tommee tippee 0499056 Notouch Forehead Infrared Thermometer
Product Information
Forehead Infrared Thermometer
The Forehead Infrared Thermometer is a no-touch thermometer that uses infrared technology to provide easy and accurate temperature readings in just one second. It is ideal for taking your child’s temperature without disturbing them. The thermometer has an LCD display and several buttons that control its functions. It can measure both object temperature and forehead temperature. The product comes with a memory function that can store up to 20 readings. The thermometer has low RF emissions and is suitable for use in all establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Product Usage Instructions
Step 1: Cleaning and Storage
- Before using the thermometer, clean the scanner lens with a soft cloth or cotton swab moistened with alcohol. The thermometer should be stored in a dry place at room temperature.
Step 2: Measuring Forehead Temperature
- To switch on the thermometer, press the power button.
- Use the mode button to select Forehead.
- Point the infrared sensor to the center of the forehead, keeping a distance of 1-3 cm between the thermometer and the forehead.
- Press the measure button.
- The forehead temperature will be displayed on the screen.
Step 3: Measuring Object Temperature
- To switch on the thermometer, press the power button.
- Use the mode button to select Object mode.
- Point the infrared sensor to the center of the object, keeping a distance of 1-3 cm between the thermometer and the object.
- Press and release the measure button.
- The object temperature will be displayed on the screen.
Step 4: Tips for Taking an Accurate Reading
The thermometer should be held steady and aimed at the center of the forehead or object. For accurate measurements, the distance between the thermometer and the forehead or object should be 1-3 cm. Be careful not to touch the forehead while taking a measurement.
Step 5: Interpreting Your Reading
- The temperature value displayed on the screen indicates the temperature of the forehead or object being measured.
Step 6: Memory Function
- When the product is switched on, press the memory button to access the last 20 readings. The screen will display 01 followed by the recorded reading. Press the memory button again to retrieve the second reading, and so on for the past 20 readings.
Step 7: Replacing the Batteries
- To replace the batteries, open the battery cover and remove the old batteries. Insert new batteries and replace the battery cover.
- If you have any further questions or need support, scan the QR code on the back of the product to visit our consumer support page at tommeetippee.com or join the conversation.
Here for you
When your baby is asleep and you need reassurance. Our no-touch thermometer uses infrared technology for easy and accurate one-second temperature readings. Making it ideal for taking your child’s temperature without disturbing them. And that’s when you can count on us.
Questions?
For further information or support, please scan below to visit our consumer support at tommeetippee.com
Or join the conversation
IMPORTANT WARNINGS
These instructions can also be found at tommeetippee.com
- This thermometer is not intended as a substitute for consultation with your doctor.
- Adult use only.
- Keep out of reach of children when not in use.
- Do not allow children to take their own temperature unattended.
- To avoid risk of choking from swallowing small parts or batteries please keep the device out of the reach of children and pets.
- The device should not be submerged into water or other liquids (not waterproof).
- The scanner lens is the most delicate part of the thermometer. Use with care when cleaning the scanner lens to avoid damage. See care and cleaning section.
- Do not use the thermometer if there are signs of damage to the lens or the thermometer itself. If damaged do not attempt to repair the product.
- No modification of this equipment is permitted.
- Intended for household use only.
- Do not use in direct sunlight.
Cleaning and storage
- Make sure the probe is clean to ensure an accurate reading.
- Use a 70% alcohol swab or cotton wool moistened with 70% alcohol to gently wipe the lens clean.
- Allow the lens to fully dry for at least 1 minute.
- Never insert a sharp object into the scanner area or any other open surface on the thermometer, because this will cause damage and affect functionality.
- Keep the thermometer dry and away from any liquids and direct sunlight.
- It should be stored at temperatures between -20ºC ~ +50ºC ,RH < 95%.
- LCD display
- Power button / Measure button
- Sound switch button
- Memory button
- Mode button
- (Object temperature, forehead temperature)
- Battery cover
- Scanner lens / Infrared sensor
Functions
- Object temperature mode
- Forehead temperature
- Mute / un-mute
- Low power indicator
- Temperature unit (˚C / ˚F)
- Temperature value
- Memory mode
Measuring forehead temperature
- A. To switch on press the power button
- B. Use the mode button to select Forehead
- C. Point the infrared sensor to the centre of the forehead
- D. Press the measure button
- E. The forehead temperature will be displayed on the screen
For effective measurement, the distance between the thermometer and the forehead must be 1-3 cm. Be careful not to touch the forehead.
Tips for taking an accurate reading
- Child must be inside for 30 minutes before taking a measurement.
- The child and device must be in the same stable ambient (room) temperature for 30 minutes before operating.
- Always make sure the scanner lens is clean and undamaged.
- Child should not drink, eat or be physically active before / while taking the measurement.
