NEUATION iFUGE L400 NXT Versatile Lab Centrifuge
INTRODUCTION
This centrifuge is equipped with a maintenance-free ·drive, a large display & simple interface for efficient operation for daily lab u.sage. The programmable ce.ntrifuge can deliver up to 4500 RPM and can accommodates different types of rotors. It features various programmable mode to save time & add convenience.
INTENDED USE
This is a large-capacity centrifuge designed to separate, sediment, spin down
aqueous solutions & solvents. suspensions of differing densities ·in
compatible sample containers.
NOTE: Before using.g the centrifuge, please read this user manual carefully.
This user manual is intended to assist with th.e operation and care of the
unit & is not a document which aids in repa·ir. For repair please contact the
supplier.
FEATURES
The centrifuge has the following features:
- Delivers up to 4500 RPM for all compatible rotors
- BLDC maintenance-free motor drive
- Imbalance detectio·n safety with auto cutoff feature
- Lid lock safety feature: Lid does not open during operation
- Program mode for customized operation
- Speed s.etting by RPM/RCF mode
- Countdown timer range from 1 to 99 minutes
- Last run mem.or feature
- Convenient and ea.sy user interface
- Emergency lid release during power cutoff
- Automatic internal diagnosis & error display
STANDARD ACCESSORIES
- PowerCord
- T -Allen key
- User manual
- Warranty card
TECHNICAL SPECIFICATIONS
Motor Type | Brushless DC Motor |
---|---|
Max capacity | 360ml {24 x 15ml) |
M ax. Speed | 4500 RPM |
Speed Accuracy & Speed Step | ± 100 RPM |
Run Time | 1 min to 99:59 mins & infinite mode |
--- | --- |
Min. Acceleration Time | < 35 seconds |
rotor
(without load)
Min. Deceleration Time| :e:; 35 seconds
Noise Level| <68 dB(A)
Ambient Temperature| s – 40 ° c
Permissible Relative Humidity| < 80%
Size (L X BX H)| 553 x 427 x 247 mm
Weight| 16.2 kg (without rotor)
Power Supply| 230 VAC, 50 Hz
Power Consumption| 391 W
Safety Fuse Rating| 2SOVAC,3A
Altitude
| Use upto an altitude of 2000 m above MSL
Pollution Degree| 2
Environment| For indoor use only
Rotor Type| Fixed Angle| Swing Angle
---|---|---
Bucket Type
| 04 x 50 ml 24x15ml 32 x 6 ml 16x15ml| 04 x 50 ml
24 x 15 ml
32 x 6 ml
16x 15 ml
Max. RCF| 2766xg| 3364xg
Max. Volume| 360 ml (24 x 15 ml)
Imbalance Cut-off| Automatic
Display Type| 4 Digit, 7 Segment
Swing out Rotor (PLASTIC}
Adaptor
Fixed Angle Rotor (PLASTIC)
Adaptor
STANDARD PARTS LISTING
INSTALLATION
The Laboratory centrifuge is supplied in packag.ing box. Open the box, remove
the packaging and gently place the, centrifuge out of the box. Before 1st time
usage,
open the centrifuge & ensure to remove all packaging from the rotor chamber.
Please keep a 11 pieces of packaging in safe storage for at least 2 yea rs for
warranty purposes.
LOCATION & MOUNTl’NG
Place the centrifuge on a flat, solid and leveled surface and ensure t,hat all
four feet of this centrifuge stand on the surface firmly. Avoid installing on
a slippery or surface prone to vibration.
- Ideal ambient temperature is 25° C ± S°C, avoid placing the ce,ntrifu,ge in direct sunlight.
- Keep clearance of at least 30 cm from all s,ide for ease of usage.
- Keep away from heat or water to avoid sample temperature issues or centrifuge failures.
- Do not place the centrifuge such that it becomes difficult to operate the device.
BALANCING THE ROTOR
USER INTERFACE & DISPLAY
LED s for LID open close indication, Run indication, RPM & RCF functions. Display Eleme.nt: 4 digit 7 segment
OPERATING THE CENTRIFUGE
SWITC’H ON THE CENTRIFUGE
After connecting the power cord. Switch ON the main power supply & then Switch
on the power switch located on the right side of the instrument. Make sure to
check the rotor fitment before use. The centrifuge will not operate with open
lid?
