AEMC 6550 10kV and 15kV Megohmmeters User Manual
- June 9, 2024
- AEMC
Table of Contents
6550 10kV and 15kV Megohmmeters
User Manual 10kV and 15kV MEGOHMMETERS
6550 , 6555
Copyright © Chauvin Arnoux® , Inc. d.b.a. AEMC® Instruments. All rights
reserved. No part of this documentation may be reproduced in any form or by
any means (including electronic storage and retrieval or translation into any
other language) without prior agreement and written consent from Chauvin
Arnoux ® , Inc., as governed by United States and International copyright
laws.Chauvin Arnoux® , Inc. d.b.a. AEMC® Instruments 15 Faraday Drive • Dover,
NH 03820 USA Tel: 800-945-2362 or
603-749-6434 • Fax:
603-742-2346
This documentation is provided “as is,” without warranty of any kind, express,
implied, or otherwise. Chauvin Arnoux® , Inc. has made every reasonable effort
to ensure that this documentation is accurate; but does not warrant the
accuracy or completeness of the text, graphics, or other information contained
in this documentation. Chauvin Arnoux® , Inc. shall not be liable for any
damages, special, indirect, incidental, or inconsequential; including (but not
limited to) physical, emotional or monetary damages due to lost revenues or
lost profits that may result from the use of this documentation, whether or
not the user of the documentation has been advised of the possibility of such
damages.
Chauvin Arnoux® , Inc, AEMC® , DataView® , AmpFlex® , MiniFlex® and PowerPad®
are registered trademarks of AEMC® Instruments.
Statement of Compliance
Chauvin Arnoux® , Inc. d.b.a. AEMC® Instruments certifies that this instrument
has been calibrated using standards and instruments traceable to international
standards.
We guarantee that at the time of shipping your instrument has met its
published specifications.
An NIST traceable certificate may be requested at the time of purchase, or
obtained by returning the instrument to our repair and calibration facility,
for a nominal charge.
The recommended calibration interval for this instrument is 12 months and
begins on the date of receipt by the customer. For recalibration, please use
our calibration services.
Refer to our repair and calibration section at
www.aemc.com.
Serial #: __
Catalog #: 2130.31 / 2130.32
Model #: 6550 / 6555
Please fill in the appropriate date as indicated:
Date Received: ____
Date Calibration Due: ___
Chauvin Arnoux® , Inc. d.b.a AEMC® Instruments
www.aemc.com
Megohmmeter Models 6550/6555
Cat. #2130.31 / Cat. #2130.32
PRODUCT PACKAGING
Ships with the following: Also Includes:
- USB drive with DataView® Software and User Manual
- 2×9.6V NiMH batteries (Cat. #2140.19 each)
Thank you for purchasing a Megohmmeter Model 6550/6555.
For best results from your instrument and for your safety, read the enclosed
operating instructions carefully and comply with the precautions for use.
These products must be only used by qualified and trained users.
| Signifies that the instrument is protected by double or reinforced
insulation.
---|---
| CAUTION – DANGER! Read the User Manual.
| Risk of electric shock. The voltage at the parts marked with this symbol may
be dangerous.
| Important instructions to read and to fully understand.
| Useful information or tip to read.
| USB socket.
| Ground/Earth.
| The CE marking guarantees conformity with European directives and with
regulations covering EMC.
****| The UKCA marking certifies that the product is compliant with the
requirements that apply in the United Kingdom, specifically regarding Low-
Voltage Safety, Electromagnetic Compatibility, and the Restriction of
Hazardous Substances.
| Chauvin Arnoux® Inc. d.b.a. AEMC® Instruments has adopted an Eco-Design
approach in order to design this instrument. Analysis of the complete
lifecycle has enabled us to control and optimize the effects of the product on
the environment. In particular this instrument exceeds regulation requirements
with respect to recycling and reuse.
| The trash can with a line through it means that in the European Union, the
product must undergo selective disposal for the recycling of electric and
electronic material, in compliance with Directive WEEE 2002/96/EC.
Precautions Before Use
This instrument and its accessories comply with safety standards IEC/EN
61010-2-030 or BS EN 61010-2-030 and IEC/EN 61010-031 or BS EN 61010-031 for
voltages of 1000V in Category IV.
Failure to observe the safety instructions may result in electric shock, fire,
explosion, and destruction of the instrument and installations.
- The operator and/or the responsible authority must carefully read and clearly understand the various precautions to be taken in use. Sound knowledge and a keen awareness of electrical hazards are essential when using this instrument.
- If the instrument is used other than as specified, the protection it provides may be compromised, thereby endangering you.
- Do not use the instrument on networks of which the voltage or category exceeds those mentioned.
- Do not use the instrument if it seems to be damaged, incomplete, or poorly closed.
- Before each use, check the condition of the insulation on the leads, housing, and accessories. Any item of which the insulation is deteriorated (even partially) must be set side for repair or scrapping.
- Use only the leads and accessories supplied. Using leads (or accessories) of a lower voltage or category reduces the voltage or category of the combined instrument and leads (or accessories) to that of the leads (or accessories).
- Use personal protection equipment systematically.
- Keep your hands away from the terminals of the instrument.
- When handling the leads, test probes, and alligator clips, keep your fingers behind the physical guard.
- As a safety measure, and to avoid interference, do not move and do not handle the leads during measurements.
Definition of Measurement Categories (CAT)
-
CAT IV – 3-Phase at utility connection, outdoor conductors:
– Origin of installation, or where low-voltage connection is made to utility power
– Electricity meters, primary overcurrent protection equipment
– Outside and service entrance, service drop from pole to building, runs between meter & panel
– Overhead line to detached building, underground line to well pump -
CAT III – 3-Phase distribution, including single-phase commercial lighting:
– Equipment in fixed installations, such as switchgear and polyphase motors
– Bus and feeder in industrial plants
– Feeders and short branch circuits, distribution panel devices
– Appliance/equipment outlets with short connections to service entrance -
CAT II – Single-phase, receptacle-connected loads:
– Appliances, portable tools, and other similar light industrial/household loads
– Outlet and long-branch circuits
– Outlets at more than 30 ft from CAT III source
– Outlets at more than 60 ft from CAT IV source
INTRODUCTION
1.1 Receiving Your Shipment
Upon receiving your shipment, make sure that the contents are consistent with
the packing list. Notify your distributor of any missing items. If the
equipment appears to be damaged, file a claim immediately with the carrier and
notify your distributor at once, giving a detailed description of any damage.
Save the damaged packing container to substantiate your claim.
Megohmmeter Model 6550 (Graphical, Analog Bargraph, Backlight, Alarm, Timer,
500V, 1000V, 2500V, 5000V, 10kV, Ramp, StepV, Variable, Auto DAR/PI/DD, USB,
w/DataView® Software) …… Cat. #2130.31
Megohmmeter Model 6555 (Graphical, Analog Bargraph, Backlight, Alarm, Timer,
500V, 1000V, 2500V, 5000V, 15kV, Ramp, StepV, Variable, Auto DAR/PI/DD, USB,
w/DataView ® Software) ….. Cat. #2130.32
Both models include set of 3 color-coded (red/blue/black) 10ft, 15kV safety
leads with clips (3000V CAT III), one 15kV jumper lead (blue), optical USB
cable, 115V US power cord, 9.6V rechargeable NiMH batteries, small classic
tool bag, and a USB stick with DataView ® software and a user manual.
1.1.1 Accessories and Replacement Parts
Replacement – Small classic tool bag ……………………………………………………………………………….. Cat.
2133.72
Replacement – Optical USB cable ……………………………………………………………………………………. Cat.
2135.41
Replacement – One 9.6V NiMH battery (two are required) ……………………………………………………
Cat. #2140.19
Replacement – Set of 3 color-coded (red/blue/black) 10ft safety leads with
clips (15kV) …………. Cat. #2151.36
Replacement – 1.5ft blue jumper lead (15kV) ……………………………………………………………………… Cat.
2151.37
Replacement – 115V US power cord ………………………………………………………………………………… Cat. #5000.14
Lead – Set of 3, 25ft, 15kV color-coded (red/blue/black) safety leads with
clips ……………………… Cat. #2151.38
Power cord – 240V EU …………………………………………………………………………………………………….. Cat. #5000.32
DataView Software Updates are Available at www.aemc.com
PRODUCT FEATURES
2.1 Description
The Megohmmeter Models 6550 and 6555 are high-end portable measuring
instruments capable of measuring very high electrical insulation and
resistance values. They are packaged in a rugged case, have a graphical LCD
screen and can operate on battery or AC power.
The Model 6550 makes insulation measurements at voltages up to 10,000V, the
Model 6555 up to 15,000V.
Their main functions are:
-
detection and measurement of input voltage, frequency, and current
-
quantitative and qualitative insulation measurement:
– measurement at a fixed test voltage of 500V, 1000V, 2500V, 5000V, 10,000V or 15,000VDC
– measurement at an adjustable test voltage from 40 to 15,000VDC
– measurement with a voltage ramp selectable in the ranges from 40 to 1100V or from 500 to 15,000V
– measurement with a voltage in steps from 40 to 15,000V with up to 10 steps and dwell times
– non-destructive (Early Break) test, test stopped at a preset current (Break at I-limit) or Burning
– calculation of the DAR, PI, and DD (dielectric discharge index) quality ratios
– calculation of the temperature corrected resistance -
measurement of the capacitance of the circuit tested
-
measurement of the residual current
These megohmmeters contribute to the safety of electrical installations and
equipment.
Their operation is managed by microprocessors that acquire, process, display,
and store the measurements.
They have many advantages, such as:
- digital filtering of insulation measurements
- measurement of the voltage
- programming of thresholds to trigger audible alarms
- time controlled measurements
- current limit programming
- plotting of resistance, voltage, and current vs. time and current vs. voltage: R(t), V(t), I(t), and I(V)
- fuse protection, with defective fuse indication
- automatic discharge of the test voltage at the end of the measurement provides operator safety
- auto power off mode to save battery power
- battery charge condition indication
- large graphical display with backlight capability
- data storage of measurements
- real-time clock, and a USB interface
- configuration, real-time testing and data export to a PC (using the included DataView® software)
2.2 Front Panel Features
- Safety connection terminals “+”, “G” and “-”.
- Graphical, digital LCD (see § 2.3).
- Power receptacle for recharging the batteries.
- USB connection for communication with a PC.
- Seven-position rotary function switch.
- Navigation buttons for moving the cursor, selecting and changing values.
- START/STOP measurement button.
- Eight function buttons (see § 2.4).
2.3 Display
The display is a graphical, digital LCD with a resolution of 320 x 240 pixels.
It has built-in backlight, which can be controlled by a long press on the
HELP/ button.
2.3.1 Example of Display Before Measurement 2.3.2 Example of Display During Measurement 2.3.3 Example of Display
After Measurement
The symbol indicates blinking.
If values are undetermined, they are represented by – – – -.
2.4 Button Functions
If the audible signal has not been deactivated in SET-UP (see § 3.4), the
instrument confirms each button press by an audible beep. If the beep has
higher frequency, this indicates that pressing the button is prohibited and
will have no effect.
A long press (press maintained for more than two seconds) is confirmed by a
second audible beep.
BUTTON | DESCRIPTION |
---|---|
TEMP | • Enter temperature and humidity information and calculate temperature |
corrected resistance
ALARM| • Enables/Disables the alarms
HELP / | • Displays Help information; Enables/Disables the backlighting of the
display
MEM| • Stores the measurements
CONFIG| • Configuration of the measurement parameters
DISPLAY| • Switch between screens
FILTER| • Smoothing of the measurements
GRAPH| • Switch graph mode ON/OFF
2.4.1 TEMP Button
This function is accessible only when the measurement is completed for V-VAR
and V-FIXED only. It is used to reference the measurement result to a
temperature other than the one at which the measurement was made. Temperature
causes the resistance to vary according to a quasi exponential law. To a rough
approximation, raising the temperature by 18°F (10°C) halves the insulation
resistance; conversely, lowering the temperature by 18°F (10°C) doubles the
insulation resistance.
Comparing measurements to a single reference temperature makes it easier to
evaluate insulation resistance measurements taken at different times and
temperatures.
Similarly, measuring the humidity improves the correlation between the various
measurements made on a given device.
Procedure:
- Make a measurement in V-FIXED or V-VAR mode.
- Press the TEMP button.
Use the ◄ ►▲▼ arrow buttons to enter the various parameters:
- Air Temperature: the ambient temperature (optional).
- Humidity: the ambient relative humidity (optional).
- Probe Temperature: the temperature of the device tested. If it has not warmed up during the measurement, it is equal to the ambient temperature.
- Rc Reference Temperature: the temperature to which the measured resistance will be referred.
- ∆T for R/2: the temperature variation, known or estimated, sufficient to halve the insulation resistance.
To facilitate the programming, the instrument proposes default values.
The instrument then displays the insulation resistance referred to the
reference temperature.If coefficient ∆T for R/2 is not known, it can
be calculated from a minimum of 3 measurements made on the same device at
different temperatures.
Detail concerning the calculation performed:
The insulation resistance varies with the measurement temperature. This
dependence can be approximated by an exponential function:
Rc = Kt * Rt where Rc: insulation resistance referred to 40°C
Rt: insulation resistance measured at temperature T
Kt: coefficient defined as follows:Kt = (1/2) ^ ((40 – T)/∆T)
where ∆T: temperature difference at which the insulation resistance is halved
2.4.2 ALARM Button
Press the ALARM button to enable the alarm defined using the CONFIG button
(see 2.4.3) or in
SET-UP (see § 3.4). The ALARM symbol is then displayed.
If the measurement is less than the alarm threshold, the instrument indicates
this by blinking the
ALARM symbol on the display and emitting an audible signal.
Press the ALARM button again to disable the alarm; the ALARM symbol disappears
from the display.
2.4.3 CONFIG Button
2.4.3.1 Before the Measurement
If the V-FIXED or V-VAR. test voltages have been chosen, there are two
configuration screens. There is only one for the V-RAMP and V-STEP test
voltages.
Press the CONFIG button (press CONFIG again to exit):
- Manual Stop: measurements are stopped manually.
- Manual Stop + DD: measurements are stopped manually and dielectric discharge ratio (DD) is calculated at the end of the programmed duration.
- Timed Run + DD: measurement is automatically stopped at the end of the programmed duration and dielectric discharge ratio (DD) is calculated.
- DAR: measurement is automatically stopped at the end of one minute (or of the programmed time, if different).
- PI: measurement is automatically stopped at the end of 10 minutes (or of the programmed time, if different).
It is always possible to stop a measurement during a test with programmed
duration by pressing the START/STOP button.
The ▲▼ arrow buttons are used to select the measurement configuration. When
Timed Run (test with programmed duration) or Timed Run + DD is selected, the
duration of the measurement (m:s) can be set. To do this, use the ◄ ►▲▼arrow
buttons.
The test will run for the time programmed. However, if, during the
measurement, the rotary switch position is changed or the START/STOP button is
pressed, the measurement will stop.
Press the DISPLAY button to see the second configuration screen.The ▲▼ arrow buttons are used to select and modify a parameter.
The second configuration screen depends on the rotary switch setting.
The V-RAMP and V-STEP settings only use the second page of the configuration
screen.
The second configuration screen is used to choose:
- The type of test (Test Type)
- Nondestructive test (Early Break)
The measurement will be stopped at the first breakdown current peak detected.
This type of test is used for non-destructive tests. The current is limited to
0.2mA.
The E-BRK symbol is displayed. ■ Stopping the test at a preset current
(Break at I-limit)
The measurement will be stopped when the current reaches the maximum value
(Maximum Output Current) defined by the user (see below). This type of test is
useful for testing varistors or other types of voltage limiter.
The I-LIM symbol is displayed. ■ Burning
The measurement is not stopped depending on the current. Depending on the
application, this type of test can be used to determine the location of
insulation faults when there is Burning: appearance of an electric arc during
the test or burn spot after the test.
The BURN symbol is displayed. ■ The Maximum Current (Maximum Output
Current)
This is the current not to be exceeded in any type of test. (In the test type
Break at I-limit the measurement is stopped if this value is reached.).
Use the ▲▼ arrow buttons to set it between 0.2 and 5mA for test types Burning
and Break at I-limit. For test type Early break this value is fixed to 0.2mA.
■ The Current Range (I-range)
This function is used to make measurements more rapidly when their order of
magnitude is already known.
Use the ▲▼ arrow buttons to set the value to Auto or Range 1, 2 or 3.
Current | < 300nA | 60nA < I < 50µA | 10µA < I < 6mA |
---|---|---|---|
Current range | 1 | 2 | 3 |
For example for VN = 10,000V :
Current range | 1 | 2 | 3 |
---|---|---|---|
Resistance | R > 30GW | 200MW < R < 16,6GW | 10MW* < R < 1 GW |
*10MΩbecause Imax = 1mA at 10,000V. _
_The fixed current range remains active until the instrument is switched off
The RANGE symbol is displayed. ■ Interference of the Signal
(Disturbance Level)
Use the ▲▼ arrow buttons to set the value, from Low to High. The DH symbol is
then displayed. The High setting is recommended when measurements
are made in the presence of strong electromagnetic fields at the network
frequency (for example near high-voltage lines).
■ In the V-FIXED and V-VAR Modes: the alarm threshold
Use the ▲▼ arrow buttons to set the alarm threshold.
The alarm threshold can also be set in SET-UP (see § 3.4).
The ALARM symbol is displayed if the alarm is enabled.
■ In the V-RAMP mode: the programming of the ramp (Set Ramp Function)
Use the ▲▼ arrow buttons to go to Set Ramp Function; the instrument displays
the voltage ramp values programming screen. This programming can also be done
in SET-UP (see § 3.4).
■ In the V-STEP mode: the programming of the step (Set Step Function)
Use the ▲▼ arrow buttons to go to Set Step Function; the instrument displays
the voltage step values programming screen. This programming can also be done
in SET-UP (see § 3.4).
2.4.3.2 During the Measurement
During the measurement (for V-VAR and V-FIXED), the CONFIG button is used to
choose the current range: automatic (default) or fixed. For more details,
refer to the previous section.
■ Once the measurement has started, press the CONFIG button. The following
screen appears:
- Use the ◄ ►▲▼ arrow buttons to modify the measurement current range. Changes are applied and saved immediately after entry.
- Confirm and exit by pressing the CONFIG button again.
- If the range is fixed, the RANGE symbol is displayed.
- The choice remains active until the rotary switch is moved to another position.
During the measurement, it is also possible to turn the analog filter ON/OFF
(Disturbance Level). For more details, refer to the previous section.
In the case of a variable test voltage, the voltage set is also displayed and
can be modified during the measurement.![AEMC 6550 10kV and 15kV Megohmmeters
-
displayed 5](https://manuals.plus/wp-content/uploads/2023/03/AEMC-6550-10kV- and-15kV-Megohmmeters-displayed-5.png) 2.4.4 DISPLAY Button
This button is used to browse through the various accessible screens containing all information available before, during or after the measurement. Depending on the measurement mode and the configuration chosen (CONFIG button), the screens are different.
2.4.5 GRAPH Button
During the measurement, and at the end of each measurement, pressing the GRAPH button displays a graphical presentation of the measurement results. On the first screen, the insulation resistance vs. time R(t) and the voltage vs. time V(t) are shown.This curve is plotted using samples recorded during the measurement.
2.4.5.1 During a Measurement
There is no cursor. Each new measurement result is automatically added to the curve and its values are shown in one line above the graphical area.
2.4.5.2 After a Measurement
The time indication in the upper right of the screen is blinking, this is the indication for cursor mode.
The ◄ ► arrow buttons can be used to move the time cursor along the curve. The minimum and maximum values at the cursor position are shown in two lines above the graph area. If the time span on the time axis is 4 minutes (which is the smallest possible) these lines are the same and represent one sample.
Depending on the range of the left vertical scale it may be possible to move the vertical scale and the corresponding curve with the ▲▼ arrow buttons.
The following is an example of the display when a measurement is performed in V-RAMP or V-STEP mode:If the interval of the scale of an axis is big enough, it may be possible to zoom.
Press the CONFIG button.- The time indication in the upper right of the screen stops blinking, this is the indication for zoom mode.
- The ◄ ► arrow buttons are used to modify the time scale of the graph.
- The ▲▼ arrow buttons are used to modify the resistance scale of the graph.
Press the DISPLAY button to view the current vs. time curve.Press the DISPLAY button again to view the current vs. voltage data points (not available for V-STEP).![AEMC 6550 10kV and 15kV Megohmmeters
-
displayed 9](https://manuals.plus/wp-content/uploads/2023/03/AEMC-6550-10kV- and-15kV-Megohmmeters-displayed-9.png)This curve is useful primarily in the case of a measurement in V-RAMP mode.
There is no cursor and it is not possible to zoom on this curve.
2.4.6 FILTER Button The FILTER button can be used to activate and deactivate a digital filter for the insulation measurements. This filter affects only the displayed results (which are smoothed), not the measured values. The recorded data therefore remains raw (no filter).
This function is useful in the case of high instability of the insulation values displayed, but it is also possible to estimate the measurement on the bargraph.
Once the measurement starts, if interference is present, press the FILTER button. Start by applying the DF10s filter. If that is not enough, go to the DF20s filter, then to the DF40s filter. The larger the time constant, the smoother (and slower) the measurement.
Successive presses on the FILTER button will modify or remove the filter:- DF 10: time constant 10 seconds
- DF 20: time constant 20 seconds
- DF 40: time constant 40 seconds
- DF: automatic filter, adapts the filter time to the resistance result changes.
- no filter
The filter is calculated as follows:
RN = RN-1 + (R – RN-1)/N
If N is set to 20, the time constant of this filter will be approximately 20
seconds.Selecting digital filtering (DF) is recommended
for measurements of fluctuating high insulation resistance values.
Such fluctuations may be due to hand movement, when touching the cables,
fluctuating capacitances in the device tested, insulation that varies because
of conducting dust, an ionizing and polarizing effect of this dust, etc., or
again to the presence of an AC voltage superposed on the measurement.
The FILTER button is active before and during the measurement (but is not
available in graph mode).
2.4.7 HELP Button
A short press on the HELP button opens the help function, in which the actions
of the buttons are explained.
This operation changes with the context: setting of the switch, operating
mode, before, during, or after a measurement. Below is an example in V-FIXED
mode:A long press on the HELP button (>2s) allows
setting of the display contrast and the backlighting (see § 3.1).
INSTRUMENT CONFIGURATION
- Press the HELP button for more than two seconds.
- Press the ◄► buttons to adjust the contrast.
- Press the ▲▼ buttons to adjust the brightness.
- Press the HELP button to exit.
These adjustments are stored even after the instrument is switched off.
3.2 Selecting the Language
NOTE: The language selection is only available in instruments
manufactured after January 2014.
To select the language, press the CONFIG button and hold it down while turning
the switch from OFF to SET-UP.
The language selection menu displays all of the languages
available. Use the ▲▼ keys to choose your language and press the ◄ key to
confirm or ◄ to cancel.
Installing the new language may take up to 30 seconds. The device then
reboots.
3.3 Choosing the Measurement of the Lead Compensation
The lead compensation is only possible if the internal firmware version allows
it and only with the red lead that is shipped with the instrument (k22 is
marked at each end). For the most recent firmware, visit
www.aemc.com.
To select the lead compensation, press the FILTER button and hold it down
while turning the switch from OFF to SET-UP.
3.4 Configuring the Instrument (SET-UP)
The Models 6550/6555 are factory configured so that they can be used without
modifying the parameters.
For most measurements, simply choose the test voltage and press the START/STOP
button.
Most parameters can be configured using the CONFIG button or in the SET-UP
function.
- The SET-UP function allows overall configuration of the instrument independently of which measurement functions are chosen.
- The CONFIG button allows configuration of the chosen measurement function before and during a measurement.
A configuration modified in one of these two ways is updated for both (SET-UP
function or CONFIG button).
3.4.1 Configuring the Instrument using the SET-UP Function
- Turn the rotary switch to the SET-UP position.
- Select and modify a parameter, using the ▲▼◄ ► buttons.
3.4.2 Configurable Parameters
- Buzzer: Sets the audible level of beeps: 1, 2, 3, or Off (no sound).
- Auto Power OFF (power management): On (instrument will turn off after 5 minutes of no activity), Off (instrument will not turn off automatically).
- Baud Rate: Sets the data rate of the serial interface to 9600, 19200, 38400 or 57600 bauds.
- Date: Sets the date in yyyy-mm-dd format.
- Time: Sets the time in hh:mm format.
- Temperature Unit: Chooses the temperature unit: Celsius or Fahrenheit.
- Instrument Number: Indicates the number of the instrument. This line is informative and cannot be modified.
- Firmware: Indicates the two version numbers of the firmware in the instrument. This line is informative and cannot be modified.
3.4.3 Resetting the Default Parameters
To return to the initial configuration, choose Set Default Parameter.
The instrument requests confirmation.If you choose OK, the following default
parameters will be reset:
- Audible level of the buzzer: 1
- Auto Power OFF: will be set to OFF
- Data rate: 38,400 baud
- Programmed measurement duration: 2 minutes
- Sampling duration: 10 seconds
- DAR: 30/60 and PI: 1/10
- Test type: Burning
- Maximum output current: 5mA
- Maximum output voltage: 10kV (6550), 15kV (6555)
- Adjustable test voltages: 50V, 800V and 7000V
- Ramp and step test voltages return to their original values, as do all alarm thresholds
- The backlight will be turned off
3.4.4 Measurement Parameters
Press the DISPLAY button to see the following:
Timed Run: Set the measurement duration (in minutes:seconds) for measurements
with programmed duration.
– The adjustment range is from 00:01 to 99:59, in 1-second steps.
DAR: Set the time at which the measurements must be recorded to calculate
the DAR (see § 4.6).
This can be used in special applications.
– The first time can be set from 10 to 90 seconds in 5-second steps.
– The second time can be set from 15 to 180 seconds in 5-second steps.
PI: Set the time at which the measurements must be recorded to calculate
the PI (see § 4.6). This can be used in special applications.
– The first time can be set from 0.5 to 30 minutes in 0.5- then 1-minute
steps.
– The second time can be set from 0.5 to 90 minutes in 0.5-, 1-, and 5-minute
steps.
Press the DISPLAY button to see the following:
Test Type: Choose the type of test: Burning, Early-Break, or
Break at I-Limit.
Maximum Output Current: Set the maximum output current, from 0.2 to 5mA.
Maximum Output Voltage: Set the maximum output voltage. This can be useful to
prevent handling errors, thus making it possible to entrust the instrument to
less experienced users for particular applications (e.g. telephone, aviation,
etc.) where it is important not to exceed a maximum test voltage.
Example: If the maximum voltage is set to 750V, the measurement will be made
at 500V for the 500V fixed voltage, and at a maximum of 750V for all other
fixed voltages.
The adjustment range is from 40 to 10,000V (15,000V for Model 6555).
3.4.5 Adjustment of the Test Voltages
Always on the third SET-UP screen.
Adjustable Voltage 1, 2, and 3: Set the values of the 3 adjustable test
voltages.
The adjustment range is from 40 to 15,000V.
Press the DISPLAY button to see the following:Set Step Function 1, 2, and 3: In the case of a measurement
with a stepped voltage, used to set the voltages and the durations of the
steps.
Pressing the ► button opens the following screen. Press the DISPLAY button to
view the rest of the steps.You can then set the voltage and duration of
each of the 10 steps. The total duration of the measurement (Total Run Time)
is calculated by the instrument.
The adjustment range of the voltages is from 40 to 15,000V.
The duration of the steps ranges from zero or 0:01 to 99:59. If a duration is
set to zero, the time displayed is -:–.
The setting should not be less than 30 seconds, because acquiring a stable
resistance result takes some time.
If either a step voltage or a step duration is set to zero, the step will be
set to zero as a whole and will be skipped during the test.
Press the ◄ button to exit the menu and return to the main SET-UP menu.
Set Ramp Function 1, 2, and 3: In the case of a measurement with a ramped
voltage, used to set the starting voltage, the slope of the ramp, and the
final voltage.
Pressing the ► button opens the following screen:You can then set the voltage and duration of the starting
level and of the final level, along with the duration of the ramp. The total
duration of the measurement (Total Run Time) is calculated by the instrument.
The voltages can be adjusted in two ranges: between 40V and 1100V or between
500V and 15,000V.
The duration of the steps can range from Start 0:30, Ramp 0:10, End 0:10 to
99:59.
Press the ◄ arrow button to exit the menu and go back to the main menu.
3.4.6 Adjustment of the Alarm Thresholds
Press the DISPLAY button to see the following:
An audible alarm is triggered for voltages below the following alarm
thresholds. There is one for each fixed or adjustable voltage and all of them
can be modified.
