
Remove a switching module
NOTE
Before you remove a switching module or begin any testing, make sure that all
the relays are open. Since some relays may be latched closed, you must open
all the relays before removing the switching module to make connections.
Additionally, if you drop your switching module, it is possible for some
relays to latch closed.
To open all channel relays, go to the CHANNEL swipe screen. Select Open All.
WARNING
To prevent electric shock that could result in injury or death, never handle a
switching module that has power applied to it. Before installing or removing a
switching module, make sure the DAQ6510 is turned off and disconnected from
line power. If the switching module is connected to a DUT, make sure power is
removed from all external circuitry.
WARNING
If a card slot is unused, you must install slot covers to prevent personal
contact with high voltage circuits. Failure to install slot covers could
result in personal exposure to hazardous voltages, which could cause personal
injury or death if contacted.
CAUTION
Before installing or removing a switching module, make sure the DAQ6510 power
is turned off and disconnected from line power. Failure to comply may result
in incorrect operation and loss of data in the memory.
Required equipment:
- Medium flat-blade screwdriver
- Medium Phillips screwdriver
To remove a switching module from the DAQ6510:
- Turn off the DAQ6510.
- Disconnect the power cord from the power source.
- Disconnect the power cord and any other cables that are connected to the rear panel.
- Position the DAQ6510 so you are facing the rear panel.
- Use the screwdriver to loosen the mounting screws that secure the switching module to the instrument.
- Carefully remove the switching module.
- Install a slot plate or another switching module in the empty slot.
- Reconnect the power cord and any other cables.
Operating instructions
CAUTION
Before installing or removing a 7710 module, make sure the DAQ6510 power is
turned off and disconnected from line power. Failure to comply may result in
incorrect operation and loss of data from 7710 memory.
CAUTION
To prevent overheating or damage to the 7710 switching module relays, never
exceed the following maximum signal levels between any two inputs or chassis:
Any channel to any channel (1 to 20): 60 VDC or 42 VRMS, 100 mA switched, 6 W,
4.2 VA maximum.
Do not exceed the maximum specifications for the 7710. Refer to the
specifications provided in the datasheet. Failure to recognize and observe
normal safety precautions could result in personal injury or death.
WARNING
When a 7710 module is inserted into the DAQ6510, it is connected to the front
and rear inputs and the other modules in the system through the instrument
backplane. To prevent damage the 7710 module and to prevent creation of a
shock hazard, the entire test system and all of its inputs should be derated
to 60 VDC (42 VRMS). Failure to recognize and observe normal safety
precautions could result in personal injury or death. Refer to the instrument
documentation for operating instructions.
WARNING
This switching module does not support current measurements. If the instrument
has the TERMINALS switch set to REAR and you are working with the slot that
contains this switching module, the AC, DC, and digitize current functions are
not available. You can measure current using the front panel or using another
slot that contains a switching module that supports the AC, DC, and digitize
current measurements.
If you use remote commands to attempt to measure current when configuring a
channel, an error is returned.
Fast scan using 7710 module with DAQ6510 mainframe
The following SCPI program demonstrates using the 7710 module and DAQ6510
mainframe to achieve fast scanning. It makes use of WinSocket control to
communicate with the 7710 mainframe.
DAQ6510 or
pseudocode| Command| Description
---|---|---
Pseudocode| int scanCnt = 1000| Create a variable to hold the scan count
| int sampleCnt| Create a variable to hold the full sample count (total number
of readings)
| int chanCnt| Create a variable to hold the channel count
| int actualRdgs| Create a variable to hold the actual reading count
| string rcvBuffer| Create a string buffer to hold extracted readings
| t imer 1 . start ( )| Start a timer to help capture elapsed time
DAQ6510| • RST| Put the instrument in a known state
| FORM: DATA ASCII| Format data as an ASCII string
| ROUT: SCAN: COUN : SCAN scanCnt| Apply the scan count
| FUNC ‘VOLT:DC’ , (@101:120)| Set function to DCV
| VOLT:RANG 1, (@101:120)| Set the fixed range at 1 V
| VOLT: AVER: STAT OFF, (@101:120)| Disable background statistics
| DISP : VOLT: DIG 4, (@101:120)| Set front panel to show 4 significant digits
| VOLT :NPLC 0.0005, (@101:120)| Set fastest NPLC possible
| VOLT:LINE:SYNC OFF, (@101:120)| Turn off line sync
| VOLT : AZER: STAT OFF, (@101:120)| Turn off auto zero
| CALC2 :VOLT :LIM1 :STAT OFF, (@101:120)| Turn off limit tests
| CALC2 :VOLT :LIM2 :STAT OFF, (@101:120)|
| ROUT : SCAN : INT 0| Set trigger interval between scans to 0 s
| TRAC:CLE| Clear the reading buffer
| DISP:LIGH:STAT OFF| Turn the display off
| ROUT :SCAN :CRE (@101:120)| Set the scan list
| chanCnt = ROUTe :SCAN:COUNt : STEP?| Query the channel count
Pseudocode| sampleCnt = scanCnt • chanCnt| Calculate the number of readings
made
DAQ6510| INIT| Initiate the scan
Pseudocode| for i = 1, i < sampleCnt| Set up a f or loop from 1 to sampleCnt .
