BRESSER 5750800 Science TFM-201 Binocular Microscope
Product Information
The Science TFM-201/301 & Infinity is a microscope with electronic parts that can be powered by an AC adapter or batteries. The product comes with a transport lock that needs to be removed before use. The user manual provides safety instructions and information on product usage. The manual is available in different languages and can be accessed via a QR code or web link.
Product Usage Instructions
- Before using the microscope, remove the transport lock by unscrewing it. Refer to the chapter “Removing the transport lock” in the user manual for more information.
- Read the safety instructions and the instruction manual carefully before using the device.
- Make sure to keep the instruction manual in a safe place for future reference.
- If you plan to sell or pass on the device, make sure to provide the instruction manual to the subsequent owner/user.
- Avoid using the device with damp body parts to prevent the risk of electric shock.
- Keep the device away from children to prevent the risk of suffocation.
- Observe the safety instructions to prevent fires and body injuries.
Visit our website via the following QR Code or web link to find further information on this product or the available translations of these instructions.
ATTENTION!
Remove the screws for transport lock before using the microscope. Refer to
chapter “Removing the transport lock” for more information.
Imprint
Bresser GmbH
Gutenbergstr. 2 46414 Rhede Germany
www.bresser.de.
For any warranty claims or service inquiries, please refer to the information on “Warranty” and “Ser-vice” in this documentation. We ask for your understanding that unsolicited returns cannot be pro-cessed. Errors and technical changes excepted.
© 2023 Bresser GmbH
All rights reserved. The reproduction of this documentation – even in extracts
– in any form (e.g. photocopy, print, etc.) as well as the use and
distribution by means of electronic systems (e.g. image file, website, etc.)
without the prior written permission of the manufacturer is prohibited. The
designations and brand names of the respective companies used in this
documentation are gen-erally protected by trade, trademark and/or patent law
in Germany, the European Union and/or other countries.
Validity note
This documentation is valid for the products with the following article numbers: 5750800 5750900 5760700
- Manual version: 0423
Manual designation:
Manual_5750800-5750900-5760700_Science-TFM-Infinity_en_BRESSER_v042023a
Always provide information when requesting service.
About this Instruction Manual
NOTICE
These operating instructions are to be considered a component of the
device. Read the safety instructions and the instruction manual carefully
before using this device. Keep this instruction manual in a safe place for
future reference. If the device is sold or passed on, the instruction manual
must be passed on to any subsequent owner/user of the product.
Intended use
- This device is not intended for use by persons (including children) with impaired physical, sensory or mental abilities, or for lack of experience and/or knowledge, unless they are supervised by or have received instructions from a person responsible for their safety on how to use this device.
- This device is for personal use only.
- It was developed for an magnified representation of nature observations.
- The device is intended only for indoor use.
General safety instructions
DANGER
Risk of an electric shock!
This device contains electronic parts that are powered by a power source (AC
adapter and/or batteries). Improper use of this product may result in electric
shock. Electric shock can cause serious or fatal injuries. It is therefore
imperative that you observe the following safety information.
- Disconnect the power supply by pulling the mains plug when the appliance is not in use, in case of a longer interruption of operation and before any maintenance and cleaning work.
- Place your device so that it can be disconnected from the power supply at any time. The power outlet should always be near your appliance and should be easily accessible, as the plug of the power cord serves as a disconnect device from the mains supply.
- To disconnect the unit from the mains, always pull the mains plug and never pull the cable!
- Check this device, cables and connections for damage before use.
- Never attempt to operate a damaged device, or a device with damaged electrical parts! Damaged parts must be replaced immediately by an authorized service agent.
- Operate the device only in a completely dry environment and do not touch the device with wet or damp body parts.
DANGER
Danger of suffocation!
Improper use of this product may result in suffocation, especially for
children. It is therefore imperative that you observe the following safety
information.
- Keep packaging materials (plastic bags, rubber bands, etc.) away from children! There is a danger of choking!
- This product contains small parts that can be swallowed by children! Choking hazard!
- If small parts are swallowed, consult a doctor immediately!
DANGER
Danger of suffocation!
Improper use of this product may result in suffocation, especially for
children. It is therefore imperative that you observe the following safety
information.
- Keep packaging materials (plastic bags, rubber bands, etc.) away from children! There is a danger of choking!
- This product contains small parts that can be swallowed by children! Choking hazard!
- If small parts are swallowed, consult a doctor immediately!
CAUTION
Risk of body injury!
This device contains components and/or accessories that can cause minor or
serious body injury if used improperly. It is therefore essential that you
observe the following safety information in order to prevent body injuries.
