BRESSER 5014000 Biolux DLX Microscope Instruction Manual

June 9, 2024
BRESSER

BRESSER 5014000 Biolux DLX Microscope-logo

BRESSER 5014000 Biolux DLX Microscope

BRESSER 5014000 Biolux DLX
Microscope-product

General Information

SERVICE AND WARRANTY:

www.bresser.de/warranty_terms

MICROSCOPE GUIDE:

www.bresser.de/guide

MICROSCOPE FAQ:

www.bresser.de/faq MANUAL DOWNLOAD:

www.bresser.de/P5014000 EXPERIMENTS:

www.bresser.de/downloads

About this Instruction Manual

Please read the safety instructions in this manual carefully. To avoid damage to the unit and the risk of injury, only use this product as described in the manual. Keep the instruction manual handy so that you can easily look up information on all the functions.

DANGER!
You will find this symbol in front of every section of text which deals with the risk of severe injury or even death in the event of improper use.

CAREFUL!
You will find this symbol before every section of text that deals with the risk of minor to severe injuries resulting from improper use.

NOTE!
You will find this symbol in front of every section of text which deals with the risk of damage to property or the environment.

Intended Use
This product is intended only for private use. It was developed for the magnified display of things in nature.

General Warning

DANGER!
Tools with sharp edges and points are often used when working with this device. For this reason, store this device and all accessories and tools in a location that is out of the reach of children. There is a RISK OF INJURY!

DANGER!
This device contains electronic components which operate via a power source (batteries). Do not leave children unattended when using the device. Only use the device as described in the manual, otherwise, you run the RISK of an ELECTRIC SHOCK.

DANGER!
Do not expose the device to high temperatures. Use only the recommended batteries. Do not short-circuit the device or batteries, or throw them into a fire. Excessive heat or improper handling could trigger a short-circuit, a fire or an explosion.
Keep batteries out of the reach of children! Make sure you insert the batteries correctly. Empty or damaged batteries could cause burns if they come into contact with the skin. If necessary, wear adequate gloves for protection.

DANGER!
Use only the recommended batteries. Always re-place weak or empty batteries with a new, complete set of batteries at full capacity. Do not use batteries from different brands or with different capacities. The batteries should be removed from the unit if it has not been used for a long time. The manufacturer is not liable for damage related to improperly installed batteries.

DANGER!
Children should only use the device under adult supervision. Keep packaging material (plastic bags, rubber bands, etc.) out of the reach of children! There is a RISK OF CHOKING!

CAREFUL!
Children must not have access to the included chemicals and liquids. Do not drink the chemicals. Wash hands thoroughly with running water after use. In the event of contact with the eyes or mouth, rinse thoroughly with water. In the event of pain, contact a doctor immediately and take the substances with you.

NOTE!
Do not disassemble the device. In the event of a defect, please contact your dealer. He will contact the Service Centre and can send the device in to be repaired, if necessary.

Operating instructions

Parts overview (Fig. 1+2):

  1. 5x WF Eypiece
  2. 16x WF Eyepiece
  3. Barlow lens
  4. Eyepiece socket
  5. Microscope head
  6. Objective-revolver
  7. Microscope stage
  8. Focus wheel
  9. LED lighting (transmitted light)
  10. On/off switch (illumination)
  11. Microscope base
  12. Battery compartment (3x AA)
  13. 5 Slides and 5 preparations
  14. Dimmer
  15. Condenser
  16. Adjusting lever for iris diaphragm
  17. Microscope tools
  18. Petri dish
  19. Empty bottles
  20. Cover platesBRESSER 5014000 Biolux DLX Microscope-fig-1BRESSER 5014000 Biolux DLX Microscope-fig-2
General/Location

First you must make sure that your microscope is on a stable and solid surface.

Electric LED lighting with dimmer

In the base of the microscope, there is a battery compartment (12). Loosen the screws at the battery compartment cover with a small screwdriver and remove the cover. Place the batteries in the compartment so that the flat minus poles (-) press against the spring terminal and the plus poles (+) are touching the flat contact sheets. Close the battery compartment with the cover and turn the microscope around again. As your device has continuously controllable lighting (dimmer) optimal illumination of the object to be viewed is guaranteed.

Microscope setup

The microscope’s eyepiece will now be prepared for the first observation. First, rotate the eyepiece into a convenient position. Begin every observation with the lowest magnification. Place the microscope stage (Fig. 1, 7) with the focus wheel (Fig 1, 8) into the lowest position and rotate the objective revolver (Fig. 1, 6) until it locks on the lowest magnification (4x).

NOTE:
First, drive the Microscope’s table (Fig. 1, 7) in its lowest position before changing the objective in order to prevent damage.

Insert the 5x eyepiece (Fig. 3, 1) in the Barlow lens (Fig. 3, 3). Take care, that the Barlow lens is inserted completely in the eyepiece socket (Fig. 3, 4).

BRESSER 5014000 Biolux DLX Microscope-fig-4

Condenser

The condenser is used to adjust the light bundling to the observed object in an optimal way. The iris diaphragm is used to increase the depth of focus. With an open aperture more light is getting through but the depth of focus is less. With a partially closed aperture the depth of focus is higher but it is necessary to increase the brightness of the illumination then. It is recommended not to close the iris diaphragm too much when working with higher magnifications. Otherwise, the image resolution will decrease. Move the lever for the iris diaphragm to right or left to change the aperture setting.

