Riester e-scope, e-xam Diagnostic Instruments Instructions

June 3, 2024
Riester

e-scope, e-xam Diagnostic Instruments

e-scope®

e-xam

Instructions

Diagnostic Instruments

Contents

  1. Important information to take note of before taking the product into operation
  2. Battery handles and start-up
  3. Otoscope and accessories
  4. Ophthalmoscope and accessories
  5. Maintenance
  6. Notes
  7. EMC requirements

1. Important information to take note of before taking the product into operation You have acquired a valuable Riester diagnostic set manufactured in compliance with  Directive 93/42/EEC for medical products and subject to continuous stringent quality control, whose excellent quality will ensure reliable diagnoses. Please read these  Operating Instructions carefully prior to startup and keep in a safe place. Should you  have any queries, please contact the Company or your Riester Agent who will be plea sed to assist you. For addresses see last page of these Operating Instructions. The  address of your authorized Riester Agent will be supplied to you on request. Please  note that any instruments described in these Operating Instructions are only suited for application by trained operators. Please also note that correct and safe operation  of instruments will only be guaranteed when Riester instruments and accessories  are used throughout.

Warning:

Please note that the proper and safe functioning of our instruments is only guaran teed if both the instruments and their accessories are exclusively from Riester. The use of other accessories may result in increased electromagnetic emissions or  reduced electromagnetic immunity of the device and may lead to incorrect operation.

Caution/Contraindications

– There may be a risk of ignition of gases when the instrument is used in the presence  of flammable mixtures or mixtures of pharmaceuticals.

– The instrument heads and battery handles must never be placed in liquids.

– In the case of a longer eye examination with the ophthalmoscope, the intensive  exposure to light can damage the retina.

– The product and the ear specula are non-sterile. Do not use on injured tissue.

– Use new or sanitized specula to limit the risk of cross-contamination.

– The disposal of used ear specula must occur in accordance with current medical practices or local regulations regarding the disposal of infectious, biological medical waste.

– Only use accessories/consumables from Riester or Riester-approved accessories/ consumables.

– Cleaning frequency and sequence must comply with regulations on the cleaning of  non-sterile products in their respective facility. Cleaning/disinfection instructions in  the operating manual must be observed.

– The product may only be used by trained personnel.

Safety instructions:

Manufacturer
CE marking
Temperature limits in °C for storage and transport
Temperature limits in °F for storage and transport
Relative humidity
Fragile, handle with care
Store in a dry place
„Green Dot“ (country-specific)
Warning, this symbol indicates a potentially dangerous situation.

 Device of protection class II
Application part type B
For single use only

Attention: Used electrical and electronic equipment should not be treated as normal household waste but should be disposed of separately in accordance with national or EU directives
Batch code
Serial number
Please observe the operating instructions

2. Battery handles and start-up

2.1. Purpose

Riester battery handles described in these Instructions for Use supply

the instrument heads with power (the lamps are included in appropriate instrument heads), also serving as a bracket.
2.2. Readiness for operation

(insertion and removal of batteries)
Turn off instrument head from handle in counter-clockwise direction. Insert two commercial type ”AA” Mignon alkaline batteries of 1.5 V (IEC standard reference LR6) into the case of the handle with the plus poles towards the upper section of the handle.

Warning:

  • Should the unit not be used for an extended period of time or whilst travelling, remove batteries from handle.
  • Insert new batteries when light intensity of the unit is reduced, thus affecting examination.
  • For maximum light yield it is recommended to always insert new high-quality batteries on replacement.
  • Ensure that no fluid or condensation penetrates into the handle.

Disposal:
Please note that batteries are subject to separate disposal. For details ask your local authority and/or your environmental officer.

2.3. Attachment of instrument heads
Turn instrument head in clockwise direction on to the handle.

2.4. Starting and stopping
When pushing the slide up, the unit is switched on, when pushing it down, the unit is off.

2.5 Instructions for care

General information

Cleaning and disinfection of the medical devices serves to protect the patient, the user and third parties and to preserve the value of the medical devices. Due to the product design and the materials used, no defined limit can be specified for the maximum number of reprocessing cycles that can be carried out. The life span of the medical devices is determined by their function and by gentle handling of the devices. Defective products must undergo the reprocessing procedure described before being returned for repair.

Cleaning and disinfection

The battery handles can be cleaned externally with a moist cloth until visually clean. Wipe-disinfection as specified by the disinfectant manufacturer. Only disinfectants with proven efficacy should be used, taking into account the national requirements. After disinfection, wipe the instrument down with a moist cloth to remove possible disinfectant residues.

PLEASE NOTE!

Never immerse the handles in liquids! Take care to ensure that no liquids get inside the casing! This item is not approved for automated reprocessing and sterilization. These procedures cause irreparable damage!

3. Otoscope and accessories

3.1. Purpose

Riester otoscopes described in these Instructions for Use have been produced for lighting and examination of the auditory canal, combined with a Riester ear speculum.

3.2. Insertion and removal of ear speculum

Position the selected speculum on the chromium plated metal socket of the otoscope. Turn speculum to the right until a resistance is felt. The size of the speculum is marked on the reverse.

3.3. Swivel lens for magnification

The swivel lens is fixed to the device and can be swivelled 360°.

3.4. Insertion of external instruments into the ear

If you wish to insert external instruments into the ear (e.g. tweezers), you have to rotate the swivel lens (approx. 3-fold magnification) located on the otoscope head by 180°.

3.5. Pneumatic test

In order to perform a pneumatic test (= examination of the ear drum), you will require a bulb which is not included in the normal scope of supply but may be ordered separately (see Spare parts and accessories). Take metal connector which is not included in the normal scope of supply but may be ordered separately (see Spare parts and accessories) and insert in recess provided on the side of the otoscope head. Attach hose of bulb to connector. Carefully introduce the required air volume into the auditory canal.

3.6. Replacement of lamp

e-scope® otoscope with direct illumination Remove the speculum socket by turning it to the left with your thumb and index finger until it stops. Pull the speculum socket forward to remove it. Unscrew the bulb counterclockwise. Screw the new bulb in clockwise and reattach the speculum socket.

e-scope® otoscope with fiber optics

Unscrew the instrument head from the battery handle. The LED/bulb is located in the lower part of the instrument head. Pull the bulb out of the instrument head using your thumb and index finger or a suitable tool. When replacing an LED with a bulb, the optionally available adapter is additionally required; when replacing a bulb with an LED, the adapter must first be removed from the bulb unit. Firmly insert the new LED/bulb.

3.7 Instructions for care

General information

Cleaning and disinfection of the medical devices serves to protect the patient, the user and third parties and to preserve the value of the medical devices. Due to the product design and the materials used, no defined limit can be specified for the maximum number of reprocessing cycles that can be carried out. The life span of the medical devices is determined by their function and by gentle handling of the devices. Defective products must undergo the reprocessing procedure described before being returned for repair.

Cleaning and disinfection

The otoscope can be cleaned externally with a moist cloth until visually clean. Wipe-disinfection as specified by the disinfectant manufacturer. Only disinfectants with proven efficacy should be used, taking into account the national requirements. After disinfection, wipe the instrument down with a moist cloth to remove possible disinfectant residues.

PLEASE NOTE!

Never immerse the otoscope in liquids! Take care to ensure that no liquids get inside the casing! This item is not approved for automated reprocessing and sterilization. These procedures cause irreparable damage!

Sterilization
a) Reusable ear specula
The ear specula can be sterilized in the steam sterilizer at 134°C with 10 minutes hold time.
Single use
ATTENTION: Repeated use could cause infection

3.8. Spare parts and accessories

Reusable ear specula

  • 2 mm Pack of 10 St. No.: 10775
  • 2.5 mm Pack of 10 St. No.: 10779
  • 3 mm Pack of 10 St. No.: 10783
  • 4 mm Pack of 10 St. No.: 10789
  • 5 mm Pack of 10 St. No.: 10795

Reusable ear specula

  • 2 mm Pack of 100 St. No.: 14061-532
    Pack of 500 St. No.: 14062-532
    Pack of 1.000 St. No.: 14063-532

  • 2.5 mm Pack of 100 St. No.: 14061-531
    Pack of 500 St. No.: 14062-531
    Pack of 1.000 St. No.: 14063-531

  • 3 mm Pack of 100 St. No.: 14061-533
    Pack of 500 St. No.: 14062-533
    Pack of 1.000 St. No.: 14063-533

  • 4 mm Pack of 100 St. No.: 14061-534
    Pack of 500 St. No.: 14062-534
    Pack of 1.000 St. No.: 14063-534

  • 5 mm Pack of 100 St. No.: 14061-535
    Pack of 500 St. No.: 14062-535
    Pack of 1.000 St. No.: 14063-535

Replacement lamps for e-scope® otoscope with direct illumination
Vacuum, 2.7 V, pack of 6 No.: 10488
XL, 2.5 V, pack of 6 No.: 10489

for e-scope® F.O. Otoscope
XL 2.5 V, Packung à 6 Stück No.: 10600
LED 3.7 V No.: 14041

Technical data of the lamp for e-scope® otoscope with direct illumination
Vacuum, 2.5 V 300 mA mean life span 15 h
XL, 2.5 V 750 mA mean life span 16.5 h

for e-scope® F.O. Otoscope
XL 2.5 V 750 mA mean life span 15 h
LED 3.7 V 52 mA mean life span 20.000 h

Other spare parts
No.: 10960 Bulb for pneumatic test
No.: 10961 Connector for pneumatic test

4. Ophthalmoscope and accessories

4.1. Purpose

Riester ophthalmoscopes described in these Instructions for Use have been designed for the examination of the eye and its background. The examination lamp e-xam was produced for the purpose of examining body orifices. Furthermore, it can also be used for pupil reaction examination. (Photobiological test report EN 62471:2008)

ATTENTION!

Because prolonged intense exposure to light can damage the retina, the use of the eye exam device should not be unnecessarily prolonged, and the brightness setting should not be set higher than needed for a clear representation of the target structures.

The irradiation dose of the photochemical exposure to the retina is the product of irradiance and duration of irradiation. If the irradiance is reduced by half, the irradiation time may be twice as long to reach the maximum limit.

Although no acute optical radiation hazards have been identified for direct or indirect ophthalmoscopes, it is recommended that the intensity of light directed into the patient‘s eye be reduced to the minimum required for examination / diagnosis. Infants / children, aphasics and people with eye diseases are at a higher risk. The risk may be increased if the patient has already been examined with this or another ophthalmological instrument during the last 24 hours. This is especially true when the eye has been exposed to retinal photography.

The light of this instrument may be harmful. The risk of eye damage increases with the duration of irradiation. An irradiation period with this instrument at maximum intensity of longer than >5 min. exceeds the guideline value for hazards.
This instrument does not pose a photobiological hazard according to DIN EN 62471 but still features a safety shutdown after 2/3 minutes.

4.2. Lens wheel and correcting lenses

The correcting lenses may be adjusted on the lens wheel. The following correcting lenses are available:

D+ 1 | 2 | 3 | 4 | 6 | 8 | 10 | 15 | 20
D- 1 | 2 | 3 | 4 | 6 | 8 | 10 | 15 | 20

Readings will be displayed on a lit panel. Plus values are displayed in black digits, minus values in red digits.

4.3. Diaphragms and filters

The following apertures and/or filters may be selected by the aperture and filter wheel:

Aperture – Function

Semi circle: For examinations with turbid lenses.
Small circle: For reduction of reflexes of small pupils.
Large circle: For standard fundus examination.
Fixation star: For definition of central and eccentric fixation.
Red-free filter: To increase contrast for assessment of (green filter) changes in fine vessels, i.e. retinal hemorrhages.
Blue filter: for improved recognition of vascular abnormalities or bleeding, for fluorescence ophthalmology.

4.4. Replacement of lamp

e-scope® ophthalmoscope

Remove the instrument head from the battery handle. The LED/bulb is located in the lower part of the instrument head. Remove the bulb from the instrument head using your thumb and index finger or a suitable tool. When replacing an LED with a bulb, the optionally available adapter is additionally required; when replacing a bulb with an LED, the adapter must first be removed from the bulb unit. Firmly insert the new LED/bulb.

CAUTION:

The pin of the bulb has to be inserted into the guide slot on the adapter and the adapter into the guide slot on the instrument head.

e-xam
Remove the instrument head from the battery grip. The XL lamp or LED is located in the lamp head.
Turn white insulation counter clockwise. Remove insulation with contact. Lamp will fall out. Insert new lamp, turn contact with insulation clockwise.

4.5 Technical data for the ophthalmoscope lamp

XL 2.5 V 750 mA avg. life 16.5 h
LED 3.7 V 38 mA avg. life 20.000 h

Technical data for the e-xam lamp
XL 2.5 V 750 mA avg. life 16.5 h
LED 2.5 V 120 mA 5.000 – 5.500 Kelvin, CRI 72 avg. life 20.000 h

4.6 Instructions for care

General information

Cleaning and disinfection of the medical devices serves to protect the patient, the user and third parties and to preserve the value of the medical devices. Due to the product design and the materials used, no defined limit can be specified for the maximum number of reprocessing cycles that can be carried out. The life span of the medical devices is determined by their function and by gentle handling of the devices. Defective products must undergo the reprocessing procedure described before being returned for repair.

Cleaning and disinfection

The ophthalmoscope can be cleaned externally with a moist cloth until visually clean. Wipe-disinfection as specified by the disinfectant manufacturer. Only disinfectants with proven efficacy should be used, taking into account the national requirements. After disinfection, wipe the instrument down with a moist cloth to remove possible disinfectant residues.

PLEASE NOTE!

Never immerse the ophthalmoscope in liquids! Take care to ensure that no liquids get inside the casing! This item is not approved for automated reprocessing and sterilization. These procedures cause irreparable damage!

4.7 Spare parts Replacement lamps

for e-scope Ophthalmoscope
XL 2.5 V, pack of 6, Art.-No.: 10605
LED 3.7 V, Art.-No.: 14051
https://www.riester.de/productdetails/d/e-scoper-pocket-instrments/e-scoper- otoscopes/

for e-xam

XL 2,5 V, Pack of 6, Art.-No.: 11178
LED 2,5 V Art.-No.: 12320
https://www.riester.de/en/productdetails/d/penlights/e-xam-penlight/

5. Maintenance

These instruments and their accessories do not require any specific maintenance. Should an instrument have to be examined for any specific reason whatsoever, please return it to the Company or an authorized Riester dealer in your area. Addresses to be supplied on request.

6. Notes

Ambient temperature: 0 ° to +40 ° C
Relative Humidity: 30 % to 70 % noncondensing
Storage location: -10° to +55°
Relative Humidity: 10 % to 95 %

CAUTION:

There is possibly a risk of ignition if the equipment is operated in the presence of flammable mixtures of substances with air or with oxygen, nitrous oxide and anesthetic gases. Safety information according to the international safety standard IEC 60601-1 Electrical safety of medical devices: Opening of the handle in patient vicinity and simultaneously touching the batteries and patient is not allowed.

7. Electromagnetic compatibility

Accompanying documents according to IEC 60601-1-2, 2014, Ed. 4.0

Attention:

Medical electrical equipment is subject to special precautions regarding electromagnetic compatibility (EMC).
Portable and mobile radio frequency communication devices can affect medical electrical equipment. The ME device is for operation in an electromagnetic environment for home health care and intended for professional facilities such as industrial areas and hospitals.
The user of the device should ensure that it is operated within such an environment.

Warning:

The ME device may not be stacked, arranged or used directly next to or with other devices. When operation is required to be close to or stacked with other devices, the ME device and the other ME devices must be observed in order to ensure proper operation within this arrangement. This ME device is intended for use by medical professionals only. This device may cause radio interference or interfere with the operation of nearby devices. It may become necessary to take appropriate corrective measures, such as redirecting or rearranging the ME device or shield.
The rated ME device does not exhibit any basic performance features in the sense of EN60601-1, which would present an unacceptable risk to patients, operators or third parties should the power supply fail or malfunction.

