POOL TECHNOLOGIE PAPI004204 ORP Controller Instruction Manual
- June 9, 2024
- Pool Technologie
Table of Contents
PAPI004204 ORP Controller
Instruction Manual
OPERATIONAL PRINCIPLE
Condition | Immediate automatic action |
---|---|
ORP measurement lower than the ORP setpoint | Switching on the electrolyser |
ORP measurement higher than the ORP setpoint | Switching off the electrolyser |
Forced operation activated | Switching on the electrolyze, for a fixed period |
and without interruption
INSTALLATION DIAGRAM
ELECTRICAL SHOCK HAZARD
Before switching on the power, make sure that all of the electrical wiring is
undertaken.
-
Electronics unit
-
Pool Terre (optional)
-
Probe holder
-
ORP probe
-
Bracket
ELEMENTS NOT SUPPLIED : -
Electrical power supply
-
Copper cable
-
Chlorinator cell
-
Electrolyze
-
Filter
-
Ground rod
-
Heat pump
-
Filtration pump
WIRING THE ELECTRONICS UNIT TO THE ELECTROLYSER
3.1. Version A (depending on model)
3.2. Version B (depending on model)
DETAILS ABOUT THE ORP CHECK
The amount of chlorine required can vary depending on several conditions :
– Covered pool (by sheeting, cover or panels)
Low chlorine requirement (because there is no UV).
– Sudden rise in the number of people using the pool
Very large amounts of chlorine needed, but on a temporary basis.
– Indoor pool or sheltered pool
Reduced need for chlorine (because of low exposure to external pollution), but
which tends to increase depending on the frequency of use of the swimming
pool.
Given this range of possible configurations, chlorine production must be
managed according to requirements. The ORP check enables you to react to each
of these situations.
The ORP measurement (in mV), reflecting the oxidation (or reduction) potential
of the water, is a major indicator of According to the WHO, an ORP measurement
of 650 mV guarantees disinfected water that is itself capable of
disinfecting. Despite the use of this value as a reference, this can only be
on a theoretical level, because ORP measurements can easily vary depending on
the following parameters :
– The ph.
– The type of chlorine (stabilised or non-stabilised).
– The presence of dissolved elements that can affect the water (metals,
phosphates, surfactants).
– The cleanliness of the filter.
– The presence of stray currents.
– The presence of flocculant (deposit on the ORP probe).
The ORP measurement :
– is not a measurement of free chlorine levels.
The ORP measurement :
– varies according to free chlorine levels and all elements in the water.
ESSENTIAL PREREQUISITES FOR AN OPTIMAL ORP CHECK
– Stable pH (with a pH regulator).
– Stabilizer level between 20 and 30 ppm.
– Earthing of the pipe where the ORP probe is installed (with a Pool Terre
kit).
– Installation of the ORP probe at least 30 cm before the chlorinator cell.
– Balanced water profile (free chlorine levels at 1 ppm, and pH at 7.2).
– ORP setpoint adjusted according to the ORP measurement displayed (a value
between 500 and 700 mV can be considered as correct).
The use of sulphates is permitted, provided they remain at levels below 360
ppm.
The use of copper sulphates is strictly forbidden.
The use of borehole water is strictly prohibited.
When using a chemical (flocculant, waterline cleaning, sequestrant), check the
ORP measurement before and after use of this product. If the ORP measurement
drops sharply, stop the electronics unit for a few days, until the effects of
the product on the ORP measurement disappear.
Influence of chloramines on the ORP measurement : as chloramine levels tend to
increase, the ORP measurement tends to decrease.
The ORP check in no case eliminates the need to regularly check free chlorine
levels.
ELECTRONICS UNIT
5.1. Interface and indicators
Visual of the non-contractual interface:
-
CONTROL KEYS.
-
RED LED.
If lit continuously : electronics unit powered off.
If flashing : alarm activated. -
SCREEN.