but leave the incrementing of 1 for later
| delay 500| Delay for 500 ms to allow readings to accumulate
DAQ6510| actualRdgs = TRACe: ACTual?| Query the actual readings captured
| rcvBuffer = “TRACe:DATA? i, actualRdgs, “defbuf ferl”, READ| Query the
readings available from i to the value of actualRdgs
Pseudocode| WriteReadings (“C: \ myData . csv”, rcvBuffer)| Write the
extracted readings to a file. myData.csv. on the local computer
| i = actualRdgs + 1| Increment i for the next loop pass
| end for| End the f or loop
| timer 1 . stop()| Stop the timer
| timerl.stop – timerl.start| Calculate the elapsed time
DAQ6510| DISP : LICH :STAT ON100| Turn the display on again
The following TSP program demonstrates using the 7710 module and DAQ6510
mainframe to achieve fast scanning. It makes use of WinSocket control to
communicate with the 7710 mainframe.
— Set up variables to be referenced during the scan.
scanCnt = 1000
sampleCnt = 0
chanCnt = 0
actualRdgs = 0
rcvBuffer = “”
— Get the initial timestamp for end-of-run comparison.
local x = os.clock()
— Reset the instrument and clear the buffer.
reset()
defbuffer1.clear()
— Set up reading buffer format and establish scan count
format.data = format.ASCII
scan.scancount = scanCnt
— Configure the scan channels for the card in slot 1.
channel.setdmm(“101:120”, dmm.ATTR_MEAS_FUNCTION, dmm.FUNC_DC_VOLTAGE)
channel.setdmm(“101:120”, dmm.ATTR_MEAS_RANGE, 1)
channel.setdmm(“101:120”, dmm.ATTR_MEAS_RANGE_AUTO, dmm.OFF)
channel.setdmm(“101:120”, dmm.ATTR_MEAS_AUTO_ZERO, dmm.OFF)
channel.setdmm(“101:120”, dmm.ATTR_MEAS_DIGITS, dmm.DIGITS_4_5)
channel.setdmm(“101:120”, dmm.ATTR_MEAS_NPLC, 0.0005)
channel.setdmm(“101:120”, dmm.ATTR_MEAS_APERTURE, 8.33333e-06)
channel.setdmm(“101:120”, dmm.ATTR_MEAS_LINE_SYNC, dmm.OFF)
channel.setdmm(“101:120”, dmm.ATTR_MEAS_LIMIT_ENABLE_1, dmm.OFF)
channel.setdmm(“101:120”, dmm.ATTR_MEAS_LIMIT_ENABLE_2, dmm.OFF)
— Dim the display.
display.lightstate = display.STATE_LCD_OFF
— Generate the scan.
scan.create(“101:120”)
scan.scaninterval = 0.0
chanCnt = scan.stepcount
— Calculate the overall sample count and use it to size the buffer.
sampleCnt = scanCnt * chanCnt
defbuffer1.capacity = sampleCnt
— Start the scan.
trigger.model.initiate()
— Loop to capture and print readings.
i = 1
while i <= sampleCnt do
delay(0.5)
myCnt = defbuffer1.n
— NOTE: Can be supplemented or replaced by writing to USB
printbuffer(i, myCnt, defbuffer1.readings)
i = myCnt + 1
end
— Turn the display on again.
display.lightstate = display.STATE_LCD_50
— Output the elapsed time.
print(string.format(“Elapsed Time: %2f\n”, os.clock() – x))
Operating considerations
Low-ohms measurements
For resistances in the normal range (>100 Ω), the 2-wire method (Ω2) is
typically used for ohms measurements.
