1075DOC_B Nitrous Oxide Separator
“`html
Product Specifications
- Material: PVC corrosion-proof plastic
- Size: Standard 2-inch no-hub coupling
- Mounting: Near vacuum pump, center of vacuum exhaust inlet no
more than 2 feet above pump bottom
Product Usage Instructions
Installation Instructions
Mounting Bracket Installation:
-
Attach the mounting bracket to the desired location using the
provided screws. -
Ensure the bracket is level using a torpedo level and then
tighten the screws securely.
Nitrous Oxide Separator Installation:
- Secure the separator to the mounting bracket through the
provided holes using a 7/16 socket and socket wrench.
Plumbing Installation:
-
Install adaptors by turning each adaptor 1/4 turn using an
adjustable wrench. -
Attach hoses by placing small hose clamps onto the hoses and
twisting them onto the PVC adaptors. -
Tighten clamps using a flathead screwdriver.
-
Connect the air discharge vent to the outside using the
provided PVC pipe and fittings.
Frequently Asked Questions (FAQ)
Q: Can the nitrous oxide separator be mounted far away from the
vacuum pump?
A: No, the separator should be mounted near the vacuum pump,
with the center of the vacuum exhaust inlet no more than 2 feet
above the bottom of the vacuum pump.
Q: What type of tools/supplies are needed for
installation?
A: You will need a socket wrench, torpedo level, 7/16 socket,
PVC primer & glue, flathead screwdriver, thread sealant tape,
and adjustable wrenches (up to 2).
“`
INSTALLATION INSTRUCTIONS
NITROUS OXIDE SEPARATOR (WET VACUUMS)
TOOLS/SUPPLIES NEEDED
Socket Wrench
Torpedo Level
7/16″ Socket
PVC Primer & Glue
Flathead Screwdriver
Thread Sealant Tape
2″ PVC Pipe & Fittings 1½” or 1″ Flexible Hose
Adjustable Wrench (Up to 2″)
AIR DISCHARGE VENT TO OUTSIDE
VACUUM EXHAUST
INLET
PARTS INCLUDED
Nitrous Oxide Separator Mounting Bracket
Hex Bolt 1/4-20 × 1½” (4) PVC Reducing Bushing
PVC Adaptor 1½” MPT ×
1½” MPT – 1″ FPT
1½” Shank
Small Hose Clamp
PVC Adaptor 1″ MPT × 1″
13/16″ – 1½” (2)
Shank (2)
Large Hose Clamp
1/16″ – 2¼”
OVERVIEW
A standard 2-inch no-hub coupling on air discharge vent eliminates need for
any adapter and allows direct hookup to a standard vent pipe.
Water trap is inside, making plugging virtually impossible.
Entire separator is made of PVC corrosion-proof plastic.
Separator should be mounted near vacuum pump, with center of vacuum exhaust
inlet no more than 2 feet above bottom of vacuum pump.
NOTE: This product should be installed by a competent technician or plumber.
The installation should meet all local health and plumbing codes.This device
requires a separate atmospheric vent for maximum safety and effectiveness.
Compatible with all single or dual water injected vacuum pumps. Parts provided
for unit to be adapted to 1″ or 1½” vacuum exhausts.
DRAIN LINE
15¼”
CLEANOUT PLUG
8″
866-DTE-INFO dentalez.com April, 2024
Page 1 of 4
PN: 1075DOC_B
NITROUS OXIDE SEPARATOR (WET VACUUMS)
INSTALL MOUNTING BRACKET
1. Attach mounting bracket to desired location.
a. If mounting to a wall stud, use two (2) installersupplied screws (FIG 1A).
b. If mounting to a wall with no stud, use three (3) installer-supplied
screws/anchors (FIG 1B).
c. Level mounting bracket using a torpedo level, then tighten screws (FIG 1A &
1B).
NOTE: Separator should be mounted near vacuum pump, with center of vacuum
exhaust inlet no more than 2 feet above bottom of vacuum pump.
STUD TORPEDO
LEVEL
FIG 1A NO STUD
TORPEDO LEVEL
FIG 1B
SCREWS (2) (INSTALLERSUPPLIED) MOUNTING
BRACKET
SCREWS/ ANCHORS (3) (INSTALLER-
SUPPLIED) MOUNTING BRACKET
April, 2024
Page 2 of 4
PN: 1075DOC_B
INSTALLATION INSTRUCTIONS
INSTALL NITROUS OXIDE SEPARATOR
1. Align four (4) through holes on top and bottom of nitrous oxide separator
with four (4) PEM® nuts on mounting bracket (FIG 2).
2. Attach nitrous oxide separator to mounting bracket with four (4) hex bolts
1/4-20 × 1½”, hand tight. Finish tightening by turning a 1/4 to 1/2 turn using
a 7/16″ socket and socket wrench (FIG 3).
FIG 2
MOUNTING BRACKET
THROUGH HOLES (4)
PEM NUTS (4)
NITROUS OXIDE SEPARATOR
7/16″ SOCKET & SOCKET WRENCH
HEX BOLTS 1/4-20 × 1½” (4)
FIG 3
866-DTE-INFO dentalez.com April, 2024
Page 3 of 4
PN: 1075DOC_B
NITROUS OXIDE SEPARATOR (WET VACUUMS)
INSTALL PLUMBING
1. Install adaptors (FIG 4).
a. Apply thread sealant tape to male threads of PVC adaptor 1″ MPT × 1″ shank.
Install into drain line PVC adaptor (bottom left). Hand tighten.
b. Apply thread sealant tape to male threads of PVC adaptor 1½” MPT × 1½”
shank. Install into vacuum exhaust inlet PVC elbow (top right). Hand tighten.
c. Complete tightening by turning each adaptor 1/4 turn using an adjustable
wrench.
NOTE: PVC reducing bushing 1½” MPT to 1″ FPT and extra PVC adaptor 1″ MPT × 1″
shank provided for vacuum exhaust inlet (top right), if needed, to match
vacuum exhaust hose size.
2. Install hoses (FIG 5).
a. Place small hose clamp onto installer-supplied 1″ hose and twist hose onto
PVC adaptor shank (bottom left).
b. Place large hose clamp on installer-supplied 1½” hose (or small hose clamp
on 1″ hose) and twist hose onto PVC adaptor shank (top right).
c. Use a flathead screwdriver to tighten both clamps.
3. Install vent (FIG 6).
a. Nitrous oxide separator air discharge vent (top left) has a 2″ no-hub
coupling.
b. Attach installer-supplied PVC pipe and fittings to vent to outside.
VACUUM EXHAUST INLET PVC ELBOW 1½”
FPT
DRAIN LINE PVC ADAPTOR
1″ FPT
FIG 4
LARGE OR SMALL HOSE
CLAMP
FIG 5
PVC ADAPTOR
SMALL HOSE CLAMP
PVC PIPE & FITTINGS (INSTALLERSUPPLIED)
AIR DISCHARGE VENT
2″ NO-HUB COUPLING
PVC ADAPTOR 1½” MPT
× 1½” SHANK
PVC ADAPTOR 1″ MPT × 1″ SHANK
1½” OR 1″ HOSE
(INSTALLERSUPPLIED)
PVC ADAPTOR
1″ HOSE (INSTALLERSUPPLIED)
April, 2024
FIG 6 Page 4 of 4
PN: 1075DOC_B
Read User Manual Online (PDF format)