Riester e-scope Diagnostic Instruments Instructions
- June 3, 2024
- Riester
Table of Contents
Riester e-scope Diagnostic Instruments
Important Information
Important information to take note of before taking the product into
operation
You have acquired a valuable Riester diagnostic set manufactured in compliance
with Directive 93/42/EEC for medical products and subject to continuous
stringent quality control, whose excellent quality will ensure reliable
diagnoses. Please read these Operating Instructions carefully prior to startup
and keep in a safe place. Should you have any queries, please contact the
Company or your Riester Agent who will be pleased to assist you. For addresses
see last page of these Operating Instructions. The address of your authorized
Riester Agent will be supplied to you on request. Please note that any
instruments described in these Operating Instructions are only suited for
application by trained operators. Please also note that the correct and safe
operation of instruments will only be guaranteed when Riester instruments and
accessories are used throughout.
Warning:
Please note that the proper and safe functioning of our instruments is only
guaranteed if both the instruments and their accessories are exclusively from
Riester.
The use of other accessories may result in increased electromagnetic emissions
or reduced electromagnetic immunity of the device and may lead to incorrect
operation.
Caution/Contraindications
- There may be a risk of ignition of gases when the instrument is used in the presence of flammable mixtures or mixtures of pharmaceuticals.
- The instrument heads and battery handles must never be placed in liquids.
- In the case of a longer eye examination with the ophthalmoscope, intensive exposure to light can damage the retina.
- The product and the ear specula are non-sterile. Do not use on injured tissue.
- Use new or sanitized specula to limit the risk of cross-contamination.
- The disposal of used ear specula must occur in accordance with current medical practices or local regulations regarding the disposal of infectious, biological me-dical waste.
- Only use accessories/consumables from Riester or Riester-approved accessories/consumables.
- Cleaning frequency and sequence must comply with regulations on the cleaning of non-sterile products in their respective facility. Cleaning/disinfection instructions in the operating manual must be observed.
- The product may only be used by trained personnel.
Safety instructions
- Manufacturer
- CE marking
- Temperature limits in °C for storage and transport
- Temperature limits in °F for storage and transport
- Relative humidity
- Fragile, handle with care
- Store in a dry place
- „Green Dot“ (country-specific)
- Warning, this symbol indicates a potentially dangerous situation.
- Device of protection class II
- Application part type B
- For single use only
- Attention: Used electrical and electronic equipment should not be treated as normal household waste but should be disposed of separately in accordance with national or EU directives
- Batch code
- Serial number
- Please observe the operating instructions
Battery handles and start-up
Purpose
Riester battery handles described in these Instructions for Use supply the instrument heads with power (the lamps are included in appropriate instrument heads), also serving as a bracket.
Readiness for operation (insertion and removal of batteries)
Turn off the instrument head from the handle in a counter-clockwise direction.
Insert two commercial-type” AA” Mignon alkaline batteries of 1.5 V (IEC
standard reference LR6) into the case of the handle with the plus poles
towards the upper section of the handle.
Warning:
- Should the unit not be used for an extended period of time or whilst traveling, remove batteries from the handle.
- Insert new batteries when the light intensity of the unit is reduced, thus affecting examination.
- For maximum light yield it is recommended to always insert new high-quality batteries on replacement.
- Ensure that no fluid or condensation penetrates into the handle.
Disposal:
Please note that batteries are subject to separate disposal. For details ask
your local authority and/or your environmental officer.
Attachment of instrument heads
Turn the instrument head in a clockwise direction onto the handle.
Starting and stopping
When pushing the slide up, the unit is switched on, when pushing it down, the
unit is off.
Instructions for care
General information
Cleaning and disinfection of the medical devices serve to protect the patient,
the user, and third parties and to preserve the value of the medical devices.
Due to the product design and the materials used, no defined limit can be
specified for the maximum number of reprocessing cycles that can be carried
out. The life span of the medical devices is determined by their function and
by the gentle handling of the devices. Defective products must undergo the
reprocessing procedure described before being returned for repair.
Cleaning and disinfection
The battery handles can be cleaned externally with a moist cloth until
visually clean. Wipe disinfection as specified by the disinfectant
manufacturer. Only disinfectants with proven efficacy should be used, taking
into account the national requirements. After disinfection, wipe the
instrument down with a moist cloth to remove possible disinfectant residues.
PLEASE NOTE!
Never immerse the handles in liquids! Take care to ensure that no liquids get
inside the casing! This item is not approved for automated reprocessing and
sterilization. These procedures cause irreparable damage!
Otoscope and accessories
Purpose
Riester otoscopes described in these Instructions for Use have been produced
for lighting and examination of the auditory canal, combined with a Riester
ear speculum.
Insertion and removal of ear speculum
Position the selected speculum on the chromium-plated metal socket of the
otoscope. Turn the speculum to the right until resistance is felt. The size of
the speculum is marked on the reverse.
Swivel lens for magnification
The swivel lens is fixed to the device and can be swiveled 360°.
Insertion of external instruments into the ear
If you wish to insert external instruments into the ear (e.g. tweezers), you
have to rotate the swivel lens (approx. 3-fold magnification) located on the
otoscope head by 180°.
Pneumatic test
In order to perform a pneumatic test (= examination of the ear drum), you will
require a bulb that is not included in the normal scope of supply but may be
ordered separately (see Spare parts and accessories). Take a metal connector
that is not included in the normal scope of supply but may be ordered
separately (see Spare parts and accessories) and insert in a recess provided
on the side of the otoscope head. Attach the hose of the bulb to the
connector. Carefully introduce the required air volume into the auditory
canal.
Replacement of lamp
e-scope® otoscope with direct illumination Removes the speculum socket by
turning it to the left with your thumb and index finger until it stops. Pull
the speculum socket forward to remove it. Unscrew the bulb counterclockwise.
Screw the new bulb in clockwise and reattach the speculum socket.
e-scope® otoscope with fiber optics
Unscrew the instrument head from the battery handle. The LED/bulb is located
in the lower part of the instrument head. Pull the bulb out of the instrument
head using your thumb and index finger or a suitable tool. When replacing an
LED with a bulb, the op-tionally available adapter is additionally required;
when replacing a bulb with an LED, the adapter must first be removed from the
bulb unit. Firmly insert the new LED/bulb.
Instructions for care
General information
Cleaning and disinfection of the medical devices serve to protect the patient,
the user, and third parties and to preserve the value of the medical devices.
Due to the product design and the materials used, no defined limit can be
specified for the maximum number of reprocessing cycles that can be carried
out. The life span of medical devices is determined by their function and by
the gentle handling of the devices. Defective products must undergo the
reprocessing procedure described before being returned for repair.
Cleaning and disinfection
The otoscope can be cleaned externally with a moist cloth until visually
clean. Wi-pe-disinfection as specified by the disinfectant manufacturer. Only
disinfectants with proven efficacy should be used, taking into account the
national requirements. After disinfection, wipe the instrument down with a
moist cloth to remove possible disinfectant residues.
PLEASE NOTE!
Never immerse the otoscope in liquids! Take care to ensure that no liquids get
inside the casing! This item is not approved for automated reprocessing and
sterilization. These procedures cause irreparable damage!
Sterilization
Reusable ear specula
The ear specula can be sterilized in the steam sterilizer at 134°C with 10
minute hold time.
Single-use
ATTENTION: Repeated use could cause infection
Spare parts and accessories
Reusable ear specula
• 2 mm | Pack of 10 St. | No.: 10775 |
---|---|---|
• 2.5 mm | Pack of 10 St. | No.: 10779 |
• 4 mm
• 5 mm
| Pack of 10 St. Pack of 10 St.| No.: 10789
No.: 10795
---|---|---
Reusable ear specula
• 2 mm
| ****
Pack of 100 St.
