easy home EHE015BLE Mini Wireless TENS Pulse Stimulator User Manual

June 12, 2024
EASY HOME

easy home EHE015BLE Mini Wireless TENS Pulse Stimulator

easy home EHE015BLE Mini Wireless TENS Pulse Stimulator-
product

Introduction

Electronic Pulse Stimulator
The EHE015BLE Electronic Pulse Stimulator is a professional model of the Easy@Home TENS Unit product line. This portable and compact device delivers electric pulses generated to the user’s skin through the electrodes and it has multiple modes of different pulse frequencies, covering
Electric Muscle Stimulation (EMS) for muscle toning, firming, and strengthening. Also, the modes of Transcutaneous Electrical Nerve Stimulation (TENS) could relieve pain associated with sore and aching muscles due to strain from exercise or normal household and work activities. It includes operating elements of ON/OFF button, intensity increase/mode selection button, and intensity decrease/timer selection button, and could be attached and detached to the electrode through the two snap-on connectors.

Indications for Use

  • TENS (Modes 1, 2, 4, 5, 6, 8,) To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, arm, and leg, due to strain from exercise or normal household and work activities.
  • It is also intended for symptomatic relief and management of chronic, intractable pain and relief of pain associated with arthritis. EMS (Modes 1, 3, 7)
  • To stimulate healthy muscles in order to improve and facilitate muscle performance. To be used for the improvement of muscle tone and firmness, and for strengthening muscles in the arms, abdomen, legs, and buttocks. Not intended for use in any therapy or
  • for the treatment of any medical conditions or diseases. It is also intended to temporarily increase local blood circulation in the healthy muscles of lower extremities.

Safety Warning

Contraindications
Do not use this device on Users who have a cardiac pacemaker, implanted defibrillator, or other implanted metallic or electronic device, because this may cause electric shock, burns, electrical interference, or death. Do not use this device on Users whose pain syndromes are undiagnosed.

Warnings

  • Do not apply stimulation over the patient’s neck because this could cause severe muscle spasms resulting in closure of the airway, difficulty in breathing, or adverse effects on heart rhythm or blood pressure.
  • Do not apply stimulation across the patient’s chest, because the introduction of electrical current into the chest may cause rhythm disturbances to the patient’s heart, which could be lethal.
  • Do not apply stimulation over, or in proximity to, cancerous lesions.
  • Do not apply stimulation when the patient is in the bath or shower.
  • If you have one of the following conditions, please consult with your physician before purchasing or using this device.
  • Acute disease, malignant tumor, infective disease, pregnant, heart disease, high fever, abnormal blood pressure, lack of skin sensation or an abnormal skin condition, any condition requiring the active supervision of a physician.

Precautions

  • Do not use this device while driving.
  • Do not use this device while sleeping.
  • Do not use this device in high humidity areas such as a bathroom.
  • Keep the device away from wet, high temperature and direct-sunlight place.
  • Keep this device out of reach of children.
  • Stop using this device at once if you feel pain, discomfort, dizziness or nausea and consult your physician.
  • Do not attempt to move the electrode pads while the device is operating.
  • Do not use the device around the heart, on the head, mouth, pudendum or blemished skin areas.
  • Do not apply stimulation of this device in the following conditions:
  1. across the chest because the introduction of electrical current into the chest may cause rhythm disturbances to the heart, which could be lethal;
  2. over painful areas. Please consult with your physician before using this device if you have painful areas;
  3. over open wounds or rashes, or over swollen, red, infected, or inflamed areas or skin eruptions (e.g., phlebitis, thrombophlebitis, varicose veins). Apply stimulation only to normal, intact, clean, healthy skin;
  4. in the presence of electronic monitoring equipment (e.g., cardiac monitors, ECG alarms). The electronic Stimulator may not operate properly when the electrical stimulation device is in use;
  5. while operating machinery, or during any activity in which electrical stimulation can put you at risk of injury;
  6. on children.