– For a test voltage of 500V, the alarm threshold is
adjustable from 10kΩ to 2TΩ.
– For a test voltage of 1000V, the alarm threshold is adjustable from 10kΩ to
4TΩ.
– For a test voltage of 2500V, the alarm threshold is adjustable from 10kΩ to
10TΩ.
– For a test voltage of 5000V, the alarm threshold is adjustable from 10kΩ to
15TΩ.
– For a test voltage of 10,000V, the alarm threshold is adjustable from 10kΩ
to 25TΩ.
– For a test voltage of 15,000V, the alarm threshold is adjustable from 10kΩ
to 30TΩ.
– For the adjustable test voltages, the alarm threshold depends on the
voltage. It is adjustable between two values that depend on the test voltage.
Press the DISPLAY button again to return to the first SET-UP screen.
OPERATION
4.1 Charging the Batteries
NOTE: When using the instrument for the first time, start by fully
charging the batteries. Charging must be conducted at a temperature between
32° and 86°F (0° and 30°C).
To charge the batteries:
- Set the rotary switch to the OFF position.
- Connect the power cord.
During the charging, the instrument displays the following
information:The percentage charge of each of the batteries,
their voltages, their charging currents, their temperatures, and the charging
times. To reduce the power to be supplied and make it possible to use the
instrument during the charging, the batteries are charged alternately at 2A
for 10 seconds. It is for this reason that the charging current varies.
The text on the side indicates:
- Charging = battery being charged
- Full = battery fully charged
- Cold = battery too cold to be charged
- Hot = battery too hot to be charged
- Defect = battery faulty (must be replaced)
Charging time: Between 6 and 10 hours, depending on the initial charge
condition.
When the instrument is switched-on, the battery state information is
accessible by pressing the HELP button followed by TEMP button. Following
prolonged storage, the batteries may be completely discharged. In this case,
the first charge may take longer.
It is possible to charge the instrument during operation. In this case the
symbol flashes.
The charging current then depends on the test voltage and on the resistance
measured. If the power necessary for the measurement approaches 10W, the
batteries are not charged.
4.2 Using the Leads
Specific leads are supplied with the instrument. To use them, attach either
the test probes or alligator clips (supplied with the instrument).
NOTE: These accessories have hand guards. For safety reasons, the user’s
hands must always be behind the hand guard.
Always keep hands behind the protective guards indicated below:Measurements of voltages ≥1000V on supply lines should be made
using the test probes only, with the user’s hands behind the hand guard on the
lead.
4.3 AC/DC Voltage Measurement
Turning the switch to any insulation measurement position (V-FIXED, V-VAR,
V-RAMP, or V-STEP) sets the instrument to AC/DC voltage measurement. The
voltage between the input terminals is measured at all times and indicated as
RMS value on the display unit: Input Voltage. Switching between AC and DC mode
is automatic.
In the case of an AC signal, the instrument measures the frequency. It also
measures the residual DC current between the terminals of the instrument. This
measurement is used to evaluate its impact on the insulation measurement about
to be made.
The insulation measurements cannot be started if there is an excessively high
external voltage (>0.4VN where VN is the test voltage, with a maximum of
1000VAC) on the terminals.When the external voltage exceeds 25V,
the blinking ( **) symbol is displayed alongside it.
NOTE:** By pressing the DISPLAY button, the screen can be switched to a big
numerical indication of the input voltage and its bargraph.
The only errors possible in a voltage measurement are:
- The frequency is outside the measurement range (see § 7.2.1)
- The voltage is outside the measurement range (see § 7.2.1)
4.4 Insulation Measurements
Insulation measurements are made on objects that are de-energized.
This measurement varies greatly with the temperature and the relative
humidity. It is therefore essential to measure them with a separate device and
to record them with the insulation value.
The ambient temperature can be entered in the instrument and
stored with the measurement results.
The value of the test voltage is generally twice the voltage at which the
object to be tested is used, unless a standard stipulates otherwise.
For example, for a motor that operates on a 230V supply, the test will be
performed at 500V.
4.4.1 Description of the Measurement Principle
The instrument generates a DC test voltage equal to the chosen nominal voltage
VN between the + and – terminals. More precisely, the value of this voltage
depends on the resistance to be measured (see the curves of § 7.2.3). The
instrument measures the voltage and the current between the two terminals and
from them deduces R=V/I.
The instrument measures the external voltage present on the terminals. It can
make the measurement if the peak voltage is less than 0.4VN or 1000VAC
maximum. Beyond this value, it does not make the measurement.
4.4.2 Using a Fixed Voltage
Set the switch to the V-FIXED position. The following screen appears:
Use the ◄ ► ▲▼ buttons to select the test voltage: 500V,
1000V, 2500V, 5000V, 10,000V or 15,000VDC.
The device generates exactly the voltage selected if the resistance to be
measured is indeed greater than RN = UN / 1mA. If the resistance measured is ≤
RN , the output voltage is less UN. In this case, use the U-VAR function and
adjust U so that the voltage displayed during the test is at the desired value
(see § 2.4.3.2).
4.4.3 Using a Variable Voltage
Set the switch to the V-VAR position. The following screen appears:
There are already 3 preset voltages that can be modified in
SET-UP (see § 3.4). Use the ▲▼ buttons to select one of them:
- Adjustable Voltage 1: 50V
- Adjustable Voltage 2: 800V
- Adjustable Voltage 3: 7000V
Otherwise, use the ◄► buttons to go to the voltage value, then use the ▲▼
buttons to adjust the value of the test voltage. The adjustment is in 10V
steps up to 1000V, and in 100V steps above 1000V. Keeping the buttons pressed
will speed up the adjustment. 4.4.4 Using a Voltage Ramp
This test is based on the principle that an ideal insulation produces the same
resistance whatever the test voltage applied.
Any negative variation of the insulation resistance therefore means that the
insulation is defective: the resistance of defective insulation decreases as
the test voltage increases. This phenomenon is barely observable with low test
voltages. At least 2500V must therefore be applied.
Since the application of the voltage is gradual, it causes no premature ageing
or deterioration of the device tested. Unlike the increase in steps, the
gradual increase of the current means that the capacitive current is constant.
A variation of the current therefore directly represents a variation of the
insulation resistance.
Evaluating the result:
- A negative slope of the resistance versus test voltage curve exceeding 500ppm/V generally indicates the presence of mildew or other deterioration.
- A larger negative slope, or a sudden drop, indicates the presence of localized physical damage (arcing, perforation of the insulation, etc).
The test with a voltage ramp is ideally suited for testing semiconductors
(diodes, transistors, and thyristors).
Take care in this case to choose a non-destructive type of test: Break at
I-limit (see § 2.4.3) and a maximum output current less than or equal to 1mA.
Set the switch to the V-RAMP position. The following screen appears:Use the ▲▼ buttons to select a preset test voltage ramp:
- Ramp function 1: 50 to 500V
- Ramp function 2: 500 to 5000V
- Ramp function 3: 1000 to 10,000V
The voltages at the beginning and end of the ramp can be programmed with the
CONFIG button (see § 2.4.3).
The duration of the test is the sum of the three durations specified: the
duration of the initial level, the duration of the ramp, and the duration of
the final level.
4.4.5 Using a Stepped Voltage
The preset stepped voltage tests have ten levels. The duration of each of the
voltage levels is identical. At the end of each level, the capacitive current
should be zero and only the measurement current remains. Unlike the ramp test,
the step test stresses the insulation and can cause a breakdown. A sudden
increase of the current (or a sudden decrease of the insulation resistance)
means that a breakdown point is near. It is then possible to discontinue the
measurement manually by pressing the START/STOP button or automatically using
E-BRK or Break at I-Limit type of test (see § 2.4.3).
A drop of 25% or more between the insulation resistance of the first level and
that of the second level is a sign of deterioration of the insulation.
Set the rotary switch to the V-STEP position. The following screen
appears:Use the ▲▼ arrow buttons to choose the preset
step type of test:
- Step function 1: 50 to 500V
- Step function 2: 500 to 5000V
- Step function 3: 1000 to 10,000V
The voltage and the duration of each step can be programmed using the CONFIG
button (see § 2.4.3).
4.4.6 Connection
Depending on the measurements to be made, there are three ways of connecting
the instrument.
In all cases, disconnect the device to be tested from the supply line.
■ Low Insulation
Connect the red high-voltage lead between earth and the + terminal of the
instrument. Connect the black highvoltage lead between one phase of the motor
and the – terminal of the instrument. ■ High Insulation
For very high insulation values, connect the small blue high-voltage lead
between the rear pick-up of the black lead and the G terminal – in this case
the shield of black lead will be connected to the ground of the instrument,
which helps to avoid leakage current and capacitive current effects.
This reduces the influence of the operator’s hands on the test leads and gives
a more stable measurement. ■ Cable
Connect the red high-voltage lead between the braid and the + terminal of the
instrument.
Connect the black high-voltage lead between the core and the – terminal of the
instrument.
Connect the blue high-voltage lead between the insulation and the G terminal
of the instrument.
Using
the guard eliminates the influence of the surface leakage current.
A conducting wire must be wrapped around the insulation.
4.4.7 Before the Insulation Measurement
It is possible to configure the measurement using the CONFIG button:
If the V-FIXED or V-VAR test voltages have been selected, it is possible to
select a measurement configuration by pressing the CONFIG button:
- Manual Stop
- Manual Stop + DD
- Timed Run
- Timed Run + DD
- DAR
- PI
Then set the type of test, the maximum current, the current range, the filtering of the measurement, and the value of the alarm threshold by pressing the DISPLAY button:
- Test Type
- Maximum Output Current
- I-range
- Disturbance Level
- Alarm
To enable the alarm, press the ALARM button. The ALARM symbol will appear in
the status line. The ALARM symbol in the status line will blink and (if the
buzzer is switched on) an audible beep will sound if the result of the
measurement is below the programmed threshold.
NOTE: The DISPLAY key is used to change between different screens of the
same menu. When re-entering a menu, the last used screen is shown.
4.4.8 During the Insulation Measurement
Press the START/STOP button to start the measurement.
The instrument generates high voltage. To indicate that the measurement is in
progress, the instrument emits an audible beep every ten seconds (if the
buzzer is switched on) and the START/STOP button lights up red.If
the test voltage generated is >5000V, the START/STOP button blinks.
After a few seconds, the measurement is displayed in digital form and in
analog form on a bargraph.If the measurement is unstable, it is possible
to apply a digital filter by pressing the FILTER button.
It is possible to view the available result values by pressing the DISPLAY
button.
In the case of a step test voltage (10 steps at most) or ramp
test voltage (3 steps), the progress of the steps will be indicated.
You can view the graphical representation of the measurement
results by pressing the GRAPH button.For V-VAR and V-FIXED it is also
possible to change measurement parameters during the measurement, by pressing
the CONFIG button. It is possible to fix the measurement range of the current,
to add an analog filter (disturbance level low/high), or to change the test
voltage if in the variable test voltage mode (V-VAR). For more details, refer
to § 2.4.3.
In the case of a ramp measurement, the resistance displayed is always greater
than the true resistance because of the permanent capacitive current due to
the permanent variation of the voltage. The value displayed will be exact only
at the end of the test, during the voltage level.
When the instrument is configured for a manual stop, once the measurement is
stable, press the START/STOP button again to stop the measurement. In the
other cases (programmed duration: Timed Run, Timed Run + DD, DAR, PI, V-RAMP,
or V-STEP), the measurement stops automatically at the end of the test.
At the end of the measurement, the instrument switches back to voltage
measurement, but usually the result of the insulation resistance measurement
remains displayed. To display the voltage, press the DISPLAY button. In case
of an external voltage >25V the instrument automatically switches to the
screen with the test description and small input voltage indication.
4.4.9 After the Insulation Measurement
Once the measurement has been stopped, the instrument discharges the device
being tested in a few seconds. For your safety, therefore, wait before
disconnecting the leads. Normally, this happens rapidly and the user is
unaware of it. But if the load is highly capacitive, the discharging time is
longer. In this case, for as long as the voltage exceeds 25V, the instrument
indicates it on the display.
The DISPLAY button is used to look up all information
available after the measurement. This information depends on the type of
measurement chosen (see § 2.4.4).
In the case of a ramp or step measurement, the measurement result is displayed
as follows:
Record the measurement and compare it to earlier measurements
in order to assess the evolution of its value.
Also record the temperature and the ambient relative humidity.
If, at equivalent temperature and humidity, the insulation resistance has
fallen significantly, the insulation is deteriorated and maintenance must be
carried out on the device tested.
The result remains displayed until another measurement is made, the rotary
switch is turned to a different position, or the measurement configuration is
changed.
After a test with programmed duration:
- Pressing the GRAPH button displays the graphical representation of the results (see § 2.4.5).
- Only for V-FIXED and V-VAR: Pressing the TEMP button opens the temperature menu (see § 2.4.1).
- Pressing the MEM button opens the recording menu (see § 5.1).
- At any time, you can press the HELP button for a reminder of the functions of the buttons.
4.5 Error Indications
The most common error in the case of an insulation measurement is the presence
of a voltage on the terminals. The instrument can make the measurement if the
peak value of this voltage is less than 0.4VN or 1000VAC max.
Above this value, it is necessary to eliminate the voltage and repeat the
measurement.
If an external voltage appears on the terminals during the measurement, and
its peak value is greater than 1.1VN, the measurement is stopped and the error
is indicated.
4.6 DAR (dielectric absorption ratio) and PI (polarization index)
For V-FIXED and V-VAR: In addition to the quantitative value of the insulation
resistance, it is very useful to calculate the quality ratios of the
insulation (the DAR and the PI) because they can eliminate the influence of
certain parameters likely to invalidate the “absolute” insulation measurement.
They also serve to predict the evolution of insulation quality over time.
The most important of the parameters influencing the measurement results are:
- temperature and relative humidity, with which insulation resistance varies according to a quasiexponential law.
- the disturbance currents (capacitive charging current, dielectric absorption current) created by the application of the test voltage. Even if they gradually fade, they interfere with the measurement at the start, for a more or less long time depending on whether the insulation is sound or degraded.
These ratios therefore complete the “absolute” insulation value and reliably
reflect the condition, good or bad, of the insulation.
In addition, long-term observation of the evolution of these ratios is a way
to monitor the aging of the insulation. For example, that of a revolving
machine or of a long cable.
The values of DAR and PI are calculated as follows:
- DAR = R 1 min/R 30s (2 values to be taken during a 1-min measurement)
- PI = R 10 min/R 1 min (2 values to be taken during a 10-min measurement)
The times of 1 and 10 minutes for the calculation of the PI and the times of
30 seconds and 1 minute for the calculation of the DAR can be modified in the
CONFIG menu or the SET-UP function (see § 3.4), to adapt to particular
applications.
4.6.1 DAR/PI Measurement
There are several ways of measuring the DAR and the PI:
■ Manual Configuration:
Press the START/STOP button. Wait one minute for the DAR or ten minutes for
the PI (if the default values are used).![AEMC 6550 10kV and 15kV Megohmmeters
- displayed 39](https://manuals.plus/wp-content/uploads/2023/03/AEMC-6550
-10kV-and-15kV-Megohmmeters-displayed-39.png)Press the START/STOP button again
to stop the measurement. ■ Automatic Configuration (preferred)
Press the CONFIG button.Use the ▲▼ arrow buttons to select DAR or PI.Press the CONFIG button to exit the configuration menu.
DAR or PI is displayed in the top left corner of the display unit to indicate the configuration chosen.
Press the START/STOP button to start the measurement. It stops automatically and the values of DAR and PI are displayed.
4.6.2 Interpretation of the Results
DAR | PI | Condition of insulation |
---|---|---|
DAR < 1.25 | PI < 1 | Poor or even dangerous |
1 £ PI < 2
1.25 £ DAR < 1.6| 2 £ PI < 4| Good
1.6 £ DAR| 4 £ PI| Excellent
A capacitance in parallel to the insulation resistance extends the settling times of the measurements. This can affect or even inhibit the measurement of DAR or PI (depending on the time set for recording the first resistance value). The table below indicates the typical values of the capacitances in parallel with the insulation resistance, making it possible to measure the DAR and the PI without changing their preset durations.
| 100kΩ| 1MΩ| 10MΩ| 100MΩ| 1GΩ| 10GΩ|
100GΩ
---|---|---|---|---|---|---|---
500V| 10µF| 10µF| 10µF| 6µF| 4µF| 2µF| 1µF
1000V| 5µF| 5µF| 5µF| 3µF| 2µF| 1µF| 0.5µF
2500V| 2µF| 2µF| 2µF| 1.2µF| 1µF| 0.5µF| 0.2µF
5000V| 1µF| 1µF| 1µF| 0.6µF| 0.4µF| 0.3µF| 0.1µF
10,000V| 0.5µF| 0.5µF| 0.5µF| 0.3µF| 0.2µF| 0.1µF| 0µF
15,000V| 0.3µF| 0.3µF| 0.3µF| 0.2µF| 0.1µF| 0.1µF| 0µF
4.7 DD (dielectric discharge index)
In the case of multilayer insulation, if one of the layers is defective but
the resistance of all the others is high, neither the quantitative insulation
measurement nor the calculation of the PI and DAR quality ratios will reveal
the problem.
This makes it judicious to perform a dielectric discharge test, from which the
DD term can be calculated. This test measures the dielectric absorption of
heterogeneous or multilayer insulation and disregards parallelsurface leakage
currents.
The dielectric discharge test is especially well suited for measuring the
insulation of revolving machines and more generally for measuring the
insulation on heterogeneous or multi-layer insulating materials containing
organic substances.
It involves applying a test voltage for long enough to electrically “charge”
the insulation to be measured. At the end of the measurement, the instrument
induces rapid discharging, during which the capacitance of the insulation is
measured, then, one minute later, it measures the residual current flowing in
the insulation. The DD term is then calculated as follows:
DD = current measured after 1 minute (mA)/[test voltage (V) x measured
capacitance (F)] 4.7.1 DD Measurement
Press the CONFIG button.Use the▲▼ arrow buttons to select Manual Stop +
DD or Timed Run + DD (manual or automatic measurement).To set the duration of the measurement, place the cursor on
Timed Run (m:s). Then use the ◄ ► and▲▼ arrow buttons to set the minutes and
seconds. The minimum setting is 0:01 but a duration below 30 seconds is hardly
useful because acquiring a stable resistance result requires some time.Once the duration has been set, move the cursor back to Timed
Run + DD.
■ Press the CONFIG button to confirm and exit the configuration menu. DD or
DD is displayed in the top left corner of the display unit to indicate the
configuration chosen.
■ Press the START/STOP button to start the measurement.
In the Manual Stop + DD configuration, wait until the elapsed time is greater
than one minute, then press the START/STOP button to stop the measurement.
In the Timed Run + DD configuration (indicated by the symbol), the
measurement stops automatically.
In both cases, it is necessary to wait one minute after the measurement is
stopped (countdown on the display unit) for the instrument to display the
result. During this time, the START/STOP button is lit but the instrument does
not emit an audible signal.
