Winsen MW-ORP101 ORP Water Quality Sensor User Manual
- June 4, 2024
- Winsen
Table of Contents
Winsen MW-ORP101 ORP Water Quality Sensor User Manual
Statement
This manual copyright belongs to Zhengzhou Winsen Electronics Technology Co.,
LTD. Without the written permission, any part of this manual shall not be
copied, translated, stored in database or retrieval system, also can’t spread
through electronic, copying, record ways.
Thanks for purchasing our product. In order to let customers use it better and
reduce the faults caused by misuse, please read the manual carefully and
operate it correctly in accordance with the instructions. If users disobey the
terms or remove, disassemble, change the components inside of the sensor, we
shall not be responsible for the loss.
The specific such as color, appearance, sizes &etc, please in kind prevail.
We are devoting ourselves to products development and technical innovation, so
we reserve the right to improve the products without notice. Please confirm it
is the valid version before using this manual. At the same time, users’
comments on optimized using way are welcome.
Please keep the manual properly, in order to get help if you have questions
during the usage in the future.
Zhengzhou Winsen Electronics Technology CO., LTD
Profile
MW-ORP101 sensor is a primary battery type ORP water quality detection sensor. Using the relationship between the ORP value of the test solution and the measured potential difference, the ORP value of the solution to be tested is determined by the potential difference measured by the working battery composed of the electrode and the counter electrode in the solution to be measured through the sensor.
Sensor characteristics
High precision, fast response, good stability, small size, low power
consumption, easy to carry.
Fig1. Sensor image
Main application
It is widely used for ORP value detection in laboratory research, water supply, waste water treatment, aquaculture, farmland irrigation and other fields
Technical indicators
Stable 1
Items
|
Parameter
---|---
Detection object
| Solution
Measure Range|
-2000 mV ~﹢2000
Resolution
| 1 mV
Response time|
≤120 s
Material
| ABS
Working temperature|
0℃~50℃
Service life
|
1year
Sensor characterization
Fig2. Sensor Structure
Fig.4. Sensor stability curve
Instructions
- Remove electrode protective cover from sensor port and soak it in 3M KCL resolution for 8 hours.
- After soaking, rinse the sensor electrode end with a small amount of pure water or de-ionized water.
- Connect the sensor port to the ORP test instrument and place it in the solution to be tested for testing after being stabilized for 3 minutes.
- After the test is completed, rinse the sensor electrode end with a small amount of pure water or deionized water.
- Dry the cleaned sensor, put a protective cover on the electrode end, and store it at room temperature.
Line connection
Stable 2
Cable color
| lead definition
---|---
Blue (center core)|
positive
Network cable
(shielded network cable)
|
negative
Precautions
- The protective cover must be removed before using the sensor
- If there is white crystal on the sensor electrode end, rinse it with de- ionized water.
- When the sensor is initial use or used again after a long term storage, the electrode port should be immersed in 3M KCL resolution for 2 hours for activation.
- If there are contaminants on the sensor electrode end, it can be treated with corresponding reagents, see Table 3 for details.
Pollutants|
Detergent
---|---
Metal oxide
| Less than 1M dilute hydrochloric acid
Organic oils|
Weakly Alkaline Detergent
Resin polymer substance
| Alcohol (Analytical pure)
Pigment substances|
Hydrogen Peroxide(3%)
Support
Zhengzhou Winsen Electronics Technology Co., Ltd
Add.: NO.299 Jin Suo Road, National Hi-TechZone, Zhengzhou, 450001 China
Tel.: 86-371-67169097
Fax: +86- 371-60932988
E-mail: sales@winsensor.com
Website: www.winsen-sensor.com
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>