DECLARATION
Thank you for choosing Coremorrow’s products!
This user manual is applicable to P12.XY500Z800 piezo nanopositioning stage.
For your smooth and correct use of this product, please read this manual
carefully before use.
Please follow the instructions in the manual installation and use procedures.
Improper operation may cause injury to the operator or damage to the product.
Therefore, the entire installation and operation process should be completed
by a professional with a certain foundation and understanding of the product
principle, or under the guidance of a professional.
If the product is disassembled or modified without permission, our company
will not be responsible for any consequences arising therefrom.
It is illegal to change the product model and other false sales of our
products. Users are advised to be vigilant. Once you find out or be deceived,
contact us directly, we will crack down on illegal activities and prevent
deception and economic losses. Coremorrow will certainly investigate legal
responsibility of the illegal subject and related parties.
With the continuous development and innovation of technology, we will add the
latest information to the manual as needed without notice. Please contact us
if you need any information.
Introduction
Purpose and content
- The manual is mainly for your smooth and correct use of piezo stage.
- The manual contains all necessary information for P12.XY500Z800 piezo stage.
- The manual describes the precautions during installation and use
Symbol and meaning
DANGER: Improper handling may could cause injury to the operator.
Attention: Improper handling can cause damage to the equipm
- No order between items.
- Please operating in order
Expected reader
The manual default user has knowledge of piezo stages and controllers, and has a basis for its installation and use.
Manual lost replacement
If the manual is accidentally lost, please contact our customer service department, we will send you the latest version of user manual.
Safety Guide
P12.XY500Z800 piezo nanopositioning stage is designed with advanced technology and safety standards, for safety and proper use of the product, please pay attention to following points when using:
**** Temperature changed and pressure will charge the piezo stage. Piezo stage will remain charged for a period of time disconnected from the controller, so the operator should pay attention when using it:
-
Do not disassemble P12.XY500Z800 piezo stage;
-
Discharge P12.XY500Z800 piezo stage before installation, it could be achieved by connecting to the controller;
-
Do not pull out the controller during operation;
-
Before using, please check whether the connecting wire of the product is in good condition, whether the product and the controller are effectively protected from grounding, in case of danger, please strictly regulate the operation during operation, do not touch it (up to 150V voltage) after power on
In order to prevent damage to the product caused by improper operation, please pay attention to following points when using: -
P12.XY500Z800 piezo stage is used as much as possible in dust-free, oil-free, lubricant-free environments;
-
Since the product adopts flexible structure design, it is recommended that the load not exceed the bearing capacity of the product. Pay attention to the torque when loading. Do not twist or twist the load surface to avoid damage to the structure;
-
Recommended voltage 0~120V for long-term operation to extend lifetime;
-
Do not disassemble the product to avoid product damage;
-
Avoid stretching and bending the cable interface to prevent damage to the cable;
-
Please use the dedicated cable provided by us to connect the piezo stage and piezo controller;
-
Please do not use cable extension wire, if need longer cable, please contact us.
Product Overview
Product Introduction
P12.XY500Z800 piezo nanopositioning stage has the following advantages:
-
High performance piezo stack:
P12.XY500Z800 piezo stage is driven by high-performance piezo stack. The piezo stage is fully insulated, so its performance and service life are far superior to those of conventional stage. The insulating layer could effectively prevent the piezo stack from being damaged by moisture. The stability of piezo stack could be ensured under extreme conditions. The piezo stage has the characteristics of no rebound, no wear. -
Flexible hinge structure:
P12.XY500Z800 adopts flexible hinge mechanism, which has the advantages of no friction and high positioning accuracy. The flexible hinge is a component that is no taffected by static and dynamic friction. It is based on the elastic deformation of solid (such as steel) and has no rolling and sliding parts. With high stiffness and load carrying capacity, the hinge guides are maintenance-free and wear-free. Suitable for a wide range of temperatures without the need for lubricants. -
SGS sensor:
SGS sensor detects position information by expanding, they are fixed at appropriate positions in transmission portion, and the displacement is measured.
P12.XY500Z800 piezo nanopositioning stage adopts frictionless flexible hinge guiding mechanism and has the characteristics of high rigidity, high resolution and fast response time.
P12.XY500Z800 is mainly used in biotechnology, interferometry, quality inspection, optical detection, nanopositioning, microscopy and imaging, image processing and other fields, and has very good development prospects
Product Features
- XYZ linear motion
- XY travel to 500μm, Z travel to 800μm
- For objective XYZ positioning adjustment
- Vacuum version available
Motion direction
P12.XY500Z800 connection mark and motion direction are as follows:
Model | Motion direction | Description | Connection Mark |
---|---|---|---|
P12.XY500Z800 | X | Drive cable | CH1 |
Sensor cable | 1 | ||
Y | Drive cable | CH2 | |
Sensor cable | 2 | ||
Z | Drive cable | CH3 | |
Sensor cable | 3 |
Operating environment
- Operating temperature: -20~80℃
- Environment free of dust, clean, and non-corrosive substances
- Relative humidity: <60%
- Recommended voltage between 0~120V for long-term and high-reliability operation
Installation and operation
Before installation, first check the safety, such as whether the cable is damaged or broken, and whether the power supply has grounding protection.
Installation and connection
P12.XY500Z800 installation diagram is as follows:
-
When connecting the load, the load must not exceed the maximum allowed load of P12. XY500Z800.
-
Grounding protection for connection, as shown below.
-
Mounting the load and tighten it properly to ensure the positioning accuracy.
-
Connect the connector with the corresponding controller according to the identification.
The wiring mode of LEMO connector is shown as below
Load
When installing the load, ensure that the load mass is less than the max load
of P12. XY500Z800, and install the load as much as possible at the center of
the moving surface.
The correct placement of the load is shown below:
Load applied to the center of the stage
For protection, we don’t recommend the load is too high. If it is used a
higher load, please reduce the position center of the load gravity. The error
placement method is as follows:
Try not to let the load center of gravity exceed the stage
It is forbidden to load the load on one side of the piezo stage, which would
cause serious damage to the product, as shown in the following
figure:
Do not load longer loads and add to the side of the piezo stage
Operating
Please confirm the following information before starting the power supply:
- P12.XY500Z800 piezo nanopositioning stage has been grounded.
- The environment meets requirements of P12.XY500Z800 working environment.
- The fixing screws must be tightened.
The above information is determined to perform controller related operations:
- Please carefully read piezo controller user manual before operating, then start the controller;
- Set the piezo controller to zero;
- After checking for no abnormalities, input voltage values according to requirements (keyboard or analog control);
- It could work normally within the rated voltage range of piezo ceramic (0~120V, max not exceeding 150V);
- Please set the piezo controller to zero after used.
Please do not touch the product during the operation, avoid affecting the piezo stage performance or even damage the piezo stage.
The resolution of piezo ceramic is not limited. It depends on the performance of piezo controller. In order to better achieve the motion accuracy, it is recommended to use CoreMorrow piezo controller. For the operation of piezo controller, please refer to piezo controller usual manua
Note:
- Please do not remove the grounding protection during operation. If it needs to be temporarily removed, reconnect the grounding protection before starting again.
- Drive voltage should not exceed 150V;
- If abnormal noise or oscillation occurs during operating, immediately turn off the power check parameter setting.
Transportation and inspection
Transportation
Piezo stage is packed in carton. The transportation is carried out under the condition of packaging. This product can be transported by various means of transportation under normal conditions. During transportation, it should avoid direct rain and snow, avoid contact with corrosive substances, and avoid strong collision, try to avoid improper behavior such as squeezing, irregular placement.
Unpacking inspection
Please check below information:
- ase check below information:
- Before opening the package, please carefully check the packaging for damage, bruising, wetting, moisture, deformation, etc.
- Please check the parts on a case-by-case basis based on the supply contract and the packing list
- Whether there is obvious damage on the surface of the product
- Whether the product identification is clear and complete
- Whether the connector is loose, the connection line is broken or not, etc.
If any of the above occurs, please make a detailed record and take a photo, andplease contact us immediately
Precautions:
- Please handle gently when unpacking to avoid permanent damage to the stage caused by bumps
- Please do not use heavy tools or use rough methods to open them
- Please do not discard all packaging materials for return shipments
Maintenance and processing
Problem Processing
Common problems are shown in the following table:
Problem Description | Possible reasons | Solutions |
---|---|---|
No displacement or displacement becomes smaller | The cable is not connected | |
correctly or is in faulty contact. | Check the cable connection | |
Excessive load | Loaded by product max load | |
Sensor zero drift | Calibrate the sensor | |
Reduced accuracy | The installation base is uneven | Mount the stage on flat |
surface
Loose connection| Fastening connection
Wiring is wrong| Check the wiring is correct
Vibration or inaccurate positioning at start-up| The parameters are
incorrect.| Stop immediately, check the parameter settings
Resonance| operating frequency less than the resonant frequency
If the user has problems that could not be solved. Please record the fault
situation and contact with us. The professional technical engineer would help
solve the problem.
Problems caused by improper operation are not covered by the warranty;
CoreMorrow is not responsible for any problems caused by dismantling the peizo
stage without permission.
Daily maintenance
If the piezo stage is not used for a long time, please store it in an room
environment with no dust, good ventilation, clean and non-corrosive
substances.
Please do not loosen the screws on the stage structure, avoiding to affect the
accuracy of the stage
Daily cleaning:
Please disconnect the piezo stage to controller, then clean the surface of it.
When cleaning, first wipe a small amount of detergent with a towel and gently
wipe the surface of the piezo stage, then wash it with alcohol. Please do not
use ultrasonic waves for cleaning.
When cleaning the surface, try not to pick up the product. If it is necessary
to pick up, please place a sponge under the piezo stage for protection, the
piezo stage should not be too high from the sponge to prevent the piezo stage
from falling off.
Connector
Driving and Sensor cable
Driving cable interface is 1-contact LEMO, model number is LEMO
ERA.00.250.CTL, as shown below:
Sensor cable interface is 4-contacts LEMO, model number is LEMO
ERA.0S.304.CLL, as shown below:
LEMO Sensors Connector
Number | Description |
---|---|
1 | +10V |
2 | Sensor feedback signal + |
3 | Sensor feedback signal + |
4 | GND |
Sensor | GND(protect) |
- Supply voltage
- Sensor signal out
Customer Service
If you have questions about the products you are currently using, please let us know the following information:
- Piezo stage model number and related number
- Controller model for this product
- Software driver version for this product
- Computer operating system for installing supporting software
Contact us
Harbin Core Tomorrow Science & Technology Co., Ltd.
Tel: +86-451-86268790
Email: info@coremorrow.com
Website: www.coremorrow.com
Address: Building I2, No.191 Xuefu Road, Nangang District, Harbin, HLJ,
China
CoreMorrow Official and CTO WeChat are below:
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>