The result is then displayed. 4.7.2 Interpretation of Results
Value of DD | Quality insulation |
---|---|
7 < DD | Very poor |
4 < DD < 7 | poor |
2 < DD < 4 | Borderline |
DD < 2 | Good |
4.8 Capacitance Measurement
The capacitance measurement is made automatically during the insulation
measurement, and is displayed after the measurement has been stopped and the
device tested has been discharged.
4.9 Measurement of the Residual Current
The measurement of the residual current flowing in the device tested is made
automatically as soon as the connection to the device tested is made, then
during and after the insulation measurement.
MEMORY FUNCTION
5.1 Recording a Measurement
Each insulation measurement can be recorded once the measurement process is
completed. It is not possible to record input voltage measurements prior to
starting a test.
The results are recorded at addresses identified by an object number (OBJ) and
a test number (TEST).
An object can contain 99 tests. An object can represent a machine or an
installation on which a certain number of measurements will be performed.
- At the end of the measurement, press the MEM button.
- The instrument proposes recording the result at the first available location in memory. It is possible to modify the proposed location using the ◄ ► and▲▼ arrow buttons.
If the screen does not display the measurement and pressing the MEM key has no
effect, press the DISPLAY key twice to restore the result screen, then press
the MEM key again.
This May happen following the discharging of a highly capacitive load.
- Press the MEM button again to confirm the storage location.
- The instrument then asks you if you want to Store Samples with the measurement.
The graph of the measurement can then be displayed by a single press on the
GRAPH button (see § 2.4.5).
If this is not required, set Store Samples to OFF.
If Store Samples are set to Yes, the Sample Time can be set using the ◄ ►
and▲▼ arrow buttons.
-
The default sampling time (the time difference between two stored samples) is the minimum, meaning that all samples acquired during the measurement are recorded.
-
The sampling time can be set to Auto (automatic), in which case the instrument itself determines the samples necessary for the plotting of the measurements while using the least possible memory. If the measurement does not vary, it will take only one value, giving a perfectly flat curve.
This value is recommended to optimize memory use.
The sampling time can also be programmed, between 1 and 25 seconds. -
The longer the measurement, the longer the sampling time can be. For example, on a measurement lasting 10 minutes, the sampling time can be 10 seconds, giving 60 points for the graph, which is sufficient.
-
Again, the more stable the measurement, the longer the sampling time can be. The more unstable the measurement, the shorter the sampling time must be in order to correctly display variations of the insulation resistance.
Press the MEM button one last time to record the measurement. The instrument confirms the storage.
The measurement is recorded with all its supporting
information.
To exit without recording and return to the last measurement, press the ◄
button.
For each new record, the instrument proposes the first free memory location
following the last stored measurement. It is also possible to record a
measurement at a memory location that has already been used.The number of measurements that can be recorded depends on
the number of samples stored for each measurement.
The instrument store 256 measurements and 80,000 samples associated with these
measurements.
5.2 Reading Recorded Values
Turn the rotary switch to the MR position.
The instrument indicates the memory used and the object number of the last
stored record, along with the lowest and highest test numbers that it
contains.Choose the object number using the ▲▼ arrow
buttons, then press the ► button.
The instrument then displays the list of records around the object chosen.
To see details of a measurement, place the cursor on the object and the test
chosen using the▲▼ arrow buttons, then press the ► button.
Press the DISPLAY button to see the rest of the recorded
information (depending on the currently used function).
When the symbol indicates that the samples have been recorded, you can press
the GRAPH button to view the graph.
Press the GRAPH button to exit from the graph.
In the case of a V-FIXED or V-VAR measurement, you can press the TEMP button
to view the temperature information. The instrument can display only the
information recorded with the measurement.
Press the TEMP button, then the ◄ arrow button, to exit and return to the list
of recorded measurements.
5.3 Erasing the Memory
Turn the rotary switch to the MR position. 5.3.1 Erasing One Record
With an object selected, use the ▲▼ arrow buttons to select the record to be
erased in the list of records in memory.
Press the CONFIG button. The instrument requests confirmation of the
deletion.Select OK to confirm or CANCEL to cancel. The
instrument will return to the top level memory screen.
NOTE: Deleting a single record doesn’t delete it physically – only the link
to it is deleted. The record number can be used again for storing another
measurement. Physical clearing of data is only accomplished when clearing the
whole memory.
5.3.2 Erasing All Records
Press the CONFIG button. The instrument requests confirmation of the deletion.
Select OK to confirm or CANCEL to cancel.The instrument in this case completely reformats the memory,
which takes a few minutes. During this time, it displays “WAIT”.
The instrument then returns to the top level memory screen. Since there is no
longer any record, it displays:
5.4 Error Codes
If an anomaly is detected when the instrument is turned on or in operation,
the display indicates a one- or twodigit number error code. This number
identifies the anomaly and states what to do to put the instrument back into
service.
■ Errors 1 to 9 concern internal circuit board problems and require a repair
by a qualified individual.
■ Error codes 20 to 25 help the repair personnel locate a malfunction
■ If the recorded data is corrupted, the only way possible to reuse the memory
is to erase it completely (see § 5.3.2). The instrument reports this problem
by displaying “CLEAR MEMORY”.
■ All other errors require returning the instrument for repair (see §
8.5).
There are three types of error messages:
■ Informative error messages:
The message appears for approximately 1 second. Depending on the error the
functionality of the device may be reduced. A repair is needed if the error
recurs.
Errors 04, 06, 07, 20, 21, 23, 30, 31, 32 (see also the second type of error),
40, 41, 42
Error 06 is preceded by an automatic reset.
Errors 04 and 07 are followed by error 06.
Error 20 indicates that a memory operation has failed.
Error 21 indicates that the settings have been automatically reset to default
settings.
Error 23 indicates that the battery management is not available and no battery
charging is possible.
Error 30 indicates that a resistance measurement has been stopped
unexpectedly; check for disturbances.
Errors 31, 32 (see also the second type of error) and 40 indicate that no
measurement is possible.
■ Recoverable error messages:
The message disappears if the rotary switch position is changed. Depending on
the error the functionality
of the device may be reduced. A repair is needed if the error recurs.
Errors 22, 32 (see also the first type of error)
Error 32 (see also the first type of error)) indicates that no measurement is
possible.46
■ Fatal error messages:
No operation is possible. Power off the device and power it back on. A repair
is needed if the error recurs.
Errors 01, 08, 09
In addition to error messages there are other indications for errors:
-
If the device shows a cross on the screen when powered on and after a few seconds additionally a horizontal bar on the top of the screen, this indicates that language data is needed.
Use the PC program from our web site (see §6.1) to update firmware and language data. -
If instead of information on the help screen just the headline “HELP” and below the number 98 or 99 is shown, this indicates that language data for help is needed.
DATAVIEW® SOFTWARE
For complete information on using the Megohmmeter with DataView, refer to the
Help Menu within the Megohmmeter Control Panel in DataView.
6.1 Installing DataView
DataView® is a registered trademark of Chauvin Arnoux® d.b.a. AEMC®
Instruments.
DO NOT CONNECT THE INSTRUMENT TO THE PC BEFORE INSTALLING THE SOFTWARE AND
DRIVERS.
-
Insert the DataView thumb drive into an available USB port on your computer. If Autorun is enabled, an AutoPlay window appears on your screen. Click “Open folder to view files” to display the DataView folder.
If Autorun is not enabled or allowed, use Windows Explorer to locate and open the USB drive labeled “DataView.” -
When the DataView folder is open, find the file Setup.exe located in the root directory of the USB drive, and double-click it to run the installation program.
-
The DataView setup screen appears. In the upper left corner of the screen, choose the language version of the Setup interface. (All Setup screens and dialogs will immediately appear in the selected language.)
In the lower left corner are the available installation options. In addition to the DataView software, you can select “Adobe Reader.” This links to the Adobe web site where you can download the latest version of Reader. This program is required to view DataView .pdf documents. The option Firmware Upgrades links to the website where you can check for new firmware updates for the instrument. Finally, User Manuals displays a list of .pdf files contained in the USB drive that accompanies DataView. (DataView also comes with a Help system that is installed with the program files.)
To install DataView, select DataView in the Options list and click Install. -
Select the language version of DataView you want to install (English, French, or Spanish) then click Next. (By default, the language selected in step 3 is highlighted.)
-
You are now prompted to select the software you want to install. Each AEMC product family has its own specially designed Control Panel. If you are performing a Complete install, by default all available Control Panels are selected (a check mark next to the Control Panel indicates it is selected). Control Panels take up disk space on the computer; so unless you have other types of AEMC instruments, we recommend that you select Megohmmeter and deselect the rest. You should also check the option DataView Core, which is a requirement if you plan to create DataView reports. After you finish selecting and deselecting Control Panels and/or DataView Core, click Next.
-
The Setup program now informs you that it is ready to install DataView. If you want to review any of your previous selections, click the Previous button to return to earlier screens. Otherwise, click Install to begin installation.
-
The InstallShield program installs the selected software. If an earlier version of the software is already installed on your computer, for each selected program the InstallShield program will:
(a) Ask you to confirm the installation of the program. Click Next.
(b) Display a status bar indicating the progress of the installation.
(c) Inform you when the program is installed. Click Finish to insta ll the next selected program.
If the software is not installed (or if the installed software is the same version as the selected software), the software is installed without requesting confirmation. When all programs are installed, a message appears informing you of this. Click Finish to return to the Setup screen -
You can now select additional Setup options to install (see step 3 above). When finished, click Exit.
-
The DataView folder now appears on your computer desktop, within which is the Megohmmeter icon and the icon(s) for any other Control Panel(s) you have installed.
6.2 Megohmmeter Control Panel
Clicking the DataView icon in the DataView folder on your desktop opens the
core DataView program. Clicking the Megohmmeter Control Panel icon opens the
Megohmmeter Control Panel.
