ZEISS 0482 Visco Ject Bio2.2 Injector Set Instructions

August 27, 2024
Zeiss

VISCOJECT-BIO 2.2 Injector Set
Handling instruction for 4-point haptic IOLs

0482 Visco Ject Bio2.2 Injector Set Instructions

Place a small amount of OVD* in the cartridge tip.
Place a line of OVD in each channel of the cartridge.
Do not touch the tip of the cartridge. Place the lens in the center of the open cartridge. Push down on the edges of the optic to ensure that the optic haptics are positioned under the ledges of the cartridge. For toric and multifocal IOLs, mind the orientation holes and the meridian marks. Maintain the IOL in place with the folding forceps while folding the cartridge. Close the cartridge until you hear a click. Place the cartridge into the injector and move it forward as far as it will go. The lens is ready for implantation. Place the bevelled edge downward and gently push the plunger to inject the IOL.

OUS 32 022 01791 CZ-8/2074 Intemational edition: Only for sale in selected countries The contents of the brochure may differ from the current status of approval of the product or service offering in your country. Please contact our regional representatives for more information. Subject to changes in design and scope of delivery and due to ongoing technical development, VISCOJECT is a trademark of Medicel Carl Zeiss Meditec AG, 2024 All rights reserved