Testboy TB 313 Handheld Digital Multimeter Instruction Manual
- June 9, 2024
- Testboy
Table of Contents
- Safety notes
- General safety notes
- Operation
- Introduction
- Included in delivery
- Safety precautions
- Switches, buttons and sockets description
- DC voltage measurement / V=
- DC voltage
- AC voltage measurement / V~
- DC current measurement / A=
- AC current measurement / A~
- Capacitance measurement / F
- Continuity testing
- Temperature
- Frequency
- Maintenance
- Cleaning
- Changing the battery
- Changing the fuse
- Technical data
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Test boy TB 313 Handheld Digital Multimeter
Safety notes
| WARNING
An additional source of danger is posed my mechanical parts which can cause
severe personal injury. Objects can also be damaged (e.g., the instrument
itself can be damaged).
---|---
| WARNING
An electric shock can result in death or severe injury. It can also lead to
property damage and damage to this instrument.
| WARNING
Never point the laser beam directly or indirectly (on reflective surfaces)
towards the eyes. Laser radiation can cause irreparable damage to the eyes.
You must first deactivate the laser beam when measuring close to people.
General safety notes
| WARNING
Unauthorized changes or modifications of the instrument are forbidden – such
changes put the approval (CE) and safety of the instrument at risk. In order
to operate the instrument safely, you must always observe the safety
instructions, warnings and the information in the “Proper and Intended Use”
Chapter
---|---
| WARNING
Please observe the following information before using the instrument:
Do not operate the instrument in the proximity of electrical welders,
induction heaters and other electromagnetic fields.
After an abrupt temperature fluctuation, the instrument should be allowed to
adjust to the new temperature for about 30 minutes before using it. This helps
to stabilize the IR sensor.
Do not expose the instrument to high temperatures for a long period of time.
Avoid dusty and humid surroundings.
Measurement instruments and their accessories are not toys. Children should
never be allowed access to them!
In industrial institutions, you must follow the accident prevention
regulations for electrical facilities and equipment, as established by your
employer’s liability insurance organization.
Please observe the following five safety rules:
- Disconnect.
- Ensure that the instrument cannot be turned back on again.
- Ensure isolation from the main supply voltage (check that there is no voltage on both poles).
- Earth and short-circuit.
- Cover neighboring parts that are under live electrical load.
Proper and intended use
This instrument is intended for use in applications described in the operation manual only. Any other usage is considered improper and non-approved usage and can result in accidents or the destruction of the instrument. Any misuse will result in the expiry of all guarantee and warranty claims on the part of the operator against the manufacturer.
Remove the batteries during longer periods of inactivity in order to avoid damaging the instrument.
We assume no liability for damages to property or personal injury caused by improper handling or failure to observe safety instructions. Any warranty claim expires in such cases. An exclamation mark in a triangle indicates safety notices in the operating instructions. Read the instructions completely before beginning the initial commissioning. This instrument is CE approved and thus fulfils the required guidelines.
All rights reserved to alter specifications without prior notice © 2014 Test boy, Germany.
Disclaimer and exclusion of liability
The warranty claim expires in cases of damages caused by failure to observe the instruction! We assume no liability for any resulting damage!
Test boy is not responsible for damage resulting from: failure to observe the instructions, changes in the product that have not been approved by Test boy, the use of replacement parts that have not been approved or manufactured by Test boy, the use of alcohol, drugs or medication.
Correctness of the operating instructions
These operating instructions have been created with due care and attention. No claim is made nor guarantee given that the data, illustrations and drawings are complete or correct. All rights are reserved in regards to changes, print failures and errors.
Disposal
For Test boy customers: Purchasing our product gives you the opportunity to return the instrument to collection points for waste electrical equipment at the end of its lifespan.
The WEEE directive regulates the return and recycling of electrical appliances. Manufacturers of electrical appliances are obliged to take back and recycle all electrical appliances free of charge. Electrical devices may then no longer be disposed of through conventional waste disposal channels. Electrical appliances must be recycled and disposed of separately. All equipment subject to this directive is marked with this logo.
Disposing of used batteries
As an end user, you are legally obliged (by the relevant laws concerning
battery disposal) to return all used batteries. Disposal with normal
household waste is prohibited!
Contaminant-laden batteries are labelled with the adjacent symbol which
indicates the prohibition of disposal with normal household waste. The
abbreviations used for heavy metals are: Cd = Cadmium, Hg = mercury,
Pb = lead.
You can return your used batteries for no charge to collection points in your
community or everywhere where batteries are sold!
Certificate of quality
All aspects of the activities carried out by Test boy relating to quality during the manufacturing process are monitored permanently within the framework of a Quality Management System. Furthermore, Test boy confirms that the testing equipment and instruments used during the calibration process are subject to a permanent inspection process.
Declaration of Conformity
The product conforms to the present directives. For more detailed information, go to www.testboy.de
Operation
Introduction
The Test boy TB 313 is a general purpose Multimeter. This measuring instrument has been manufactured to the latest safety specifications, and guarantees safe and reliable use. The Multimeter is a valuable aid for all standard measurement tasks in trade and industrial applications as well as for the hobby electrician interested in electronics.
Included in delivery
TB 313 Safety testing leads CAT III 600 V Operating instructions Holster Temperature probe
Safety precautions
The TB 313 left the factory in proper and safe working order. In order to maintain this condition, the user must observe the safety notes contained in this manual.
| CAUTION!
Use only the enclosed safety measurement leads or equivalent measurement leads
that meet the correct measuring category CAT III 600 V.
---|---
In order to avoid an electrical shock hazard you must observe the specified
precautionary measures when working with voltages greater than 120 V (60 V) DC
or 50 V (25 V) eff AC. These values represent the specified limits of safe-to-
touch voltages in accordance with DIN VDE (values given in brackets apply to
medical or agricultural applications).
Before taking each measurement, ensure that the test leads and the measuring
instrument are in a flawless condition.
The test leads and test probes must only be handled using the isolated grips.
Avoid touching the tips of the test probes under all circumstances.
The test instrument must only be used for the specified measurement range.
According to the standard EN 61010-1 the following measurement categories are defined:
Measurement Category CAT II
Measurements on circuits that are electrically connected directly to the
network, via plugs in the home, office and laboratory. Measurement Category
CAT III
Measurements on the building installations:
Fixed consumer units, distributor connection, equipment fitted permanently to
the distributor.
Measurement Category CAT IV
Measurements at the source of the low voltage installation: Meter, primary
surge protection, main connection.
Switches, buttons and sockets description
-
ON/OFF switch
The device is turned on and off using the “POWER” push-button switch. -
Lighting switch ( )
Press this button to turn the torch function on and off. -
Memory log button (H)
Press this button to store the actual measurement value. -
Measuring function selector switch
Use the rotary selector switch to select the various measurement modes. -
10 A socket (left)
The 10 A socket must be used for current measurements above 200 mA. -
mA socket
For current measurements up to 200 mA. -
Input socket V//TEMP/CAP
Red test lead for all types of signals supported by the instrument. -
Ground socket
Black test lead for all types of signals supported by the instrument.
DC voltage measurement / V=
Use the selector switch to set the appropriate range. Insert the black test lead into the `COM’ socket andthe red test lead into the V//TEMP/CAP socket. Using the test probes, touch the test points of the test object. Read measurement value on the display. The polarity of the voltage will also be displayed.
DC voltage
Measurement range | Resolution | Accuracy |
---|---|---|
200 mV | 0.1 mV | ± 0.5 % of reading + 1 digits |
2 V | 0.001 V | ± 0.5 % of reading + 3 digits |
20 V | 0.01 V | |
200 V | 0.1 V | |
600 V | 1 V | ± 0.8 % of reading + 3 digits |
– Input resistance: 10 MΩ
– Max. input voltage: 600 V DC
AC voltage measurement / V~
Use the selector switch to set the appropriate range. Insert the black test lead into the `COM’ socket and the red test lead into the V//TEMP/CAP socket. Using the test probes, touch the test points of the test object. Read measurement value on the display.
AC voltage
Measurement range | Resolution | Accuracy |
---|---|---|
200 mV | 1 mV | ± 1.2 % of reading + 5 digits |
2 V | 0.001 V | ± 1 % of reading + 5 digits |
20 V | 0.01 V | |
200 V | 0.1 V | |
600 V | 1 V | ± 1.2 % of reading + 5 digits |
– Input resistance: 10 M
– Max. input voltage range of 200 mV to 250 V
– Max. input voltage: 600 V AC RMS, frequency range: 40-400 Hz
DC current measurement / A=
Use the selector switch to set the appropriate range. Connect the black test lead into the “COM” socket, and the red test lead with the mA / 10A socket (the appropriate socket opens automatically depending on the measuring range). Using the test probes, touch the test points of the test object. Read measurement value on the display. The current direction is indicated by the sign.
You must use the ’10 A’ socket when measuring currents above 200 mA.
Direct current
Measurement range | Resolution | Accuracy |
---|---|---|
2 mA | 1 µA | ± 1.0 % of reading + 3 digits |
20 mA | 0.01 mA | ± 1.0 % of reading + 3 digits |
200 mA | 0.1 mA | ± 1.5 % of reading + 5 digits |
10 A* | 0.01 A | ± 2.0 % of reading + 10 digits |
**** Overload protection: mA range is protected with a 200 mA self-resetting
fuse.
– 10A range is protected with F 10 A / 1000 V.
– In the 10A range observe the maximum operating times!
*To protect against overheating, measurements should not be taken for more than 10 seconds. Following this the device should be allowed to cool down for at least 15 minutes.
AC current measurement / A~
Use the selector switch to set the appropriate range. Connect the black test lead into the “COM” socket, and the red test lead with the mA / 10A socket (the appropriate socket opens automatically depending on the measuring range). Using the test probes, touch the test points of the test object. Read measurement value on the display.
You must use the `10 A’ socket when measuring currents above 200 mA.
Alternating current
Measurement range | Resolution | Accuracy |
---|---|---|
2 mA | 1 µA | ± 1.2 % of reading + 5 digits |
200 mA | 0.1 mA | ± 2.0 % of reading + 5 digits |
10 A* | 0.01 A | ± 3.0 % of reading + 10 digits |
Overload protection: mA range is protected with a 200 mA self-resetting fuse.
– 10A range is protected with F 10 A / 1000 V.
– In the 10A range observe the maximum operating times! – Frequency range:
40-400 Hz
*** To protect against overheating, measurements should not be taken for more than 10 seconds. Following this the device should be allowed to cool down for at least 15 minutes.
Capacitance measurement / F
Use the selector switch to set the appropriate range. Insert the black test lead into the `COM’ socket and the red test lead into the V//TEMP/CAP socket. Using the test probes, touch the test points of the test object. Read measurement value on the disp
Measurement range | Resolution | Accuracy |
---|---|---|
2 nF | 1 pF | ± 4 % + 3 digits |
20 nF | 10 pF | |
200 nF | 0.1 nF | |
2 µF | 1 nF | |
20 nF | 10 nF |
**** Discharge the capacitors before every measurement.
Resistance measurement / Ω
Use the selector switch to set the appropriate range. Insert the black test
lead into the `COM’ socket and the red test lead into the V//TEMP/CAP socket.
Using the test probes, touch the test points of the test object. Read
measurement value on the display.
Measurement range | Resolution | Accuracy |
---|---|---|
200 Ω | 0.1 Ω | ± 1 % + 5 digits |
2 kΩ | 1 Ω | ± 1 % + 5 digits |
20 kΩ | 10 Ω | |
200 kΩ | 100 Ω | |
2 MΩ | 1 kΩ | |
20 MΩ | 0.01 MΩ | ± 1.8 % + 5 digits |
– Measuring voltage: 0.25 V.
Diode test
With the selector switch set to ” / ))”. Insert the black test lead into the `COM’ socket and the red test lead into the V//TEMP/CAP socket. Using the test probes, touch the test points of the test object. Red test lead = anode, black test lead = cathode. The forward voltage drop is displayed.
Measurement range | Resolution | Display |
---|---|---|
1 mV | Forward voltage |
– Forward current: approx. 25 µA, reverse voltage: approx. 2.8 V.
Continuity testing
With the selector switch set to ” / ))”. Insert the black test lead into the `COM’ socket and the red test lead into the V//TEMP/CAP socket. Using the test probes, touch the test points of the test circuit. An acoustic signal is emitted if a resistance under 70 is measured.
**** Important: Isolate from the power supply and discharge capacitors in the circuit to be measured.
Measurement range | Function |
---|---|
◦)) | The integrated buzzer signals up to a resistance of 70 Ω |
– Measuring voltage: approx. 2.8 V.
Temperature
Set the selector switch to “TEMP / °C”. Place the supplied temperature probe into the sockets “COM” and V//TEMP/CAP. Touch the object to be measured with the measuring tip. Read measurement value on the display.
Measurement range | Resolution | Accuracy |
---|---|---|
-20 to 0 °C | 1°C | ± 5 % + 5 digits |
0 to 400 °C | ± 2 % + 3 digits | |
400 to 1000 °C | ± 1 % + 3 digits |
Frequency
Set the selector switch to “Hz”. Insert the black test lead into the `COM’ socket and the red test lead into the V//TEMP/CAP socket. Using the test probes, touch the test points of the test circuit. Read measurement value on the display.
Measurement range | Resolution | Accuracy |
---|---|---|
20 kHz | 10 Hz | ± 1.5 % + 10 digits |
Maintenance
The instrument does not require special maintenance when used as specified in this operation manual.
Cleaning
Use a damp cloth and mild household cleaning agent to clean the instrument should it become soiled through daily use. Never use aggressive cleaning agents or solvents to clean the instrument.
Changing the battery
Change the battery when the battery symbol is displayed. Remove the test leads from the measuring instrument before changing the battery or fuse! Remove the two screws on the rear of the instrument, open the battery compartment and remove the used batteries. Insert new battery (1 × 9 V Block 6F22). Replace battery compartment cover and screw tight.
**** Only use the specified batteries!
Do not dispose of batteries in normal household rubbish! Observe statutory
regulations pertaining to disposal!
Changing the fuse
When changing the fuse, remove the measurement leads from the device and
remove the screws on the rear of the device (two at the bottom, one at the
top). Carefully remove the rear of the housing and replace the fuse with fuse
of the same type (Fuse F 10 A / 1000 V). Replace screws and screw tight.
The F200 mA fuse is self-resetting and therefore maintenance free.
**** Only use the specified fuse!
Technical data
The accuracy relates to 1 year used in temperatures between 18 °C 28 °C and
75% relative humidity. Further calibration is also offered. Max. voltage
between the connection socket and ground:
600 V AC / DC
Fuses | F 200 mA self-resetting (maintenance free) F 10 A 1000 V quick-blow |
---|---|
Max. operating height: | 2000 m above MSL |
Height of display: | 37 mm LCD |
Display | Max. 1999 (3½ ) |
Polarity indicator | Automatic |
Overrange indicator: | “1” is displayed |
Sampling rate | Approx. 0.4 sec. |
Low battery status: | Battery symbol is displayed |
Power supply | 1 × 9 V block battery |
Operating temperature | 0 °C to 40 °C |
Storage temperature | -10 °C to 50 °C |
Dimensions | 191 × 82 × 37 mm (without holster) |
200 × 89 × 38 mm (with holster)
Weight| 280 g incl. battery
Category| CAT III 600 V
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>