Observation

After you have set up the microscope with the corresponding illumination, the following principles are important:

  • Begin each observation with a simple observation at the lowest magnification, so that the center and position of the object to be viewed is in focus.
  • The higher the magnification the more light is required for good picture quality.
  • Place a permanent slide culture (Fig. 4) directly under the microscope lens on the plate (Fig. 4, 7) and clamp it tight. The specimen to be examined must be directly over the lighting.
  • Look through the eyepiece (Fig. 1, 1/2) and turn carefully the focusing wheel (Fig. 1, 8) until you can see a sharp picture.
  • Now you can get a higher magnification, while you pull out slowly the Barlow lens (Fig. 5, 3) of the eyepiece socket (Fig. 5, 4). With a nearly entirely pulled-out Barlow lens the magnification is raised to 2x.
  • For still higher magnification you can put the 16x eyepiece (Fig. 5, 2) into the objective revolver (Fig. 1, 6) and set it on a higher position (10x / 40x).BRESSER 5014000 Biolux DLX Microscope-fig-5

TIP:
Depending on the preparation higher magnifications do not always lead to better pictures.

With changing magnification (eyepiece or objective lens changes, pulling out of the Barlow lens) the sharpness of the image must be newly defined by turning the focusing wheel (Fig. 1, 8).

NOTE:
Please be very careful when doing this. When you move the mechanical plate upwards to fast the objective lens and the slide can touch and become damaged.

Viewed Object condition and preparation

Condition

With the Barlow lens nearly fully extended magnification can be doubled. Both transparent and non-transparent specimens can be examined with this microscope, which is a transmitted light model. If translucent (transparent) specimens are examined the light from below goes through the specimen, lens, and eyepiece to the eye and is magnified en route. Many small organisms of the water, plant parts and finest animal components have now from nature this transparent characteristic, other ones must be accordingly prepared. Is it that we make it by means of a pre-treatment or penetration with suitable materials (media) transparent or thus that we cut the finest wafers off of them (hand cut, MicroCut) and these then examine? With these methods will us the following part make familiar.

Creation of thin preparation cuts

Specimens should be sliced as thin as possible, as stated before. A little wax or paraffin is need-ed to achieve the best results. A candle can be used for the purpose. The wax is put in a bowl and heated over a flame. The specimen is then dipped several times in the liquid wax. The wax is finally allowed to harden. Use a MicroCut or knife/scalpel (carefully) to make very thin slic-es of the object in its wax casing.

DANGER!
Be extremely careful when using the knives/scalpels or the MicroCut. There is an increased risk of injury due to the sharp edges!

These slices are then laid on a glass slide and covered with another.

Creation of an own preparation

Put the object which shall be observed on a glass slide and give with a pipette (Fig. 6, 17a) a drop of distilled water on the object (Fig. 6). Set a cover glass (in each well sorted hobby shop available) perpendicularly at the edge of the water drop, so that the water runs along the cover glass edge (Fig. 7). Lower now the cover glass slowly over the water drop.

Experiments

Use the following web link to find interesting experiments you can try out http://www.bresser.de/downloads

Care and maintenance

Separate the device from the power supply before cleaning (remove batteries)! Only use a dry cloth to clean the exterior of the device.

NOTE:
Do not use any cleaning fluid to avoid damaging the electronics.

Clean the lenses (eyepiece and/or lens) with a soft and lint-free cloth only (e.g. microfibre).

NOTE:
Do not apply excess pressure to the cloth so as to avoid scratching the lenses.

To remove more stubborn dirt, moisten the cleaning cloth with an eyeglass- cleaning solution and wipe the lenses gently.

Protect the device from dust and moisture! After use – in particular in situations of high humidity – let the device acclimatize for a short period of time so that the residual moisture can dissipate.

Troubleshooting

Error
No picture visible

Solution

  • switch light on
  • put condenser lens in place
  • adjust focus

Specifications

Magnification table

Eyep.      Object.     Magn.     with Barlow l.

5x            4x            20x          40x

5x            10x          50x          100x

5x            40x          200x        400x 16x

4x           64x           128x        16x

10x         160x          320x

16x          40x          640x         280x

EC Declaration of Conformity

Bresser GmbH has issued a „Declaration of Conformity“ in accordance with applicable guidelines and corresponding standards . The full text of the EU declaration of conformity is available at the follow-ing internet address:
www.bresser.de/download/5014000/ CE/5014000_CE.pdf

UKCA Declaration of Conformity

Bresser GmbH has issued a „Declaration of Conformity“ in accordance with applicable guidelines and corresponding standards. The full text of the UKCA declaration of conformity is available at the following internet address: www.bresser.de/download/5014000/ UKCA/5014000_UKCA.pdf

Bresser UK Ltd.
Suite 3G,
Eden House,
Enterprise Way,
Edenbridge,
Kent TN8 6HF,
Great Britain

Disposal

Dispose of the packaging materials properly, according to their type (paper, cardboard, etc). Contact your local waste disposal service or environmental authority for information on the proper disposal.

Do not dispose of electronic devices in the household garbage! As per Directive 2002/96/EC of the European Parliament on waste electrical and electronic equipment and its adaptation into German law, used electronic devices must be collected separately and recycled in an environmentally friendly manner. In accordance with the regulations concerning batteries and rechargeable batteries, disposing of them in normal household waste is explicitly forbidden. Please make sure to dispose of your used batteries as required by law — at a local collection point or in the retail market. Disposal in domestic waste violates the Battery Directive. Batteries that contain toxins are marked with a sign and a chemical symbol.

Warranty & Service

The regular warranty period is 2 years and begins on the day of purchase. To benefit from an extended voluntary warranty period as stated on the gift box, registration on our website is required. You can consult the full warranty terms as well as information on extending the warranty period and details of our services at www.bresser.de/warranty_terms.

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

BRESSER User Manuals

Related Manuals