Warning:

Portable RF communications equipment (radios) including accessories, such as antenna cables and external antennas, should not be used in closer proximity than 30 cm (12 inches) to parts and cables of the e-scope® instrument head with hand grips specified by the manufacturer. Failure to comply may result in a reduction in the device‘s performance features.

Guidance  and  manufacturer’s  declaration  -­ electromagnetic  emission


The  e-­scope®  instrument  is  intended  for  use  in  the  electromagnetic environment  specified  below.  The   customer  or  the  user  of  the e-­scope®  should  ensure  that  it  is  used  in  such  an  environment

Emission  test

|

Compliance

|

Electromagnetic  environment  – guidance

RF  emissions
HF-­ emissions pursuant  to CISPR  11

|

Group  1

|

The  e-­scope® uses  RF energy exclusively  for  an  internal  function. Therefore his  RF  transmission  is  very  low  and  it  is unlikely  to  be adjacent electronic  devices  are  disturbed.

RF  emissions
HF-­ emissions pursuant  to CISPR  11

|

Class  B

|

The  e-­scope®  is  intended  for  use  in  all   establishments,  including residential  areas  and   those directly  connected  to  a  public  supply network   that  also  supplies  buildings  used  for  residential   purposes.

Emissions  of harmonics IEC 61000-­3-­2

|

Not applicable

Emissions  of voltage fluctuations, flicker

IEC  61000-­3-­3

|

Not applicable

Guidance  and  manufacture’s  declaration  – electromagnetic  immunity


The  e-­scope®  instrument  is  intended  for  use  in  the  electromagnetic environment  specified  below.  The  customer  or  the   user  of  the e-­scope®  should  ensure  that  it  is  used  in  such  an  environment

Immunity  testing

|

IEC  60601  test  level

|

Compliance

|

Electromagnetic environment  -­ Instructions

Electrostatic discharge   (ESD)
IEC  61000-­4-­2

|

Con: ±8  kV
Air: ±2,4,8,15  kV

|

Con: ±8  kV

Air: ±2,4,8,15  kV

|

Floors  should  be  wood,  concrete  or   ceramic  tile.  If floor  are covered  with  synthetic  material, the relative  humidity  should  be  at least  30%.

Fast  transient electrical   disturbances  / bursts IEC   61000-­4-­4

|

5/50  ns,  100  kHz  ,  ±2  kV

|

Not  applicable

|

The  quality  of  the  supply voltage   should  be  that  of  a typical business   or  hospital environment.

Surge voltage

IEC  61000-­4-­5

|

±  0.5  kV  voltage

Phase-­to-­phase conductor
±  2  kV  voltage
Line-­to-­earth

|

Not applicable

|

The quality of the supply  voltage should be that  of  a  typical  business or hospital  environment.

Voltage  dips,  short-­term   interruptions  and

fluctuations  in  the  supply   voltage  acc.  to

IEC  61000-­4-­11

|

<0%  UT  0.5  period  at  0.45,  90,  135,   180,  225,  270  and  315 degrees

0%  UT  1  period  and  70%  UT  25/30   periods  single-­phase:  at  0 degrees   (50/60  Hz)

|

Not applicable

|

The  quality  of  the  supply  voltage   should  be  that  of  a  typical business   or  hospital  environment.

Magnetic  field  with

efficiency-­rated

frequencies

IEC  61000-­4-­8

|

30A/m

50/60  Hz

|

30A/m

50/60  Hz

|

Mains  frequency  magnetic  fields   should  be  at  a  level  characteristic of   a  typical  location  in  a  typical   commercial  hospital  environment.

NOTE UT is the AC source. Mains voltage before the application of the test level.

Directives  and  manufacturer’s  declaration  -­ Electromagnetic immunity


The e-­scope®  instruments  are  intended  for  use  in  the  electromagnetic environment  specified  below.  The  customer  or   user  of  the  e-­scope® should  ensure  that  it  is  used  in  such  an  environment.

Immunity  testing

|

IEC  60601  test  level

|

Compliance

|

Electromagnetic  environment  -­ Instructions

Guided  RF
Diturbances acc. to

IEC61000-­4-­6

|

3 Vrms

0,5  MHz  bis  80MHz 6  V  in  ISM  frequency   bands  between  0.15   MHz and  80  MHz

80%  AM  at 1  kHz

|

Not  applicable

|

Portable  and  mobile  RF  communications  equipment   should  not  be  used closer  to  any  part  of  the  non-­contact   e-­scope®,  including  the cables,  than  the  recommended   distance,  which  is  calculated  using  the equation   applicable  to  the  transmitter  frequency.

Recommended  separation  distance

d=  1.2√P  80  MHz  to  800  MHz

d=  2.3√P  800  MHz  to  2.7  GHz

Where  P is  the  maximum  output  power  rating  of  the transmitter  in watts  (W)  according  to  the  transmitter   manufacturer  and  d is  the recommended  separation   distance  in  metres  (m).

Field  strengths  from  fixed  RF  transmitters,  as  determined   by  an electromagnetic  site  survey,a should  be  less  than   the  compliance level in  each  frequency  range.b

Interference  may  occur  in  the  vicinity  of  equipment   marked  with  the following  symbol:

|

|

Radiated  RF

IEC  61000-­4-­3

Proximity  fields   from  RF  wireless   communications   equipment

|

3  V/m

80  MHz  to  2.7  GHz

380  -­ 390  MHz

27  V/m;;  PM  50%;;  18  Hz 430  -­ 470  MHz

28  V/m;;  (FM  ±5  kHz,  1   kHz  sine)

PM;;  18  Hz11

704  -­ 787  MHz

9  V/m;;  PM  50%;;  217  Hz 800  -­ 960  MHz

28  V/m;;  PM  50%;;  18  Hz 1700  -­ 1990  MHz

28  V/m;;  PM  50%;;  217   Hz

2400  -­ 2570  MHz

28  V/m;;  PM  50%;;  217   Hz

5100  -­ 5800  MHz

9  V/m;;  PM  50%;;  217  Hz

|

10  V/m

27  V/m

28  V/m

9  V/m

28  V/m

28  V/m

9  V/m

NOTE  1 At  80  MHz  and  800  MHz,  the  higher  frequency  range applies.

NOTE  2 These  guidelines  may  not  apply  in  all  situations. Electromagnetic  propagation  is  affected  by  absorption  and  reflection from  structures,  objects  and  people.

a. field  strengths  of  fixed  transmitters,  such.  B.  Base  stations  for radio  (cellular  / cordless)  telephones  and  land   mobile  radios, amateur  radio,  AM  and  FM broadcasting  and  television  broadcasting  can not  be  predicted   theoretically accurately.  To  assess  the electromagnetic  environment  due  to  fixed  RF  transmitters, an electromagnetic  examination  should  be  considered.  If  the  measured field strength  at  the  location  where  the  e-­ scope®  is  used  exceeds the  above  RF compliance  level,  the  e-­scope®  should  be  observed  to verify  normal   operation.  If abnormal  performance  is  observed, additional  measures  may  be  required,  such  as reorienting  or   shifting the  e-­scope®.

b With  a  frequency  range  over  150  kHz  to  80  MHz,  the  field strengths  should  be  less  than  3  V/m.

Recommended  distances  between  portable  and  mobile  RF  communications equipment and  the e-­scope®

The  e-­scope®  is  intended  for  use  in  an  electromagnetic  environment in  which  RF emissions  are  controlled.  The  customer  or  user  of  the e-­scope®  can  help  to  avoid electromagnetic interference  by  observing the  minimum  distance  between  portable and mobile  RF communications equipment  (transmitters)  and  the  e-­scope®  in  accordance with  the maximum output  power  of  the  communication  equipment.

References