- Sharp-edged and pointed tools are often used for working with this device. Therefore, keep this device and all accessories and tools in a place inaccessible to children. Danger of body injury!
- Chemicals and liquids do not belong in children’s hands! Don’t drink! Clean hands thoroughly under running water after use. In case of accidental contact with eyes or mouth, rinse with water. In case of complaints, consult a doctor immediately and show the substances.
CAUTION
Fire hazard!
Improper use of this product may result in fire. It is essential that you
observe the following safety information in order to avoid fires.
- Never cover the ventilation slots or cooling fins of the device during operation or until it has cooled down completely!
NOTICE
Danger of material damage!
Improper handling may result in damage to the unit and/or accessories.
Therefore, use the device only in accordance with the following safety
information.
- Do not disassemble the device! In the event of a defect, please contact your dealer. They will contact the Service Center and can arrange the return of this device for repair if necessary.
- Do not expose the device to high temperatures and protect it from water and high humidity.
- Do not subject the device to excessive vibrations.
- Only use accessories and spare parts for this device that comply with the technical specifications.
Product images
Parts overview
1 | Trinocular extension / Photo adapter |
---|---|
2 | Eyepiece |
3 | Eyepiece holder |
4 | Microscope head |
5 | Objective nosepiece |
6 | Objective |
7 | Object stage |
8 | Adjusting lever for iris diaphragm |
9 | Condenser |
10 | Lighting unit |
11 | Microscope foot |
12 | Rotary wheel for brightness control |
13 | Fine focus |
14 | Coarse focus |
--- | --- |
15 | Condenser height adjustment |
16 | Object clamping |
17 | Cross table longitudinal adjustment (Y-axis) |
18 | Cross table transverse adjustment (X-axis) |
19 | On/Off switch |
20 | DC connection socket |
21 | Universal USB power adapter |
22 | USB power cable with barrel connector |
23 | Mains adapter attachment (EU) |
24 | Mains adapter attachment (UK) |
*not for model TFM-201
Scope of delivery
Microscope (A); 10x WF Eyepiece (B); 4 Objectives (pre-assembled) (C); Condenser (pre-assembled) (D); Power adapter with plug and cable (E)
Choice of location
- Choose a suitable location before installation and commissioning.
- Place the device on a stable, level and vibration-free surface.
Removing the transport lock
Your BRESSER microscope is secured with one or more screws, depending on the
model, to prevent transport damage.
- NOTICE! REMOVE THE TRANSPORT SCREWS (T1 OR T1,T2 AND T3) BEFORE USE.
- NOTICE! Please refer to the following illustrations for the position of the screws. Use the Allen key included in the scope of delivery for this purpose.
- Model “Science TFM-201” (Art. No. 5750800)
- Model “Science TFM-301” (Art. No. 5750900)
-
Pull the lighting unit (10) out of the microscope base (11).
-
Loosen the screw (T1) in the rear area of the microscope foot and remove it completely.
Model “Science Infinity” (Art. No. 5760700) -
Pull the lighting unit (10) out of the microscope base (11).
-
Loosen the screw (T1) in the rear area of the microscope foot and remove it completely.
-
Loosen the two screws (T2 and T3) below the object stage (7) and remove them completely.
Installing the USB power adapter
- USB power adapter
- Adapter attachment (EU)
- Adapter attachment (UK)
- USB power cable with coaxial/barrel connector
- Place suitable adapter attachment with its notches on the pins of the USB power adapter.
- Turn adapter attachment in LOCK direction until it is locked.
- Put the USB plug of the power cable into the USB port of the power adapter.
- To remove the adapter attachment, push the PUSH button and turn the attachment in OPEN direc-tion.
NOTICE
Using the USB power cable with a computer
Alternatively the device can be powered by a computer via the included USB
power cable: Put the USB plug into a free USB port of the computer.
Setting up power supply
DC connection socket for coaxial/barrel connector
- On/Off switch
- DC connection socket
- Put the DC connector of the power cable into the DC connection socket of the device.
- Put the mains plug of the power cable into the power outlet.
Turning on/off the device
- Move On/Off switch to position ‘I’ to turn on the device.
- Move On/Off switch to ‘O’ position to turn off the device.
Adjusting the lighting
Lighting unit
- Rotary knob for brightness adjustment
- Illumination
This device provides an illumination with a permanently installed LED light source.
NOTICE
Do not operation the device on highest brightness over a longer time! This
will reduce the lifetime of the LED.