Note: Maintain a gap of 3 seconds between switch OFF and switch ON again. DO
NOT switch OFF’ and ON again instantly.
RPM/RCF Setting:
- After closing the centrifuge lid, press ”RPM/RCF” button to select speed setting in RPM/RCF mode. Now press{+) button to increase the”RPM/RCF” value and press{-) button to decrease the ”RPM./RCF” value.
- Minimum and Maximum RPM of the centrifuge is 500 rpm to 4500 rpm respectively. No roll over mode.
- The flashing of RPM value will stop and the RPM value will be stored automatically if no button is pressed after 5 blinks of the adjustment.
- ”RPM/RCF” can also be changed while the centrifu.ge is under operation. Press the ”RPM/RCF” button & u.se setting(+) and(-) to change value. Changing the ”RPM/RCF” value between the ongoing centrifugation will run the centrifuge at updated speed for the rest of ti me as indicated by the ti mer.
- To set RCF value, long Pre.ss ”RPM/RCF” button to change the RCF setting. Press the(+) & (-) button to increase. or decrease t·he value. The flashing of RCF value will stop and the RCF value will be. stored automatically if no button is pressed after 5 blinks of the adjustment.
- The var ue will blink 5 time to indicate acceptance.
Timer Setting:
- Operating time can be selected from 1 min to 99:59 min or ”continuous”.
- Press TIME button to change the time setting.
- Press the(+) and (-)button to increase or decrease the value. and after 5 time blink of the set value.
- The centrifuge time set for run between lm.in to 99:59 minutes or operated in infinite time mode or roll over mode.
- lnfinite timer will be indicated by][ in display.
- The timer in the centrit·uge ‘s countdown timer in the display will be in ”Min/Sec” mode.
- Time will .showed on after 5 sec.
- Time indicat.e on Tirne LE0 light.
- The input will be ·accepted if we leave the setting for display to blink .5 time. The value will blink 5 time to indicate acceptance.
Ace/Dec:
This button w·ill set the parameters of acceleration or deceleration of the
rotor ramp. Press ace/dee button once to· set the acceleration from 1-9 by
.using +/
button for increment/decrement respectively. Similarly press onc:e again t.he
acc./dcc button to set deceleration from 1-.9 by using +/- button for
increment/decrement respectively. The values of acceleration & deceleration
timings are listed be.low.
Acceleration & **Deceleration Time**
Rotor Type : Swing Out| Max. RPM :- 4500
Tube Capacity| 4x50ml| 16x15m l| 24x15ml| 32x6ml
Mode Ace/ Dec| Time (S)| Time (S)| Time (S)| Time (S)
DL| 478| 478| 481| 478
DO| 293| 293| 297| 293
Al/D1| 181/178| 184/177| 181/183| 188/180
A2/D2| 172/164| 174/164| 172/170| 173/168
A3/D3| 152/143| 155/145| 151/158| 152/146
A4/D4| 131/125| 133/126| 131/139| 131/128
A5/D5| 110/103| 112/107| 111/109| 113/108
A6/D6| 90/86| 93/87| 91/90| 91/89
A7/D7| 71/65| 73/68| 72/71| 74/69
A8/D8| 51/47| 52/46| 52/49| 52/46
A9/D9| 34/30| 38/30| 33/32| 35/30
Acceleration & **Deceleration Time**
Rotor Type : Fixed Angle| Max. RPM :- 4500
Tube Capacity| 4x50ml| 16x15ml| 24x15ml| 32x6ml
Mode Ace/ Dec| Time (S)| Time (S)| Time (S)| Time (S)
DL| 483| 484| 484| 484
DO| 301| 302| 301| 301
Al/D1| 182/187| 181/188| 182/188| 181/188
A2/D2| 173/174| 172/176| 172/174| 172/176
A3/D3| 152/153| 150/153| 151/152| 151/153
A4/D4| 131/133| 129/134| 130/133| 130/134
AS/D5| 111/112| 110/113| 111/113| 110/113
A6/D6| 91/93| 89/94| 90/93| 90/94
A7/D7| 71/73| 68/74| 71/72| 70/73
A8/D8| 51/48| 50/49| 51/49| 51/49
A9/D9| 33/34| 32/34| 33/34| 32/34
Note: The readings were taken under No Load conditions, so the results may vary when the equipment is loaded.
PRG:
- Press PRG button to select program from PO:t-P·99.
- Each stored program used for direct operation.