| ****
No.: 14061-532
• 2.5 mm
• 3 mm
• 4 mm
• 5 mm
| Pack of 500 St. Pack of 1.000 St. Pack of 100 St. Pack of 500 St. Pack of 1.000 St. Pack of 100 St. Pack of 500 St. Pack of 1.000 St. Pack of 100 St. Pack of 500 St. Pack of 1.000 St. Pack of 100 St. Pack of 500 St. Pack of 1.000 St.| No.: 14062-532
No.: 14063-532
No.: 14061-531
No.: 14062-531
No.: 14063-531
No.: 14061-533
No.: 14062-533
No.: 14063-533
No.: 14061-534
No.: 14062-534
No.: 14063-534
No.: 14061-535
No.: 14062-535
No.: 14063-535
Replacement lamps for e-scope® otoscope with direct illumination
- Vacuum, 2.7 V, pack of 6 No.: 10488
- XL, 2.5 V, pack of 6 No.: 10489
for e-scope® F.O. Otoscope
- XL 2.5 V, Packung à 6 Stück No.: 10600
- LED 3.7 V No.: 14041
Technical data of the lamp for e-scope® otoscope with direct illumination
- Vacuum, 2.5 V 300 mA mean life span 15 h
- XL, 2.5 V 750 mA mean life span 16.5 h
for e-scope® F.O. Otoscope
- XL 2.5 V 750 mA mean life span 15 h
- LED 3.7 V 52 mA mean life span 20.000 h
Other spare parts
- No.: 10960 Bulb for pneumatic test
- No.: 10961 Connector for pneumatic test
Ophthalmoscope and accessories
Purpose
Riester ophthalmoscopes described in these Instructions for Use have been
designed for the examination of the eye and its background. The examination
lamp exam was produced for the purpose of examining body orifices.
Furthermore, it can also be used for pupil reaction examination.
(Photobiological test report EN 62471:2008).
ATTENTION!
Because prolonged intense exposure to light can damage the retina, the use
of the eye exam device should not be unnecessarily prolonged, and the
brightness setting should not be set higher than needed for a clear
representation of the target structures. The irradiation dose of the
photochemical exposure to the retina is the product of irradiance and duration
of irradiation. If the irradiance is reduced by half, the irradiation time may
be twice as long to reach the maximum limit.
Although no acute optical radiation hazards have been identified for direct or indirect ophthalmoscopes, it is recommended that the intensity of light directed into the patient‘s eye be reduced to the minimum required for examination/diagnosis. Infants/children, aphasics, and people with eye diseases are at a higher risk. The risk may be increased if the patient has already been examined with this or another ophthalmo-logical instrument during the last 24 hours. This is especially true when the eye has been exposed to retinal photography.
The light of this instrument may be harmful. The risk of eye damage increases with the duration of irradiation. An irradiation period with this instrument at the maximum intensity of longer than >5 min. exceeds the guideline value for hazards. This instrument does not pose a photobiological hazard according to DIN EN 62471 but still features a safety shutdown after 2/3 minutes.
Lens wheel and correcting lenses
The correcting lenses may be adjusted on the lens wheel. The following
correcting lenses are available:
- D+ 1 | 2 | 3 | 4 | 6 | 8 | 10 | 15 | 20
- D- 1 | 2 | 3 | 4 | 6 | 8 | 10 | 15 | 20
Readings will be displayed on a lit panel. Plus values are displayed in black digits, minus values in red digits.
Diaphragms and filters
The following apertures and/or filters may be selected by the aperture and
filter wheel:
Aperture/Function
- Semi circle: For examinations with turbid lenses.
- Small circle: For reduction of reflexes of small pupils.
- Large circle: For standard fundus examination.
- Fixation star: For definition of central and eccentric fixation.
- Red-free filter: To increase contrast for assessment of (green filter) changes in fine vessels, i.e. retinal hemorrhages.
- Blue filter: for improved recognition of vascular abnormalities or bleeding, for fluorescence ophthalmology.
Replacement of lamp
e-scope® ophthalmoscope
Remove the instrument head from the battery handle. The LED/bulb is located
in the lower part of the instrument head. Remove the bulb from the instrument
head using your thumb and index finger or a suitable tool. When replacing an
LED with a bulb, the optionally available adapter is additionally required;
when replacing a bulb with an LED, the adapter must first be removed from the
bulb unit. Firmly insert the new LED/bulb.
CAUTION:
The pin of the bulb has to be inserted into the guide slot on the adapter and
the adap-ter into the guide slot on the instrument head.
exam
Remove the instrument head from the battery grip. The XL lamp or LED is
located in the lamp head.
Turn the white insulation counterclockwise. Remove insulation with contact.
The lamp will fall out. Insert a new lamp, and turn contact with insulation
clockwise.
Technical data for the ophthalmoscope lamp
- XL 2.5 V 750 mA avg. life 16.5 h
- LED 3.7 V 38 mA avg. life 20.000 h
Technical data for the exam lamp
- XL 2.5 V 750 mA avg. life 16.5 h
- LED 2.5 V 120 mA 5.000 – 5.500 Kelvin, CRI 72 avg. life 20.000 h
Instructions for care
General information
Cleaning and disinfection of the medical devices serve to protect the patient,
the user, and third parties and to preserve the value of the medical devices.
Due to the product design and the materials used, no defined limit can be
specified for the maximum number of reprocessing cycles that can be carried
out. The life span of the medical devices is determined by their function and
by the gentle handling of the devices. Defective products must undergo the
reprocessing procedure described before being returned for repair.
Cleaning and disinfection
The ophthalmoscope can be cleaned externally with a moist cloth until visually
clean. Wipe disinfection as specified by the disinfectant manufacturer. Only
disinfectants with proven efficacy should be used, taking into account the
national requirements. After disinfection, wipe the instrument down with a
moist cloth to remove possible disinfectant residues.
PLEASE NOTE!
Never immerse the ophthalmoscope in liquids! Take care to ensure that no
liquids get inside the casing! This item is not approved for automated
reprocessing and sterilization. These procedures cause irreparable damage!
Spare parts Replacement lamps
for e-scope Ophthalmoscope
- XL 2.5 V, pack of 6, Art.-No.: 10605
- LED 3.7 V, Art.-No.: 14051
https://www.riester.de/productdetails/d/e-scoper-pocket-instrments/e-scoper- otos-copes/
for e-xam
- XL 2,5 V, Pack of 6, Art.-No.: 11178
- LED 2,5 V Art.-No.: 12320
https://www.riester.de/en/productdetails/d/penlights/e-xam-penlight/.
Maintenance
These instruments and their accessories do not require any specific maintenance. Should an instrument have to be examined for any specific reason whatsoever, please return it to the Company or an authorised Riester dealer in your area. Addresses to be supplied on request.
Notes
- Ambient temperature: 0 ° to +40 ° C
- Relative Humidity: 30 % to 70 % non-condensing
- Storage location: -10° to +55°
- Relative Humidity: 10 % to 95 %
CAUTION:
There is possibly a risk of ignition if the equipment is operated in the
presence of flammable mixtures of substances with air or with oxygen, nitrous
oxide, and anesthetic gases. Safety information according to the international
safety standard IEC 60601-1 Electrical safety of medical devices: Opening of
the handle in the patient vicinity and simultaneously touching the batteries
and patient is not allowed.
Electromagnetic compatibility
Accompanying documents according to IEC 60601-1-2, 2014, Ed. 4.0
Attention:
Medical electrical equipment is subject to special precautions regarding
electromagnetic compatibility (EMC). Portable and mobile radio frequency
communication devices can affect medical electrical equipment. The ME device
is for operation in an electromagnetic environment for home health care and is
intended for professional facilities such as industrial areas and hospitals.
The user of the device should ensure that it is operated within such an
environment.
Warning:
The ME device may not be stacked, arranged or used directly next to or with
other devices. When the operation is required to be close to or stacked with
other devices, the ME device and the other ME devices must be observed in
order to ensure proper operation within this arrangement. This ME device is
intended for use by medical professionals only. This device may cause radio
interference or interfere with the operation of nearby devices. It may become
necessary to take appropriate corrective measures, such as redirecting or
rearranging the ME device or shield.
The rated ME device does not exhibit any basic performance features in the sense of EN60601-1, which would present an unacceptable risk to patients, operators or third parties should the power supply fail or malfunction.
Warning:
Portable RF communications equipment (radios) including accessories, such
as antenna cables and external antennas, should not be used in closer
proximity than 30 cm (12 inches) to parts and cables of the e-scope®
instrument head with hand grips specified by the manufacturer. Failure to
comply may result in a reduction in the device‘s performance features.
Guidance and manufacturer’s declaration - electromagnetic emission
The e-scope® instrument is intended for use in the electromagnetic
environment specified below. The customer or the user of the e-scope® should
ensure that it is used in such an environment
Emission test| Compliance| Electromagnetic environment – guidance
RF emissions
HF- emissions pursuant to CISPR 11
|
Group 1
| The e-scope® uses RF energy
exclusively for an internal function. Therefore his RF transmission is very low and it is unlikely to be adjacent
electronic devices are disturbed.
RF emissions
HF- emissions pursuant to CISPR 11
|
Class B
| The e-scope® is intended for use in all establishments, including
residential areas and those directly connected to a public supply network that
also supplies buildings used for residential purposes.
Emissions of harmonics IEC 61000-3-2| Not applicable
Emissions of voltage fluctuations, flicker
IEC 61000-3-3
|
Not applicable
Guidance and manufacture’s declaration – electromagnetic immunity
The e-scope® instrument is intended for use in the electromagnetic
environment specified below. The customer or the user of the e-scope® should
ensure that it is used in such an environment
Immunity testing| IEC 60601 test level| Compliance| Electromagnetic
environment - Instructions
Electrostatic discharge (ESD)
IEC 61000-4-2
| Con: ±8 kV
Air: ±2,4,8,15 kV
| Con: ±8 kV
Air: ±2,4,8,15 kV
| Floors should be wood, concrete or ceramic tile. If floor are covered with
synthetic material, the relative humidity should be at least 30%.
Fast transient electrical disturbances / bursts IEC 61000-4-4| 5/50 ns, 100
kHz , ±2 kV| Not applicable| The quality of the supply voltage should be that
of a typical business or hospital environment.
Surge voltage IEC 61000-4-5| ± 0.5 kV voltage
Phase-to-phase conductor
± 2 kV voltage Line-to-earth
| Not applicable| The quality of the supply voltage should be that of a
typical business or hospital environment.
Voltage dips, short-term interruptions and fluctuations in the supply voltage
acc. to
IEC 61000-4-11
| <0% UT 0.5 period at 0.45, 90, 135,
180, 225, 270 and 315 degrees
0% UT 1 period and 70% UT 25/30 periods single-phase: at 0 degrees (50/60 Hz)
| Not applicable| The quality of the supply voltage should be that of a
typical business or hospital environment.
Magnetic field with efficiency-rated frequencies
IEC 61000-4-8
| 30A/m
50/60 Hz
| 30A/m
50/60 Hz
| Mains frequency magnetic fields should be at a level characteristic of a
typical location in a typical commercial hospital environment.
NOTE UT is the AC source. Mains voltage before the application of the test
level.
Directives and manufacturer’s declaration - Electromagnetic immunity|
---|---
The e-scope® instruments are intended for use in the electromagnetic
environment specified below. The customer or user of the e-scope® should
ensure that it is used in such an environment.|
Immunity testing| IEC 60601 test level| Compliance| Electromagnetic
environment - Instructions|
Guided RF Diturbances acc. to
IEC61000-4-6
|
3 Vrms
0,5 MHz bis 80MHz 6 V in ISM frequency bands between 0.15 MHz and 80 MHz 80% AM at 1 kHz
|
Not applicable
| Portable and mobile RF communications equipment should not be used closer to any part of the non-contact e-scope®, including the cables, than the recommended distance, which is calculated using the equation applicable to the transmitter frequency.
Recommended separation distance
d= 1.2√P 80 MHz to 800 MHz d= 2.3√P 800 MHz to 2.7 GHz
|
| | | Where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer and d is the recommended
separation distance in metres (m).|
| | | Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey,a should be less than the compliance level in each
frequency range.b|
| | |
Interference may occur in the vicinity of equipment marked with the following
symbol:|
| | |
Radiated RF| 3 V/m| 10 V/m| |
IEC 61000-4-3| 80 MHz to 2.7 GHz| | |
Proximity fields from RF wireless communications equipment
|
380 - 390 MHz
27 V/m;; PM 50%;; 18 Hz
430 - 470 MHz
28 V/m;; (FM ±5 kHz, 1
kHz sine) PM;; 18 Hz11
704 - 787 MHz
9 V/m;; PM 50%;; 217 Hz
800 - 960 MHz
28 V/m;; PM 50%;; 18 Hz
1700 - 1990 MHz
28 V/m;; PM 50%;; 217
Hz
2400 - 2570 MHz
28 V/m;; PM 50%;; 217
Hz
5100 - 5800 MHz
9 V/m;; PM 50%;; 217 Hz
|
27 V/m
28 V/m
9 V/m
28 V/m
28 V/m
9 V/m
| |
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
|
a
b
| field strengths of fixed transmitters, such. B. Base stations for radio (cellular / cordless) telephones and land mobile radios, amateur radio, AM and FM broadcasting and television broadcasting can not be predicted theoretically accurately. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic examination should be considered. If the measured field strength at the location where the e- scope® is used exceeds the above RF compliance level, the e-scope® should be observed to verify normal operation. If abnormal performance is observed, additional measures may be required, such as reorienting or shifting the e-scope®.
With a frequency range over 150 kHz to 80 MHz, the field strengths should be less than 3 V/m.
|
Recommended distances between portable and mobile RF communications
equipment and the e-scope®|
---|---
The e-scope® is intended for use in an electromagnetic environment in which
RF emissions are controlled. The customer or user of the e-scope® can help to
avoid electromagnetic interference by observing the minimum distance between
portable and mobile RF communications equipment (transmitters) and the
e-scope® in accordance with the maximum output power of the communication
equipment.
Rated maximum output power of the transmitter (W)
| Separation distance according to the frequency of the transmitter (m)
150 KHz to 80 MHz| 80 MHz to 800 MHz| 800 MHz to 2.7 GHz
0.01| 0.12| 0.12| 0.23| 0,23
0.1| 0.38| 0.38| 0.73| 0,73
1| 1.2| 1.2| 2.3| 2,3
10| 3.8| 3.8| 7.3| 7,3
100| 12| 12| 23| 23
For transmitters with a maximum output power not listed above, the recommended
distance d|
in metres (m) can be estimated using the equation for the transmitter
frequency, where P is the|
maximum output power of the transmitter in watts (W). according to the
transmitter|
manufacturer.|
NOTE 1 At 80 MHz and 800 MHz, the separation distance applies to the higher
frequency|
range.|
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
propagation is|
affected by the absorption and reflection of structures, objects, and people.|
WARRANTY
This product has been manufactured under the strictest quality standards and has undergone a thorough final quality check before leaving our factory. We are therefore pleased to be able to provide a warranty of 2 years from the date of purchase on all defects, which can verifiably be shown to be due to material or manufacturing faults. A warranty claim does not apply in the case of improper handling. All defective parts of the product will be replaced or repaired free of charge within the warranty period. This does not apply to wearing parts.
For r1 shock-proof, we grant an additional warranty of 5 years for the calibration, which is required by CE certification. A warranty claim can only be granted if this Warranty Card has been completed and stamped by the dealer and is enclosed with the product. Please remember that all warranty claims have to be made during the warranty period. We will, of course, be pleased to carry out checks or repairs after the expiry of the warranty period at a charge. You are also welcome to request a provisional cost estimate from us free of charge.
In case of a warranty claim or repair, please return the Riester product with the completed Warranty Card to the following address:
Rudolf Riester GmbH
Dept. Repairs RR Bruckstr. 31
72417 Jungingen
Germany
- Serial number or batch number
- Date, Stamp and signature of the specialist dealer
Rudolf Riester GmbH
Bruckstraße 31 | 72417Jungingen | Germany
Tel.: (+49) 7477-9270-0 | Fax.: (+49) 7477-9270-70
info@riester.de |
www.riester.de.
References
- Rudolf Riester GmbH - Medical devices of the highest standards.
- Rudolf Riester GmbH – Medical devices of the highest standards.
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>