For low ohms (≤100 Ω), the signal path resistance in series with the DUT could
be high enough to adversely affect the measurement. Therefore, the 4-wire
method (Ω4) should be used for low-ohms measurements. The following discussion
explains the limitations of the 2-wire method and the advantages of the 4-wire
method.
Two-wire method
Resistance measurements in the normal range (>100 Ω) are generally made using
the 2-wire method (Ω2 function). The test current is forced through the test
leads and the resistance being measured (RDUT). The meter then measures the
voltage across the resistance value accordingly.
The main problem with the 2-wire method, as applied to low-resistance
measurements is the test lead resistance (RLEAD) and the channel resistance
(RCH). The sum of these resistances typically lies in the range of 1.5 to 2.5
Ω.
Therefore, it is difficult to get accurate 2-wire ohms measurements below 100
Ω.
Due to this limitation, the 4-wire method should be used for resistance
measurements ≤100 Ω.
Four-wire method
The 4-wire (Kelvin) connection method using the Ω4 function is generally
preferred for low-ohms measurements.
The 4-wire method cancels the effects of channel and test lead resistance.
With this configuration, the test current (ITEST) is forced through the test
resistance (RDUT) through one set of test leads (RLEAD2 and RLEAD3), while the
voltage (VM) across the device under test (DUT) is measured through a second
set of leads (RLEAD1 and RLEAD4) called the sense leads.
With this configuration, the resistance of the DUT is calculated as follows:
RDUT = VM / ITEST
Where: I is the sourced test current and V is the measured voltage.
As shown in the figure in Maximum test lead resistance (on page 17), the
measured voltage (VM) is the difference between VSHI and VSLO. The equations
below the figure show how test lead resistance and channel resistance are
canceled out of the measurement process.
Maximum test lead resistance
The maximum test lead resistance (RLEAD), for specific 4-wire resistance
ranges:
- 5 Ω per lead for 1 Ω
- 10% of range per lead for 10 Ω, 100 Ω, 1 kΩ, and 10 kΩ ranges
- 1 kΩ per lead for 100 kΩ, 1 MΩ, 10 MΩ, and 100 MΩ ranges
Assumptions:
- Virtually no current flows in the high-impedance sense circuit because of the high impedance of the voltmeter (VM). Therefore, the voltage drops across Channel 11 and test lead 1 and 4 are negligible and can be ignored.
- The voltage drops across Channel 1 Hi (RCH1Hi) and test lead 2 (RLEAD2) are not measured by the voltmeter (VM).
RDUT = VM/ITEST
Where:
- VM is the voltage measured by the instrument.
- ITEST is the constant current sourced by the instrument to the DUT.
- VM = VSHI − VSLO
- VSHI = ITEST × (RDUT + RLEAD3 + RCH1Lo)
- VSLO = ITEST × (RLEAD3 + RCH1Lo)
- VSHI − VSLO = ITEST × [(RDUT + RLEAD3 + RCH1Lo) − (RLEAD3 + RCH1Lo)]
- = ITEST × RDUT
- = VM
Voltage measurements
Path resistance can adversely affect low-ohms measurements (see Low-ohms
measurements (on page 16) for more information). Series path resistance can
cause loading problems for DC voltage measurements on the 100 V, 10 V, and 10
mV ranges when the 10 MΩ input divider is enabled. High signal path resistance
can also adversely affect AC voltage measurements on the 100 V range above 1
kHz.
Insertion loss
Insertion loss is AC signal power lost between the input and the output. In
general, as frequency increases, insertion loss increases.
For the 7710 module, insertion loss is specified for a 50 Ω AC signal source
routed through the module to a 50 Ω load. Signal power loss occurs as the
signal is routed through the signal paths of the module to the load. Insertion
loss is expressed as dB magnitudes at specified frequencies. The
specifications for insertion loss are provided in the data sheet.
As an example, assume the following specifications for insertion loss:
<1 dB @ 500 kHz 1 dB insertion loss is approximately 20% loss of signal power.
<3 dB @ 2 MHz 3 dB insertion loss is approximately 50% loss of signal power.
As signal frequency increases, power loss increases.
