AEMC MR525,MR526 AC DC Current Probes User Manual
- September 6, 2024
- AEMC
Table of Contents
AEMC MR525,MR526 AC DC Current Probes
Important Information
Copyright© Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments. All rights
reserved.
No part of this documentation may be reproduced in any form or by any means
(including electronic storage and retrieval or translation into any other
language) without prior agreement and written consent from Chauvin Arnoux®,
Inc., as governed by United States and International copyright laws.
Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments
15 Faraday Drive • Dover, NH 03820 USA
Tel: 800-945-2362 or
603-749-6434 • Fax:
603-742-2346
This documentation is provided as is, without warranty of any kind, express,
implied, or otherwise.
Chauvin Arnoux®, Inc. has made every reasonable effort to ensure that this
documentation is accurate; but does not warrant the accuracy or completeness
of the text, graphics, or other information contained in this documentation.
Chauvin Arnoux®, Inc. shall not be liable for any damages, special, indirect,
incidental, or inconsequential; including (but not limited to) physical,
emotional or monetary damages due to lost revenues or lost profits that may
result from the use of this documentation, whether or not the user of the
documentation has been advised of the possibility of such damages.
Statement of Compliance
Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments certifies that this instrument
has been calibrated using standards and instruments traceable to international
standards.
We guarantee that at the time of shipping your instrument has met the
instrument’s published specifications.
An NIST traceable certificate may be requested at the time of purchase, or
obtained by returning the instrument to our repair and calibration facility,
for a nominal charge.
The recommended calibration interval for this instrument is 12 months and
begins on the date of receipt by the customer. For recalibration, please use
our calibration services. Refer to our repair and calibration section at
www.aemc.com/calibration.
INTRODUCTION
Thank you for purchasing an AEMC® Instruments AC/DC Current Probe Model MR525
or MR526.
For the best results from your instrument and for your safety, you must read
the enclosed operating instructions carefully and comply with the precautions
for use. Only qualified and trained operators should use this product.
SYMBOLS AND DEFINITIONS
| WARNING , risk of DANGER! The operator must refer to these
instructions whenever this symbol appears
---|---
| Refers to a Type A current sensor per IEC/EN 61010-2-032 or BS EN
61010-2-032. This symbol signifies that application around and removal from
conductors carrying dangerous voltages is authorized
| Equipment is protected by double insulation
| Battery
| USB
| Useful information or tip
| Direction of the current
| The instrument follows recycling directives
| The instrument complies with US directives
| Sorting for the recycling of electric and electronic waste
DEFINITION OF MEASUREMENT CATEGORIES (CAT)
CAT IV : Corresponds to measurements performed at the primary electrical
supply (< 1000 V).
Example: primary overcurrent protection devices, ripple control units, and
meters.
CAT III : Corresponds to measurements performed in the building
installation at the distribution level. Example: hardwired equipment in fixed
installation and circuit breakers.
CAT II : Corresponds to measurements performed on circuits directly
connected to the electrical distribution system. Example: measurements on
household appliances and portable tools.
PRECAUTIONS FOR USE
These instructions are intended to ensure the safety of users and proper
operation of the instrument. This instrument is compliant with the IEC
61010-2-032 or BS EN 61010-2-032 safety standards for voltages of 300 V in
measurement CAT IV or 600 V in CAT III.
Failure to observe these safety instructions may result in electric shock,
fire, explosion, and destruction of the instrument and installations.
- The operator and/or responsible authority must read and understand the various precautions to take when using the instrument.
- Do not use the instrument on networks that exceed the instrument’s specifications for voltage or category.
- Never exceed the protection limits stated in the specifications.
- Observe the environmental conditions of use, including relative humidity, altitude, degree of pollution, and location of use.
- Do not use the instrument if it appears to be damaged, incomplete, or improperly closed.
- Before each use, check the condition of the insulation on the leads, housing, and accessories. Any component with deteriorated insulation (even partially) must be set aside for repair or scrapping.
- When handling the instrument, keep your fingers behind the physical guard.
- Use suitable personal protective equipment when appropriate.
- All troubleshooting and metrological checks must be performed by competent and accredited personnel.
