56x
Infrared Thermometers
Users Manual
August 2010
©2010 Fluke Corporation, All rights reserved. Specifications are subject to
change without notice.
All product names are trademarks of their respective companies.
LIMITED WARRANTY AND LIMITATION OF LIABILITY
Each Fluke product is warranted to be free from defects in material and
workmanship under normal use and service. The warranty period is two years and
begins on the date of shipment. Parts, product repairs, and services are
warranted for 90 days. This warranty extends only to the original buyer or
end-user customer of a Fluke authorized reseller, and does not apply to fuses,
disposable batteries, or to any product which, in Fluke’s opinion, has been
misused, altered, neglected, contaminated, or damaged by accident or abnormal
conditions of operation or handling. Fluke warrants that software will operate
substantially in accordance with its functional specifications for 90 days and
that it has been properly recorded on non-defective media.
Fluke does not warrant that software will be error free or operate without
interruption.
Fluke authorized resellers shall extend this warranty on new and unused
products to end-user customers only but have no authority to extend a greater
or different warranty on behalf of Fluke. Warranty support is available only
if product is purchased through a Fluke authorized sales outlet or Buyer has
paid the applicable international price. Fluke reserves the right to invoice
Buyer for importation costs of repair/replacement parts when product purchased
in one country is submitted for repair in another country.
Fluke’s warranty obligation is limited, at Fluke’s option, to refund of the
purchase price, free of charge repair, or replacement of a defective product
which is returned to a Fluke authorized service center within the warranty
period.
To obtain warranty service, contact your nearest Fluke authorized service
center to obtain return authorization information, then send the product to
that service center, with a description of the difficulty, postage and
insurance prepaid (FOB Destination). Fluke assumes no risk for damage in
transit. Following warranty repair, the product will be returned to Buyer,
transportation prepaid (FOB Destination). If Fluke determines that failure was
caused by neglect, misuse, contamination, alteration, accident, or abnormal
condition of operation or handling, including overvoltage failures caused by
use outside the product’s specified rating, or normal wear and tear of
mechanical components, Fluke will provide an estimate of repair costs and
obtain authorization before commencing the work. Following repair, the
product will be returned to the Buyer transportation prepaid and the Buyer
will be billed for the repair and return transportation charges (FOB Shipping
Point).
THIS WARRANTY IS BUYER’S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER
WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED
WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. FLUKE SHALL
NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES
OR LOSSES, INCLUDING LOSS OF DATA, ARISING FROM ANY CAUSE OR THEORY.
Since some countries or states do not allow limitation of the term of an
implied warranty, or exclusion or limitation of incidental or consequential
damages, the limitations and exclusions of this warranty may not apply to
every buyer. If any provision of this Warranty is held invalid or
unenforceable by a court or other decision-maker of competent jurisdiction,
such holding will not affect the validity or enforceability of any other
provision.
Fluke Corporation
P.O. Box 9090
Everett, WA 98206-9090
U.S.A.
Fluke Europe B.V.
P.O. Box 1186
5602 BD Eindhoven
The Netherlands
Introduction
The 561, 566 and 568 Infrared Thermometers (“the thermometers” or “the
Product”) are for non-contact temperature measurement. These thermometers
determine an object’s surface temperature by measuring the amount of infrared
energy radiated by the object’s surface. The thermometers also support
contacttemperature measurement via K-type thermocouple.
Note that the Japanese models indicate Celsius only.
Contacting Fluke
To contact Fluke, call one of the following telephone numbers:
- Technical Support USA: 1-800-44-FLUKE (1-800-443-5853)
- Calibration/Repair USA: 1-888-99-FLUKE (1-888-993-5853)
- Canada: 1-800-36-FLUKE (1-800-363-5853)
- Europe: +31 402-675-200
- Japan: +81-3-3434-0181
- Singapore: +65-738-5655
- Anywhere in the world: +1-425-446-5500 Or, visit Fluke’s website at www.fluke.com.
To register your product, visit http://register.fluke.com.
To see, print, or download the latest manual supplement, visit
http://us.fluke.com/usen/support/manuals.
Safety Information
A Warning identifies conditions and actions that pose hazard(s) to the user; A
Caution identifies conditions and procedures that could cause Product damage,
equipment under test damage, or permanent loss of data.
