AEMC PA32ER Power Adapter Power Supply User Manual
- June 9, 2024
- AEMC
Table of Contents
AEMC PA32ER Power Adapter Power Supply
Copyright © Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments. All rights
reserved.
No part of this documentation may be reproduced in any form or by any means
(including electronic storage and retrieval or translation into any other
language) without prior agreement and written consent from Chauvin Arnoux®,
Inc., as governed by United States and International copyright laws.
Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments
15 Faraday Drive • Dover, NH 03820 USA
Tel: 800-945-2362 or
603-749-6434 • Fax:
603-742-2346
This documentation is provided “as is,” without warranty of any kind, express, implied, or otherwise. Chauvin Arnoux®, Inc. has made every reasonable effort to ensure that this documentation is accurate; but does not warrant the accuracy or completeness of the text, graphics, or other information contained in this documentation. Chauvin Arnoux®, Inc. shall not be liable for any damages, special, indirect, incidental, or inconsequential; including (but not limited to) physical, emotional or monetary damages due to lost revenues or lost profits that may result from the use of this documentation, whether or not the user of the documentation has been advised of the possibility of such damages.
INTRODUCTION
Thank you for purchasing this PA32ER power supply and battery charger. This
accessory is designed to be used with the PowerPad® IV Model 8345, the
Oscilloscope 9000 Series and the C.A 6117.
For the best results from your instrument and for your safety, you must read
the enclosed operating instructions carefully and comply with the precautions
for use. These products must be used only by trained and qualified users.
Signifies that the instrument is protected by double or reinforced insulation
CAUTION – Risk of Danger! Indicates a WARNING. Whenever this symbol
is present, the operator must refer to the user manual before operation
Suitable for indoor use
Chauvin Arnoux® has adopted an Eco-Design approach in order to design this
appliance. Analysis of the complete lifecycle optimized the effects of the
product on the environment. In particular this appliance exceeds regulation
requirements with respect to recycling and reuse
Important instructions to read and understand completely
Important information to acknowledge
USB socket
This product complies with the Low Voltage & Electromagnetic Compatibility
European directives (73/23/CEE & 89/336/CEE)
The UKCA marking certifies that the product complies with the requirements
that apply in the United Kingdom, specifically regarding Low-Voltage Safety,
Electromagnetic Compatibility, and the Restriction of Hazardous Substances
In the European Union, this product is subject to a separate collection system
for recycling electrical and electronic components in accordance with
directive WEEE 2002/96/EC
Definition of Measurement Categories (CAT)
-
CAT IV corresponds to measurements performed at the primary electrical supply (<1000 V).
Example: primary overcurrent protection devices, ripple control units, and meters. -
CAT III corresponds to measurements performed in the building installation at the distribution level.
Example: hardwired equipment in fixed installation and circuit breakers. -
CAT II corresponds to measurements performed on circuits directly connected to the electrical distribution system.
Example: measurements on household appliances and portable tools.
PRECAUTIONS FOR USE
For personal safety and for the safety of the environment of use, you must read and comply with the precautions for use. If you do not comply with these safety instructions, there is a risk of electric shock, explosion, or fire. If you use the instrument in an unspecified manner or do not observe elementary safety rules, the protection that they provide could be compromised and endanger you.
This instrument complies with safety standard IEC/EN 61010-1 and the leads are compliant with IEC/EN 61010-031, for voltages up to 1,000 V in overvoltage category IV. The protection provided by the instrument may be impaired if it is used other than as specified by the manufacturer.
- Comply with the rated maximum voltage and current and the overvoltage category.
- Never exceed the protection limits stated in the specifications (§ 4.2).
- Observe the environmental conditions of use (§ 4.3).
- Use only the leads and accessories supplied. The use of leads (or accessories) of a lower voltage or category limits the voltage or category of the combined instrument and leads (or accessories) to that of the leads (or accessories).
- Before each use, check the condition of the insulation for the cables, case, and accessories—anything that appears damaged (even partially) must be reported for repair or scrapping.
- Do not expose the housing to a source of heat or to direct sunlight. Let air flow freely around it to ensure sufficient cooling.
- When handling the alligator clips, keep your fingers behind the physical guard.
- Use protective equipment when required.
- Do not modify the instrument and use only original replacement parts. Repairs or adjustments must be performed by authorized personnel.
- If hazardous voltages are present at the location of measurement, use suitable personal protective equipment.
RECEIVING YOUR SHIPMENT
Upon receiving your shipment, make sure that the contents are consistent with
the packing list. Notify your distributor of any missing items. If the
equipment appears to be damaged, file a claim immediately with the carrier and
notify your distributor at once, giving a detailed description of any damage.
