etac IMM1049_IFU Immedia AutoTurn Swivel Cushion Instructions
- October 30, 2023
- Etac
Table of Contents
etac IMM1049_IFU Immedia AutoTurn Swivel Cushion Instructions
General
Thank you for choosing an Etac product.
To avoid accidents and injury when moving and handling the products, please read this manual carefully.
The person described as “the user” in this “Instruction for Use” is the person who is lying or sitting on the product. The carers are the people who manoeuvre the product.
This symbol appears alongside the text in the manual. It draws the reader’s attention to points at which there may be a risk to the health and safety of the user or carer.
These products comply with the standards applicable for Class 1 products in the Medical Device Regulation (EU) 2017/745 on medical devices.
At Etac we strive to continually improve our products and therefore we reserve the right to make changes to products without prior warning. All measurements given on illustrations and similar material are for guidance only and Etac cannot be held liable for errors and defects.
The information given in this manual, including recommendations, combinations and sizing, does not apply to special orders and modifications. If the customer makes adjustments, repairs or uses combinations not predetermined by Etac, the Etac CE certification and Etac warranty will not be valid. If in doubt, please contact Etac.
Warranty: Two-year warranty on material and manufacturing defects, provided that the product is used correctly.
Expected lifetime: The device has an expected service life of 1 to 2 years under normal use. The service life of the device varies depending on usage frequency, loads and how often and how it is washed.
For further information on the Etac Patient handling, see www.etac.com.
In case of an adverse event occurred in relation to the device, incidents should be reported to your local dealer and the national competent authority in a timely manner. The local dealer will forward information to manufacturer.
Symbols in general
Washing
Do not bleach
Tumbledry
Iron
Do not dryclean
Medical device
Disinfect / Wipe off
Never leave on the floor
Risk of sliding down
Keep out of the sun
Read the user manual
Warning
Production year, month and day
Batch no./Lot no.
User mass limit = maximum rated load
The product can be scrapped according to national regulations
Manufacturer
CE -certification
Keep dry
Non sterile
Article number
Intended use
AutoTurn is specially developed for use with car seats to make it easy to enter and exit a vehicle. The two nylon sides facing each other reduce friction during turning. The underside of the AutoTurn pillow has a non-slip surface, to ensure that the device rests firmly on the car seat.
AutoTurn is fastened to the seat back.
AutoTurn features a functional handle which a carer can use to pull the user
into place on the car seat.
Intended environment
Acute care, Long-term care, Home care
Practical handling
Positioning
Position AutoTurn centrally on the car seat. Fit the straps, fasten and tighten securely at seat back. Fit the functional locking system (next to the functional handle) to the safety belt. When the user turns in the seat, the belt will automatically glide slightly forward.
The safety belt can be secured on the opposite side using the functional locking system **(Illustration 4).
**
Use
The user seats him/herself on the centre of AutoTurn. The user then raises either both legs simultaneously or one leg at a time inside the vehicle, thus turning the body inside the car (AutoTurn is also perfect for turning the user inside the vehicle) (Illustrations 1, 2 and 3 ). This avoids unnecessary strain on the body. The carer can pull on the functional handle to help the user glide into place.
To exit the car, the user swings the legs out of the vehicle, preferably in stages, before raising him/herself to the standing position.
Notice
Special features
Always check the product before use and after washing.
Never use a defective product. If the product shows signs of wear-and-tear, it must be scrapped.
Read these instructions carefully.
It is important that carers receive instruction in manual handling. Etac offers advice and training for carers. For further information, contact Etac.
Always use the correct manual handling techniques.
Encourage the user to assist where possible.
To ensure that the user feels safe and that every manual handling is smooth, always plan the manual handling in advance.
Never leave the product on the floor.
Always conduct a risk assessment, and ensure that the assistive product can be used for the individual user and, in combination with other devices, that it is safe for the user and carers.
Always fit the strap, fasten and tighten securely at the seat back.
If in any doubt – please contact Etac.
Material and cleaning
Material:
Top side: Polyester
Inner surface: Nylon
Back side: Non-Slip.
The product should be checked regularly, preferably each time it is used, and especially after washing.
Check that there is no damage to seams or fabric.
Never try to repair a product yourself. Never use a defective product.
Washing instructions:
Do not use fabric softener.
This product can be cleaned with a solvent-free cleaning agent with a pH level
between 5 and 9, or a 70% disinfection solution.
Products and accessories
Article no. | Product Description | Size (mm) |
---|---|---|
IM99999 | AutoTurn w/belt | Ø450 |
Service information
Inspection
First inspection:
- Is the packaging intact?
- Read the label on the packaging and check the article no. and product description
- Check that Short Instruction is enclosed – updated User Manual/Instruction for use can be downloaded from www.etac.com or contact your Etac Customer Service or the local dealer.
- Check the label on the product – does it include article no., product description, lot/batch no., cleaning instructions and supplier name?
Periodic Inspection:
- Make sure that your Manual transfer system always is in a perfect condition.
- After wash – always check materials, stitching, handles, buckles.
- If the product shows signs of wear and tear, it must be removed from service immediately.
Visual inspection/Check the product:
Ensure that the materials, seams, stitching, handles, buckles are
intact/faultless.
Mechanical load or stress/Test
handles:
Draw hard in the handles in opposite directions and control the material,
seams and stitching. Test buckles:
Lock the buckle and draw in opposite directions, control the buckle and the
stitching
Stability test:
Try to bend the product verify that it feels firm, stable and solid.
Test low/high friction:
Place the product on a firm surface or a bed, place your hands on the material
and test the friction by pushing your hands down into the product.
Low friction – slides effortlessly
High friction – no sliding/moving
Check:
Manual transfer aids with handles, straps and buckles:
Manual transfer aids with for sitting and lying transfer and with high or low
friction:
Production year, month and day
The product can be scrapped in accordance with national regulations
Read the user manual
References
Read User Manual Online (PDF format)