ZEISS VISCOJECT-BIO 1.8 Injector Set Instructions

August 27, 2024
Zeiss

ZEISS VISCOJECT-BIO 1.8 Injector Set Instructions
ZEISS VISCOJECT-BIO 1.8 Injector Set

Handling instruction for non-preloaded MICS IOLs

  1. Place a small amount of OVD* in the cartridge tip. Place a line of OVD in each channel of the cartridge. Do not touch the tip of the cartridge.
    Handling instruction

  2. Place the lens in the center of the open cartridge. Push down on the edges of the optic to ensure that the optic haptics are positioned under the ledges of the cartridge.
    Handling instruction

  3. For toric and multifocal IOLs, mind the orientation holes and the meridian marks.
    Handling instruction

  4. Maintain the IOL in place with the folding forceps while folding the cartridge. Close the cartridge until you hear a click.
    Handling instruction

  5. Place the cartridge into the injector and move it forward as far as it will go. The lens is ready for implantation.
    Handling instruction

  6. Place the bevelled edge downward and gently push the plunger to inject the IOL.
    Handling instruction

*Carl Zeiss Meditec recommends the use of ZEISS OVDs.

OUS_32_022_0178I CZ-II/2024 International edition: Only for sale in selected countries.
The contents of the brochure may differ from the current status of approval of the product or service offering in your country. Please contact our regional representatives for more information. Subject to changes in design and scope of delivery and due to ongoing technical development. VISCOJECT is a trademark of Medicel. © Carl Zeiss Meditec AG, 2024. All rights reserved.

0482
VISCOJECT-Bio 1.8 Injector Set

Medicel AG
Dornierstrasse 11
9423 Altenrhein
Switzerland
www.zeiss.com/injectors
www.zeiss.com/med/contacts

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals