ZEISS VISCOJECT-BIO 1.8 Injector Set Instructions
- August 27, 2024
- Zeiss
Table of Contents
ZEISS VISCOJECT-BIO 1.8 Injector Set Instructions
Handling instruction for non-preloaded MICS IOLs
-
Place a small amount of OVD* in the cartridge tip. Place a line of OVD in each channel of the cartridge. Do not touch the tip of the cartridge.
-
Place the lens in the center of the open cartridge. Push down on the edges of the optic to ensure that the optic haptics are positioned under the ledges of the cartridge.
-
For toric and multifocal IOLs, mind the orientation holes and the meridian marks.
-
Maintain the IOL in place with the folding forceps while folding the cartridge. Close the cartridge until you hear a click.
-
Place the cartridge into the injector and move it forward as far as it will go. The lens is ready for implantation.
-
Place the bevelled edge downward and gently push the plunger to inject the IOL.
*Carl Zeiss Meditec recommends the use of ZEISS OVDs.
OUS_32_022_0178I CZ-II/2024 International edition: Only for sale in selected
countries.
The contents of the brochure may differ from the current status of approval of
the product or service offering in your country. Please contact our regional
representatives for more information. Subject to changes in design and scope
of delivery and due to ongoing technical development. VISCOJECT is a trademark
of Medicel. © Carl Zeiss Meditec AG, 2024. All rights reserved.
0482
VISCOJECT-Bio 1.8 Injector Set
Medicel AG
Dornierstrasse 11
9423 Altenrhein
Switzerland
www.zeiss.com/injectors
www.zeiss.com/med/contacts
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>