RECEIVING YOUR SHIPMENT
Upon receiving your shipment, make sure that the contents are consistent with the packing list. Notify your distributor of any missing items. If the equipment appears to be damaged, file a claim immediately with the carrier, and notify your distributor at once with a detailed description of any damage. Save the damaged packing container to substantiate your claim.
ORDERING INFORMATION
AC/DC Current Probe Model MR525……………………………………… Cat. #1200.81
Includes 9 V battery, multi-language safety data sheet, and user manual.
AC/DC Current Probe Model MR526……………………………………… Cat. #1200.83
Includes 9 V battery, multi-language safety data sheet, and user manual.
Accessories / Replacement Parts
Cable – 6 ft USB type A to Micro type B……………………………………. Cat. #2138.66
Adapter – US Wall plug to USB……………………………………………….. Cat. #2153.78
PRODUCT FEATURES
DESCRIPTION
The Models MR525 and MR526 are clamp-on current probes that measure DC
currents up to 1400 A, AC currents up to 1000 Arms (1400 Apeak), and combined
AC+DC currents without opening the circuit. They indicate the shape and
amplitude of the current measured in the form of a voltage.
These instruments can be used with a multimeter, wattmeter, recorder, and
other instruments. They can be powered by a battery or with 5 Vdc via the
optional micro-USB cable.
The MR525 and MR526 include the following features:
- Range overage indicator
- Power supply indicator
- Zero adjustment
- Auto Standby feature
- One or two ranges, depending on the model
- 1 mV/A (MR525 and MR526)
- 10 mV/A (MR526 only)
- Micro-USB connector for an external power supply
INTERFACE
Item | Function |
---|---|
1 | Fixed jaw (does not move when the trigger is pressed) |
2 | Banana plugs (4 mm) |
3 | Arrow indicating current flow direction |
4 | DC Zero button |
5 | OL (overload) and ON indicators. ON is green when Auto Standby |
is enabled, and it is yellow when disabled.
6| P (Permanent mode) indicator. Hold down the DC Zero button while
turning ON the instrument enables Permanent mode. In this mode, Auto Standby
is disabled (see § 3.4).
7| Slide switch (2-position for MR525, 3-position for MR526)
8| USB port
9| Trigger
10| Hand guard
11| Mobile jaw (moves when the trigger is pressed)
OPERATION
BATTERY INSTALLATION
NOTE : Before changing the batteries, set the switch to OFF, and remove the clamp from the circuit under measurement.
- Using a screwdriver, remove the battery compartment cover (1) from the back of the housing (see Figure 2).
- Connect the battery to the snap-on connector (2) while making sure that the polarity of the battery terminals match the snap-on connector.
- Place the battery into the battery compartment (3).
- Replace the battery compartment cover, and screw it onto the housing.
EXTERNAL POWER (OPTIONAL)
For long-term measurements, you can connect the clamp to external power via
any micro-USB adapter that delivers 100 mA or more. If the external power is
disconnected, the clamp will automatically switch to battery operation.
Between the type B micro-USB connector and the measurement output, the
isolation is 600 V CAT III. This enables you to safely connect the clamp to
measuring instruments with uninsulated inputs. The type B micro‑USB connector
must not contact conductors or uninsulated parts at dangerous voltage.
When operating on external power, the Auto Standby feature is disabled. The
color of the ON indicator signals whether automatic standby is enabled (green)
or disabled (yellow).
TURNING ON THE INSTRUMENT
Turn on the clamp by pushing the slide switch to the 1 mV/A or 10 mV/A setting:
- 1 mV/A corresponds to the 1400 A range (MR525 and MR526)
- 10 mV/A corresponds to the 150 A range (MR526)
- The green ON indicator should light up:
- If the indicator blinks, the battery has less than 4 hours remaining.
- If the indicator fails to light up, replace the battery (see § 5.2).
AUTO STANDBY
After 10 minutes of operation without any user action (such as pressing the DC
Zero button), the clamp will automatically enter Auto Standby mode. In this
mode, the ON indicator will turn OFF.
To reactivate the clamp, press the DC Zero button, or change the switch to any
setting other than OFF.
To disable automatic standby, press and hold the DC Zero when turning the
instrument ON. The ON indicator will blink to indicate that the request has
been applied. When you release the DC Zero button, the indicator will glow
steady yellow.
DC ZERO ADJUSTMENT
NOTE : You must adjust the DC Zero before a series of measurements and whenever the instrument reconnects after a disconnection.
- With the clamp connected to the measuring instrument, select the desired measurement range (or sensitivity) with the switch.
- Make sure that there is no conductor in the clamp and that the jaws are closed correctly.
- Press the DC Zero button.
- The OL indicator will light up for approximately three seconds to indicate that the zero adjustment is in progress.
- The OL indicator will turn off to indicate that the operation has succeeded.
- If the OL indicator stays on, the zero could not be adjusted.
- Before repeating the operation, check that the jaws are closed correctly (air gaps clean, no dust, no oxidation, etc.) and that there is no conductor in the clamp.
- Press the DC Zero button again.
- In the event of failure, or if the clamp is switched off (selector set to OFF), the instrument keeps the last successful DC Zero adjustment.
WARNING : If the measurement range (or sensitivity) is changed, the DC Zero must be adjusted before any further measurements are made.
MEASUREMENTS
Making a Measurement
After adjusting the DC Zero:
-
Press the clamp’s trigger to open the jaws.
-
Clamp the jaws around the conductor to be measured. Use the centering marks on the jaws to position the clamp around the conductor. If the measurement is for a power calculation, ensure that the arrow on the clamp jaws (see Figure 3) points in the direction of the current flow:
source load. -
Release the trigger while ensuring that the jaws are completely and correctly closed.
-
Observe the measurement displayed on the instrument.
-
If the OL indicator turns on, the current is too high to be measured. If you are using the MR526 and the sliding switch is set to the 10 mV/A range, change the setting to 1 mV/A.
Converting to Current
Both models can measure current up to 1400 A with 1 mV corresponding to 1 A.
In addition, the MR526 provides a second measurement range up to 150 A with 10
mV corresponding to 1 A.
To convert the clamp output to current, divide the voltage reading on the
connected measuring instrument by the V/A coefficient.
SPECIFICATIONS
REFERENCE CONDITIONS
Quantities of Influence | Reference Conditions |
---|---|
Temperature | 73 °F ± 9 °F (23 °C ± 5 °C) |
Relative humidity | (20 to 75) % RH |
Position of the conductor | Centered on the marks of the jaws |
Measurement frequency | DC to 65 Hz sine wave |
External electrical field | Zero |
External DC magnetic field (earth) | <40 A/m |
External AC magnetic field | Zero |
Input impedance | ≥1 MΩ and ≤100 pF |
The intrinsic uncertainty is the error defined under the reference conditions.
It is expressed as a percentage of the output signal (R) plus an offset in mV:
±(a % R + b)
ELECTRICAL SPECIFICATIONS
Electrical Specifications, 1 mV/A Sensitivity
Output impedance: 215 Ω
Specified Measurement Range| (0.5 to 100) Aac/dc| (100 to
800) Aac/dc| (800 to 1000) Aac/dc| (1000 to 1400) Adc
---|---|---|---|---
Intrinsic uncertainty| ≤ ±(2 % R + 1.5 mV)| ≤ ±2.5 % R| ≤ ±4 % R| ≤ ±5 % R
Phase error (45 to 65) Hz
Specified Measurement Range | (3 to 200) Aac | (200 to 1000) Aac |
---|---|---|
Phase shift | ≤ -2 ° | ≤ -1.5 ° |
Typical amplitude error curve at 60 Hz
Typical amplitude error curve in DC
Typical phase error curve at 60 Hz
Frequency Specifications, 1 mV/A Sensitivity
Bandwidth -3 dB: DC to 30 kHz
Frequency | 50 Hz | 400 Hz | 1 kHz | 10 kHz |
---|---|---|---|---|
Insertion impedance | <0.01 mΩ | 0.05 mΩ | 0.14 mΩ | 3.4 mΩ |
Typical amplitude error versus frequency curve at 60 A
Typical phase versus frequency error curve at 60 A
ELECTRICAL SPECIFICATIONS, 10 mV/A SENSITIVITY (MR526)
Output impedance: 215 Ω
Specified Measure ‑ ment Range| (0.5 to 40) Aac/dc|
(40 to 100) Aac/dc| (100 to 150) Adc
---|---|---|---
Intrinsic uncertainty| ≤ ±(3 % R + 8 mV)| ≤ ±1.5 % R| ≤ ±1.5 % R
Phase error (45 to 65 Hz)
Specified Measurement Range | (1 to 100) Aac |
---|---|
Phase shift | ≤ -2 ° |
Typical amplitude error vs current curve at 60 Hz
Typical amplitude error vs current curve in DC
Typical phase vs current error curve at 60 Hz
Frequency Specifications, 10 mV/A Sensitivity
Bandwidth -3 dB: DC to 30 kHz
Frequency | 50 Hz | 400 Hz | 1 kHz | 10 kHz |
---|---|---|---|---|
Insertion impedance | <0.01 mΩ | 0.05 mΩ | 0.14 mΩ | 3.4 mΩ |
Typical amplitude error versus frequency curve at 30 A
Typical phase versus frequency error curve at 30 A
OPERATING LIMITS
- In DC: 3000 A permanent
- In AC: 1000 A permanent up to 1 kHz from 1 kHz, IMAX = 1000/f (kHz)
- Conductor temperature: ≤194 °F, 230 °F peak (90 °C, 110 °C peak)
- Temperature of the jaws: ≤176 °F (80 °C)
Curve of derating versus frequency
VARIATIONS IN THE RANGE OF USE
Quantity of influence| Range of influence| Error in % of
reading
---|---|---
Typical| Maximum
Temperature| (14 to 131) °F
(-10 to +55) °C
| | Drift of the zero
±55.56 mA/°F
(±100 mA/°C)
| Drift of the gain 3 %
Relative humidity| (10 to 85) % RH| | 0.5 %
Frequency| from (10 to 400) Hz
from 400 Hz to 7 kHz
from (7 to 30) kHz| | 1 %
3.5 % see curves
Position of the conductor 20 mm in diameter| | | 0.5 %
Adjacent conductor carrying a 50 Hz AC current| Conductor 23 mm from the
clamp| | 10 mA/A
External 400 A/m field at 50 Hz| Cable centered| | 1.3 A
Common mode rejection| 600 V between the jacket and the secondary| | 90 dB A/V
at 50 Hz
Remanence| | 100 Adc: 2.8 A
200 Adc: 3.5 A
400 Adc: 5 A
800 Adc: 5.3
1200 Adc: 5.7 A
1400 Adc: 5.8 A|
POWER SUPPLY
The instrument is powered by a 9 V battery (type 6LR61, 6LF22, or NEDA 1604).
The average battery life is 50 hours with an alkaline battery.
The instrument can also be powered by an external supply (5 Vdc, 100 mA) via
the type B micro-USB connector.
ENVIRONMENTAL CONDITIONS
The instrument must be used in the following environmental conditions.
1 = Range of reference
2 = Operating range
3 = Storage range
Indoor use
Degree of pollution: 2
Altitude: <6500 ft (2000 m)
Transport altitude: ≤40,000 ft (12,000 m)
MECHANICAL SPECIFICATIONS
Dimensions (L x W x H): 9.3 x 3.8 x 1.7 in (237 x 97 x 44 mm)
Weight: 18.3 oz (520 g)
Cable: 4.9 ft (1.50 m)
Maximum Conductor Size:
Cables: One 1.54 in (39 mm) or two 1 in (25.4 mm)
Bus Bar: One 1.97 x 0.49 in (50 x 12.5 mm); two 0.98 x 0.2 in (50 x 5 mm);
two 1.24 x 0.30 in (31.5 x 10 mm); or three 0.98 x 0.31 in (25 x 8 mm)
Housing Protection
Protection index:
- IP 40 per IEC 60529
- IK 06 per IEC 62262
Drop test per IEC/EN 61010-2-032 or BS EN 61010-2-032
INTERNATIONAL STANDARDS
The instrument is compliant with IEC/EN 61010-2-032 or BS EN 61010-2-032,
300 V in CAT IV or 600 V in CAT III.
Double or reinforced insulation
Type of current sensor per IEC/EN 61010-2-032 or BS EN 61326-1: (type A)
ELECTROMAGNETIC COMPATIBILITY
The device is in conformity with standard IEC/EN 61326-1 or BS EN 61326-1.
