AND GX-AE/GX-A/GX-AWP/GX-AWP Series Multi Function Balance User Guide
- June 4, 2024
- AND
Table of Contents
1WMPD4004249B
GX-AE/GX-A/GF-A
GX-AWP/GX-AWP
Quick Start Guide
GX-AE/GX-A/GF-A/GX-AWP/GF-AWP Series
Multi-Function Balance
© 2020 A&D Company, Limited. All rights reserved.
No part of this publication may be reproduced, transmitted, transcribed, or
translated into any language in any form by any means without the written
permission of A&D Company, Limited.
The contents of this publication and the specifications of the instrument
covered by this publication are subject to change for improvement without
notice.
Introduction
This manual describes the basic functions and operations of the balance.
For more details, refer to each manual listed in “1-1. List of Functions
Explained in Separate Manuals”.
1-1 List of Functions Explained in Separate Manuals
The functions and operating procedures of the balance are explained in detail
in other documents
besides this manual.
The following three manuals can be downloaded from our website
https://www.aandd.jp.
-
[Communication Manual] Using the communication function of the balance, a personal computer (PC) or printer
This is a supplementary manual for connecting to peripheral devices. -
[GX-AE/GX-A/GF-A/GX-AWP/GF-AWP Series Instruction Manual] This is an instruction manual for understanding and fully utilizing the precision electronic balance GX-AE/GX-A/GF-A/GX-AWP/GF-AWP series.
-
[Flow Rate Display (FRD) Function] This is a supplementary manual for using the flow rate measurement function (FRD: Flow rate display).
Installation and Precautions
The balance is a precision instrument. Unpack it carefully. Keep the packing material to be used for transporting the balance in the future. The packing contents depend on the balance model. See the illustrations to confirm that everything is contained.
Note
- Please confirm that the AC adapter type is correct for your local voltage and receptacle type.
- Please use the specified dedicated AC adapter for the balance.
- Do not use the included AC adapter for models that are not considered compatible with the AC adapters.
- If you use the wrong AC adapter, the balance and other equipment may not operate properly.
2-1 Precautions
To get the optimum performance from the balance and acquire accurate weighing
data, note the following:
-
Install the balance in an environment where the temperature and humidity are not excessive.
The best operating temperature is about 20°C±2°C at about 45 ~ 60%RH relative humidity. -
Install the balance where it is free of dust.
-
The weighing table should be solid, free from vibration and drafts, and as level as possible. (An anti-vibration table or stone table is ideal)
-
Install the balance in a stable place avoiding vibration and shock. The corners of rooms on the first floor are best, as they are less prone to vibration.
-
Install the balance where it is not affected by heaters or air conditioners.
-
Install the balance where it is not exposed to direct sunlight.
-
Install the balance away from equipment that produces magnetic fields.
-
Level the balance by adjusting the leveling feet and confirm it using the bubble spirit level.
-
Warm up the balance for at least 30 minutes. Plug in the AC adapter as usual.
-
Adjust the sensitivity of the balance before use or after moving it to another location. Refer to the separate instruction manual “7. Sensitivity Adjustment”.
About the GX-AWP/GF-AWP series
- The dustproof and waterproof level of the balance is equivalent to IP65, and its second digit, “5”, corresponds to “having no harmful influence by receiving a direct jet of water”. Washing with strong water pressure or submersion in water may cause water to enter the balance and cause a malfunction.
- When installing and using the balance under conditions requiring dustproof and waterproof performance, make sure that the AC adapter plug is fully inserted into the AC adapter jack and that the terminal cover is attached to the RS-232C interface or the waterproof RS-232C cable (AXKO2737-500) is used.
- If the RS-232C terminal cover is removed or the waterproof RS-232C cable (AX-KO2737-500) is not used, protection against dust and water is not provided.
Caution
Do not install the balance where flammable or corrosive gas is present.
How to adjust the bubble spirit level
2-2 Connecting to a PC
At the factory settings, you can transfer weighing data from the balance to a
PC by following the steps below (Quick USB mode).
- Connect the balance and PC with the supplied USB cable.
- When connecting for the first time, the PC will automatically start installing the driver.
- Start up the PC software (Excel, etc.) for transmitting the weighing data.
- Set the keyboard input mode to half-width. It is not entered correctly in the full-pitch setting.
- Move the cursor to the place you want to input the weighing data.
- When you press the PRINT key on the balance, weighing data will be transmitted from the balance and input at the location of the cursor.
- Disconnect the USB cable when finished.
For details, download the communication manual from the A&D website (https://www.aandd.jp/) and refer to “5. Connecting to a PC”.
Display Symbols and Key Operation
Display symbols
- Number of statistical data (Statistical calculation mode)
- Displays the weight data relative to the weighing capacity, in percentage, in the weighing mode
Blinking display contents
Key operation
Key operations affect how the balance functions. Normal key operation during
measurement is “Press and release the key immediately” or “Press the key”.
Please do not “Press and hold the key (for 2 seconds)” unless required.
Key | When pressed and released | When pressed and held (for 2 seconds) |
---|---|---|
Turns the display ON: and of. The standby indicator is displayed when the |
display is turned off. The weighing mode is enabled when the display is turned
on.
When the password function is enabled, the password input display will be
displayed. Refer to the separate instruction manual “19-4 How to Input The
Password At The Start Of Weighing”
This ON: OFF key is available anytime. Pressing the ON: OFF key during
operation will interrupt operation and turn the display OFF. *
| In the weighing mode, turns the digit for readability on and off.
In the counting or percent mode, enters the sample storing mode.| – Enters
the function table mode. Please refer to the separate instruction manual “9.
Function Table”.
– Runs the repeatability check function when pressed and held for another 2
seconds after the function table menu is displayed.
(GX-AE/GX-A/GX-AWP series only) Please refer to the separate instruction
manual “20. Repeatability Check Function”.
| Switches the weighing units stored in the function table. Refer to “4.
Weighing”.
Displaying of the unit mg is available for 0.0001g models only.| Displays
other items on the calibration menu. Please refer to the separate instruction
manual “6-2 Self Check Function/Automatic Setting Of Minimum Weighing Value by
ECL”.
| Performs sensitivity adjustment of the balance using the internal mass. (GX-
AE/GX-A/GX-AWP series)| Displays other items of the sensitivity adjustment
menu.
| Stores the weighing data in memory or outputs to a printer or personal
computer depending on the function table settings. (Factory setting = output)|
Enters mode to change the unit mass registration number in counting mode. By
changing the function table:
– Outputs “Title block” and “End block” for GLP, GMP report.
– Displays the data memory menu.
– Enters mode for reading density number in flow measurement.
| Sets the display to zero.
- When the “Gross net tare function” is selected, the display is turned off by pressing and holding (for 2 seconds). Please refer to the separate instruction manual “14. Gross Net Tare Function”.
Sensitivity Adjustment
Since the balance’s resolution is high, weighing values may change due to gravity and daily environmental changes.
4-1 Automatic Sensitivity Adjustment (GX-AE/GX-A/GX-AWP Series Only)
The GX-AE/GX-A/GX-AWP series automatically adjusts the balance sensitivity using the built-in weight according to the temperature change of the operating environment or the set time and interval time. For details, refer to “7-1 Automatic Sensitivity Adjustment” in the separate instruction manual.
4-2 Sensitivity Adjustment Using The Internal Mass (GX-AE/GX-A/GX-AWP Series Only)
Sensitivity adjustment using the internal mass can be performed with one key press.
- Connect the AC adapter and warm up the balance for at least 30 minutes with nothing on the weighing pan.
- Press the CAL key. The balance displays
- The balance performs sensitivity adjustment using the internal mass. Do not allow vibration or drafts to affect the balance.
- After sensitivity adjustment, if the “GLP output parameter of the function table is set, the balanced outputs a sensitivity adjustment report.
- The balance will automatically return to weighing mode.
4-3 Sensitivity Adjustment Using an External Weight
This function adjusts the sensitivity of the balance using an external weight.
- Connect the AC adapter and warm up the balance for at least 30 minutes with nothing on the weighing pan.
- Press and hold the CAL key for 2 seconds until is displayed, then release the key.
- Make sure that nothing is on the weighing pan and press the PRINT key to weigh the zero point. Do not apply vibration, etc.
- Place the external weight on the weighing pan and press the PRINT key. Do not apply vibration etc.
- Remove the external weight from the weighing pan.
- After sensitivity adjustment, if GLP output is set, a “sensitivity adjustment report” is output or stored in data memory.
- The display automatically returns to the weighing display.
- Place the external weight again and check that the set value is ± 2 digits.
If it is out of range, pay attention to the surrounding environment and start from “1”.
4-4 Error Codes
Display Error code | Description |
---|
EC, E11
| Stability error
The balance cannot stabilize due to an environmental problem. Check around
the pan. Prevent vibration, drafts, temperature changes, static electricity,
and magnetic fields, from influencing the balance. Refer to the separate
instruction manual “2-3 During Use”.
To return to the weighing mode, press the CAL key.
| Out of the setting range
The data to be stored is out of the setting range.
| Malfunction of the internal memory element of the balance
If this error continues to be displayed, please request a repair.
EC, E16
| Internal mass error
Applying the internal mass does not yield a change in the mass value as
specified.
Confirm that there is nothing on the pan and perform the weighing operation
from the beginning again.
EC, E17
| Internal mass error
The internal mass application mechanism does not function properly. Perform
the weighing operation from the beginning again.
| Abnormality in the internal memory data of the balance
If this error continues to be displayed, please request a repair.
| Abnormality in the internal memory data of the balance
If this error continues to be displayed, please request a repair.
EC, E20
| Sensitivity adjustment weight error (Positive value)
The sensitivity adjustment weight is too heavy. Confirm the mass value of
the weight. Press the CAL key to return to the weighing mode.
EC, E21
| Sensitivity adjustment weight error (Negative value) The sensitivity
adjustment weight is too light. Confirm the mass value of the weight. Press
the CAL key to return to the weighing mode.
| Overload error
A sample beyond the balance weighing capacity has been placed on the pan.
Remove the sample from the pan.
| Weighing pan error
The mass value is too light.
Confirm that the weighing pan is properly installed and adjust the sensitivity
of the balance.
Display Error code| Description
---|---
| Sample mass error
The balance cannot store the sample for the counting mode or for the
percent mode because it is too light.
Use a larger sample.
| Unit mass error
The sample unit mass for the counting mode is too light. Storing and using
it for counting will cause a counting error. Add samples to reach the
specified number and press the PRINT key. Pressing the PRINT
key without adding samples will shift the balance to the counting mode. But,
for accurate counting, be sure to add samples.
| Clock battery error
The clock backup battery has been depleted.
Press any key and set the time and date. The clock and calendar function works
normally as long as the AC adapter is connected to the balance. If this error
appears frequently, contact the local A&D dealer.
| Power supply voltage fault
The voltage supplied from the AC adapter is abnormal.
Please check if the problem is the AC adapter supplied with the balance.
| ECL repeatability
With the self-check function, the standard deviation (SD) of repeatability
due to electronically controlled load (ECL) exceeded 50 digits.
Please revise the installation environment of the balance.
This is displayed when repeatability is displayed by ECL.
This is displayed when the minimum weighing value (reference value) by ECL is
displayed.
Refer to the separate instruction manual “6-2 Self Check
Function / Automatic Setting Of Minimum Weighing Value by ELC.
| Full memory
The maximum number of stored weighing values has been reached. In order to store more weighing values, it is necessary to delete the data. Refer to the separate instruction manual “11 Data Memory”.
| Full memory
The stored sensitivity adjustment and calibration test history have reached
50 instances. If more is stored, the old history will be deleted. Refer to the
separate instruction manual “11 Data Memory”
Display Error code| Description
---|---
EC, E00| Communications error
A protocol error occurred in communications. Confirm the format, baud rate,
and parity.
EC, E01| Undefined command error
An undefined command was received. Confirm the command.
EC, E02| Not ready
A received command cannot be processed.
Example: The balance received a “Q” command, but not in the weighing mode.
The balance received a “Q” command while processing a RE-ZERO command.
Adjust the delay time to transmit a command.
EC, E03| Timeout error
If the timeout parameter is set to, the balance did not receive the next
character of command within the time limit of one second. Confirm the
communication.
EC, E04| Excess characters error
The balance received excessive characters in a command. Confirm the
command.
EC, E06| Format error
A command includes incorrect data. Example: The data is numerically
incorrect. Confirm the command.
EC, E07| Parameter setting error
The received data exceeds the range that the balance can accept. Confirm
the parameter range of the command.
4-5 Other Display
Advice
This is the automatic sensitivity adjustment notice (the indicator blinking).
If the balance is not used for several minutes with this indicator blinking,
the balance automatically performs sensitivity adjustment using the internal
mass. The blinking period depends on the operating environment. The balance
can be used when this indicator is blinking, but we recommend that you perform
sensitivity adjustment before weighing.
Specifications
| GX-124AE GX-124A GF-124A| GX-224AE GX-224A GF-224A| GX-324AE GX-324A GF-324A
---|---|---|---
Weighing capacity| 122g| 220g| 320g
Maximum display| 122.0084g| 220.0084g| 320.0084g
Readability| 0.0001g
Repeatability (Standard deviation)| 0.0001g| 0.0002g (300g)
0.0001g (200g)
Linearity| ± 0.0002g| ± 0.0003g
| GF-123A| GX-203A GF-203A
GX-203AWP2 GF-203AWP2| GX-303A GF-303A| GX-403A GF-403A
GX-403AWP2 GF-403AWP2| GX-603A GF-603A
GX-603AWP2 GF-603AWP2| GX-1003A GF-1003A| * GX-1603A GF-1603A
---|---|---|---|---|---|---|---
Weighing capacity| 122g| 220g| 320g| 420g| 620g| 1100g| 1620g
Maximum display| 122.084g| 220.084g| 320.084g| 420.084g| 620.084g| 1100.084g|
1620.084g
Readability| 0.001g
| | 0.002g
Repeatability (Standard deviation)| 0.001g| (1600g) 0.001g (1000g)
Linearity| ±0.002g| ±0.003g
Accuracy after| |
sensitivity adjustment with internal mass 1| ±0.010g| ±0.010g (1000g)
| GF-1202A| GX-2002A GF-2002A
GX-2002AWP2 GF-2002AWP2| ** GX-3002A GF-3002A| GX-4002A GF-4002A
GX-4002AWP2 GF-4002AWP2| GX-6002A GF-6002A
GX-6002AWP2 GF-6002AWP2| GX-10002A GF-10002A
---|---|---|---|---|---|---
Weighing capacity| 1220g| 2200g| 3200g| 4200g| 6200g| 10200g
Maximum display| 1220.84g| 2200.84g| 3200.84g| 4200.84g| 6200.84g| 10200.84g
Readability| 0.01g
Repeatability
(Standard deviation)| 0.01g| 0.02g (10000g) 0.01g (5000g)
Linearity| ±0.02g| ±0.03g
Accuracy after sensitivity adjustment with internal mass *1| ±0.10g| ±0.15g|
±0.15g (5000g)
1 The operating environment does not include an excessive change in ambient temperature, humidity, vibration, drafts, magnetic fields, and static electricity.
2 Compliant with IP65
| GX-6001A GF-6001A
GX-6001AWP2 GF-6001AWP2| GX-10001A GF-10001A
---|---|---
Weighing capacity| 6200g| 10200g
Maximum display| 6208.4g| 10208.4g
Readability| 0.1g
Repeatability (Standard deviation)| 0.1g
Linearity| ±0.1g
Accuracy after sensitivity adjustment with internal mass *1| ±0.5g (5000g)
1 The operating environment does not include an excessive change in ambient temperature, humidity, vibration, drafts, magnetic fields, and static electricity.
2 Compliant with IP65
A &D Company, Limited
3-23-14 Higashi-Ikebukuro, Toshima-ku, Tokyo 170-0013, JAPAN
Telephone: [81] (3) 5391-6132
Fax: [81] (3) 5391-1566
A &D ENGINEERING, INC.
1756 Automation Parkway, San Jose, California 95131, U.S.A.
Telephone: [1] 408-263-5333
Fax: [1] (408)263-0119
A &D INSTRUMENTS LIMITED
Unit 24/26 Blacklands Way, Abingdon Business Park, Abingdon, Oxfordshire OX14
1DY United Kingdom
Telephone: [44] (1235) 550420
Fax: [44] (1235) 550485
A &D AUSTRALASIA PTY LTD
32 Dew Street, Thebarton, South Australia 5031, AUSTRALIA
Telephone: [61] (8) 8301-8100
Fax: [61] (8) 8352-7409