BRASSELER Brioprep Sterile Diamond Burs Instruction Manual
- September 23, 2024
- BRASSELER
Table of Contents
Brioprep Sterile Diamond Burs
“`html
Specifications:
- Product Name: BRIOPREPTM STERILE DIAMOND BURS
- Models: Sterile models with various head diameters, shapes, and
working lengths
Product Information:
Description: The BRIOPREPTM STERILE DIAMOND
Dental Bur family includes Clinical Diamond Burs. These burs are
rotary abrading devices made of stainless steel coated with diamond
particles on the working end, designed to fit into a dental
handpiece.
Intended Use: BRIOPREPTM STERILE DIAMOND BURS
are used with a dental handpiece to abrade hard structures like
teeth, bone, metals, plastics, and porcelains. They are single-use
devices and not meant for reprocessing.
Usage Instructions:
Warnings and Precautions:
-
Ensure the device is single-use only and used on one
patient. -
Operate the bur at the recommended RPM mentioned on the
packaging to prevent breakage or harm. -
Proper irrigation is necessary during use to avoid heat
generation and patient discomfort. -
Avoid excessive pressure on the bur to prevent injury.
-
Do not swallow the bur.
Contraindications:
Avoid using these burs on individuals allergic to nickel.
Precautions:
-
Inspect burs before use to ensure they are not dull.
-
Move the bur continuously while in use to prevent localized
heating. -
Avoid removing the bur at sharp angles to prevent
breakage. -
Maintain handpieces properly for maximum effectiveness.
FAQ:
Q: What is the maximum permissible speed (RPM) for different
head diameters?
A: The recommended RPM varies based on the head diameter. Refer
to the provided table for guidance.
“`
BRIOPREPTM STERILE DIAMOND BURS Instructions for Use
BRASSELER USA BRIOPREPTM STERILE DIAMOND BURS
BRIOPREPTM STERILE DIAMOND BURS are available in Sterile models with numerous
head diameters, shapes and working lengths.
Description
The BRIOPREPTM STERILE DIAMOND Dental Bur family includes Clinical Diamond
Burs. A DIAMOND dental bur is a rotary abrading device made of stainless steel
which is coated with diamond particles on the working end, and which is
designed to fit into a dental handpiece.
Intended Use
BRIOPREPTM STERILE DIAMOND BURS fit into a dental handpiece, which provides
the rotation, allowing the user to abrade hard structures in the mouth, e.g.,
teeth or bone. BRIOPREPTM STERILE DIAMOND BURS can also be used to abrade hard
metals, plastics, porcelains and similar materials. These devices are not
meant for reprocessing.
Warnings and Precautions
BRIOPREPTM STERILE DIAMOND BURS are Single use. Devices marked as single use
are not intended to be used on more than one patient.
The device is to be used on the instruction of, or by a dentist or other
licensed practitioner. Attention should be paid to the speed of work (RPM)
o Always refer to the product packaging for the Maximum RPM. Use of the bur
beyond the RPM range may cause the bur to break and result in patient or user
harm.
o Operating a bur with too high of an RPM may generate undesirable heat and
cause patient discomfort, tooth or tissue necrosis, or patient burns.
Proper irrigation is required while using the device. Inadequate use of
irrigation may generate undesirable heat and cause patient discomfort, tooth
or tissue necrosis, or patient burns.
Do not apply excessive pressure on the bur as this could cause undesirable
heat or may cause the bur to fail and cause patient or user injury. BRIOPREPTM
STERILE DIAMOND BURS are labeled “Sterile” and require no further action prior
to first use.
Use a rubber dental dam while using BRIOPREPTM STERILE DIAMOND BURS to avoid
possible aspiration or swallowing of the device. Always wear gloves when
handling contaminated instruments to avoid possible infection/cross-
contamination. Carefully read package labels to ensure use of the appropriate
device. Failure to do so may cause patient or user injury. Failure to follow
these instructions may cause the following: preparation site damage, injury to
the patient or user, or possible aspiration or
swallowing of the BRIOPREPTM STERILE DIAMOND BUR.
Contraindications
BRIOPREPTM STERILE DIAMOND BURS contain nickel and should not be used for
individuals with known allergic sensitivity to this metal as it may cause
hypersensitivity.
Precautions
Always inspect the burs before use. Use of dull burs could cause undesirable
heat or may cause the device to fail. Move the bur continuously when in use to
avoid localized heating and/or damage to the bur. Undesirable heat generation
can cause patient
discomfort, tooth or tissue necrosis, or patient burns. Avoid removing the bur
at too sharp an angle to avoid leverage and breakage and cause patient or user
injury. Maintain handpieces in good working condition to ensure maximum
effectiveness of the device. Failure to properly maintain handpieces may
lead to injury of the patient or user, aspiration or swallowing of the device
or damage to the preparation site due to vibration of a worn chuck or turbine.
Ensure the bur is fully seated and securely gripped in the handpiece collet
prior to use. Failure to do so may cause the device to “walk out” of the
handpiece and may lead to injury of the patient or user or aspiration or
swallowing of the device. Never force a bur or disc into a handpiece as this
could cause damage to the handpiece collet. Eye protection must be worn to
protect against eject particles. Surgical masks must be worn to avoid
inhalation of any aerosol or dust generated. BUSA Bur Blocks used to hold the
devices for storage and steam sterilization are not intended to maintain
sterility of the device.
Page 1 of 2
IFU-0013 REV D EDCO 24.0127 2024-04-29
BRIOPREPTM STERILE DIAMOND BURS Instructions for Use
The bur pattern selected will be chosen to cut a specific material in a specific application. The following table gives guidance and provides recommended RPM.
Instrument head diameter 1/10 mm
007 – 010 011 – 014 015 – 018 019 – 023 024- 027 028 – 031 032 – 040 041 – 054
055 – 070 080 – 100 019 – 023 024- 027 028 – 031 032 – 040
Maximum permissible speed (RPM) 450,000 450,000 450,000 300,000 160,000 140,000 120,000 95,000 60,000 45,000 300,000 160,000 140,000 120,000
Glossary of Symbols: Brasselerusadental.com/resources/
Manufactured for Brasseler U.S.A. Dental, LLC One Brasseler Blvd Savannah, GA 31419 U.S.A. Telephone 800-841-4522
Recommended operational speed (RPM)
100,000 – 220,000 70,000 – 220,000 55,000 – 160,000 40,000 – 120,000 35,000 –
110,000 30,000 – 95,000 25,000 – 75,000 15,000 – 60,000 12,000 – 40,000 10,000
– 20,000 40,000 – 120,000 35,000 – 110,000 30,000 – 95,000 25,000 – 75,000
Page 2 of 2