To control the brightness proceed as follows:
- Turn on the device.
- Turn/Move the dimmer control to adjust the brightness of the light source.
NOTICE
Illuminating the object optimally
Adjust the brightness of the light source so that the illumination works at
its optimum. Depending on light influences from outside and the object beeing
observed, the illumination needs regular adjustments.
Inserting slides
- Move the object plate downwards completely by turning the focus wheel.
- Press the object clamping outwards.
- Insert the slide straight into the fixture’s angle.
- Let the object clamping fall back carefully until the slide is fixed in the fixture.
Observation
NOTICE
Using eyepieces and objectives
When using a specific eyepiece or objective, you will get the individual
magnification of an object being observed. The magnification is imprinted on
the eyepiece and objective with a large number (e.g. 10x for 10-fold
magnification, 4 for 4-fold magnification).
- Move the object plate downwards completely by turning the focus wheel.
- Insert a slide and place it in the middle under the objective.
- Set an objective with lowest magnification below the object plate.
- Insert an eyepiece with lowest magnification in the eyepiece holder.
- Turn on the device and the illumination.
- Look through the eyepiece and move the object plate upwards by turning the focus wheel until a sharp image is visible.
- If neccessary, adjust the image sharpness by turning the fine focus adjustment.
NOTICE
Avoid damages to the specimen!
Make sure you turn the focus wheel very slowly, to avoid the objective touch
the slide or specimen, and thus to avoid damages to either of them, or to
both. Only turn the focus wheel until it stops and do not apply excessive
pressure.
Adjusting the eye distance
Distance between the eyepieces and diopter adjustment
- With the eyes look through both eyepieces.
- Change the distance between the eyepieces by pushing them together or pulling apart until a circular image of the obeserved object is visible.
Diopter adjustment
- Turn the diopter adjustment ring to position ‘0’.
- With the left eye, look through the left eyepiece and turn the focusing wheel until a sharp image of the observed object is visible.
- With the right eye, look through the right eyepiece and turn the diopter adjustment ring until a sharp image of the observed object is visible too.
The diopter adjustment can also be done in the opposite way on the left eyepiece.
Using
Using the iris diaphragm
NOTICE
The iris diaphragm is used to increase the depth of focus. With an open
aperture more light is getting through but the depth of focus is less. With
partially closed aperture the depth of focus is higher but it is neccessary to
increase the brightness of the illumination then.
It is recommended not to close the iris diaphragm too much when working with higher magnifications. Otherwise the image resolution will decrease.
Move the lever for the iris diaphragm to right or left to change the aperture setting.
Using the Köhler illumination
Proceed as follows to adjust the Köhler illumination for an optimal
illumination of the field of view:
- Focus on the object with the 10x objective.
- Close the condensor and the light field aperture. For microscopes with additional lens, fold it out.
- Move the condensor up and down until the edge of the light field aperture is in focus.
- If neccessary, use the adjustment screws to move the condensor until the light field aperture is centered in the field of view.
- Open the light field aperture as far as it is not visible in the field of view anymore.
Oil immersion
The oil immersion can be done with a suitable 100x objective only. For the
purpose proceed as follows:
- Focus on the object with the 40x objective.
- Turn the 40x objective sideways.
- Drop a drop of immersion oil onto the cover glass of the preparation.
- Retract the 100x objective so that it is immersed in the oil.
- Refocus the image.
- Clean the 100x objective after each use.
Using the trinocular extension
Trinocular extension with pre-assembled photo adapter
1 | Trinocular extension | 2 | Protection cap |
---|---|---|---|
3 | Fixing screw | 4 | Photo adapter thread |
5 | Photo adapter |
The trinocular extension of the microscope was specially designed for C-mount threads. For the camera mounting proceed as follows:
- Remove the dust protection cap from the trinocular head.
- Attach the camera* with its C-mount thread to the photo adapter thread and turn it on.
- Loosen the clamping screw of the trinocular extension to move the photo adapter with the camera* within the extension tube. In this way the image sharpness and the position of the camera can be adjusted. Most camera models can be adjusted parfocal to the eyepieces when doing so.
- Re-tighten the screw to fix the camera in the desired position.
not included
Cleaning and maintenance
- Protect the device from dust and moisture!
- Avoid fingerprints and similar contamination of the optical surfaces.
- The residual moisture must be completely degraded before use.
- Disconnect the power supply by pulling the mains plug when the appliance is not in use, in case of a longer interruption of operation and before any maintenance and cleaning work.
- Only clean the device externally using a dry cloth. Do not use cleaning solution to prevent damage to the electronic parts.