Save the damaged packing container to substantiate your claim.
ORDERING INFORMATION
Adapter – 1000 V PA32ER Power Supply for Model 8345 ………….. Cat.
5100.15Includes Power Supply, (2) 1000 V CAT IV safety leads, two 600 V CAT
IV 10 ft black leads, (2) 1000 V CAT IV black alligator clips, (1) power plug adapter with banana jacks, (1) 115 V power cord, and user manual.
Accessories and Replacement Parts
-
Adapter – 110 V Outlet Adapter w/4 mm Banana Plugs ………….. Cat. #2118.49
-
Lead – One 10 ft (3 m) Black Lead w/Black Alligator Clip
{Lead rated 1000 V CAT IV 10 A, Clip rated 1000 V CAT IV 15 A,UL} for PowerPad® & PEL series. ……………………………………………Cat. #2140.44 -
Lead – Replacement 10 ft (3 m) Black Stackable 600 V CAT IV for PA31ER & PA32ER ………………………………………………………… Cat. #5000.88
-
Adapter – Replacement Power Plug Adapter for PA32ER ………… Cat. #5100.14
Order Accessories and Replacement Parts Directly Online Check our Storefront at www.aemc.com/store for availability
PRODUCT FEATURES
DESCRIPTION
The PA32ER power adapter and battery charger are used to power the attached
instrument for long-term applications and thereby optimize the instrument’s
internal battery. It also serves to recharge the battery.
The power supply may come from the measurement source or from the wall plug.
OVERALL VIEW
OPERATING INSTRUCTIONS
POWER FROM THE MEASUREMENT SOURCE
Using the leads provided and alligator clips, connect the adapter to a line
voltage. Either between a phase (L1, L2, or L3) and neutral or between 2
phases.
In the image above, the adapter is connected between the neutral (N) and phase L1.
- Open the elastomer cap and connect the 4-point connector to the external power supply input of the instrument.
The indicator light on the front of the adapter displays:
- A steady green when the power is on
- A steady orange when the battery is charging
- If it blinks, the battery is missing and the instrument is powered by the external power supply alone.
CHARGE DIRECTLY FROM THE WALL PLUG
- Connect the adapter banana plug to the PA32ER power adapter.
- Connect the plug adapter cord between the adapter and the power socket.
- Open the elastomer cap on the instrument and connect the 4-point connector of the power supply to the instrument.
The indicator light on the front of the adapter displays:
- A steady green when the power is on
- A steady orange when the battery is charging
- If it blinks, the battery is missing and the instrument is powered by the external power supply alone.
SPECIFICATIONS
REFERENCE CONDITIONS
Quantity of influence | Reference values |
---|---|
Ambient temperature | 20 °C ± 3 °C |
ELECTRICAL SPECIFICATIONS
Input voltage characteristics
- Voltage range:100 V to 1000 V at a frequency from 40 Hz to 400 Hz 150 V to 1,000 Vdc
- Overload tolerance: ± 10% permanent, ± 15% temporary
- Frequency range: DC, 40 Hz – 70 Hz, 340 Hz – 440 Hz
- Input current: 0.5 Arms max
- No-load consumption:
Supply voltage (V)| Input current (mA)| Apparent power (VA)| RMS power
(W)
---|---|---|---
115| 11.5| 0.92| 0.65
230| 6| 1.265| 0.6
400| 5.3| 1.920| 1
1,000| 3.5| 3.3| 1.4
Ouput voltage characteristics
Characteristics | Values |
---|---|
Permanent output power | 23 W |
Output voltage @ 30 W | 15 V ± 8% |
Maximum no-load output voltage | 16.2 V |
Ripple of the output voltage | 550 mVrms |
Maximum short-circuit duration | Permanent |
Starting-up time at 110 V | < 13 s |
Efficiency
The efficiency depends on the supply voltage and the load. In general, it is
68%.
Charger
Charging current: 1A
ENVIRONMENTAL CONDITIONS
Range of operation: from (-4 to 113) °F (-20 to +45) °C, from (30 to 95) %RH
without condensation
Range of storage: from (-13 to 158) °F (-25 to +70) °C, from (10 to 90) %RH
without condensation
Electric field: < 1 V/m
Magnetic field: < 40 A/m
Indoor use:
Altitude : < 2000 m
Degree of pollution: 3
MECHANICAL CHARACTERISTICS
-
Dimensions (L x D x H): (8.7 x 4.4 x 2.1) in (220 x 112 x 53) mm
-
Lead: 5 ft (1.50 m), terminated by a specific 4-point connector
-
Weight: Approximately 2 lbs (930 g)
-
Inrush protection: IP54 when not in operation,
IP50 in operation per IEC 60529
IK08 per IEC 62262 -
Drop Test: 3.3 ft (1 m) per IEC 60068-2-31
COMPLIANCE WITH INTERNATIONAL STANDARDS
-
Electrical Safety
The instrument is compliant with IEC/EN 61010-1 and IEC/EN 61010-031 or BS
EN 61010-031 for a voltage of 1000 V CAT IV, degree of pollution 3.
The instruments are compliant with FCC Rules part 1 under number 83TD07571. -
Electromagnetic compatibility (EMC)
The instrument is compliant with standard IEC/EN 61326-1.
MAINTENANCE
The instrument contains no parts that can be replaced by untrained or unaccredited personnel. Any unapproved work or part replacement using equivalents may gravely compromise safety.
CLEANING
- Disconnect the unit completely.
- Use a soft cloth, moistened with soapy water.
- Rinse with a damp cloth.
- Dry rapidly with a dry cloth or forced air.
REPAIR AND CALIBRATION
To ensure that your instrument meets factory specifications, we recommend that
it be sent back to our factory Service Center at one-year intervals for
recalibration, or as required by other standards or internal procedures.
For instrument repair and calibration:
You must contact our Service Center for a Customer Service Authorization
Number (CSA#). Send an email to repair@aemc.com requesting a CSA#. You will be
provided a CSA Form and other required paperwork along with the next steps to
complete the request. Then return the instrument along with the signed CSA
Form. This will ensure that when your instrument arrives, it will be tracked
and processed promptly. Please write the CSA# on the outside of the shipping
container. If the instrument is returned for calibration, we need to know if
you want a standard calibration or a calibration traceable to N.I.S.T.
(includes calibration certificate plus recorded calibration data).
Ship To : Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments
15 Faraday Drive, Dover, NH 03820 USA Phone:
800-945-2362 (Ext. 360)
E-mail: repair@aemc.com
(Or contact your authorized distributor.)
Contact us for the costs for repair, standard calibration, and calibration
traceable to N.I.S.T.
NOTE: You must obtain a CSA# before returning any instrument.
TECHNICAL ASSISTANCE
If you are experiencing any technical problems, or require any assistance with
the proper operation or application of your instrument, please call, mail,
fax, or e-mail our technical support team:
Chauvin Arnoux®, Inc. d.b.a. AEMC®
Instruments Phone: 800-343-1391 (Ext.
351)
Fax: 603-742-2346
E-mail: techsupport@aemc.com
www.aemc.com
LIMITED WARRANTY
The instrument is warrantied to the owner for a period of two years from the date of original purchase against defects in manufacture. This limited warranty is given by AEMC® Instruments, not by the distributor from whom it was purchased. This warranty is void if the unit has been tampered with, abused, or if the defect is related to service not performed by AEMC® Instruments.
Full warranty coverage and product registration is available on our website at
www.aemc.com/warranty.html
Please print the online Warranty Coverage Information for your records.
What AEMC® Instruments will do:
If a malfunction occurs within the warranty period, you may return the
instrument to us for repair, provided we have your warranty registration
information on file or a proof of purchase. AEMC® Instruments will repair or
replace the faulty material at their discretion.
REGISTER ONLINE AT: www.aemc.com/warranty.html
WARRANTY REPAIRS
What you must do to return an Instrument for Warranty Repair:
First, request a Customer Service Authorization Number (CSA#) by email from
our Service Department at repair@aemc.com (see link
below). Then, return the instrument along with the signed CSA Form. Please
write the CSA# on the outside of the shipping container. Return the instrument
with postage or shipment pre-paid to:
Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments 15 Faraday Drive
Dover, NH 03820 USA
Phone: 800-945-2362 (Ext. 360)
603-749-6434 (Ext. 360)
Fax: 603-742-2346 or
603-749-6309
E-mail: repair@aemc.com
Caution: To protect yourself against in-transit loss, we recommend you
insure your returned material.
NOTE: You must obtain a CSA# before returning any instrument.
Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments
15 Faraday Drive • Dover, NH 03820 USA
Phone: 603-749-6434 •
800-343-1391 • Fax:
603-742-2346
www.aemc.com
References
- Get Electrical Testers – Power Analyzers | AEMC Instruments
- Online Ecommerce Store | AEMC Instruments
- AEMC Warranty Registration