Be aware of the following

  1. to consult with your physician before using this device. The simulation with the device may:
  2. cause lethal rhythm disturbances to the heart in susceptible individuals, and,
  3.  disrupt the healing process after a recent surgical procedure;
  4. that the device is not effective for pain of central origin, including headache;
  5. that the device is not a substitute for pain medications and other pain management therapies;
  6. that the device has no curative value;
  7. that the device is a symptomatic treatment and, as such, suppresses the sensation of pain that would otherwise serve as a protective mechanism;
  8. that the long-term effects of electrical stimulation are unknown;
  9. that the user may experience skin irritation, burns or hypersensitivity due to the electrical stimulation or electrical conductive medium (gel);
  10. if the user has suspected or diagnosed epilepsy, the user should follow precautions recommended by his or her physician;
  11. to use caution if the user has a tendency to bleed internally, such as following an injury or fracture;
  12. use caution if stimulation is applied over the menstruating uterus;
  13. use caution if stimulation is applied over areas of skin that lack normal sensation;
  14. stop using the device if the device does not provide pain relief; and,
  15. use this device only with the leads, electrodes, and accessories that the manufacturer recommends.
  16. Do not share the use of the electrode pads with others.
  17. Do not use the device while it’s charging.
  18. The device contains the lithium battery. If overheating of the device occurred during the charging, stop the charging or operation immediately and report to the distributor/seller.
  19. Dispose of the battery-containing device according to the local, state, or federal laws.
  • The long-term effects of electrical stimulation are unknown.
  • Since the effects of stimulation of the brain are unknown, stimulation should not be applied across the head, and electrodes should not be placed on opposite sides of the head.
  • The safety of electrical stimulation during pregnancy has not been established.
  • Some Users may experience skin irritation or hypersensitivity due to the electrical stimulation or electrical conductive medium (gel).
  • Users with suspected or diagnosed heart disease should follow precautions recommended by their physicians.
  • Users with suspected or diagnosed epilepsy should follow precautions recommended by their physicians.
  • Use caution if stimulation is applied over the menstruating or pregnant uterus.

Adverse Reactions

  • Users may experience skin irritation and burns beneath the stimulation electrodes applied to the skin;
  • Users may experience headaches and other painful sensations during or following the application of electrical stimulation near the eyes and to the head and face.
  • Users should stop using the device and should consult with their physicians if they experience adverse reactions from the device.

Environmental Condition for Transport and Storage

  • Normal working ambient temperature: 5-40°C
  • Normal working ambient humidity: 15-90%RH
  • Store and transport ambient temperature: -25 -70°C
  • Store and transport ambient humidity: 0-90%RH
  • Atmospheric pressure:(70-106)kpa

Symbol Interpretation

Information essential for proper use shall be indicated by using the corresponding symbols. The following symbols may be seen on the device and labeling.

easy home EHE015BLE Mini Wireless TENS Pulse Stimulator-
fig1

Safety Test Standards

  • Medical Devices Directive 93/42/EEC
  • IC 60601-1:2005+A1:2012/EN 60601-1:2006 Medical electrical equipment – Part 1: General requirements for basic safety and essential performance
  • IC 60601-1-2:2007/EN 60601-1-2:2007 Medical electrical equipment – Part 1-2: General requirements for safety – Collateral standard: Electromagnetic compatibility – Requirements and tests
  • IC 60601-2-10:2012/EN 60601-2-10:2000+A1:2001 Medical electrical equipment – Part 2-10:
  • Particular requirements for the safety of nerve and muscle stimulators
  • IC 60601-1-11:2010 Medical electrical equipment – Part 1-11: General requirements for basic safety and essential performance – – Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
  • EN ISO 15223-1:2016: Medical devices – Symbols to be used with medical device labels, labelling and information to be supplied – Part 1: General requirements
  • EN 1041 Information supplied by the manufacturer with medical devices
  • IC/60601-1-6/ EN 60601-1-6 Medical electrical equipment – Part1-6: General requirements for basic safety and essential performance – Collateral standard: Usability
  • IC 62304/ EN 62304 Medical device software – Software life-cycle processes
  • IEC 62366/ EN 62366 Medical devices – Application of usability engineering to medical devices
  • ISO 10993-1 Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process

Operating Instruction

Setup
Unpack the box of the product, take the product and accessories out, and connect the electrode pad onto the device.

easy home EHE015BLE Mini Wireless TENS Pulse Stimulator-
fig2

Instructions for use without APP

The following steps are used to guide the device operation, and the details about each step are listed in the following table.

  • 1st Step Check the battery power for the Electronic Pulse Stimulator
  • 2nd Step Install the electrode pad onto the Electronic Pulse Stimulator
  • 3rd Step Put the electrode pad-installed Stimulator on the stimulation-needed body area
  • 4th Step Press the “ON/OFF” to turn on the power
  • 5th Step Select one of the stimulation modes
  • 6th Step Choose the stimulation time
  • 7th Step Adjust the stimulation intensity
  • 8th Step Stay with the stimulation
  • 9th Step Press the “ON/OFF” to turn off the power after doneeasy home EHE015BLE Mini Wireless TENS Pulse Stimulator-fig3 easy home EHE015BLE Mini Wireless TENS Pulse Stimulator-fig4The electrode pads could be used until they lose stickiness and/or conductivity. Their lifetime greatly depends on the body skin condition and protection. Please see the following for the Use Directions, Removal, and Storage for the electrode pad and conductive accessories.

Directions for Use

  • Clean skin thoroughly prior to each application of electrodes, which will not stick well if any lotion, make-up, or dirt is left on the body skin.
  • Turn the Electronic Pulse Stimulator off before applying the electrode to the body skin.
  • Install the pad included onto the back side of the Electronic Pulse Stimulator.
  • Apply the pad-installed device firmly to the skin.

Removal and Storage

  • Turn the device off before removing the electrode from the body skin.
  • Lift at the edge of the electrode and peel.
  • When not in use, store the electrode in the re-sealable bag in a cool place, out of direct sunlight and Gust.

Recommended Practice

  • Duration suggested for each skin area is 20 min and 2 times per day. Consult with your physician for longer and more frequent uses.
  • Start from the lowest intensity and gradually adjust the intensity to a comfortable level at a scale from 1 to 20.
  • Good skin care is important for the comfortable use of device. Be sure the treatment site is clean of dirt and body lotion.
  • Keeping the electrode in the storage bag after use will extend its lifespan. The electrode is disposable and should be replaced when it loses adhesiveness. To purchase additional electrodes, please contact the seller.

Product Specifications

Accessories Included in the Package

  • Tens unit controller * 1pc
  • Electrode gel pad * 1pc
  • USB line * 1pc
  • Manual * 1 pc

Technical Information

Model/type EHE015BLE

Power supply

| Powered by internal 3.7V Ii-ion battery| ****

Automatic shutoff

| ****

10 minutes


Waveform and wave shape

| Biphasic rectangular wave puls| Degree of protection against electric shock|


Type BF applied part


Pulse width

| ****

50~500µsec

| Type of protection against electric shock| Internally powered equipment
Pulse frequency| 1~500Hz (Hz=vibra-

lion per second)

| Grade of

waterproof

| IP22
Output Voltage| Max. 60Vpp ±20%(at 500ohm load)| Product life| 3years



Treatment time

| 10,20,30,40,50,60

minutes

| Lifetime for electrode| Storage for 2 years no use)  , Times of

reusable: 30 times


Output intensity

| 0 to 20 levels, adjustable| Mode of operation| Continuous operation
Modes| 8 auto modes| Software version| AO
Typical operation time of Battery| If to use at highest level , the battery can be used for about 150 minutes after fully charged.| The time required for equipment to warm from the minimum storage ternperature between uses until it is ready for

intended use

| 30 minutes
Behavior of equipment while the rechargeable internal electrical power source

is charging:

| The indication light will flash during charging and will be still with full capacity.| The time required for equipment to cool from the maximum storage temperature between uses until it is ready for

intended use

| 15 minutes
---|---|---|---
Typical service Iife of Battery| 300 times of recharging| Adapter for charging| Please use output DC5V and output current

0.3-2.0A adapter for charging

Size| 51.54312.3mm| Weight| 20g
Electrode Size| | Effective conductive|
Note: Not intended to be sterilized.
Not for use in an OXYGEN RICH ENVIRONMENT

Product Programs

Program name

|

Time min.

|

Frequency (Hz)

|

Pulse Width (µs)