NOTE
The insertion loss values used in the above example may not be the actual
insertion loss specifications of the 7710. The actual insertion loss
specifications are provided in the datasheet.
Crosstalk
An AC signal can be induced into adjacent channel paths on the 7710 module. In
general, crosstalk increases as frequency increases.
For the 7710 module, crosstalk is specified for an AC signal routed through
the module to a 50 Ω load. Crosstalk is expressed as a dB magnitude at a
specified frequency. The specification for crosstalk is provided in the
datasheet.
As an example, assume the following specification for crosstalk:
<-40 dB @ 500 kHz -40 dB indicates that crosstalk into adjacent channels is
0.01% of the AC signal.
As signal frequency increases, crosstalk increases.
NOTE
The crosstalk values used in the above example may not be the actual crosstalk
specification of the 7710. The actual crosstalk specification is provided in
the datasheet.
Heat sink temperature measurements
Measuring the temperature of a heat sink is a typical test for a system that
has temperature measurement capability. However, the 7710 module cannot be
used if the heat sink is being floated at a dangerous voltage level (>60 V).
An example of such a test is shown below.
In the following figure, the heat sink is floating at 120 V, which is the line
voltage being input to a +5V regulator.
The intention is to use channel 1 to measure the temperature of the heat sink,
and use channel 2 to measure the +5 V output of the regulator. For optimum
heat transfer, the thermocouple (TC) is placed in direct contact with the heat
sink. This inadvertently connects the floating 120 V potential to the 7710
module. The result is 115 V between channel 1 and channel 2 HI, and 120 V
between channel 1 and chassis. These levels exceed the 60 V limit of the
module, creating a shock hazard and possibly causing damage to the module.
WARNING
The test in the following figure demonstrates how a dangerous voltage can
unintentionally be applied to the 7710 module. In any test where floating
voltages >60 V are present, you must be careful not to apply the floating
voltage to the module. Failure to recognize and observe normal safety
precautions could result in personal injury or death.
CAUTION
Do not use the 7710 module to perform this type of test. It exceeds the 60 V
limit creating a shock hazard and could cause damage to the module. Excessive
voltages:
The voltage differential between Ch 1 and Ch 2 HI is 115 V.
The voltage differential between Ch 1 and Ch 2 LO (chassis) is 120 V.
Module
handling precautions
The solid state relays used on the 7710 module are static sensitive devices.
Therefore, they can be damaged by electrostatic discharge (ESD).
CAUTION
To prevent damage from ESD, only handle the module by the card edges. Do not
touch the backplane connector terminals. When working with the quick-
disconnect terminal blocks, do not touch any circuit board traces or other
components. If working in a high-static environment, use a grounded wrist
strap when wiring the module.
Touching a circuit board trace may contaminate it with body oils that can
degrade the isolation resistance between circuit paths, adversely affecting
measurements. It is good practice to handle a circuit board only by its edges.
Solid state relay precautions
To prevent damage to the module, do not exceed the maximum signal level
specification of the module. Reactive loads require voltage clamping for
inductive loads and surge current limiting for capacitive loads.
Current limiting devices can be resistors or resettable fuses. Examples of
resettable fuses are polyfuses and positive temperature coefficient (PTC)
thermistors. Voltage clamping devices can be Zener diodes, gas discharge
tubes, and bidirectional TVS diodes.
Limiting resistor usage
Cabling and test fixtures can contribute considerable capacitance to the
signal path. Inrush currents may be excessive and require current limiting
devices. Large inrush currents can flow when incandescent lamps, transformer
and similar devices are initially energized and current limiting should be
used.
Use current limiting resistors to limit inrush current caused by cable and DUT
capacitance.
Clamp voltage
Voltage clamping should be used if power sources have the capability of
creating transient voltage spikes.
Inductive loads like relay coils and solenoids should have voltage clamping
across the load to suppress counter electromotive forces. Even if transient
voltages generated at the load are limited at the device, transient voltages
will be generated by inductance if circuit wires are long. Keep wires as short
as possible to minimize inductance.
Use a diode and Zener diode to clamp voltage spikes generated by counter
electromotive forces at the relay coil. Use a gas discharge tube to prevent transient
spikes from damaging the relay. If the device under test (DUT) changes impedance states
during testing, excessive currents or voltages may appear at the solid state
relay. If a DUT fails due to low impedance, current limiting may be required.
If a DUT fails due to high impedance, voltage clamping may be required.
Calibration
The following procedures calibrate the temperature sensors on the 7710 plug-in
modules.
WARNING
Do not attempt to perform this procedure unless you are qualified, as
described by the types of product users in the Safety precautions. Do not
perform these procedures unless qualified to do so. Failure to recognize and
observe normal safety precautions could result in personal injury or death.
Calibration setup
To calibrate the module, you need the following equipment.
- Digital thermometer: 18 °C to 28 °C ±0.1 °C
- Keithley 7797 Calibration/Extender Board
Extender board connections
The extender board is installed in the DAQ6510. The module is connected to the
extender board externally to prevent heating of the module during calibration.
To make extender board connections:
- Remove power from the DAQ6510.
- Install the extender board into Slot 1 of the instrument.
- Plug the module into the P1000 connector on the rear of the 7797 Calibration/Extender Board.
Temprerature calibration
NOTE
Before calibrating the temperature on the 7710, remove power from the module
for at least two hours to allow module circuitry to cool down. After turning
on the power during the calibration procedure, complete the procedure as
quickly as possible to minimize module heating that could affect calibration
accuracy. Initially allow the DAQ6510 to warm up for at least one hour with
the 7797 calibration card installed. If calibrating multiple modules in a row,
power off the DAQ6510, quickly unplug the previously calibrated 7710, and plug
in the next one. Wait three minutes before calibrating the 7710.
Set up calibration:
- Turn on the DAQ6510 power.
- To ensure the instrument is using the SCPI command set, send: *LANG SCPI
- On the front panel, verify that TERMINALS is set to REAR.
- Allow three minutes for thermal equilibrium.
To calibrate temperature:
-
Accurately measure and record the cold temperature of the 7710 module surface at the center of the module with the digital thermometer.
-
Unlock calibration by sending:
:CALibration:PROTected:CODE “KI006510” -
Calibrate temperature on the 7710 with the following command, where
is the cold calibration temperature measured in step 1 above:
:CALibration:PROTected:CARD1:STEP0 -
Send the following commands to save and lock out calibration:
:CALibration:PROTected:CARD1:SAVE
:CALibration:PROTected:CARD1:LOCK
Errors that can occur during calibration
If calibration errors occur, they are reported in the event log. You can
review the event log from the front panel of
the instrument by using the SCPI :SYSTem:EVENtlog:NEXT? command or the TSP
eventlog.next()
command.
The error that can occur on this module is 5527, Temperature Cold Cal error.
If this error occurs, contact Keithley
Instruments. Refer to Factory service (on page 24).
Factory service
To return your DAQ6510 for repair or calibration, call
1-800-408-8165 or complete the form at
tek.com/services/repair/rma-request. When you request service, you need the serial number and firmware or
software version of the instrument.
To see the service status of your instrument or to create an on-demand price
estimate, go to tek.com/service-quote.
Safety precautions
The following safety precautions should be observed before using this product
and any associated instrumentation. Although some instruments and accessories
would normally be used with nonhazardous voltages, there are situations where
hazardous conditions may be present.
This product is intended for use by personnel who recognize shock hazards and
are familiar with the safety precautions required to avoid possible injury.
Read and follow all installation, operation, and maintenance information
carefully before using the product.
Refer to the user documentation for complete product specifications. If the
product is used in a manner not specified, the protection provided by the
product warranty may be impaired.
The types of product users are:
Responsible body is the individual or group responsible for the use and
maintenance of equipment, for ensuring that the equipment is operated within
its specifications and operating limits, and for ensuring that operators are
adequately trained. Operators use the product for its intended function. They
must be trained in electrical safety procedures and proper use of the
instrument. They must be protected from electric shock and contact with
hazardous live circuits.
Maintenance personnel perform routine procedures on the product to keep it
operating properly, for example, setting the line voltage or replacing
consumable materials. Maintenance procedures are described in the user
documentation. The procedures explicitly state if the operator may perform
them. Otherwise, they should be performed only by service personnel.
Service personnel are trained to work on live circuits, perform safe
installations, and repair products. Only properly trained service personnel
may perform installation and service procedures.
Keithley products are designed for use with electrical signals that are
measurement, control, and data I/O connections, with low transient
overvoltages, and must not be directly connected to mains voltage or to
voltage sources with high transient overvoltages.
