SDI 8252 Hand Held Digital Monometer Instruction Manual

June 9, 2024
SDI

Duct Detector Test Tool
Instruction Manual
Hand-Held Digital
Monometer
Model: 8252 

SDI 8252 Hand Held Digital Monometer -

Congratulations on your purchase of the ductchecker™ Duct Detector Test Tool. This digital manometer is a portable, battery operated pressure measuring device.
The ductchecker™ is ideal for Fire Testing Technicians measuring pressure levels i n HVAC/R ducts containing smoke detectors.

INTRODUCTION

The meter will display all LCD segments when it is first turned on for approx. 3 seconds.
Note: Though you might see DATALOGGER, Y/M/D, REL, AVG, these are not available on this meter.
The LCD is divided in to two distinct sections: One large (primary) top screen and one smaller right – bottom screen (relative Clock). The two display areas keep you constantly updated with the pressure measurements.
The meter measures:
Gauge pressure – a measurement of pressure that is referenced to ambient pressure.
Differential pressure – a measurement of difference of two pressures.
The meter has five selectable units of measure: InH2O, psi, mbar, bar, mmHz2O.
Please check tubing is not leaking or damaged before using.

CONVERSION & RESOLUTION

1 mbar= Resolution
Inch of H2O 0.401
psi 0.0145
mbar 1
mm of H2O 10.

MANOMETER QUICK START

Unscrew the battery compartment on the rear of instrument and fit PP3 (or equivalent) battery. Replace the cover and secure with the screw.
Press to switch the instrument on.
Press to select the unit of pressure measurement required. Zero by pressing and hold for 3 seconds. The instrument now reads gauge pressure.
Press for differential pressure measurement.
Press  to freeze reading. Press again to cancel feature.
Press to start clock. Press  again to see time (since start of recording). For MAX reading press again to see time (since start of recording). For MIN reading pressing again returns to real time recording mode.
Press and hold for 3 seconds to turn clock feature off.
Note:
Clock feature available with gauge pressure only not differential. The instrument will automatically switch off after 20 minutes unless sleep mode is disabled, see page 5  UTO POWER OFF.
Press GD for display backlight -illuminates for approximately 30 seconds and automatically switches off.

CONTROL & INDICATORS

  1. Primary Data Screen displays pressure value.
  2. “-“ Minus pressure display.
  3. MAX MIN Pressure recorded.
  4. REC Starts recording mode and displays max/min pressure recorded.
  5. AVG Average records (N/A).
  6. DC Power i n Jack.
  7. RS232 Output port
  8. H/M/S 88:88:88 Displays time for Hour/Minute/Second.
  9. Pressure unit indication
  10. BAT Battery low indicator.
  11. DIF Differential pressure mode
  12. “+” Positive pressure connection
  13. “-“ Negative pressure connection
  14. HOLD Freezes pressure reading.
  15. REL(N/A) Establish a relative zero for the primary screen information.

SDI 8252 Hand Held Digital Monometer - Fig2

AUTO POWER OFF (SLEEP FUNCTION)

This instrument will shut off automatically after approx. 20 minutes for every power on. For recording o r operating over longer periods of time, you can disable the sleep mode by pressing and simultaneously before power on.
An “n” will appear in the middle of the screen at which time you can release the button (see Fig. A). The disable sleep mode will be invalid after power on.

SDI 8252 Hand Held Digital Monometer - FigA

MODE OPTIONS

Delete and replace with programmable user selectable start-up mode. The display will default to the mode last used.
The following table lists the modes of operation that can be invoked by pressing the button indicated.

Turns instrument on (default setting) and off.
Press momentarily and relative clock starts in the lower right screen.
REC is displayed i n the middle left of screen (Fig. B). Other button functions are locked out except Power, Unit and Backlight.
Press momentarily again and the unit cycles through MAX ( F i g . C) and MIN (Fig. D) and back to current pressure; the record mode is displayed on the LCD. Press and hold ® for 3 seconds to turn off the record function and return to normal mode.

SDI 8252 Hand Held Digital Monometer - FigC

Press momentarily to freeze the pressure recording ( F i g . E).

Press momentarily, DIF appears on top of the LCD and the display indicates the relative zero (relative zero causes the value of the display to show as “0.0”) – only the amount of pressure change will be indicated. Press momentarily again and the unit returns to the normal mode of pressure differential (see Fig. F).

Differential Pressure: a measurement of the difference between two pressures, i.e. use differential pressure sensor to measure gauge pressure by leaving one process connection open to the atmosphere and connecting the second sensor port to your system.
Press momentarily and the unit will cycle through “InH2O0”, “psi”, “mbar”, “mmHz2O” which are indicated on the bottom of the display (see Fig. G & H).

SDI 8252 Hand Held Digital Monometer - FigG

Press momentarily and the backlight illuminates for approx. 30 seconds then turns off automatically. Or press momentarily to decrease the figure when calibration is being performed.

MAINTENANCE

The meter is calibrated in-house before shipping.
When properly maintained, the meter will maintain its accuracy to specification. To ensure your meter is performing at its peak, send it to the factory or a qualified
instrument calibration facility for annual calibration.
Recommended – always set to zero before measurement. Refer to the zero setting procedure on page 11.
Cleaning:
Use a damp cloth and mild soap to clean the case of the Monometer. Do not use harsh detergents o r abrasives as these may mar the finish o r damage the unit’s case with an adverse chemical reaction.

CALIBRATION MODE

Calibration mode is only applicable for a standard manometer calibrator or any qualified calibration facility for annual calibration.

  1. Manually set the display to zero (no pressure applied to the connector), refer to the manual zero procedure (see page 11).

  2. Turn the meter off.

  3. Press & simultaneously. “CA” appears on the display. (see F i g . 1 ) and the meter enters the calibration mode. Make sure the unit is set on “PSI” to start positive (+) pressure calibration.

  4. The meter has defaulted to 1.6 psi calibration point, the adjustable pressure range is from 1.5 to 1.7 psi. If calibration pressure source is not 1.6 psi, increase by pressing key, or decrease by pressing key to set calibration point as required.

  5. Save the calibration point by pressing  key. “SA” and small “CA” appears on the display (see Fig. J). After 2 seconds, the meter auto-skips to the negative pressure
    (-) point for ther next calibration mode.

  6. Follow the same procedure as in step 4 for the negative pressure calibrator point. The LCD now displays “- 1.600” and a small “CA” (see Fig. K). Do the necessary calibration figure referring to your pressure standard if needed.

  7. Again save the calibration point by pressing key, “SA” and “CA” appears in 2 seconds and then pressing “End” and “CA” appears i n another 2 seconds. The meter turns back to the normal mode (see Fig. L).

If you can’t save by pressing the key, i.e. no “SA” appeared, please check: (a) the calibration pressure source is between 1.5 and 1.7, or (b) if you entered the correct positive pressure (+) or negative pressure (-).
If you want to skip positive (+) calibration when i n calibration mode, press to skip to negative (-) calibration point.
Calibration point reference

psi Calibration Recommended
range point(+) (+)
0~+2 1.6 1.5~1.7

MANUAL ZERO SETTING

When you set the display to zero (no pressure applied to the connector), press for 2 seconds. Now the meter displays “-0.000” from right to left (see Fig. M), then
the LCD display shows a normal mode.

TROUBLESHOOTING

? Power on but no display. Check the battery connections. Replace with new battery or attach optional AC adaptor.
? BAT indication. Replace with a new battery when LCD displays BAT at the middle bottom.
? No display. Make sure battery is not empty. I f the display disappears, check sleep mode is active. Refer to the disable sleep mode function for a long time on page 4. Or check the tubing is connected to the meter tightly.
? Err.1. For the pressure value exceeding the maximum range, “Err.1” appears on the display (see Fig. O). Do not exceed above the rated pressure range of Manometer. The sensor will be damaged.

? Err.2. For the measurement of pressure less than minimum range, “Err.2″ will appear (see F i g . P).

? Err.3. For a differential pressure value larger than maximum display, “Err.3″ appears on the display (see Fig. Q).

? Err.4. When you set zero, make sure you have disconnected the tubing. I f “Err.4″ appears on the display, it means the Manometer is damaged (see Fig. R).
Note: “Err.4” will also appear if the tubing is connected during Zero setting.

? E1OL or E2OL. When you see these errors while operating the RS232 software, it means the pressure source is less than or over the range of the instrument.

REPLACING THE BATTERY

Replace your 9-volt battery when:

The BAT icon appears o n the right of the screen.
The meter will not power on.
Use of the backlight causes the BAT icon to appear.
Even if the battery was recently replaced, check its voltage level if you get no response from the instrument.
To replace the battery:

  1. Remove the tubing from the instrument.
  2. Lay the instrument face-down on a clean, flat surface.
  3. Remove the battery cover.

If you do not intend to use the instrument for a month or more remove the battery. Do not leave battery i n the instrument.

OPERATING CONDITIONS

Compensated temperature range: 0~50°C (32~122°F)
Operating temperature range: 0~50°C (32~122°F)
Storage temperature range: -20-60°C (-4~140°F)
Operating Humidity M a x . 80% RH
Power: One 9.0 volt battery
Exceeding maximum pressure will Cause permanent sensor damage.

MATERIAL SUPPLIED

This package contains:

The Manometer
Battery (9.0 volt)
Operation manual
Connection tube 6mm (ID) x 9mm (OD) x 500mm length x 2 pcs.

OPTIONAL ACCESSORIES

RS232 software CD with D-sub connector
Rubber boot

SPECIFICATION

| Pressure
---|---
Range| 0- ± 138
= 0- ± 55.4
= 0- ± 1410
= 0- ± 0.138
= 0- ± 2| mbar
inH2O
mm inH2O
bar
psi
Resolution| See Page 2 data sheet
Accuracy| ±0.3%of full scale at ±25°C
Dimensions| 72 x 182 x 30mm (meter)
Unit Weight| Approz. 220 gram (with
battery)
Response time| 0.5 seconds
Format| Baud Rate: 2400 bit/sec
Data Bit: 8, Stp Bit: 1
P XXXXX, P – XXXXX (unit)

RS232 DATA OUTPUT

This Manometer can link with a personal computer to capture online data, display pressure records with real-time output. You can retrieve files, save the data for operating data analysis, record statistics, multi- file display on the screen etc. – versatile functions of your choice.
Connection Procedures:

  1. Plug the optional accessory RS232 cable into the DC jack port (at the right side of the meter).
  2. Insert the D-sub 9P type connector into the computer’s Com. 1 or 2 port.
  3. Start the set up of RS232 software by inserting the CD-ROM
  4. When installing the RS232 software, please follow the operation manual procedure in the software package.

SDI 8252 Hand Held Digital Monometer - Fig5

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals