Zeiss 5000 CIRRUS HD-OCT Scan Acquisition User Manual
- June 2, 2024
- Zeiss
Table of Contents
- CIRRUS™ HD-OCT Scan Acquisition
- 1. Find the pupil and focus the image in the iris viewport.
- Advanced Tips
- 2. Make adjustments for optimal fundus and B-scan images.
- 3. Capture the fundus image and high-density scan
- 4. Review the captured scan, then save or try again
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
CIRRUS™ HD-OCT Scan Acquisition
Quick Reference Guide
This guide is an overview of scan acquisition for scans available* with CIRRUS HD-OCT software version 8.0. See the CIRRUS HD-OCT Instructions for Use for safe and effective operation.
1. Find the pupil and focus the image in the iris viewport.
Adjust table height.
Place patient’s chin in the left chin cup (marked blue) to scan the right eye
or the right chin cup (marked white) to scan the left eye. Patient’s forehead
should make contact with the headrest. Adjust the chinrest height to align the
canthus marker with the patient’s eye.
Adjust the patient position to find the iris and pupil. Use the chinrest
controls or click on the pupil in the iris viewport. Instruct the patient to
move with the chinrest. The patient should not hold on to instrument. Click
within the center of the pupil for final alignment.
Use to obtain a clear focus on the iris. This helps the system set the
position for the OCT scan.
Advanced Tips
- Click on the icon to adjust the brightness, contrast, and illumination. Lowering the illumination may increase pupil size.
- The mouse scroll wheel can be used to make fine changes to the iris image in/out adjustment.
- During subsequent steps, monitor: 1) the patient’s head to ensure they maintain contact with the headrest and, 2) the position of the red scan
circle in the pupil to ensure it remains within the pupil.
2. Make adjustments for optimal fundus and B-scan images.
Click on Auto Focus. Instruct patient to move with the chinrest, fixate on
the target and to not blink while this function runs. Adjust manually if
necessary. Proper focus can increase the signal for the OCT scans and improve
ability to track.
Click on Optimize to center and enhance the B-scan images. To sharpen the
fundus image, mouse over the fundus image and click on the wand that appears
in the upper right corner to perform auto B/C.
Performing a scan with FastTrac™
-
When the border around the Capture button is green, you are ready to acquire the scan
with FastTrac. If the border is red, further actions can be taken: -
Center the scan with the Center controls or hold Shift and roll the mouse wheel.
-
Click Auto Focus or manually adjust focus to sharpen fundus image quality.
-
Change the prior scan or disable Track to Prior button for this scan.
-
You can turn off tracking by clicking the FastTrac button .
Advanced Tips
-
The B-scan images should show a range of colors. Faint green B-scans indicate that the signal is low.
-
To modify brightness or contrast for either the B-scans or fundus images, click near the image of interest and adjust the controls.
Note: The setting changes made for the B-scans here will persist for future scans. -
If the OCT image appears inverted, click the Center up/down buttons several times until the image flips and is centered it in the window.
-
Center the B-scan images and ensure all B-scan images are not too high or too low. This could result in missing data in the scan.
-
For highly myopic patients it may be necessary to click the Auto Focus button a second time.
-
Moving the scan in the pupil in the iris viewport can help if: 1) there are media opacities, 2) the OCT B-scans appear tilted.
-
Instead of using Auto Focus, enter spherical equivalent information on the patient demographic page for CIRRUS to pre-focus to this value.
-
When tracking on the macula with patients with tilted retinas, you can improve success using tracking by turning off the system’s center
position monitoring. Click on the wand next to the bottom B-scan, then in the dialog box, uncheck Monitor Z position.
3. Capture the fundus image and high-density scan
When the fundus image and B-scans are optimized, tell the patient to blink
once or twice.
Then tell the patient to open wide and maintain focus on the center of the
green fixation
target.
Click Capture.
Avoid a patient blink right at the moment you intend to click Capture.
If tracking is on, you can tell the patient to blink naturally during the
scan.
If tracking is off, remind the patient to avoid blinking and to focus on the
target until the capture is completed.
FastTrac Scan in Progress Screen. When FastTrac is interrupted, the progress bars turn red and processing stops. Click Cancel to return to the Acquire Screen to make adjustments to center the scan, focus the fundus or iris images, or turn off FastTrac for a non-tracked scan.
4. Review the captured scan, then save or try again
On the Review screen, evaluate the B-scan quality and the Signal Strength (6
or greater). Check that the OCT B-scans are approximately centered in the
windows without missing data.
If scan quality, position, or eye movements are unacceptable, click Try
Again to retake the scan.
Adjust brightness and contrast by right clicking on the fundus or OCT image.
The B/C values set on the review screen are saved and used for analysis for
this scan only.
Click Save.
If you perform another scan on the same eye, the patient can stay positioned.
Otherwise, the patient should sit back.
Add a Smart HD Scan
With the Smart HD scan workflow, you can automatically track HD scans to the
fovea. Acquire a Macular Cube 512×128 or Macular Cube 200×200 scan. Auto
FoveaFinder™ automatically locates the fovea. Alternately, target a specific
region of interest with a Smart HD scan by moving the slice navigators.
On the Review screen, click the Save and Add Smart HD Scan button. The
Macular Cube scan is automatically saved and an HD scan is initiated and
tracked to the fovea location.
On the Acquire screen, click Capture , and then on the Review screen,
click Save.
Note: To make adjustments to line spacing, rotation, or length of the Smart HD
scan, first turn off Track to Prior , make adjustments, and then turn
Track to Prior back on before clicking Capture.
Note: The last Smart HD scan used for the Smart HD scan workflow becomes the
default scan.
Advanced Tips
-
Examine the high definition en face scan closely for any subtle eye movement. Also slide the Transparency control back and forth to compare the images.
If noteworthy movement is detected, click Try Again. -
If the next scan will be on the other eye, choose the scan pattern for that eye before the patient returns their head to the appropriate chin cup and headrest. This can help avoid
contact with the patient’s nose. -
During a scan with FastTrac, individual B-scans in a cube may be acquired at slightly different positions, resulting in horizontal lines or bands on the fundus image. If there are no saccades, a scan with bands in the OCT fundus should be acceptable for analysis.
-
Smart HD Scans automatically find the fovea based on the Macular Cube scan information.
You can also use the Macular Cube to target a specific feature for a Smart HD Scan.
Example: Saccades (vessel mismatch)
Example: Banding (no vessel mismatch)