HUATO S380-WS Explosion-Proof Temperature Humidity Data Logger User Manual

October 30, 2023
HUATO

HUATO S380-WS Explosion-Proof Temperature Humidity Data Logger

S380WS Data Logger Hardware

Introduction

The HUATO S380-WS data logger has a large capacity of up to 120,000 readings and can record data for up to one year based on a sampling frequency of 10 minutes. It comes with matched software for data download and transfer to a computer. The user can set the log time, sampling interval, and logging interval by software.

Product Usage Instructions

  1. Remove the back cover of the data logger and install two ER14505 (3.6V) Lithium batteries.
  2. Connect the data logger to the computer using the USB port.
  3. Install the driver and Logpro software on the computer.
  4. Run Logpro software and select Connect from the toolbar.
  5. Read the logger’s properties and set them accordingly.
  6. Synchronize the properties and time.
  7. Press Log button on Menu to start recording.
  8. Select Disconnect from the toolbar and unplug the logger from the computer USB port when finished.
  9. Refer to Chapter 3 and Chapter 4 in the user manual for troubleshooting tips.

Features

  • Logging interval: 60 seconds to 6 hours
  • Logging protection: Logging will stop when the battery is low. The value can be saved forever safely.
  • Display Resolution: temperature: 0.1 / humidity: 0.1%RH
  • Interface: USB Port
  • Dimension: 97 68 32 mm
  • Wireless antenna: Built-in
  • Installation: Wall-mounted with fixed support, easy to use and convenient to maintain
  • Battery: 2 x ER14505 (3.6V AA Size) Non-rechargeable Lithium Battery

Model

Model| Measurement Accuracy| Measurement Range| Probe Type (T:Temperature; H:Humidity)| Capacity of Records
---|---|---|---|---
-20~70 ℃ / 0~100%RH| N/A| -20~70 ℃ / 0~100%RH| T&H| 120,000

Appearance (S380WS)

The S380WS data logger has an LCD display with ON/OFF(Log) and ON/OFF(Power) options, a USB interface, a wall-mounted back cover, a power back cover, and various symbols for power, signal, wireless, display RH, DC power, LOG (Recording) status, selectable/humidity symbols.

  1. LCD Display
  2. ON/OFF(Log)
  3. ON/OFF(Power)
  4. USB Interface
  5. Wall-mounted Back Cover
  6. Power Back Cover

Screen (S380WS)

The S380WS screen displays various symbols for power, signal, wireless, display RH, DC power, LOG (Recording) status, selectable/humidity symbols.

  1. Power Symbol
  2. Signal Symbol
  3. Wireless Symbol
  4. ℃/℉ Display
  5. RH Display
  6. DC power Symbol
  7. LOG (Recording) Status Symbol
  8. Selectable ℃/℉ Symbol
  9. Humidity Symbol

Install battery

  1. Take out the back cover as indicated by the symbol.
  2. Install 2 x ER14505 (3.6V) Lithium batteries.

Logpro Software

The Requirement of the Computer Hardware

  1. OS: Windows XP/Vista/7/8 (32/64bit)
  2. CPU: 1.6GHz
  3. Physical Memory: 512MB
  4. Hard-drive Space: 4 GB
  5. USB Port: 1

Install Driver & Software

  1. Install Driver
    Notice : Windows 64bit please install 64bit-driver.

  2. Install Logpro Software
    Notice : Please Install Logpro Software in D:

Setting the Logger’s Properties

  1. Connect the data logger to the computer.
  2. Run Logpro software on the PC.
  3. From the toolbar, select Connect.
  4. Read the logger’s properties.
  5. Set the properties.
  6. Sync the properties & time. (The PC time will be synchronized to the logger as well.) Now, the data logger is in Standby mode.
  7. Press Log button on Menu, start recording.
  8. From the toolbar, select Disconnect.
  9. Unplug the logger from the Computer USB Port.

The windows operating system cannot handle USB devices being unplugged and plugged back too fast. When unplugging the logger, wait for about 5 seconds before plugging it in again. If you unplug and plug back a device too quickly, the computer may stop recognizing any USB devices on that port. If this happens you will have to restart the computer. This is a windows USB problem and is not related to Logpro.  If more than one logger is connected to your computer at the same time, the program will ask you to choose COM  port manually.  The logger has three modes: 1.LOG: In the mode, the logger samples and records data timely. 2.Standby: In the  mode, the logger stops to sample and record, and the LCD display is ON. 3.OFF: In the mode, the logger stops to sample and record, and the LCD display is off. Logpro cannot connect to the logger also.

Properties Description

Property Text
Name English Letters or Numbers
SN 10 Characters
Property Text Description
Name English Letters or Numbers Name of the logger
SN 10 Characters Must be the same as the one in the label of

logger.

Sampling Interval (s)| Number from 1 to 240| Sampling frequency in LCD screen
Logging Interval (s)| Number from 2 to 86400| Logging frequency / S400 Upload frequency
Down Mode|  | (Cannot be changed)
Start Mode|  | (Cannot be changed)
Delay Time|  | (Cannot be changed)
Stop Mode| First In First Out| When recording memory is full, the logger

will delete the earliest data

Stop When Full| When recording memory is full, the logger

will stop record new data

Logs|  | The count of records in the memory.
Total|  | The total capacity of the logger’s storage.
Alarm|  | Turn ON / OFF

the Data Logger’s Over Limit Alarm

LCD|  | Please choose ON
Power|  | Please choose Normal
High & Low Limit| CH1: Temperature

CH2: Humidity / Temperature2

| When the value over the limit, the buzz will

sound / red LED will flash an alarm.

Offset6|  | Input positive number to decrease the value.

Input negative number to increase the value.

Calibrate the logger: The logger is factory calibrated to an accuracy given in the device specifications. However, there may be times when you wish to adjust the calibration of your logger. Logpro provides you with the ability to perform a single point offset calibration. This calibration can be used to increase the accuracy of the logger for a restricted data range.

Please refer to Chapter 3 (Attention) and Chapter 4 (FAQ) in the user manual for additional information and troubleshooting  tips.

Attention

  • S400W / S380WS Series Data Logger are not waterproof. Do not connect loggers with computer when the logger is on water.
  • S400W / S380WS Series Data Logger plastic shell is ABS, flame retardant, not resistant to acid and alkali.
  • If repair is needed, only authorized technician could do the repair.
  • AA alkaline battery (1.5V) is not rechargeable.
  • ER14505 lithium battery (3.7V) is not rechargeable.
  • USB power supply (5V) cannot let the Data Logger work.
  • LCD Screen Dim
  • Recording (LOG) are Automatically Stop
  • Software Runtime Error
  • Check COM Port Number

FAQ

LCD Screen Dim

Reason: Insufficient battery or the environment temperature is too low or too high.

Solution: In the case of insufficient battery, please replace the battery. If resulted from environment temperature, please immediately take the logger out of the environment.

Recording (LOG) are Automatically Stop

Reason: When the battery level is low, there will not be sufficient power to support recording function of the logger.

Solution: Replace the battery.

Software Runtime Error

Reason: OS forbid software creating files.

Solution:

  • Run the program (software) as an administrator.
  • Install software in Disk D:\
  • A data logger name cannot contain any of the following characters: \ / : * ? ” < > |
  • Software’s installation path cannot include Chinese character or garbage character.

Check COM Port Number
Press “Win” + “R” in keyboard -> Run “devmgmt.msc” to Open “Device Manager” in Windows-> Expand “Ports (COM & LPT)” -> “USB-SERIAL CH340 (COM No.)” is the Data Logger

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>