- Do not use organic solvents (e.g. alcohol, ether, acetone, xylol or other dilutions) to clean lacquered parts or plastic parts!
- Disassemble optical elements (e.g. objective, eyepiece, etc) before cleaning.
- Blow away loose dust from the lens surfaces first.
- Use high quality lens cleaning towels or soft cloth and moisten it with a little bit of pure alcohol (available in a drug store). Wipe the lens surface with it.
- Pull a dust protection cover* over the microscope and store it in a closed container at a dry and mould-free place. Store the microscope and the accessories in the relevant containers when they are not used for a longer time. It is recommended to store also objectives and eyepieces in closed containers with desiccant.
(*depending on the model a dust protection cover and box can be part of the content)
NOTICE
Remember:
A well maintained microscope will keep its optical quality for years and thus
maintain its value.
Disposal
Dispose of the packaging materials according to its type. Information on
proper disposal can be obtained from the municipal waste disposal service
provider or environmental agency.
Do not dispose of electronic devices in the household garbage! According to the European Directive 2012/19/EU on Waste Electrical and Electronic Equipment and its transposition into national law, used electrical equipment must be collected separately and recycled in an environmentally sound manner.
Batteries and rechargeable batteries must not be disposed of with household waste. You are legally obliged to return used batteries and accumulators and can return them after use either at our sales outlet or in the immediate vicinity (e.g. in the trade or in municipal collection points) free of charge. Batteries and accumulators are marked with a crossed-out dustbin and the chemical symbol of the pollutant, “Cd” stands for cadmium, “Hg” stands for mercury and “Pb” stands for lead.
Technical data
Science TFM-201 | Science TFM-301 | Science Infinity | |
---|---|---|---|
Article No. | 5750800 | 5750900 | 5760700 |
Microscope head | binocular | trinocular (incl. photo adapter) | trinocular |
(incl. photo adapter)
Cross table| Yes, with Nonius scale| Yes, with Nonius scale| Yes, with
Nonius scale
Eyepiece(s)| 1 Pair of DIN WF 10x| 1 Pair of DIN WF 10x| 1 Pair of DIN
WF 10x
Objective(s)| 4x / 10x / 40x / 100x| 4x / 10x / 40x / 100x| 4x / 10x /
40x / 100x
Magnification| 40x / 100x / 400x /
1000x
| 40x / 100x / 400x /
1000x
| 40x / 100x / 400x /
1000x
---|---|---|---
Power supply| DC 5V / 1000mA| DC 5V / 1000mA| DC 5V / 1000mA
Illumination| LED| LED| LED*
The LED illumination used in this device has a very long lifespan and do not need to be changed. LEDs can not be changed due to technical reasons!
Warranty
The regular guarantee period is 10 years and begins on the day of purchase. You can consult the full guarantee terms and details of our services at www.bresser.de/warranty_terms.
Declaration of Conformity
EC Declaration of Conformity
Bresser GmbH has issued a “Declaration of Conformity” in accordance with
applicable guidelines and corresponding standards. This can be viewed any time
upon request. www.bresser.de •
info@bresser.de.
UKCA declaration of conformity
A “Declaration of Conformity” in accordance with the applicable directives and
relevant standards has been issued by Bresser GmbH The full text of the UKCA
Declaration of Conformity is available at the following web address:
https://www.bresser.de/download/science/ukca/science_ukca.pdf.
Bresser UK Ltd.
Suite 3G, Eden House, Enterprise Way, Edenbridge, Kent TN8 6HF, Great Britain.
Please contact the service centre first for any questions regarding the product or claims, preferably by e-mail.
- E-Mail: service@bresseruk.com.
- Telephone*: +44 1342 837 098
BRESSER UK Ltd.
Suite 3G, Eden House Enterprise Way Edenbridge, Kent TN8 6HF Great Britain.
Number charged at local rates in the UK (the amount you will be charged per phone call will depend on the tariff of your phone provider); calls from abroad will involve higher costs.
Bresser GmbH
Gutenbergstraße 2 46414 Rhede · Germany.
www.bresser.de
@BresserEurope
References
- konsoleH :: Login
- IBERIA.COM en United States - the best prices for Iberia flights - Iberia USA
- Bresser | Startseite | Expand Your Horizon mit BRESSER
- Bresser | BRESSER Mikroskop Science TFM-201 Bino (30) | Expand Your Horizon
- Bresser | BRESSER Science Infinity Mikroskop | Expand Your Horizon
- bresser.de/warranty_terms
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>