BONDLINE CWST3 Dual Wrist Strap and Footwear Tester Instruction Manual
- August 28, 2024
- BONDLINE
Table of Contents
BONDLINE CWST3 Dual Wrist Strap and Footwear Tester
Introduction
- Thank you for purchasing a Dual Wrist Strap and Footwear Tester.
- Please read this instruction manual in great detail before operating the device.
- Keep this manual readily accessible for later reference.
Description
- A unique checker that quickly and easily measures the resistance of fitted wrist straps at 750KQ to 35MQ and footwear/heel straps at 100KQ to 100MO.
- There are three settings: wrist strap-only test, footwear-only test, and also dual test feature where you can test your wrist strap and footwear at the same time.
- The unit is under the current IEC 61340-5-1 specification. CE and UKCA approved.
Included In The Kit
- Test unit (CWST3)
- Wall mounting board
- 6ft (1.8m) cord (0 meg-ohm)
- 6 inch (150mm) cord (0 meg-ohm)
- Dual footplate
- 9 volt PP3 alkaline battery
Wrist-Strap Test Guidance ONLY
Firstly, ensure the black toggle switch located on the front of the tester is
moved DOWN for wrist-strap test only. When testing the wrist strap, the
operator simply connects to the wrist strap via the 4mm banana plug socket or
10mm stud. To test, make finger contact with both stainless steel discs that
are located on the left-hand side of the front of the meter. If the wrist
strap is working correctly within the specification, the green LED for the
wrist strap pass will illuminate. An audible buzzer will sound if the red fail
LED illuminates. If the resistance through the coil cord, wristband, and
person exceeds 35 meg-ohms, the high red fail LED will light. If the
resistance is below 750 kilohms, the low red fail LED will light.
Dual Footwear Test Guidance ONLY
Firstly, ensure the black toggle switch located on the front of the tester is
moved all the way UP for the dual footwear test only. Ensure that the
footplate lead is connected to the footplate socket located at the bottom of
the meter. To test, make finger contact with both stainless steel discs that
are located on the left-hand side of the front of the meter whilst standing on
the left and right footplates. If your footwear is working correctly and
within the specification, the green LED for the footwear pass will illuminate.
If your footwear is not working within the specification, the red LED for the
footwear either left, right or both will illuminate for failure. An audible
buzzer will sound if any of the red fail LEDs illuminate. If the resistance
through the footwear and the person exceeds 100 meg-ohms, the red high fail
LED will light. If the resistance is below 100 kilohms, the red low-fail LED
will illuminate.
Wrist-Strap and Footwear Test Guidance TOGETHER
Ensure the black toggle switch is moved into the middle for dual testing.
Repeat the steps above to test both at the same time. Powered by a 9-volt PP3
alkaline battery. If the 9-volt PP3 battery falls below 6.5 volts during the
test, the yellow battery low LED will not illuminate.
CONTACT
- sales@bondline.co.uk
- www.bondline.co.uk
- +44 (0)1793 511000
The information contained within this instruction manual is for guidance only. We make no warranties expressed or implied and assume no liability regarding any use of this information. (Review February 23rd, 2024)
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>