MAINTENANCE
WARNING : Except for the battery, the instrument contains no parts that can be replaced by personnel who have not been specially trained and accredited. Any unauthorized repair or replacement of a part by an equivalent may gravely impair safety.
CLEANING
- Disconnect anything connected to the clamp.
- Use a soft cloth, dampened with soapy water.
- Rinse with a damp cloth and dry rapidly with a dry cloth or forced air.
- Do not use alcohol, solvents, or hydrocarbons.
- Keep the clamp jaws as clean as possible.
BATTERY REPLACEMENT
The battery must be replaced if the ON indicator remains unlit when the instrument is turned ON.
- Disconnect the instrument completely and set the switch to OFF.
- Remove the battery compartment cover from the instrument casing (see § 3.1).
- Remove the old battery.
- Insert the replacement battery into the snap-in battery connector, and place it into the battery compartment.
- Replace the battery compartment cover.
NOTE : Depleted batteries must not be treated as ordinary household waste. Take them to the appropriate collection point for recycling.
REPAIR AND CALIBRATION
To ensure that your instrument meets factory specifications, we recommend that the instrument be sent back to our factory Service Center at one-year intervals for recalibration or as required by other standards or internal procedures.
For instrument repair and calibration:
You must contact our Service Center for a Customer Service Authorization Number (CSA#). Send an email to repair@aemc.com requesting a CSA#, you will be provided a CSA Form and other required paperwork along with the next steps to complete the request. Then return the instrument along with the signed CSA Form. This will ensure that when your instrument arrives, it will be tracked and processed promptly. Please write the CSA# on the outside of the shipping container. If the instrument is returned for calibration, we need to know if you want a standard calibration or a calibration traceable to N.I.S.T. (includes calibration certificate plus recorded calibration data).
Ship To: Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments
15 Faraday Drive • Dover, NH 03820 USA
Phone: 800-945-2362 (Ext. 360) /
603-749-6434 (Ext. 360)
Fax: 603-742-2346
E-mail: repair@aemc.com
(Or contact your authorized distributor.)
Contact us for the costs for repair, standard calibration, and calibration traceable to N.I.S.T.
NOTE: You must obtain a CSA# before returning any instrument.
TECHNICAL ASSISTANCE
If you are experiencing any technical problems or require any assistance with
the proper operation or application of your instrument, please call, e-mail or
fax our technical support team:
Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments
Phone: 800-343-1391 (Ext. 351)
Fax: 603-742-2346
E-mail: techsupport@aemc.com
www.aemc.com
LIMITED WARRANTY
The instrument is warrantied to the owner for a period of two years from the date of original purchase against defects in manufacture. This limited warranty is given by AEMC® Instruments, not by the distributor from whom it was purchased. This warranty is void if the unit has been tampered with, abused, or if the defect is related to service not performed by AEMC® Instruments.
Full warranty coverage and product registration is available on our website at
www.aemc.com/warranty.html.
Please print the online Warranty Coverage Information for your records.
What AEMC® Instruments will do:
If a malfunction occurs within the warranty period, you may return the instrument to us for repair, provided we have your warranty registration information on file or a proof of purchase. AEMC® Instruments will repair or replace the faulty material at our discretion.
REGISTER ONLINE AT: www.aemc.com/warranty.html
Warranty Repairs
What you must do to return an Instrument for Warranty Repair:
First, send an email to repair@aemc.com requesting a
Customer Service Authorization Number (CSA#) from our Service Department. You
will be provided a CSA Form and other required paperwork along with the next
steps to complete the request. Then return the instrument along with the
signed CSA Form. Please write the CSA# on the outside of the shipping
container. Return the instrument, postage or shipment pre-paid to:
Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments
15 Faraday Drive, Dover, NH 03820 USA
Phone: 800-945-2362 (Ext. 360)
603-749-6434 (Ext. 360)
Fax: 603-742-2346
E-mail: repair@aemc.com
Caution : To protect yourself against in-transit loss, we recommend that you insure your returned material.
NOTE: You must obtain a CSA# before returning any instrument.
Customer Support
AEMC® Instruments
15 Faraday Drive • Dover, NH 03820 USA
Phone: +1 603-749-6434 • +1
800-343-1391 • Fax: +1
603-742-2346
www.aemc.com