---|---|---|---
Mode 1: TENS and EMS (combination mode)|

10, 20 , 30 , 40, 50 , 60

|

1.2-100

|

100

Mode 2:TENS

|

10, 20, 30, 40, 50, 60

|

62.5

|

100

Mode 3:EMS

|

10, 20, 30, 40, 50 , 60

|

12.5-55.5

|

100

Mode 4:TENS

|

10, 20 , 30, 40, 50 , 60

|

1.2

|

100

Mode 5:TENS

|

10, 20, 30, 40, 50, 60

|

100

|

100

Mode 6:TENS

|

10, 20, 30, 40 , 50 , 60

|

100

|

100

Mode 7:EMS

|

10, 20 , 30, 40, 50 , 60

|

20

|

100

Mode 8:TENS

|

10, 20, 30, 40, 50 , 60

|

160

|

100

Cleaning and Maintenance

Please use the moisturized cloth of water or neutral detergent to clean the device first, and then use the dry cloth to wipe it again. The electrode pads coming with the device are disposable and should be replaced when their adhesiveness becomes worse. Contact the seller for replacements. Do not let the sticky side of the pad touch anything, including the greasy fingertips.

Trouble Shooting

If your device is not operating properly, please check below for common problems and suggested solutions. If the recommended action does not solve the problem, please contact the seller.

Problem Possible Cause Solution
One pad feels stronger than the other This is normal. Different area of your
body will react differently Nothing needs to be done. Make sure the pads are

moist and making good contact.

The intensity is not felt or with a very weak intensity level

| Pads are not attached to the body firmly| Attach both pads firmly to the skin
The transparent films are still stuck to the pads| Peel off film on the adhesive surface of pads
The pads stack together or overlap| Do not stack pads together or overlap pads
The intensity setting is getting weak| Increase the intensity level
The battery capacity is low| Charge the battery

The skin turns red or the skin feels irritated

| The adhesive surface of the pads is dirty or dry| Wash adhesive surface of pads

gently with your fingertips for about 3 seconds under slow running water

The therapy time is too

long or the intensity is set too high

| Reduce the application time or reduce the intensity
The electrode pad surface if worm out| Replace electrode pad
No power outpu; no display on LCD.| The battery capacity is depleted| Charge the battery
Power cuts off during use| The battery is weak| Charge the battery

It is difficult to attach the pad to the skin

| Have you removed the transparent film from the pad?| Peel off film on the adhesive surface of pads
---|---|---
Was the pad applied immediately after washing?| Dry the pad
Is the adhesive surface of the pad damaged?| Replace the pad

Adhesive surface of pad is not sticky

| Pads get deteriorative| Contact the vendor for replacements
Were the pads stored under high temperature, high humidity, or direct sunshine?| Replace the pad

Recommended Use Positions

easy home EHE015BLE Mini Wireless TENS Pulse Stimulator-
fig5

Warranty
This device carries a limited warranty of one year from the date of delivery. The warranty applies to the device only, and the accessories are not covered by this warranty.
During the warranty period, defective items will be repaired or replaced at no charge. Any evidence of misuse, abuse, alternations, or externally caused damage may have this warranty invalid.
For more information, please contact the manufacturer.

Electromagnetic Compatibility and FCC Compliance Statement

  1. This product needs special precautions regarding electromagnetic compatibility (EMC) and needs to be installed and put into service according to the EMC information provided, and this unit can be affected by portable and mobile radio frequency (RF) communications equipment.
  2. Do not use a mobile phone or other devices that emit electromagnetic fields, near the unit. This may result in incorrect operation of the unit.
  3. Caution: This unit has been thoroughly tested and inspected to assure proper performance and operation!
  4. Caution: This machine should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, this machine should be observed to verify normal operation in the configuration in which it will be used.

FCC ID 2AMJI-TENS

Guidance and manufacture’s declaration – electromagnetic immunity

The deviceis intended for use in the electromagnetic environment specified below. The customer or the user of deviceshould assure that it is used in such an environment.
Immunity test| EC 60601

test level

| Compliance level| Electromagnetic environment – guidance

Electrostatic discharge (ESD) IEC 61000-4-2

|

±6 kV contact

±8 kV air

|

±6 kV contact

±8 kV air

|

Floors should be wood, concrete or ceramic tile. If floor are covered with synthetic material, the relative humidity should be at least 30%.

Electrical fast transient/burst IEC 61000-4-4

|

±2 kV for power supply lines

±1 kV for input/ output lines

|

Not applicable (internal battery powered)

|

Mains power quality should be that of a typical commercial or hospital environment.

|

± 1 kV line(s)

|

Not applicable (internal battery powered)

|

Mains power quality should be

Surge| to line(s)| that of a typical commercial or
IEC 61000-4-5| ± 2 kV line(s)| hospital environment.
| to earth|

Voltage dips, short interruptions and voltage variations on power supply input lines

IEC 61000-4-11

|

<5%UT

(>95% dip in UT) for 0.5 cycle 40%UT

(60% dip in UT) for 5 cycles 70%UT

(30% dip in UT) for 25 cycles

<5%UT

(>95% dip in UT) for 5 sec

|

Not applicable (internal battery powered)

|

Mains power quality should be that of a typical commercial or hospital environment. If the user of the device requires continued operation during power mains interruptions, it is recommended that the devicebe powered from an uninterruptible power supply or a battery.

Power frequency (50Hz/60Hz)

magnetic field IEC 61000-4-8

|

3A/m

|

3A/m

| Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial

or hospital environment.

---|---|---|---
NOTE: UT is the a.c. mains voltage prior to application of the test level.

Guidance and manufacture’s declaration – electromagnetic immunity

The device is intended for use in the electromagnetic environment specified below.

The customer or the user ofthe device should assure that it is used in such an environment.

| | |
Immunity test| IEC 60601

test level

| Compliance level| Electromagnetic environment – guidance
Conducted RF IEC 61000-4-6| 3Vrms 150 kHz to

80 MHz

| 3Vrms| Portable and mobile RF communica­ tions equipment should be used no closer to any part of the device, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.

Recommended separation distance

d=1,2Jfi

Radiated RF IEC 61000 -4-3

|

3V/m

80 MHz to

2.5 GHz

|

3V/m

| d = 1,2 Jfi 80 MHz to 800 MHz

d =2,3 Jfi 800 MHz to 2,5 GHz Where Pis the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and dis the recommended separation distance in metres (m).

Field strengths from fixed RF transmitters, as determined by an

| | |

electromagnetic site survey,a should be less than the compliance level in each frequency range.b

Interference may occur in the vicinity of equipment marked with the

following symbol:                   {(<•>)}

I

---|---|---|---

NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

a: Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and lV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the device is used exceeds the applicable RF compliance level above, the device shouldbe observed to verify normal operation. If abnormal performance is observed, additional measures may benecessary, such as re-orienting or relocating the device.

b: Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.

Recommended separation distances between

portable and mobile RF communications equipment and the device .


The device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the device as recommended below, according to the maximum output power of the communications equipment.

Rated maximum output power of transmitter (W)| Separation distance according to frequency of transmitter(m)
150 KHz to 80 MHz| 80 MHz to 800 MHz| 800 MHz to 2 . 5 GHz
0.01| 0.12| 0.12| 0.23
0.1| 0.38| 0.38| 0.73
| 1.2| 1.2| 2.3
10| 3.8| 3.8| 7.3
100| 12| 12| 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance dinmetres (m) can be estimated using the equation applicable to the frequency of the transmitter, where Pis the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.

FCC

This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions:

  1. This device may not cause harmful interference, and
  2. This device must accept any interference received, including interference that may cause undesired operation.

The subject device has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation.
The product generates, uses, and can radiate radio frequency energy and, if not installed and used accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that the interference will not occur in particular installation. If the product does cause harmful interference to radio or television reception, which can be determined by turning the product on or off, the user is encouraged to try to correct the interference by one or more of the following measures:

  • a) Reorient or relocate the receiving antenna;
  • b) Increase the separation between the product and the receiver;
  • c) Consult the dealer or an experienced radio / TV technician for help.
  • d) Connect the equipment into an outlet on a circuit different from that to which the receiver is 0connected.

Changes or modifications to this product not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment.

Contact Information

Healthcare-Manager.com
or Questions or comments?
Please contact us via email : service@healthcare- manager.com
Manufactured for Easy Healthcare Corporation 360 Shore Dr., Burr Ridge, IL USA 60527 Made in China

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals