CEM DT-180A True RMS Autoranging Digital Multimeter Instruction Manual

June 8, 2024
CEM

CEM DT-180A True RMS Autoranging Digital Multimeter

Safety

international Safety Symbols

This symbol, adjacent to another symbol or terminal, indicates the user must refer to the manual for further information.

This symbol, adjacent to a terminal, indicates that, under normal use, hazardous voltages may be present.

Safety Notes
  • Do not exceed the maximum allowable input range of any function.

  • Do not apply voltage to meter when resistance function is selected.

  • Set the function switch OFF when the meter is not in use.
    WARNINGS

  • Set function switch to the appropriate position before measuring.

  • When measuring volts do not switch to current/resistance modes.

  • Do not measure current on a circuit whose voltage exceeds 240V.

  • When changing ranges using the selector switch always disconnect the test leads from the circuit under test.

  • Do not exceed the maximum rated input limits.
    CAUTIONS

  • Improper use of this meter can cause damage, shock, injury or death.

  • Read and understand this user manual before operating the meter.

  • Always remove the test leads before replacing the battery.

  • Inspect the condition of the test leads and the meter itself for any damage before operating the meter, repair or replace any damage parts before use.

  • Use carefully when making measurements if the voltages are greater than 25VAC rms or 35VDC, these voltages are considered a shock hazard.

  • Remove the battery if the meter is to be stored for long periods.

  • Always discharge capacitors and remove power from the device under test before performing Diode, Resistance or Continuity tests.

  • Voltage checks on electrical outlets can be difficult and misleading because of the  uncertainty of connection to the recessed electrical contacts, other means should be used to ensure that the terminals are not “live··.

  • If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.

input Limits
Function Maximum Input
mA AC/DC 800mA 600V fast acting fuse
V DC, V AC 600V DC/600V AC
Frequency, Duty Cycle, Resistance, Diode, Continuity, CAP 250V DC/AC
  • When the measured voltage exceeds the range, the buzzer will continue to sound as a warning of exceeding the range.

Description

Meter Description

Meter Description

  1. Non-Contact Voltage Indicator
  2. LCO Display
  3. MOOE Button
  4. 0ata Hold/Backlight Button
  5. Rotary Function Switch
  6. COM Input Jack
  7. Positive Input Jack
  8. Lanyard Hole
  9. Battery Cover
Symbols on the Display

Symbols on the Display

  1. Auto Power Off
  2. Auto Ranging
  3. Display Hold
  4. Diode Test
  5. Continuity 12-Minus Sign
  6. Percent [Duty Cycle]
  7. Non-Contact AC Voltage Detection
  8. Measurement Reading
  9. Units of Measure List
  10. Firewire Identification Measurement
  11. Low Battery

Operation

Notices: Read and understand all warning and precaution statements listed in the safety section of this operation manual prior to using this meter. Set the function select switch to the OFF Position when the meter is not in use.

AC/DC Voltage (Frequency/Duty Cycle) Measurement
  1. Set the function switch to the VAC/DC/Hz/% Position.
  2. Inset the black test lead into the COM Input Jack and the red test lead into the Positive Input Jacks.
  3. Use the MODE Button to select AC or DC Voltage.
  4. Position the black test probe tip and red test probe tip respectively in the spots of the circuit to be measured.
  5. Read the voltage in the display.
  6. ln the VAC mode, press and hold the MODE Button for >1 second to indicate “HZ” read the frequency in the display.
  7. Press the MODE Button again to indicate “%” read the % of duty cycle in the display.
    0peration
DC Current Measurement
  1. Insert the black test lead into the COM Input Jack and the red test lead into the Positive Input Jacks.
  2. For current measurements up to 6000pA DC, set the function switch to the A AC/DC/Hz/% Position.
  3. For current measurements up to 600mA DC, set the function switch to the mA AC/DC/Hz/% Position.
  4. Press the MODE Button to indicate “DC” on the display.
  5. Remove power from the circuit under test. then open up the circuit at the point where you wish to measure current.
  6. Touch the black test probe tip to the negative side of the circuit; Touch the red test probe tip to the positive side of the circuit.
  7. Apply power to the circuit.
  8. Read the current in the display.
    DC Current Measurement
AC Current (Frequency/Duty Cycle) Measurement
  1. Insert the black test lead into the COM Input Jack and the red test lead into the Positive Input Jacks.
  2. For current measurements up to 6000pA AC, set the function switch to the pA AC/DC/Hz/% Position.
  3. For current measurements up to 600mA AC. set the function switch to the mA AC/DC/Hz/% Position.
  4. Press the MODE Button to indicate “AC” on the display.
  5. Remove power from the circuit under test. then open up the circuit at the point where you wish to measure current.
  6. Touch the black test probe tip to the negative side of the circuit; Touch the red test probe tip to the positive side of the circuit.
  7. Apply power to the circuit.
  8. Read the current in the display.
  9. Press and hold the MODE Button for >1 second to indicate “Hz”. read the frequency in the display.
  10. Press the MODE Button again to indicate “%”, read the % of duty cycle in the display.
    0peration
Resistance Measurement
  1. Set the function switch to the CAP Position.
  2. Insert the black test lead into the COM Input Jack and the red test lead into the Positive Input Jacks.
  3. Press the MODE Button to indicate ” “on the Display.
  4. Touch the test probe tips across the circuit or component under test. it is best to disconnect one side of the device under test so the rest of the circuit will not interfere with the resistance reading.
  5. For Resistance tests. read the resistance on the LCD display.
Diode Measurement
  1. Set the function switch to the CAP Position.
  2. Insert the black test lead into the COM Input Jack and the red test lead into the Positive Input Jacks.
  3. Press the MODE Button until ” ”  and ” ” appears in the display.
  4. Touch the test probes to the diode under test. forward voltage will indicate 0.4V to 0.7V; Reverse voltage will indicate ·al.””; Shorted devices will indicate near OmV and an open device will indicate ·al”· in both polarities.
    0peration
    0peration
Continuity Check
  1. Set the function switch to the I-CAP Position.
  2. Insert the black test lead into the COM Input Jack and the red test lead into the Positive Input Jacks.
  3. Press the MODE Button until” ” and “ ” appears in the display.
  4. Touch the test probe tips to the circuit or wire you wish to check.
  5. If the resistance is less than approximately 50 , the audible signal will sound: If the circuit is open. the display will indicate “OL”.
    0peration
Capacitance Measurement

WARNING: To avoid electric shock, disconnect power to the unit under test and discharge all capacitors before taking any capacitance measurements. Remove the batteries and unplug the line cords.

  1. Set the function switch to the CAP Position.
  2. Insert the black test lead into the COM Input Jack and the red test lead into the Positive Input Jacks.
  3. Press the MODE Button to indicate · NF · on the display.
  4. Touch the test leads to the capacitor to be tested.
  5. The test may take up to 3 minutes or more for large capacitors to charge, wait until the readings settle before ending the test. 6.Read the capacitance value in the display.
Non-Contact AC Voltage Measurement

WARNING: Risk of Electrocution. Before use. always test the Voltage Detector on a known live circuit to verify proper operation test the Voltage Detector on a known live circuit to verify proper operation.

  1. Set the function switch to the NCV/Live Position.
  2. Press the MODE Button to indicate “NCV” on the display.
  3. Hold the detector close to the AC voltage being tested.
  4. If no signal is detected. the LCD will show “EF”, NCV indicator light doesn’t flashes and the buzzer is no sound; According to the detected signal strength. LCD displays different horizontal lines; When the signal is strongest. LCD displays four horizontal lines. when the signal is weakest. only one horizontal line, meanwhile NCV indicator light flashes, buuer make different sound. The meter flashes three different colors of light depending on the strength of the ac signal it detects: bright green when the LCD displays one horizontal line. bright yellow when the LCD displays two horizontal lines. and bright red when the LCD displays three or four horizontal lines.

Note: The conductors in electrical cord sets are often twisted. for best results. rub the probe tip along a length of the cord to assure placing the tip in close proximity to the live conductor.
Note: The detector is designed with high sensitivity, static electricity or other sources of energy may randomly trip the sensor. this is normal operation.
0peration

Live Measurement

WARNING: Fi rewire test is to facilitate the use of the firewire with the distinction function, through this function can quickly locate the firewire, However safety should be paid attention to in terms of electricity consumption, if necessary please test the voltage with ac voltage file to confirm the voltage.

  1. Set the function switch to the NCV/Live Position.
  2. Press the MODE Button to indicate ·1ive” on the display.
  3. Insert the red test lead into the Positive Input Jacks.
  4. By inserting the red test probe tip into the electrical outlet hole alone, the red test probe tip touches the line under test. if the main display displays a large ‘”LIVE” and a red light with LED flashing and buuer sound, it is a Live wire.

Function

MODE Button

To select /Diode/Continuity/CAP or DC/AC or Hz/% or NCV/Live.

HOLD Mode
  • The hold function freezes the reading in the display.
  • Press the HOLD/Backlight Button momentarily to activate or to exit the HOLD function.
    Note: The HOLD feature will activate when the Backlight is turned on. press the HOLD/Backlight Button again to exit Hold.
Display Backlight
  • Press and hold the HOLD/Backlight Button for> 1 second to turn on or off the display backlight function.
  • The backlight will automatically turn off after approx. 5 minutes.
Auto Power Off
  • The auto off feature will turn the meter off after approx. 15 minutes.
  • To disable the auto power off feature. hold down the MODE Button and turn the meter on.

Battery Replacement

  1. Remove the one rear Phillips head screw.
  2. Open the battery compartment.
  3. Replace the requires two 1.5V “”AAK’ Batteries.
  4. Re-assemble the meter.

Care and Maintenance

  • Keep the meter dry, if it gets wet, wipe it off.
  • Keep the meter clean. Wipe the case occasionally, do not use chemicals, cleaning solvents, or detergents.
  • Handle the meter gently and carefully, dropping it can damage the electronic parts or the case.
  • Use and store the meter in normal temperatures, temperature extremes can shorten the life of the electronic parts and distort or melt plastic parts.
  • Use only fresh batteries of the recommended size and type, remove old or weak batteries so they do not leak and damage the unit.
  • If the meter is to be stored for a long period of time, the battery should be removed to prevent damage to the unit.

Specifications

Function Range & Resolution Accuracy
DC Voltage 600.0mV ±[0.5% + 10 digits)

6.000V
60.00V
600.0V
AC Voltage| 6.000V| ±[1.0% + 4 digits)
60.00V
600.0V
All AC voltage ranges are spec1f1ed from 5% of range to 100% of range .
AC voltage bandwidth: 50Hz to lkHz [Sine); 50Hz to 60Hz [All wave).
DC Current| 600.0µA| ±[1.0% + 3 digits)
6000µA
60.00mA
600.0mA
AC Current| 600.0µA| ±[1.5% + 3 digits)
6000µA
60.00mA
600.0mA
All AC Current ranges are spec1f1ed from 5% of range to 100% of range .
AC Current 50Hz to 1 kHz [Sine); 50Hz to 60Hz [All wave).
Resistance| 600.0Q| _±[1.0% + 3 _digits)
6 000kQ
60.00kQ| ±[0 8% + 2 digits)
600.0kQ
6 000MQ| ±[2 0% + 3 digits)
60.00MQ
Capacitance| 99.99nF| ±[5 0% + 20 digits)
999.9nF|

±[4% + 5 digits)

9.999µF
99.99µF
999.9µF
9.999mF| ±[10% reading)
99.99mF
Frequency [Electrical)| 10Hz to 100kHz| ± 1.2% + 5 digits
Sensitivity: ACV Range >6% Range ; 6000μA/600.0mA Range >6% Range ; 600.0μA/60.00mA Range [>60% Rangel.
Duty Cycle| 10.5% to 99.9%| 1±[1.2% + 2 digits]
Pulse width: 0.1 ms to 1 00ms; Frequency: 5Hz to 1 0kHz. Sensitivity: ACV Range [>6% Range); 6000μA/600.0mA Range [>6% Range); 600.0μA/60.00mA Range [>60% Rangel.

Note: Accuracy specifications consist of two elements:

  • (% reading)-This is the accuracy of the measurement circuit.
  • (+ digits)-This is the accuracy of the analog to digital converter.

Note: Accuracy is given at 18 to 28°C [65 to 83°Fl. less than 70 % RH.

General Specifications
Insulation Class 2
Display 6000 counts Backlight LCD
Polarity Automatic [No indication for positive polarity) Minus symbol”-” is

displayed for negative polarity
Overrange Indication| “Or is displayed
Diode Test| Test current of 0.63mA typical: Open circuit voltage 3.2V DC typical.
Continuity Check| Threshold <50 : Test current< 1 mA
Low Battery Indication| ; : is displayed
Measurements Rate| 3 per second, nominal
Auto Power Of| 10M O [VDC and VAC)
Input Impedance| True RMS
AC Response| True RMS
AC Voltage Bandwidth| 50Hz to 1 kHz [VAC)
Power Source [Battery)| Two 1.5V “”AAA”” Batteries
Operating Temperature| 0 to 50°C [32 to 122°F1
Storage Temperature| 10 to 60°C [14 to 140°Fl
Relative Humidity| <70% Operating: <80% Storage
Operating Altitude| 2000 meters [7000ft.) maximum.
Over Voltage| Category Ill 600V
Safety| For indoor use and in accordance with Overvoltage Category II, Pollution Degree 2. Category II includes local level, appliance, portable equipment, etc., with transient overvoltage’s less than Overvoltage Category Ill. Probe assemblies to be used for mains measurements shall be rated as appropriate for measurement Category Ill according to IEC 61010-31 and shall have a voltage rating of at least the voltage of the circuit to be measured. Measurement category Ill [CAT Ill) is applicable to test and measuring circuits connected to the distribution part of the buildings low-voltage mains installation.

Read User Manual Online (PDF format)