If display flashing : information awaiting confirmation, or alarm activated.DESCRIPTION OF THE DEFAULT DISPLAYSetting| Displayed value| Meaning
ORP measurement| From CO to 99| From 0 to 99 mV
From is to 9a| From 100 to 990 mV -
GREEN LED.
If lit continuously : electronics unit in operation.
If flashing : electrolytic cell under voltage.
5.2. Basic operations
Switching on and off : long press on the left key.
Selecting a value or data : top and bottom keys.
Confirmation of an entry / Entering a menu : right-hand key.
Cancelling an entry / Returning to the previous menu : left-hand key.
5.3. Meaning of the menus
MENU | FEATURE |
---|---|
CR | Calibrating the ORP probe |
Rd | ORP measurement adjustment |
SE | Setting the ORP setpoint |
nR | Forced operation |
5.4. Features
5.4.1. Calibrating the ORP probe
To ensure an optimal ORP check, it is imperative that the ORP probe is
calibrated :
– when first commissioning the equipment.
– at the start of each season when it is commissioned.
– each time the probe is replaced.
-
Turn off the filtration (and therefore the electronics unit).
-
If the probe is already installed :
a) Remove the probe from the probe holder, without disconnecting it.
b) Remove the probe holder nut and replace it with the stopper supplied.
If the probe is not already installed :
Connect the probe to the electronics unit. -
Insert the probe into the ORP 470 mV calibration solution.
-
Wait a few moments, without touching the probe.
-
Turn on the electronics unit.
-
Press and hold down the right-hand key, until the menu flashes.
-
Press the top or bottom key x times until menu flashes.
-
Confirm by pressing the right-hand key : the message flashes.
-
Press on the right-hand key.
-
Wait until one of the messages below is displayed.
-
If the message is displayed : calibration was successful.
a) Press 3 times on the left-hand key : the default display reappears.
b) Rinse the probe under running water.
c) Drain the probe without wiping it.
d) Install the probe into the probe connector.
If the message is displayed : the calibration failed.
a) Press 3 times on the left-hand key : the default display reappears.
b) Visually check the condition of the probe.
c) Try the calibration again, several times if necessary. If the calibration still fails, change the probe and recalibrate.
5.4.2. ORP measurement adjustment
The adjustment of the ORP measurement requires a measuring device (not
supplied) to obtain an actual ORP value.
- Prerequisite : the ORP measurement must be between 200 and 900 mV.
- Possible settings : from – to + 100 mV compared to the ORP measurement, in steps of 10 mV.
- Default setting : ORP measurement.
- Press and hold down the right-hand key, until the menu flashes.
- Press the top or bottom key x times until menu flashes.
- Confirm by pressing the right-hand key.
- Select a value with the top/bottom keys.
- Confirm by pressing the right-hand key.
- If the message is displayed : the adjustment was successful.
Press 2 times on the left-hand key : the default display reappears.
The message is displayed : the calibration has failed.
a) Press 2 times on the left-hand key : the default display reappears.
b) Visually check the condition of the ORP probe.
c) Try the adjustment again, several times if necessary. If the adjustment still fails, replace the ORP probe and carry out a calibration of the ORP probe.
5.4.3. Setting the ORP setpoint
Possible settings : from to (from 200 to 900 mV, in steps of 10 mV).
Default setting :
- Press and hold down the right-hand key, until the menu flashes.
- Press the top or bottom key x times until menu flashes.
- Confirm by pressing the right-hand key.
- Select a setpoint with the top/bottom keys.
- Confirm by pressing the right-hand key : the selected setpoint freezes briefly.
- Press on the left-hand key : the default display reappears.
5.4.4. Forced operation
- Possible settings : from to (i.e. from 1 to 24 h, in steps of 1 h).
- Default setting :
To activate forced operation :
- Press and hold down the right-hand key, until the menu flashes.
- Confirm by pressing the right-hand key.
- Select a duration with the top/bottom keys.
- Confirm by pressing the right-hand key : the timer countdown is displayed in real time.
To stop forced operation :
Press on the left-hand key : the default display reappears.
5.5. Alarm
Flashing display| Fault detected| Automatic action immediate| Checks and
solutions| Dismissal
---|---|---|---|---
RS| Deviation of + or – 400 mV between the ORP measurement and the ORP
setpoint for 48 hours| Switching off the electrolyser| •Check the ORP
measurement in the swimming pool with a recent analysis kit.
•Check that the [electrolier + cell] assembly is functioning correctly.
•Check all the settings :
– Calibrating the ORP probe.
– ORP measurement adjustment.
– Setting the ORP setpoint.
– Forced operation not activated.| Press on the right-hand key
GUARANTEE
Before contacting your dealer, please have the following to hand :
– your purchase invoice.
– the serial no. of the electronics unit.
– the installation date of the equipment.
– the parameters of your pool (salinity, pH, chlorine levels, water
temperature, stabilizer level, pool volume, daily filtration time, etc.).
Every effort and all our technical experience has gone into designing this
equipment. It has been subjected to quality controls. If, despite all the
attention and expertise involved in its manufacture, you need to make use of
our guarantee, it only applies to free replacement of the equipment’s
defective parts (excluding shipping costs in both directions).
Guarantee period (proven by date of invoice)
Electronics unit : 2 years.
ORP probe : depending on model.
Repairs and spare parts : 3 months.
The periods indicated above correspond to standard guarantees. However, these
can vary depending on the country of installation and the distribution
network.
Scope of the guarantee
The guarantee covers all parts, with the exception of wearing parts that must
be replaced regularly.
The equipment is guaranteed against all manufacturing defects within the
strict limitations of normal use.
After-sales services
All repairs will be carried out in the workshop.
Shipping costs in both directions are at the user’s own expense.
Any downtime and loss of use of a device in the event of repairs shall not
give rise to any claim for compensation.
In all cases, the equipment is always sent at the user’s own risk. Before
taking delivery, the user must ensure that it is in perfect condition and, if
necessary, write down any reservations on the shipping note of the carrier.
Confirm with the carrier within 72 hours by recorded letter with
acknowledgement of receipt.
Replacement under guarantee shall in no case extend the original guarantee
period.
Guarantee application limit
In order to improve the quality of their products, the manufacturer reserves
the right to modify the characteristics of the products at any time without
notice.
This documentation is provided for information purposes only and is not
contractually binding with respect to third parties.
The manufacturer’s guarantee, which covers manufacturing defects, should not
be confused with the operations described in this documentation.
Installation, maintenance and, more generally, any servicing of the
manufacturer’s products should only be performed by professionals.
This work must also be carried out in accordance with the current standards in
the country of installation at the time of installation. The use of any parts
other than original parts voids the guarantee ipso facto for the entire
equipment.
The following are excluded from the guarantee :
– Equipment and labor provided by third parties in installing the device.
– Damage caused by installation not in compliance with the instructions.
– Problems caused by modifications, accidents, misuse, negligence of
professionals or end users, unauthorised repairs, fire, floods, lightning,
freezing, armed conflict or any other force-majeure events.
Any equipment damaged due to non-compliance with the instructions regarding
safety, installation, use and maintenance contained in this documentation will
not be covered by the guarantee.
Every year, we make improvements to our products and software. These new
versions are compatible with previous models. The new versions of hardware and
software cannot be added to earlier models under the guarantee.
Implementation of the guarantee
For more information regarding this guarantee, contact your dealer or our
After-Sales Service. All requests must be accompanied by a copy of the
purchase invoice.
Legislation and disputes
This guarantee is subject to French law and all European directives or
international treaties in force at the time of the claim, applicable in
France. In case of disputes concerning its interpretation or execution, the
High Court of Montpellier (France) shall have exclusive jurisdiction.
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>