In general, core DataView features are for creating, viewing, editing, and
storing DataView reports; while the Control Panel is for connecting to,
configuring, viewing measurements on, and downloading data from the
instrument. You can access all DataView features through either the DataView
icon or the Control Panel icon.
For users who interact with megohmmeter instruments, we recommend primarily
using the Control Panel.
However, there are situations where using the core DataView icon may be more
convenient for some users, such as when viewing multiple archived reports from
different AEMC product families.
For further information about using the Megohmmeter Control Panel, consult the
Help system that comes with the product. Access this Help by clicking the
option Help in the Control Panel’s menu bar at the top of the screen.
SPECIFICATIONS
7.1 Reference Conditions
Influence Quantities | Reference values |
---|---|
Temperature | 23 ± 3°C |
Relative Humidity | 45 to 55% RH |
Supply Voltage | 9 to 12V |
Frequency Range | DC and 15.3 to 65Hz |
Capacitance in Parallel on Resistance | 0µF |
Electric Field | null |
Magnetic Field | <40A/m |
The inherent accuracy is the error specified for the reference conditions.
The operating accuracy includes the inherent uncertainty plus variations of
the quantities of influence (supply voltage, temperature, interference, etc.)
as defined in standard IEC-61557.
7.2 Electrical Specifications
7.2.1 Voltage
Measurement Range| 1.0 to 99.9V| 100 to 999V| 1000 to
2500V| 2501 to 4000V
---|---|---|---|---
Resolution| 0.1V| 1V| 2V| 2V
Accuracy| ±(1% +5cts)| ±(1% +1ct)
Frequency range| DC or 15 to 500Hz| DC
■ Input impedance: 3MW
7.2.2 Current
Measurement Range (DC)| 0.000 to 0.399nA| 0.400 to 3.999nA|
4.00 to 39.99nA| 40.0 to 399.9nA| 400 nA to 3.999µA
---|---|---|---|---|---
Resolution| 1pA| 1pA| 10pA| 100pA| 1nA
Accuracy| ±(15% + 10cts)| ±10%| ±5%
Measurement Range (DC)| 4.00 to 39.99µA| 40.0 to 399.9µA|
400 µA to 3.999mA| 4.00 to 9.999mA
---|---|---|---|---
Resolution| 10nA| 100nA| 1µA| 10µA
Accuracy| ±5%
7.2.3 Insulation Resistance
■ Method: Voltage-current measurement per IEC-61557-2 from 300 to 10,000V and
per DIN VDE 0413 Part 1/09.80)
■ Nominal output voltage: 500V, 1000V, 2500V, 5000V, 10,000V, and 15,000VDC
for the Model 6555 or adjustable from 40 to 10,000VDC and 15,000VDC for the
Model 6555
Inherent accuracy ±1%
adjustable from 40 to 10,000VDC in 10-V steps
adjustable from 1000 to 15,000VDC in 100-V steps
■ Maximum current: ≤1mADC from 40 to 999V
5 to 0.2mADC from 1000 to 15,000V. The user can adjust this current.
■ Maximum acceptable peak AC voltage at terminals during measurement:
0.4VN or 1000VAC maximum
■ Short-circuit current: ≤5mADC ±5%. This current can be limited in SET-
UP (setting “maximum output current”), to between 0.2 and 5mA. It can also be
limited by the maximum output power, which is 10W.
■ Maximum output current as a function of the test voltage
U N (V)| 50| 100| 200| 300| 1100|
1200| 1300| 5000| 10,000| 15,000
---|---|---|---|---|---|---|---|---|---|---
I (mA)| 0.22| 0.46| 0.93| 1.07| 1.07| 5| 5| 2| 1| 0.5
P (W)| £ 1| 10
If the current is limited in SET-UP , the values mentioned above that
exceed the limit will be lowered.
■ Fixed Test Voltage
Test Voltage (V) | 500 – 1000 – 2500 – 5000 – 10,000 – 15,000 |
---|---|
Specified measurement range | 10 to 999kW |
1.000 to 3.999MW | 4.00 to 39.99MW |
39.99GW
Resolution| 1kW| 10kW| 100kW| 1MW| 10MW
Accuracy| ±(5% + 3cts)
Operating error| ±(10% + 6cts)
Test Voltage (V)| 500 – 1000 – 2500 – 5000
10,000 – 15,000| ³ 1000| ³ 2500| ³ 5000
---|---|---|---|---
Specified measurement range| 40.0 to 399.9GW| 400 to 999GW
1.000 to 1.999TW| 2.000 to 3.999TW| 4.00 to 10.00TW| 4.00 to 15.00TW
Resolution| 100MW| 1GW| 1GW| 10GW| 10GW
Accuracy| ±(15% + 10cts)| ±(20% + 10cts)
Operating error| ±(20% + 15cts)| ±(30% + 15cts)
Test Voltage (V)| ³ 10,000| 15,000 (6555 only)
---|---|---
Specified measurement range| 4.00 to 25.00TW| 4.00 to 29.00TW
Resolution| 10GW| 10GW
Inherent accuracy| ±(20% + 10cts)| ±(20% + 10cts)
Operating error| ±(30% + 15cts)
Variable test voltage
Minimum resistance measured = 10kΩ
Maximum resistance measured = to be interpolated from the values in the tables
of fixed test voltages above. The inherent uncertainty depends on the test
voltage and on the resistance measured. It can be interpolated from the tables
of fixed test voltages.
■ Typical discharge time of a capacitive element to reach 25 VDC
Test Voltage| 50V| 100V| 250V| 500V| 1000V|
2500V
---|---|---|---|---|---|---
Discharge Time (C at µF)| 0.25 s x C| 0.5 s x C| 1 s x C| 2 s x C| 4 s x
C| 7 s x C
Test Voltage| 5000V| 10,000V| 15,000V
---|---|---|---
Discharge Time (C at µF)| 14 s x C| 27 s x C| 57 s x C
Typical curves of the test voltages at the instrument terminals as a function of the load resistance
7.1.1 DAR, PI, and DD
■ Calculation of the DAR and PI terms
Measurement Range | 0.02 to 50.00 |
---|---|
Resolution | 0.01 |
Accuracy | ± (5% + 1ct) |
Calculation of the DD term
Measurement Range | 0.02 to 50.00 |
---|---|
Resolution | 0.01 |
Accuracy | ±( 10% + 1ct) |
7.2.5 Capacitance
Capacitance measurement
This measurement is made following the discharging of the element tested,
after each measurement
Measurement Range | 0.005 to 9.999µF | 10.00 to 19.99µF |
---|---|---|
Resolution | 1nF | 10nF |
**Accuracy*** | ± (10% + 1 ct) | ± 10% |
7.3 Power Supply
The instrument is powered by two rechargeable 9.6V, 4Ah NiMH battery packs.
Charging is carried out by connecting the instrument to line voltage of 90 to
260V and a frequency of 50/60 Hz, with an ambient temperature of 32° to 86°F
(0° to 30°C).
7.3.1 NiMH Technology
The NiMH technology has many advantages, such as:
-
long life between charges with limited bulk and weight
-
recharging capability
-
a very small memory effect: you can recharge your battery even if it is not fully discharged, without reducing its capacity
-
environmental protection through the elimination of polluting materials such as lead and cadmium
The NiMH technology allows a limited number of charging/discharging cycles
that depends on the conditions of use and the charging conditions. Under
optimum conditions, this number of cycles is 200.
7.3.2 Battery Charger
The built-in charger manages the charging current, the battery voltage, and
the internal temperature of the battery simultaneously. This optimizes the
charging, while ensuring a long battery life.
The day before you use your device, check its charge condition. If the battery
level indicator shows less than three bars, charge the device overnight (see
§4.1).
The charging time varies between 6h and 10h.
A half-hour charge restores 10% of the capacity of the battery, enough to make
a few measurements.
It is possible to recharge the batteries while making insulation
measurements, provided that the voltages used are not too high and the
measured values are high enough. In this case, the recharging time will exceed
6 hours;
it will depend on the frequency of the measurements made. Otherwise, the
battery will be discharged faster than it is charged.
In order to extend the life of your battery:
- charge your device only between 10° and 30°C
- observe the conditions of use and storage stated in this manual
A new battery becomes fully effective only after several complete
charging/discharging cycles. This will not however prevent you from using your
device when it has been charged for the first time. However, we recommend
making the first charge a full charge (at least 10 hours).
If the instrument indicates that charging is over, do not hesitate to
disconnect the charger for a few seconds, then reconnect it to top up the
charge.
The battery in your instrument, like any rechargeable battery, is subject to
significant residual discharging, even when the instrument is off. If your
device has not been used for several weeks, it is probable that the battery
will be partially discharged, even if it had been fully recharged just before
going into storage.
In this case, before using it again, you should fully recharge the battery (at
least 10 hours).
The longer your battery is stored, the more it is discharged. After three
months’ storage of the battery without periodic recharging, the battery is
probably fully discharged.
Possible consequences are:
- Failure of the instrument to switch on, as long as the power cord is not connected.
- A loss of the instrument’s date and time.
7.3.3 Optimize Battery Charging
During charging, the temperature of the battery rises substantially,
especially towards the end. A safety device, built into the battery, checks
constantly that the battery temperature does not exceed an acceptable maximum.
If this maximum is exceeded, the charger switches off automatically, even if
charging is not complete. Above 86°F (30°C), it is not possible to charge the
battery fully because the charging will cause overheating.
7.3.4 Battery Life
The mean battery life depends on the measurement and on how the device is
used.
Test Voltage (V)| 500| 1000| 2500| 5000|
10,000| 15,000| Voltmeter
---|---|---|---|---|---|---|---
Battery Life (h)| 15| 12| 2| 2| 2| 2| 25
How long your device can operate when the battery is fully charged depends on several factors:
- The consumption of the device, which depends on the measurements you make.
- The capacity of the battery is greatest when the battery is new, and declines as the battery ages.
Here are a few ways to extend battery life between charges:
- Use the back-lighting only when it is strictly necessary.
- Set the brightness of the back-lighting to the lowest level at which you can still read the display unit.
- Enable the Auto Power OFF function (see SET-UP § 3.4).
- During insulation measurements conducted in MANUAL mode, with high test voltages, stop the measurement by pressing the START/STOP button when the measurement is complete.
7.3.5 “Defect” message
When a battery is deeply discharged or its storage temperature is low, the
charger may execute a reactivation stage prior to fast charge. That means that
the charger applies a slow charge until the battery reaches either a minimum
temperature threshold or a minimum charge voltage threshold.
If the battery is in good condition, this reactivation stage ends after about
45 minutes then charger switches over to fast charging.
However, if the maximum time allowed for the reactivation stage is exceeded or
the internal resistance of a attery at the end of its life is high, the
instrument declares the battery defective in the form of a “Defect” message on
the instrument screen.
The instrument must then be sent in for repair (see § 8.5).
7.4 Environmental Specifications
-
Range of use
The relative humidity can significantly affect insulation. Take care not to make an insulation resistance measurement if the temperature is below the dew point.
32° to 113°F (0° to 45°C); 0 to 90% RH -
Specified domain of use 32° to 95°F (0° to 35°C); 0 to 75% RH
-
Storage (without the batteries) -40° to 158°F (-40° to 70°C); 10 to 90% RH
-
Altitude: <2000m
-
Degree of pollution: 2
7.5 Mechanical Specifications
- Dimensions: (LxWxH): 13.4 x 11.8 x 7.9″ (340 x 300 x 200mm)
- Weight: approximately 13.7 lb (6.2kg)
7.6 Safety Specifications
- Electrical safety per: IEC/EN 61010-2-030 or BS EN 61010-2-030, IEC/EN 61010-031 or BS EN 61010031, IEC-61557 parts 1 and 2 (up to 10kV) or VDE 0413.
- Double insulation
- Degree of pollution: 2
- Voltage measurement category: 1000V CAT IV
- Maximum voltage with respect to earth: 1000Vrms CAT IV
- Maximum voltage between guard terminal G and the – terminal: 30Vrms
7.6.1 Electromagnetic Compatibility
Emissions and immunity in an industrial environment per IEC/EN 61326-1 or BS
EN 61326-1.
7.6.2 Mechanical Protections
IP 65 according to IEC-60529 with the housing closed and IP 54 with the
housing open.
IK 04 according to IEC-50102.
7.7 Variations in the Domain of Use
Influence quantity| Range of influence| Quantity influenced
(1)| Influence
---|---|---|---
Typical| Maximum
Battery voltage| 9 to 12V| V
MW| < 1ct
< 1ct| 2cts 3cts
Temperature| -10 to +55°C| V MW – GW
U >7.5kV and R < 10TW| ±0.15%/10°C
±0.2%/10°C
±1.5%/10°C
| ±(0,3%/10°C + 1ct)
±(1%/10°C + 2cts)
±(3%/10°C + 2cts)
Humidity| 10 to 75% RH
with t £ 35°C| V MW (10kW to 40GW) MW (40GW to 10TW)
U > 7.5 kV and 3TW < R < 10TW| ±0.2%
±0.2%
±0.3%
±(15% + 5cts)| ±(1% + 2cts)
±(1% + 5cts)
±(15% + 5cts)
±(30% + 5cts)
Frequency| 15 to 500Hz| V| ±3%| ±(0.5% + 1ct)
AC voltage superimposed on test voltage| 0 to 20%Vn| MW| ±0.1%/%Vn| ±(0.5%/%Vn
- 5cts)
(1): The DAR, PI and DD terms and the capacitance and leakage current
measurements are included in the quantity “MΩ”.
7.8 Inherent and Operating Accuracy
The Megohmmeter Models 6550 and 6555 comply with standard IEC-61557, which
requires that the operating accuracy, called B, be less than 30%.
In insulation measurements, B = ± ( |A| + 1,15 √ E1² + E2² + E3² ) with A =
inherent accuracy
E1 = influence of the reference position ±90°
E2 = influence of the supply voltage within the limits indicated by the
manufacturer
E3 = influence of the temperature between 0° and 35°C
Specifications are subject to change without notice.
MAINTENANCE
8.1 Recharging the Battery
Refer § 4.1 for full details on recharging the battery.
The Megohmmeter Model 6550/6555 is equipped with a rechargeable NiMH battery.
This technology offers several advantages:
- Long battery charge life for a limited volume and weight
- Possibility of quickly recharging your battery
- Significantly reduced memory effect: you can recharge your battery even if it is not fully discharged
- Environmental protection through the elimination of polluting materials such as lead and cadmium
After prolonged storage, the battery may be completely discharged. If so, it
must be completely recharged.
Your instrument may not function during part of this recharging operation.
Full recharging of a completely discharged battery may take several hours.
In this case, at least 5 charge/discharge cycles will be necessary for your
battery to recover 95% of its capacity.
To make the best possible use of your battery and extend its effective service
life:
- Only charge your instrument at temperatures between 32° and 86°F (0° and 30°C)
- Comply with the conditions of use defined in the operating manual
- Comply with the storage conditions specified in the operating manual
Before first use, charge and discharge the instrument one or two cycles to
ensure the proper level display of the battery indicator.
8.2 Replacing the Battery
The batteries can be replaced only by the manufacturer’s service center or by
an approved repair center.
WARNING: Replacing the battery may result in the loss of the stored data.
Back up all stored data before sending the instrument in for repair.
To return the instrument for battery replacement see § 8.5.
When the repaired instrument is returned:
- Erase the memory completely (see § 5.3.2) to be able to use the MEM/MR functions again
- If necessary, reset the date and time of the instrument (see § 3.4)
- Fully recharge the battery
AEMC® will not be held responsible for any accident, incident, or malfunction following a repair done other than by its service center or by an approved repair center.
8.3 Replacing the Fuse
If the GUARD FUSE message appears on the display unit, the guard terminal fuse
must be replaced. The fuse can be replaced only by competent, accredited
personnel.
8.4 Cleaning
Disconnect the instrument from any source of electricity.
- Use a soft cloth, lightly dampened with soapy water
- Wipe with a damp cloth and then dry with a dry cloth
- Do not splash water directly on the clamp
- Do not use alcohol, solvents or hydrocarbons
8.5 Repair and Calibration
To ensure that your instrument meets factory specifications, we recommend that
it be scheduled back to our factory Service Center at one-year intervals for
recalibration, or as required by other standards or internal procedures.
For instrument repair and calibration:
You must contact our Service Center for a Customer Service Authorization
Number (CSA#). This will ensure that when your instrument arrives, it will
be tracked and processed promptly. Please write the CSA# on the outside of the
shipping container. If the instrument is returned for calibration, we need to
know if you want a standard calibration, or a calibration traceable to
N.I.S.T. (Includes calibration certificate plus recorded calibration data).
Ship To: Chauvin Arnoux® , Inc. d.b.a. AEMC ® Instruments
15 Faraday Drive
Dover, NH 03820 USA
Phone: 800-945-2362 (Ext. 360)
603-749-6434 (Ext. 360)
Fax: 603-742-2346 or
603-749-6309
E-mail: repair@aemc.com
(Or contact your authorized distributor)
Costs for repair, standard calibration, and calibration traceable to N.I.S.T.
are available.
NOTE: You must obtain a CSA# before returning any instrument.
8.6 Technical and Sales Assistance
If you are experiencing any technical problems, or require any assistance with
the proper operation or application of your instrument, please call, fax or
e-mail our technical support team:
Chauvin Arnoux®, Inc. d.b.a. AEMC ® Instruments
Phone: 800-343-1391 (Ext. 531)
Fax: +1 603-749-9153
E-mail: techsupport@aemc.com
8.7 Limited Warranty
The Models 6550/6555 are warranted to the owner for a period of two years from
the date of original purchase against defects in manufacture. This limited
warranty is given by AEMC® Instruments, not by the distributor from whom it
was purchased. This warranty is void if the unit has been tampered with,
abused or if the defect is related to service not performed by AEMC®
Instruments.
Full warranty coverage and product registration is available on our website
at:www.aemc.com/warranty.html.
Please print the online Warranty Coverage Information for your records.
What AEMC® Instruments will do:
If a malfunction occurs within the warranty period, you may return the
instrument to us for repair, provided we have your warranty registration
information on file or a proof of purchase. AEMC® Instruments will, at its
option, repair or replace the faulty material.
REGISTER YOUR PRODUCT ONLINE AT: www.aemc.com
8.8 Warranty Repairs
What you must do to return an Instrument for Warranty Repair:
First, request a Customer Service Authorization Number (CSA#) by phone or by
fax from our Service Department (see address below), then return the
instrument along with the signed CSA Form. Please write the CSA# on the
outside of the shipping container.
Return the instrument, postage or shipment pre-paid to:Ship To: Chauvin Arnoux
® , Inc. d.b.a. AEMC® Instruments
15 Faraday Drive
Dover, NH 03820 USA
Phone: 800-945-2362 (Ext. 360)
603-749-6434 (Ext. 360)
Fax: 603-742-2346 or
603-749-6309
E-mail: repair@aemc.com
Caution: To protect yourself against in-transit loss, we recommend you insure
your returned material.
NOTE: You must obtain a CSA# before returning any instrument.
Chauvin Arnoux® , Inc. d.b.a. AEMC® Instruments
15 Faraday Drive
Dover, NH 03820 USA
Phone: 603-749-6434
Fax: 603-742-2346
www.aemc.com
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>