dynamic BIOSENSORS AS-2-Gb Adapter Strand 2 User Manual

September 9, 2024
dynamic BIOSENSORS

dynamic BIOSENSORS AS-2-Gb Adapter Strand 2

Key Features

  • Adapter strand 2 for functionalization of heliX ® Adapter Chip Spot 2.
  • Compatible with heliX ® Adapter Chip.
  • Includes Adapter strands for 50 regenerations.
  • Ideal for MIX & RUN sample preparation.
  • Adapter strand 2 carries a moderately hydrophilic green dye (Gb) with a negative net charge.

heliX ® Adapter Chip Overview

2 spots with 2 different anchor sequences for DNA-encoded addressing.

Product Description

Order Number: AS-2-Gb

Table 1. Contents and Storage Information

Material Cap Concentration Amount Buffer Storage
Adapter strand 2 – Gb White 400 nM 5 x 100 µL TE40 [1] -20°C

For research use only.

This product has a limited shelf life, please see expiry date on label.
To avoid many freeze thaw cycles please aliquot the nanolever.

Preparation | MIX & RUN

In-solution hybridization of adapter and ligand strands:

  1. Mix Adapter strand 1 – Gb (400 nM) and conjugated Ligand strand with ligand 1 (500 nM) at 1:1 ratio (v/v).
  2. Mix Adapter strand 2 – Gb (400 nM) and conjugated Ligand strand with ligand 2 (500 nM) at 1:1 ratio (v/v).
  3. Incubate separately the two solutions of step 1 and 2 at RT at 600 rpm for 30 min to ensure complete hybridization.
  4. Mix solution of step 1 and 2 at 1:1 ratio (v/v).

Solution is ready to use for biochip functionalization.

Stability of the solution is related to the stability of the ligand molecules.

Table 2. Additional material for functionalization of spot 1 and reference spot 2.

Material Concentration Buffer Related Product Name Order No
Adapter strand 1 – Gb 400 nM TE40 [ 1] ** _Adapter strand
2_ with green dye Gb** AS-1-Rb

Ligand strand
carrying the conjugated ligand 1| 500 nM| P 2] E40 [| heli X ® Amine Coupling Kit 1| HK-NHS-1
Ligand strand
carrying the conjugated ligand 2| 500 nM| PE40 [ 2]| heli X ® Amine Coupling Kit 1| HK-NHS-1

Example

Required volume for 3 functionalizations: 100 µL with a final concentration of 100 nM.

Vial 1 Vial 2
Adapter strand 1 – Gb (400 nM) Conjugated Ligand strand with
ligand 1 (500 nM) Adapter strand 2 – Gb (400 nM)

Ligand strand + with ligand 2 (500 nM)
25 µL| 25 µL| 25 µL| 25 µL

Contact

  1. TE40: 10 mM Tris, 40 mM NaCl, 0.05 % Tween 20, 50 μM EDTA, 50 μM EGTA
  2. If the protein is not stable in PE40 (TE40, HE40), please check buffer compatibility with the switchSENSE® compatibility sheet.

Dynamic Biosensors GmbH
Perchtinger Str. 8/10 81379 Munich Germany
Dynamic Biosensors, Inc.
300 Trade Center, Suite 1400 Woburn, MA 01801 USA
Order Information order@dynamic-biosensors.com
Technical Support support@dynamic-biosensors.com
www.dynamic-biosensors.com
Instruments and chips are engineered and manufactured in Germany.
©2024 Dynamic Biosensors GmbH | Dynamic Biosensors, Inc. All rights reserved.

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

dynamic BIOSENSORS User Manuals

Related Manuals