Measurement Category II (as referenced in IEC 60664) connections require
protection for high transient overvoltages often associated with local AC
mains connections. Certain Keithley measuring instruments may be connected to
mains. These instruments will be marked as category II or higher.
Unless explicitly allowed in the specifications, operating manual, and
instrument labels, do not connect any instrument to mains. Exercise extreme
caution when a shock hazard is present. Lethal voltage may be present on cable
connector jacks or test fixtures.
The American National Standards Institute (ANSI) states that a shock hazard
exists when voltage levels greater than 30 V RMS, 42.4 V peak, or 60 VDC are
present. A good safety practice is to expect that hazardous voltage is present
in any unknown circuit before measuring.
Operators of this product must be protected from electric shock at all times.
The responsible body must ensure that operators are prevented access and/or
insulated from every connection point. In some cases, connections must be
exposed to potential human contact. Product operators in these circumstances
must be trained to protect themselves from the risk of electric shock. If the
circuit is capable of operating at or above 1000 V, no conductive part of the
circuit may be exposed.
For maximum safety, do not touch the product, test cables, or any other
instruments while power is applied to the circuit under test. ALWAYS remove
power from the entire test system and discharge any capacitors before
connecting or disconnecting cables or jumpers, installing or removing
switching cards, or making internal changes, such as installing or removing
jumpers.
Do not touch any object that could provide a current path to the common side
of the circuit under test or power line (earth) ground. Always make
measurements with dry hands while standing on a dry, insulated surface capable
of withstanding the voltage being measured.
For safety, instruments and accessories must be used in accordance with the
operating instructions. If the instruments or accessories are used in a manner
not specified in the operating instructions, the protection provided by the
equipment may be impaired.
Do not exceed the maximum signal levels of the instruments and accessories.
Maximum signal levels are defined in the specifications and operating
information and shown on the instrument panels, test fixture panels, and
switching cards. Chassis connections must only be used as shield connections
for measuring circuits, NOT as protective earth (safety ground) connections.
The WARNING heading in the user documentation explains hazards that might
result in personal injury or death. Always read the associated information
very carefully before performing the indicated procedure.
The CAUTION heading in the user documentation explains hazards that could
damage the instrument. Such damage may
invalidate the warranty.
The CAUTION heading with the symbol in the user documentation explains
hazards that could result in moderate or minor injury or damage the
instrument. Always read the associated information very carefully before
performing the indicated procedure.
Damage to the instrument may invalidate the warranty.
Instrumentation and accessories shall not be connected to humans.
Before performing any maintenance, disconnect the line cord and all test
cables.
To maintain protection from electric shock and fire, replacement components in
mains circuits — including the power transformer, test leads, and input jacks
— must be purchased from Keithley. Standard fuses with applicable national
safety approvals may be used if the rating and type are the same. The
detachable mains power cord provided with the instrument may only be replaced
with a similarly rated power cord. Other components that are not safety-
related may be purchased from other suppliers as long as they
are equivalent to the original component (note that selected parts should be
purchased only through Keithley to maintain accuracy and functionality of the
product). If you are unsure about the applicability of a replacement
component, call a Keithley office for information.
Unless otherwise noted in product-specific literature, Keithley instruments
are designed to operate indoors only, in the following environment: Altitude
at or below 2,000 m (6,562 ft); temperature 0 °C to 50 °C (32 °F to 122 °F);
and pollution degree 1 or 2.
To clean an instrument, use a cloth dampened with deionized water or mild,
water-based cleaner. Clean the exterior of the instrument only. Do not apply
cleaner directly to the instrument or allow liquids to enter or spill on the
instrument. Products that consist of a circuit board with no case or chassis
(e.g., a data acquisition board for installation into a computer) should never
require cleaning if handled according to instructions. If the board becomes
contaminated and operation is affected, the board should be returned to the
factory for proper cleaning/servicing.
Safety precaution revision as of June 2018. ![KEITHLEY 7710 Multiplexer Module
- bar code](https://manuals.plus/wp-content/uploads/2023/07/KEITHLEY-7710 -Multiplexer-Module-bar-code.jpg)
References
- Keithley Instruments & Products | Tektronix
- Keithley Instruments & Products | Tektronix
- Service Quote | Tektronix
- Request Service for Tektronix & Keithley Instruments | Tektronix
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>