- Remove hats and wait 10 minutes before taking a measurement.
- Always hold the thermometer and forehead steady when taking a reading.
- Do not move the thermometer until you hear the final beep.
- Don’t take a measurement during or immediately after nursing a baby.
- Wait at least 60mins after bathing to take a reading.
- Make sure the forehead is clean and free from moisture.
How to interpret your reading
Child (from Newborn)
Memory function
The memory function is great if you want to look back on your last 20 readings, maybe to show a healthcare professional.
When the product is switched on, press the memory button. The screen will display “01” followed by the recorded reading. Press the memory button again to retrieve the second reading, and so on for the past 20 readings.
Measuring object temperature
- A. To switch on press the power button
- B. Use the mode button to select Object mode
- C. Point the infrared sensor to the centre of the object 1-3 cm away, then press and release the Measure button. The object temperature will be displayed on the screen.
Troubleshooting
- In Forehead mode, a temperature reading of more than 42.2ºC (108.0ºF) and in Object mode, a temperature reading of more than 100ºC (212.0ºF), the display will show “HI”.
- In Forehead mode, a temperature reading of less than 35.0°C (95.0°F) and In Object mode, a temperature reading of less than 0ºC (32.0ºF), the display will show “LO”.
- Switching between ºC and ºF – Pressing and holding the measurement/power button for about 3 seconds allows you to switch units between ºC and ºF.
- How to mute – Press the sound switch button to switch the sound on or off. A long beep will be sounded when the sound is turned on, and a mute symbol will be displayed when the sound is turned off.
Replacing the batteries
- Slide the battery cover along the marked direction, the battery cover will flip open.
- Insert the two alkaline AAA batteries into the compartment.
- Make sure that the batteries are installed correctly.
- If the low-battery symbol is displayed on the screen, replace the batteries.
- Batteries of the same type should be used. Dispose the used batteries in accordance with the local environmental policies.
- The thermometer is supplied with batteries. Not suitable for rechargeable batteries.
FIND OUT HOW TO USE
- SCAN QR CODE ON THE BACK OF THE PRODUCT.
Product Specifications
- Product Name: Infrared Thermometer
- Product Model: JPD-FR203
- Power Supply Mode: Internal power supply
- Operating Voltage: DC 3V
- Battery Model: AAA x 2
- Operating Mode: Continuous operating
- Display: Segment LCD
- Measure time: About 1 second
- Latency Time: About 3second
- Measuring Distance: 1 to 3 cm
- Measuring Range: Forehead: 35.0°–42.2° (95.0°F–108.0°F)
- Object: 0.0°C–100.0°C (32.0°F–212.0°F)
- Accuracy (Laboratory): ±0.4°F/±0.2°C from 35.0°C–42.2°C (95.0°F–108.0°F)
- ±1.8°F/±1.0°C, outside the range of 35.0°C–42.2°C (95.0°F–108.0°F)
- Accuracy: (Clinical): ±0.3° C (±0.6° F)
- Resolution: 0.1°C (0.1°F)
- Measuring site: Forehead (keep distance 1~3 cm from forehead)
- Reference body site: Armpit
- Mode of operation: Adjusted mode
- Memory: 20 temperature readings
- Low-battery Alert: The low-battery symbol is displayed if the power voltage is lower than 2.5 V±0.1V.
- Automatic Power-off: The thermometer automatically powers off if it is not used in 10±1 seconds.
- Dimensions (mm): 151.8×36.2×35.3
- Weight (g): 79 g (with batteries)
- Manufacturing date: see the label
- Service life: 2 years
- Operating Environment:
- Temperature: 10°C–40°C (50°F–104°F)
- Humidity: 15%–95% RH, non-condensing
- Atmospheric pressure: 86–106 kPa
- Storage and Transportation:
- Temperature: -20°C to 50°C (-4°F–131°F)
- Humidity: 0- 95% RH, non-condensing
- Atmospheric pressure: 50–106 kPa
The infrared thermometer has been tested and conforms to the standard ASTM E1965-98. ASTM laboratory accuracy requirements in the display range of 98°F to102°F (37°C-39°C) for skin IR thermometers is ±0.5°F (±0.3°C). Note that for mercury glass and electronic thermometers, the requirement per ASTM Standards E667-86 and E1112-86 is ±0.2°F (±0.1°C).
Security Class
- Type of protection against electric shock: internally powered equipment.
- Degree of protection against electric shock:
- Type BF applied part.
- Degree of protection against ingress of water: IP22
- Safety degree of using in flammable anesthetic gas blending with air, oxygen or nitrous oxide: Non-AP/APG
- No application parts of the thermometer prevents defibrillation charge effect.
- No application parts of the thermometer output signal.
- The thermometer is impermanent installed device.
Storage and Transportation
The thermometer can be transported using general transportation tools. Severe vibration, shock, or rain must be avoided during transportation. The thermometer must be packaged and then stored in a well-ventilated room without corrosive gas. The ambient temperature must be between –20°C and +50°C (-4°F–131°F), the relative humidity must be lower than 95% (non-condensing), and the atmospheric pressure must be 50–106 kPa.
EMC Information-Guidance and Manufacture’s Declaration
CAUTION
- The Infrared Thermometer JPD-FR203 needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided for in the ACCOMPANYING DOCUMENTS.
- Portable and mobile RF communications equipment can affect Infrared Thermometer JPD-FR203.
- The Infrared Thermometer JPD-FR203 should not be used adjacent to or stacked with other equipment.
Guidance and manufacturer´s declaration – Electromagnetic emission for all equipment and systems
The Infrared Thermometer JPD-FR203 is intended for use in the electromagnetic environment specified below. The customer or the user of the Infrared Thermometer JPD-FR203 should assure that it is used in such an environment.
Emissions test | Compliance | Electromagnetic environment guidance |
---|
RF
emissions CISPR 11
|
Group 1
| The Infrared Thermometer JPD-FR203 uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference
in nearby electronic equipment.
RF
emissions CISPR 11
|
Class B
| The Infrared Thermometer JPD-FR203 is suitable for use in all establishments, including domestic establishments and those directly connected to the
public low-voltage power supply network that supplies buildings used for domestic purposes.
Guidance and manufacturer’s declaration – Electromagnetic immunity for all equipment and systems
The Infrared Thermometer JPD-FR203 is intended for use in the electromagnetic environment specified below. The customer or the user of the Infrared Thermometer JPD-FR203 should assure that it is used in such an environment.
Immunity test| IEC 60601
test level
| Compliance level| Electromagnetic environment guidance
---|---|---|---
Electrostatic discharge (ESD) IEC 61000-4-2
|
±6 kV
contact
±8 kV air
|
±6 kV
contact
±8 kV air
| Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material,
the relative humidity should be at least 30 %.
Power frequency (50/60 Hz)
magnetic field IEC 61000-4-8
|
3 A/m
|
3 A/m
| Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment.
Guidance and manufacturer´s declaration – Electromagnetic immunity for equipment and systems that are not life-supporting
The Infrared Thermometer JPD-FR203 is intended for use in the electromagnetic environment specified below. The customer or the user of the Infrared Thermometer JPD-FR203 should assure that it is used in such an environment.
- Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcasts and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the JPD-FR203 is used exceeds the applicable RF compliance level above, the JPD-FR203 should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the JPD-FR203.
- Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Recommended separation distances between portable and mobile RF communications equipment and the EQUIPMENT or SYSTEM -for EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING
The Infrared Thermometer JPD-FR203 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Infrared Thermometer JPD-FR203 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the Infrared Thermometer JPD-FR203 as recommended below, according to the maximum output power of the communications equipment.
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
- NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
- NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
-
CAUTION
-
BF type applied part.
-
Classification for water ingress and particulate matter.
-
Indicates this device is subject to the Waste Electrical and Electronic Equipment Directive in the European Union. To protect the environment dispose of the useless device at appropriate collection sites according to national or local regulations.
-
Please read instructions.
Manufactured by
- Shenzhen Jumper Medical Equipment Co., Ltd.
- Address: D Building, No. 71, Xintian Road, Fuyong Street, Baoan, Shenzhen, Guangdong, China
- Distributed by: Mayborn (UK) Limited, Balliol Business Park, Newcastle upon Tyne, NE12 8EW.
- Mayborn France Sarl, 92100, France.
- Mayborn USA Inc. CT 06901, USA.
MedPath GmbH
- Mies-van-der-Rohe-Strasse 8, 80807 Munich, Germany
Information on the Disposal of Waste Electrical and Electronic Equipment (WEEE)
This symbol means that your product shall be disposed of separately from household waste in accordance with local laws and regulations. When this product is ready for disposal, take it to a collection point designated by local authorities. Separating and recycling of your product at the time of disposal will help conserve natural resources and ensure that the product is recycled in a manner that protects human health and the environment. Penalties may be applicable for the incorrect disposal of this product in accordance with national legislation.
Mayborn (UK) Limited
- Balliol Business Park, Newcastle Upon Tyne, NE12 8EW, UK.
Mayborn USA Inc.
- 1010 Washington Boulevard, 11th Floor, Stamford, CT 06901, USA
EU Representative
- Mayborn France Sarl, 56 Rue de paris, Boulogne Billancourt, 92100, France.
MADE IN CHINA.
TOMMEE TIPPEE®. NOTOUCH™. Copyright© 2022.
All IP rights held by Mayborn (UK) Limited and/or its affiliates.
Any questions?
Visit our website: tommeetippee.com
0499056