HEALTHTIME OC10 Oxygen Concentrator User Guide
- June 4, 2024
- HEALTHTIME
Table of Contents
- GENERAL INFORMATION
- SAFETY GUIDELINES
- STARTING-UP / INSTALLATION
- OPERATING PRINCIPLES
- PREVENTIVE MAINTENANCE
- TROUBLESHOOTING
- AIRFLOW DIAGRAM
- TECHNICAL CHARACTERISTICS
- ACCESSORIES – PACKAGE INCLUDED
- ELECTROMAGNETIC COMPATIBILITY (EMC) DECLARATION
- GUARANTEE TERMS & CONDITIONS
- CERTIFICATES
- MANUFACTURER
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
“OXYGEN CONCENTRATOR / USER GUIDE“
Good Design Award|
---|---
Please read this manual thoroughly before use of the Device
GLOSSARY OF SYMBOLS
| CAUTION CONSULT ACCOMPANYING DOCUMENTS| | CLASS II EQUIPMENT
---|---|---|---
| DATE OF MANUFACTURE| IPX1| PROTECTION AGAINST VERTICALLY FALLING
DROPS
| MANUFACTURE| | TYPE B
| SERIAL NUMBER| | SHOULD BE DISPOSED OF ACCORDING TO THE WEEKLY
DIRECTIVES
| CONFORMITY EUROPE| | TYPE BF
GENERAL INFORMATION
Before operating the device, make sure that you have carefully read and
understood all the topics written in this manual.
Intended Use of the Device: It is used to provide oxygen therapy to the
patient oxygen therapy is required (such as heart, lung, and other respiratory
diseases) who determined and recommended by the doctor.
Working Principle: This oxygen concentrator is designed for personal use
at home by providing separate oxygen from the air in the room. This device
obtains 90% pure oxygen from the environment at the rate of 21% oxygen by
using PSA technology. Thanks to the cannula attached to the nose, allows the
patient to receive highly concentrated oxygen directly.
Side Effects: As with any treatment, oxygen therapy has risks and
unwanted side effects. The most important risks in oxygen treatment are
accidents and fires related to the transportation, filling, and use of oxygen
devices. Since oxygen is a flammable and explosive substance, the oxygen
source should be placed away from heat generating systems and the patient
should never smoke during oxygen therapy. Potential side effects of exposure
to high concentrations of oxygen are changes in normal physiological
functions, oxygen-related tissue damage/oxygen toxicity, and carbon dioxide
accumulation. Carbon dioxide accumulation is mostly seen as a result of using
high levels of oxygen, and it is usually manifested by drowsiness in patients.
Patients Need Oxygen Therapy: Patients with Chronic Obstructive
Pulmonary, Pulmonary Hypertension, Congestive Heart Failure, Cystic fibrosis,
Lung cancer, Alpha 1, and similar patients.
Patient Population: There are approximately 210 million COPD patients in
the world and 3.5 million in our country. Especially, over the age of 40 who
smoke, this disease is seen in one out of every 5 people.
General Features
The HEALTH TIME is easy to use.
It has a Safety and Solid body, 2×16 LCD screen
Digital timer to show total using time.
Electrical cut-off warning for safety against high pressure, low pressure
&high temperature.
Thermo-protected compressor, Light and easy to carry.
Audio-visual warning (alarm) system for periodic changes of Nasal Cannula,
Pull Filter & Humidifier.
SAFETY GUIDELINES
-
The HEALTH TIME must only be used for purposes mentioned in this manual!
-
Do not use this device as Life Supporting or Life Sustaining unit. Use your device according to your doctor’s prescription and advice only, otherwise, it may present a risk to the health of you (patient).
-
The use of highly concentrated oxygen could be dangerous if not prescribed by an authorized person. Please keep spare oxygen sources such as oxygen tubes or bags ready for the patient considering possible malfunctioning of the device may occur anytime. Keep the device upright position during transportation to avoid damage to the cabinet. Switch off the device when not used by the patient.
-
During transportation or when not in use unplug the power supply cord. Do not allow an unauthorized person to open the cabinet of the device. If the electric supply line suffers from frequent voltage fluctuation, we suggest using a Voltage Regulator for the device to avoid unwanted damage to the sensitive electronic parts of the device.
-
Do not unplug the power cable by hanging it on the device or by pulling the cable.
-
Do not leak water on the device or emerge the device into water. If you think there is a chance of water may come in contact with the device, please unplug the power cable and inform the technical service.
-
Do not open the cabinet of the device when the power cable is connected. Only authorized technical service personnel is permitted to open the device. There are no parts inside the device which you can repair.
-
Do not put the nasal cannula under heavy material like a bed or cushion.
-
Keep the device under surveillance when used by an aged, child, or disabled patient who may not be able to inform any malfunctioning of the device may take place during the treatment.
-
Call your doctor immediately in case of the slightest disturbance seen in the patient during treatment.
-
HEALTH TIME device produces highly concentrated oxygen which is not flammable but accelerates the combustion of the material. To avoid all risks of fire do not smoke nearby the device and other accessories. HEALTH TIME should be kept away from all flames, incandescent sources, and sources of heat, as well as combustible products like oil, grease, solvents, aerosols, etc.
-
The manufacturer or the seller of the device is not responsible for the problems that may occur due to wrong, unnecessary or unintended use of the device or use of wrong electrical voltage which does not comply with electrical safety rules.
-
Calibration requirement in the device during use of the device:
During use of the device every 2000 hours or once in a year following calibration controls to be done by man face. Oxygen Concentrator percentage
b. Oxygen Flow rate
c. Oxygen outlet pressure -
Flow Cut-off Valve Against Fire: This feature available as standard in our device has double protection facility.
The thermostat located in the compressor detects the increase in Motor temperature and shut down the motor when the temperature reaches a deadline a second thermostat located inside the device continuously measures the temperature inside the device and if any fire risk is detected immediately stops the device and gives an alarm. -
Children, Lactating Women, and Pregnant must use the oxygen concentrator under the control of the specialist doctor. Oxygen Concentrator use requires a physician’s prescription and is not intended for life support use. Although oxygen therapy can be prescribed for patients of all ages, the typical oxygen therapy patient is older than 65 years of age and sufferers of chronic obstructive pulmonary disease (COPD). Patients typically have good cognitive abilities and must be able to communicate easily. If the user is unable to communicate discomfort or unable to read and understand the concentrator labeling and instructions for use, then use is recommended only under the supervision of one who can. If any discomfort is felt while using the concentrator, patients are advised to contact their doctors. Patients are also advised to have backup oxygen available in the event of a power outage or concentrator failure. . The device’s lifetime is 5 years.
.Should be disposed of according to WEEE ( Waste Electrical and Electronic Equipment) Directive
STARTING-UP / INSTALLATION
Please control the packing and the device on arrival. If there is any
deformation or damage in the packing or in the inside materials, please inform
the courier or the seller of the goods immediately. Take out from the package
all components carefully controlling crash, breakage, or any other damages.
NOTE: If HEALTH TIME Oxygen concentrator will not be used immediately, we
suggest keeping the components inside the package until use.
Let stand four (4) hours in the user environment before starting using the
device.
Fill the water bowl with boiled and cooled water and close the lid. Take
care of not overfilling the water bowel out of its maximum mark. The water
bowl must be cleaned with soap & water and rinsed properly once a week. The
water label of the bowel should control frequently and not permit it to go
down to its minimum mark by adding water regularly.
Connect the intermediate cannula between the device and the patient unit (
flowmeter unit ). Place the water bowel in its position in the patient’s unit
as shown in the picture. Plug in the electrical cable. Start the device by
pushing the OFF-ON switch to the ON position. Adjust the flow rate of
concentrate oxygen with the help of an inbuilt flow meter within the range of
0-10 lt./min. as per the doctor’s prescription. The ball located inside the
flow meter will indicate the adjusted flow rate of oxygen. Please consult with
the ‘’Troubleshooting’’ section of this manual if the device is not running
properly.
To use a new device for the first time it is recommended to run the device for
30 min. without patient. This period is necessary to achieve a 90% oxygen
concentration level at the initiation. Repeat the same procedure if the device
is not used for a time period of 15 days or more. For daily use, there is no
need of following this instruction.
OPERATING PRINCIPLES
Please the device at the most suitable place in the room or near to the
patient.
The device can be moved easily with the help of a wheel fitted with the
cabinet.
Do not cover the concentrator with any covering material. There should be at
least 30 cm. distances between the device & the furniture in the room and the
walls to avoid blocking in environmental air inlet to the device. Do not block
the air inlet to the oxygen concentrator by any means. Do not place the device
on soft floors like beds, sofas, etc. it may block air flow to the device.
Keep the device always vertical position and do not use the device during
bath. Do not smoke near the device when running. Keep the device away from
sources of fire and heat. It is dangerous to use a device with a dirty filter
or without a filter. Never run the device with a wet filter. Control the nasal
cannula at a regular interval against twisting, leaking, or bursting. Replace
the damaged cannula with a new one.
Like most other electronic devices HEALTH TIME too is very sensitive towards
radio frequency waves (parasites), please be aware of using mobile phones when
running the device. To get the best efficiency from the device we suggest not
shutting down and opening the device frequently in a short interval which may
cause a loss in the life period of the device too.
PREVENTIVE MAINTENANCE
Please switch off the device and unplug the power supply cable before the
maintenance operation start.
Please switch off the device and unplug the power supply cable before the
maintenance operation start.
Do not use any liquid cleaning agent directly on the device. Petroleum-based
solvent & cleaner are not to be used. Wipe the outer face of the cabinet with
a wet sponge and dry properly once in 30 days.
CLEANING OF DUST FILTER
Take out the dust filer placed at the rear side of the device once every 7
days. The black color dust filter may clean from dust by simply shaking it and
then washing it with water only. Put in place after drying
properly.
CHANGING OF PALL FILTER
Replace the Pall Filter with a new one when there is a warning on the LCD
screen or once every 60 days.
CHANGING OF NASAL CANNULA
Replace the Nasal Cannula with a new one if there is a warning on the LCD
screen or once every 30 days.
CLEANING AND CHANGING OF HUMIDIFIER
Every 7 days clean with soap and water and rinse properly and replace with a
new one when there will be a warning on the LCD screen or once every 90 days.
TROUBLESHOOTING
OBSERVATION| ALARM SOUND| SCREENSHOT| PROBABLE
REASON/CAUSES| SOLUTION
---|---|---|---|---
On opening, there is no alarm and the device is not running| OK| NO IMAGE| The
electrical power supply cord is not properly connected either with the device
or in the plug
There is no electricity in the plug. fuse burnt out| Connect the electrical
cord prop- early both ends Test if there is electricity in the plug or put the
cord in another plug Change the fuse.
Alarm due to high tern- perature| OK| HIGH
TEMPERATURE| The system may be tuned to high-temperature protection mode|
Clean the black filter located at the rear side of the device
Check whether the ventilation is sufficient all around the device Cooling fan
may not work Call authorized technical service.
Keeps the device waiting for 40-50 mins in switched position then reopen, if
the problem continues to call technical service
High-pressure alarm| OK| HIGH
PRESSURE| The system may turn to high pressure protecting mode| Control the
Nasal Cannula against possible breaking. twisting Control
the water bowel against clog-ging Check whether the flow meter is open or
closed If the situations still continue please call Authorized Technical
Service
Low-Pressure Alarm| OK| LOW
PRESSURE| The system may turn to low pressure protecting mode| Control the
dust filter, if required
clean it Flow meter may setted at 5 It /min reset it to a lower value
Please check. the intermediate connecting hose may not be connected- ed. If
the problems still continue call technical service
Low Oxygen Alarm| OK| LOW OXYGEN| A low oxygen condition may have occurred in
the system.| Control the dust filter. pall filter, nasal cannula, and
flowmeter If the problems still continue call technical service.
Call technical service if the suggested troubleshooters are not helpful for
the solution of problems. For problems not mentioned in this table, please
consult with an authorized technical service.
AIRFLOW DIAGRAM
TECHNICAL CHARACTERISTICS
OXYGEN FLOW RATE | 0-10 PM |
---|---|
OXYGEN CONCENTRATION | % 93 ± 3 ( 10 lt/DK ) |
OUTLET PRESSURE | 8 ± 1 PASI |
NOISE LEVEL | < 55 dBA |
VOLTAGE / CURRENT | 230 VAC / 50 Hz / 2,60 A max. |
POWER | Max. 600 Watt |
WEIGHT | 18 kg |
BOX DIMENSIONS | 38 49 61 cm |
WORKING CONDITIONS | 10 – 40 º C, % 20 – % 60 RH, 100kPa – 80 kPa |
STORAGE CONDITIONS | 10 – 60 º C , % 10 – % 95 RH |
ELECTRICAL SAFETY | TYPE B, CLASS II |
FUSE | 4 A |
CERTIFICATE | ISO 9001, ISO 13485, 1984 |
ACCESSORIES – PACKAGE INCLUDED
Package Included: 1 pc Oxygen Concentrator, 1 pc Oxygen Cannula, 1 pc
Humidifier, 1 pc Power Cord
Accessories: The cannula, intermediate tube, and humidifier which are to
be used with the oxygen concentrator must comply with medical device
directives and should be CE certified.
The use of inappropriate parts and accessories may cause a lack of performance
in the device.
ELECTROMAGNETIC COMPATIBILITY (EMC) DECLARATION
This equipment generates, uses, and can radiate radio frequency (RF) energy.
This equipment may cause electromagnetic interference if it is not installed
and used in accordance with the manual. This device has been tested for
Medical Products in accordance with EN-60601-1-2:2015 and has been determined
to comply with acceptable limits. These limits indicate that the device
provides an acceptable level of protection against electromagnetic
interference (EMC) if the device is used as specified in the manual. This
device may be affected by portable and mobile RF communication devices. This
equipment must not be stored together with other equipment. See the following
tables for more information about this device and EMC.
GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC EMISSIONS
The device is intended for use in the electromagnetic environment specified
below. The customer or the user of the device should assure that it is used in
such an environment
Emission test| Compliance| Electromagnetic environment –
guidance
---|---|---
Emissions RF CISPR 11|
Group 1
| The device uses RF energy only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause any interference in nearby
electronic equipment.
Emissions RF CISPR 11| Class B| The device is suitable for use in all
establishments including domestic and those directly connected to the public
low-voltage power supply network that supplies buildings used for domestic
purposes.
Harmonic emissions are used for domestic purposes. IEC 61000-3-2| Class A
Voltage fluctuations/ flicker emissions IEC 61000-3-3| Complies
GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC EMISSIONS
The device is intended for use in the electromagnetic environment specified
below. The customer or the user of the device should assure that it is used in
such an environment
Immunity test| IEC 60601 Test level| Compliance level|
Electromagnetic environment – guidance
---|---|---|---
Electrostatic discharges IEC 61000-4-2| ± 8 kV contact
±2 kV, ±4 kV,
±8 kV,±15 kV air| ± 8 kV contact
±2 kV, ±4 kV,
±8 kV,±15 kV air| Floors should be wood, concrete or ceramic tile. If floors
are covered with synthetic material, the relative humidity should be at least
30
Electrical fast transient/burst IEC 61000-4-4| Power supply lines: ±2 kV
Input/output lines:±1 kV 100 kHz repetition frequency| Power supply lines: ±2
kV Input/output lines:±1 kV 100 kHz repetition frequency| The main power
quality should be that of a typical commercial or hospital environment.
Surges Line-to-line IEC 61000-4-5| ± 0,5 kV, ± 1 kV| ± 0,5 kV, ± 1 kV| The
main power quality should be that of a typical commercial or hospital
environment.
Surges Line-to-ground IEC 61000-4-5| ± 0,5 kV, ± 1 kV, ± 2 kV| ± 0,5 kV, ± 1
kV,± 2 kV
Voltage dips IEC 61000-4-11| 0% UT;0.5 cycle
0º, 45 º, 90 º, 135 º,
180 º,225 º, 270 º, and 315 º| 0% UT;0.5 cycle
0º, 45 º, 90 º, 135 º,
180 º,225 º, 270 º, and 315 º| The main power quality should be that of a
typical commercial or hospital environment.
0% UT; 1 cycle and 70% UT;25/30 cycle Single phase: 0º| 0% UT; 1 cycle and 70%
UT;25/30 cycle Single phase: 0º
Voltage interruptions IEC 61000-4-11| 0% UT;250/300 cycle| 0% UT;250/300 cycle
Power frequency (50/60Hz) magnetic field IEC 61000-4-8| 30 A/m 50Hz/60Hz| 30
A/m 50Hz/60Hz| Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial or hospital
environment.
Note: UT is the a.c. mains voltage prior to application of the test
level.
Conducted RF
IEC 61000-4-6
Radiated RF IEC 61000-4-3
| 150kHz to
80MHz:
3Vrms
6 Vrms (in ISM and amateur radio bands) 80% Am at 1kHz
10V/m
80% Am at 1kHz
| 150kHz to
80MHz:
3Vrms
6 Vrms (in ISM and amateur radio bands) 80% Am at 1kHz
10V/m
80% Am at 1kHz
| Portable and mobile RF communications equipment should be used no closer to
any part of the device including cables, than the recommended separation
distance calculated from the equation applicable to the frequency of the
transmitter. Recommended separation distance:
d= 0.35√
d= 1.2√ 80 MHz le 800 MHz
d= 1.2√ 800 MHz le 2.7 GHz d= 2.3√
Where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer and d is the recommended separation
distance in meters (m). Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey, should be less than the
compliance level in each frequency range. Interference may occur in the
vicinity of equipment marked with the following symbol:
Note 1: 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects,
and people.
a. Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and
FM radio broadcasts, and TV broadcasts cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the device is used exceeds
the applicable RF compliance level above, the device should be observed to
verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as re-orienting or relocating the device.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less
than 3 V/m.
c. Calibration for current injection clamps shall be performed in a 150 Ω
system.
d. If the frequency stepping skips over an ISM or amateur band, as applicable,
an additional test frequency shall be used in the ISM or amateur radio band.
This applies to each ISM and amateur radio band within the specified frequency
range.
e. Applicable to ME EQUIPMENT and ME SYSTEMS with RATED input current less
than or equal to 16 A / phase and ME EQUIPMENT and ME SYSTEMS with RATED input
current greater than 16 A / phase.
Recommended separation distances between portable and mobile RF communications
equipment and the device
The device is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the
device can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment
(transmitters) and the device as recommended below, according to the maximum
output power of the communications equipment.
The rated maximum output power of the transmitter (W)| Separation distance
according to the frequency of the transmitter(m)
150 kHz to 80 MHz (out ISM and amateur radio bands) d= 1.2 √ P| 150 kHz to 80
MHz (out ISM and amateur radio bands)d= 0.6 √ P| 80 MHz to 800 MHz d= 1.2 √ P|
800 MHz to 2.7 GHz d= 2.3 √ P
0,01| 0,12| 0,06| 0,12| 0,23
0,1| 0,38| 0,19| 0,38| 0,73
1| 1,2| 0,6| 1,2| 2,3
10| 3,8| 1,9| 3,8| 7,3
100| 12| 6| 12| 23
For transmitters rated at a maximum output power not listed above, the
recommended separation distance in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W)according to the
transmitter manufacture Note 1: At 80 MHz and 800 MHz, the separation distance
for the higher frequency range applies.
Note 2: The ISM (industrial, scientific and medical) bands between 0,15 MHz and 80 MHz are 6,765 MHz to 6,795 MHz; 13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70 MHz. The amateur radio bands between 0,15 MHz and 80 MHz are 1,8 MHz to 2,0 MHz, 3,5 MHz to 4,0 MHz, 5,3 MHz
to 5,4 MHz, 7 MHz to 7,3 MHz, 10,1 MHz to 10,15 MHz, 14 MHz to 14,2 MHz, 18,07 MHz to 18,17
MHz,21,0 MHz to 21,4 MHz, 24,89 MHz to 24,99 MHz, 28,0 MHz to 29,7 MHz and 50,0 MHz to 54,0 MHz.
Note 3: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.
Radiated RF| Test Frequency (MHz)| Band a) (MHz)| Service a)| Modulation b)|
Modu- lation b) (W)| Distance (m)| Immunity Test Level(V/m)
IEC61000-4-3
(Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment
| 385| 380-390| TETRA 400| Pulse modulation b) 18 Hz| 1,8| 0,3| 27
450| 380-390| GMRS 460, FRS 460| FM c) ± 5 kHz deviation 1 kHz sine| 2| 0,3|
28
710 745 780| 704-787| LTE BANDI 13, 17| Pulse modulation b) 217 Hz| 0,2| 0,3|
9
810 870 930| 800-960| GSM 800/900, TETRA 800, İDEN 820, LTE Band 5| Pulse
modulation b) 18 Hz| 2| 0,3| 28
1720 1845 1970| 1700-1990| GSM 1800, CDMA 1900, GSM 1900, DECT, LTE
Bant1,3,4,25, UMTS| Pulse modulation b) 217 Hz| 2| 0,3| 28
2450| 2400-2570| Bluetooth,WLAN 802.11b/g/n, RFID 2450, LTE Band 7| Pulse
modulation b) 217 Hz| 2| 0,3| 28
5240 5240 5785| 5100-5800| WLAN 802.11a/n| Pulse modulation b) 217 Hz| 0,2|
0,3| 9
Note: If necessary to achieve the IMMUNITY TEST LEVEL, the distance between
the transmitting antenna and the ME EQUIPMENT or ME SYSTEM may be reduced to 1
m. The 1 m test distance is permitted by IEC 61000-4-3.
a. For some services, only the uplink frequencies are included.
b. The carrier shall be modulated using a 50 % duty cycle square wave signal.
c. As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be
used because while it does not represent actual modulation, it would be the
worst case.
The MANUFACTURER should consider reducing the minimum separation distance,
based on RISK MANAGEMENT, and using higher IMMUNITY TEST LEVELS that are
appropriate for the reduced minimum separation distance. Minimum separation
distances for higher IMMUNITY TEST LEVELS shall be calculated using the
following equation:
E= 6 / d √ P
Where P is the maximum power in W, d is the minimum separation distance in m,
and E is the IMMUNITY TEST LEVEL in V/m.
GUARANTEE TERMS & CONDITIONS
- The guarantee period commences from the date of delivery of the product to the client.
- In case of malfunctioning within the warranty period, repair time is added to the warranty period. The maximum repairing period is 30 (Thirty) working days.
- If you have any objections regarding the device, please return the device to the company from where you bought the device or to the manufacturer by prepaid cargo along with the sales certificate and the guarantee certificate.
- Problems appeared due to wrong or misuse of the device are not covered under warranty.
- Opening the device by an unauthorized person will cause the device to remain outside of the scope of the warranty.
- The guarantee for the device will only be valid once the consumer accepts after reading all terms and conditions mentioned here.
- Consumable parts are not covered under warranty.
CERTIFICATES
- CE SERTİFİKASI
- ISO 9001 BELGESİ
- ISO 13485 BELGESİ
MANUFACTURER
ELMASLAR A.S.
ASO 1. OSB BABURSAH CD. NO: 17 SINCAN
ANKARA / TURKEY
Tel: +90 312 394 80 01 Fax: +90 312 394 80 04
info@elmaslarmedikal.com.tr
www.elmaslarmedikal.com.tr
REV 05, 01.09.2020
References
- Elmaslar Medikal A.Ş. - Hayat Sağlık ile Başlar Sevdiklerinizle Devam Eder
- Elmaslar Medikal A.Ş. - Hayat Sağlık ile Başlar Sevdiklerinizle Devam Eder
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>