AEMC 2137.77 Power Adapter PEL User Manual

June 9, 2024
AEMC

AEMC 2137.77 Power Adapter PEL

AEMC 2137.77 Power Adapter PEL

ORDERING INFORMATION

Adapter-600 V CAT III Power Adapter for use with Models PEL102 and PEL103 ….Cat. #2137.98

Accessories and Replacement Parts

Set of 12 Color-coded Input ID Markers ….Cat. #2140.45

USE

WARNING: Use Only on CAT III Pollution 2 Location.

The power adapter is used to supply the device when it is used for an extended period. The internal battery of the PEL remains charged, enabling the device to continue to operate in the event of a power outage up to 30 minutes.

  • Connect the adapter to the voltage inputs of the PEL.
  • Connect the C7 plug from the adapter to the  input of the PEL.
  • Connect the voltage inputs to the adapter, following the markings on the adapter. Connect the current inputs directly to the PEL, per normal operation (see the illustration on page 5).

The L1 and N or L1 and L2 voltage inputs must be connected for the adapter to operate.

The adapter can operate on (50 and 60) Hz networks.

SPECIFICATIONS

Electrical

Range of use| Phase-to-neutral voltage: (100 to 346) Vac
Phase-to-phase voltage: (100 to 600) Vac
---|---
Max. input voltage| Permanent: 660 Vac
Max. output voltage| 360 V peak
Frequency| (50 and 60) Hz
Consumption| 100 VA max at (50 and 60) Hz

Environmental

Operation: (32 to 122) °F (0 to 50) °C and (10 to 85) % RH
Storage: (-4 to 158) °F (-20 to +70) °C and (0 to 95) % RH
Indoor use; Level of pollution: 2; IEC 61010-1 600 V CAT III
Altitude: < 2000 m; Storage: < 10,000 m

Mechanical

Dimensions (L x H x D): (5 x 3.8 x 1.5) in (127 x 97 x 39) mm
Weight: 7 oz (200 g) approximately
Cord: 11 in (30 cm) approximately
Ingress protection: IP 40 per IEC 60529 except for terminals (IP20) IK 04 per IEC 50102
Drop Test : 3.3 ft (1 m) per IEC/EC 6101-2-030 or BS EN 61010-2-030

Electromagnetic compatibility: Emissions and immunity in an industrial setting compliant with IEC 61326-1 or BS IEC/EN 61326-1

Safety: 600 V CAT III IEC/EN 61010-2-30 or BS EN 61010-2-030

Specifications are subject to change without notice.

MAINTENANCE

This instrument contains no parts that can be replaced by personnel who have not been specially trained and accredited. Any unauthorized repair or replacement of a part by an equivalent may impair the safety rating.
Cleaning: Disconnect all inputs to the device and remove from the PEL. Clean with a soft cloth moistened with soapy water. Wipe dry completely.
Repair: For all repairs, please contact the AEMC® Instruments or your local distributor.

DIMENSION

Dimension

Copyright © Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments. All rights reserved. No part of this documentation may be reproduced in any form or by any means (including electronic storage and retrieval or translation into any other language) without prior agreement and written consent from Chauvin Arnoux®, Inc., as governed by United States and International copyright laws.

This documentation is provided “as is,” without warranty of any kind, express, implied, or otherwise. Chauvin Arnoux®,

Inc. has made every reasonable effort to ensure that this documentation is accurate; but does not warrant the accuracy or completeness of the text, graphics, or other information contained in this documentation. Chauvin Arnoux®, Inc. shall not be liable for any damages, special, indirect, incidental, or inconsequential; including (but not limited to) physical, emotional or monetary damages due to lost revenues or lost profits that may result from the use of this documentation, whether or not the user of the documentation has been advised of the possibility of such damages.

CUSTOMER SUPPORT

Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments
15 Faraday Drive • Dover, NH 03820 USA
Phone: 603-749-6434800-343-1391 • Fax: 603-742-2346
www.aemc.com

AEMC Logo

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals