SHINING 3D Aoralscan 3 Intraoral Scanner Device Instructions
- June 3, 2024
- SHINING 3D
Table of Contents
SHINING 3D Aoralscan 3 Intraoral Scanner Device
Disinfection Guide for Scanner Tips
Cautions
- Before EACH use, the scanner tip must be disinfected and (or) sterilized.
- The scanner tip is of medium-danger, and should be sterilized or disinfected at high level.
- Take protective measures when disinfecting the scanner tip in the disinfection liquid.
Operation Steps
1. After intraoral scanning, separate the scanner tip and the Scanner body.
Put the scanner tip in the instrument tray.
2.Clean the scanner tip: Wipe the tip with a piece of alcohol gauze or
CaviWipes.
3. Ensure there is no damage to the scanner tip and the tip has been deaned
thoroughly.
High-level disinfection with ortho-phthalaldehyde
- Wear gloves before performing disinfection operation.
- Separate the scanner tip and the scanner body.
- Pre-clean the scanner tip.
- Put ortho-phthalaldehyde in a cup. The liquid should completely cover the scanner tip (5.5 g/L pH 7.0-8.0, 20-25°C). Note: f the frequency of recycled disinfection is not high, one barrel of the ortho-phthalaldehyde can be used for up to 14 days (the concentration should be higher than 0.3%6).
- The scanner tip should be immersed for 12-30 mins.
- Clean the scanner tip carefully with distiled water
- Dry the scanner tip with lint-free doth thoroughly to ensure it’s totally dry.
The scanner tip cleaning and disinfection process confoms to the Regulation for Disinfection and Sterilization Technique of Dental Instruments WS 506-2016
CautionS
- Before EACH use, the scanner tip must be disinfected and (or) sterilized.
- The scanner tip is of medium-danger, and should be sterilized or disinfected at high level. The scanner tip is consumable. After disinfecting the tip under high temperature and air pressure for 100 times, discard it and use a new one.
Operation Steps
- After intraoral scanning, separate the scanner tip and the scanner body. Put the scanner tip in the instrument tray tor later cleaning.
- Clean the scanner tip: Wipe the tip with a piece of alcohol gauze or CaviWipes to ensure its totaly dry-inside and outside. See Figure 1.
- Fold the gauze into a strip similar to the scanner tip size. See Figure 2.
- Fill the lens with gauze. See Figure 3.
- Seal the tip into the sterilization bag (see Figure 4) and put the bag into the autodave.
- Set autocdave parameters: 134C, relative pressure 205.8 kPa and 4 mins; or 121C, relative pressure 1029 kPa and 30 mins. Implement the step in vacuum.