- Different parameters of RPM/RCF, l11ME, ACC/DCC can be set in each program.
- All the param,eters saves automatically after blink of 5 times.
- RPM./RCF, TIME, ACC/DCC parameters value entered respectively.
- After entering each values it will blink for 5 times to in· indicate acceptance:e.
Quick Spin:
Press hold t.he quick spin button for immediate center bifurcation of the
samples up to F) previous run RPM. Releasi.ng th.e hold will stop of the rotor
gradually.
Rotor:
Press rotor button to select· rotors as listed be.low. The rotor numbers can
be changed by u·si ng + /- for increment or decrement of the rotor nu m.ber.
START &STOP
Press ”Start” button to start operation and press ”Stop” button to stop the
ongoing operation. When the centrifuge is running will indicated the RUN LED
light. Pressing the ”Stop/Open” button will stop the operation .Once operation
is sto,pped automatically lid opens.
If the set time gets over, the centrifuge will stop automatically and lid
opens automatically.
SAFETY PRECAUTIONS
Read all safety & usage information provided in this manual carefully before using the device.
-
Never us·e. the centrifuge in any manner not specified in this manual .
-
Always use recommended original rote.rs and spare parts for best results & product safety.
-
The rotors must be loaded symmetrically. Each tube should be counterbalanced by another tube of same weight.
-
Do not use centrifuge or rotor that have not been correctly installed or shows any si,gn of damage.
-
The rotor must always be securely fastened. If the centrifuge makes unusual noise during operation, the rotor fitment needs to be checked. Switch OFF the device immediately by pressing STOP, check fitment & fasten it we,11.
-
Never move the centrifuge during its operation.
-
Prior to centrifugation, the tubes should be visually inspected for material damage. Damaged tubes must not be centrifuged. This is because broken tubes can result in sample loss and can create imbalance which can result in
-
further damage to the centrifuge and accessories.
-
Do not fill tubes while they are in the rotor. Liquid spillage may harm the device. If liquids are spilled on the rotor or rotor chamber, the centrifuge, must
be cleaned carefully and properly before being used again. -
The centrifuge may be used for the specified applications only. It must not be operated in a hazardous or flammable environment and must not be used to , centrifuge explosive or hi,ghly reactive substances. Also do not place the potential hazardous material within the clearance area/envelope.
-
Equipment if used in any manne,r not specified iri this manual or by the manufacturer can re-suit in the lapse of the product warranty.
-
Repairs must only be performed by the authorized service technician.
-
Do not lean on the equipment. It may damage the equipment or even harm the operator.
-
In the event of contamination caused by aggressive agents, the rotor must be cleaned immediately using a natural cleaning. liquid. This is particularly important for the bores of the tubes. If any damage is seen, contact the service technician.
-
Before using cleaning or decontamination methods other than those mentioned by the manufacturer, contact the manufacturer to ensure that the intended method will not d.amage the centrifuge.
-
For safety we have provided protective earthing with power supply. Make sure power supply is earthed,
MAINTENANCE AND Cl.JEAN ING
-
The rotor and the outside of the centrifuge should be cleaned regularly with a mild wet (w’ith water) cloth.
-
Ensure that while cleaning the unit is not plugged in.
-
Wear protective glove & safety glass while operating & cleaning the device.
-
The brushless motor in the centrifuge requires no ro,utine maintenance,. Any required service should be performed by authorized, qualified personnel only. Repairs performed by unauthorized personnel may void the warranty.
-
Always keep the centrifuge housing, rotor chamber and rotor clean. All parts should b.e wip.ed down periodica1lly with a soft cloth.
Note: Liquid should not co.me into contact with the motor. -
After cleaning, ensure that all parts are dry before re-u,se.
-
Regul·arly cleaning of the· rotor is important.
-
If t,he rotor chamber needs cleaning, clean w·ith cloth or sponge moistened w.ith a neutral detergent solution.
-
Do not place the rotor into the cleaning solution.
-
If corrosive, toxic or pathogenic bac.teria are accidentally spilled i.n the roto.r or rotor chamber the centrifuge must be decontaminated thr.oughly.
-
Apply min.or Grease on the Motor sh.aft (Threaded Parts.,Grooves) at regular intervals
TROUBLESHOOTING
This centrifuge has a self – diagnostic function. If a problem occurs, an
error/warning code will be displayed o,n the display screen and the operator
can determine t.he malfunction with the warning code below.
ERROR | PROBLEM | SOLUTION |
---|
No display
| No main power connection.| Power check & proper plug-in of mains cable at
both ends.
Power failure| Check the mains fuse of the lab.
Improper conn ect io n.| Connect adaptor properly.
| Lid not closed correctly.| Close lid correctly.
Error with lid closing and opening mechanism.|
Contact service.
Err 55
| Rotor not loaded symmetrically
with equal weights
|
Load rotor symmetrically & restart ce nt rif uge.
Centrifuge lid cannot be opened
| Rotor is still spinning.| Wait for the rotor to come to a stop.
Power failure
| Emergency lid release after rotor stop
ERROR| PROBLEM| SOLUTION
---|---|---
Centrifuge shakes during acceleration & exceptional running noi•se
| Rotor not loaded symmetrically.| Load rotor symmetrically &
restart operation
Either a broken tube, damage to the rotor or motor is caused for run
noi•se.
| Replace broken tube. For damaged rotor/motor contact service representative.
Rotor damaged.| Remove & change rotor
Display error| Loose connection of display.| Contact service representative.
Err 1
| Latch is damaged, Latch jammed| ****
Contact service representative.
Lid not closed/locked| Close/lock the lid
Err 52
| ****
Rotor stuck or incorrect operating voltage
| Turn OFF the centrifuge, Check rotor fitment or apply
correct 230VAC ± lOVAC operating voltage
Power tripping| Cable not fit properly.| Remove cable and connect properly.
IMPORTANT NOTE:
- If the system get hangs or gets heated due to over current, switch OFF & switch ON (restart) the centri,fuge and check it again.
- Maintain 3 seconds gap between!J switch OFF and switch ON. Instant ON-OFF can lead to a reset, erasing last run memory.
- If the motor gets hot due to which there will be a fluctuation in speed value then allow the centrifuge to get cool for atleast 30 minutes. Do not do any operation for 30 minutes.
WARRANTY STATEMENT
This product is warranted to be free from de,fects in material and workmanship
for a period of Two (2) year from date of purchase. Your product will be duly
repaired upon prompt notification in compliance with the following
c.onditi’ons:
This warranty is vali,d only if the product is used for its intended purpose
and
w·ithin the guidelines specified in this instruction manual. This warranty
does not cover damage caused by accident, neglect, mi·suse, improper service,
natural forces or other causes not arising from defects in original material
or workmanship. This warranty does not cover any incidental or consequential
damages, commercial loss or any other damages from the use of this product.
The warranty is invalidated by any non,-factory modification, which will
immediately terminate all liabilities on us for the products or damages caused
by its use. The buyer and its customer shall be responsible for the product or
use of products as well as any supervision required for safety. If requested
the products must be returne.d to the distributor in well .Packed and insured
manne.r and all shipping charges must be paid.
Some states do not allow limitations on the length of implied warranties or
the exclusion or limitation of incidental or consequential damages. This
warranty gives you specific legal rights. This warranty is given expressly in
lieu of all other warranties, expressed or implied.
The purchaser agrees that there is no warranty of merchantability or·of
fitness for any intended purpose and that there are no other remedies or
warranties, expressed or implied, which extend beyond the description on the
face of the agreement. This warranty is only applicable to the original
purchaser.
Products received wit·hout proper authorization will not be entertained. All
items returned for service should be sent postage prepaid in the original pa.
packaging or another suitable carton, padded to avoid damage. We will not be
responsible for damage incurred by improper packaging.
All items returned for service should be set postage prepaid in the original
packaging or another suitable carton, added to avoid damage.
This warranty i.s valid only if the warranty is registered with the suppli·er
within 30 days from the date of purchase.
In case the product is to be disposed of, the relevant legal regulations are
to be observed.
PRODUCT DISPOSAL
Information on the disposal of electrical and electr·onic devices in the
European Co.mm unity
The disposal of electrical devices is regulated within the European Community
by national regulations based on ElJ Directive 2012/19/EU on waste electrical
and electronic equipment (WEE·E). According to these regulations, any devices
supplied after 13.06.05 in the business-to-business sphere, to which this
product is assigned, may no longer be disposed of.off in municipal or domestic
waste. They are_ marked with the following symbol to indicate this.
As disposal regulations within the EU may vary from country to country,
ple.ase contact your supplier if necessary.