Symbols used on the Product and in this manual are explained in Table 1 and
Figures 1 and 2.
Warning
To prevent eye damage and personal injury:
-
Read all safety Information before you use the Product.
-
Do not look directly into the laser with optical tools (for example, binoculars, telescopes, microscopes). Optical tools can focus the laser and be dangerous to the eye.
-
Do not look into the laser. Do not point laser directly at persons or animals or indirectly off reflective surfaces.
-
Do not use laser viewing glasses as laser protection glasses. Laser viewing glasses are used only for better visibility of the laser in bright light.
-
Do not open the Product. The laser beam is dangerous to eyes. Have the Product repaired only through an approved technical site.
-
Replace the batteries when the low battery indicator shows to prevent incorrect measurements.
-
The battery door must be closed and locked before you operate the Product.
-
Do not use the Product if it operates incorrectly.
-
Do not use the Product around explosive gas, vapor, or in damp or wet environments.
-
Do not connect the optional external probe to live electrical circuits.
-
See emissivity information for actual temperatures.
Reflective objects result in lower than actual temperature measurements. These objects pose a burn hazard. -
Do not leave the thermometer on or near objects of high temperature.
-
Use of controls or adjustments or performance of procedures other than those specified herein may result in hazardous laser radiation exposure.
-
Use the Product only as specified, or the protection supplied by the Product can be compromised.
Caution
To avoid damaging the thermometer or the equipment under test, protect them
from the following:
- EMF (electro-magnetic fields) from arc welders, induction heaters, etc.
- Static electricity
- Thermal shock (caused by large or abrupt ambient temperature changes- for highest accuracy, allow 30 minutes for thermometer to stabilize before use).
Table 1. Symbols
Symbol | Explanation |
---|---|
Hazardous voltage. Risk of electrical shock. | |
**** | Risk of danger. Important information. |
Warning. Laser. | |
Conforms to requirements of European Union and European Free Trade |
Association (EFTA)
° C
| Celsius
° F
| Fahrenheit
__| Battery
| Do not dispose of this product as unsorted municipal waste. Go to Fluke’s
web site for recycling information.
| Battery
| China metrology certification mark for measuring instruments manufactured in
the Peoples Republic of China (PRC).
Features
561:
- Single-spot laser sighting
- Backlit display
- Current Temperature plus MAX, MIN, DIF, AVG temperature displays
- Two AA batteries
- Hard case
- 80PK-1 and 80PK-11 K-type thermocouple probe
- Adjustable emissivity and predefined emissivity table
- Infrared and thermocouple temperature display
- Celsius or Fahrenheit temperature display
- Tripod mount
- Auto off
- Standard miniature K-type thermocouple connector input
- Printed 56x Getting Started Guide
- 56x Manuals CD
566 and 568 include items listed above plus:
- 12 or 24 hour clock
- Last reading Hold (20 seconds)
- Multi-language interface
- High and low alarm
- Data storage and review
- Trigger lock
- USB 2.0 computer interface cable (568)
- FlukeView Forms Documenting Software (568)
561 Display
The primary temperature display reports the current or last IR temperature
read until the 7-second hold time elapses.
The secondary temperature display reports current thermocouple temperature
when a type-K thermocouple is attached. When a thermocouple is not connected,
the small temperature display reports a choice of maximum, minimum, or
difference between maximum and minimum temperature.
You can toggle through the minimum, maximum, and difference IR temperatures
anytime the display is on. The MIN, MAX, and DIF temperatures are constantly
calculated and updated when the trigger is pressed. After the trigger is
released, the MIN, MAX, DIF temperatures are held for 7 seconds.
Note
When the battery is low, appears on the display.
The last selection (MIN/MAX/DIF) is maintained on the secondary display even
after the Thermometer has been turned off, providing the batteries have not
failed. See Figure 3.
Number| Description
①| Laser “On” symbol
2| SCAN or HOLD
3| °C/°F symbol (Celsius/Fahrenheit)
4| Primary temperature display
⑤| Secondary temperature display
6| Emissivity LO, MED, HI
7| Temperature values for the MIN, MAX, DIF, KTC.
KTC indicates the thermocouple temperature.
8| Low Battery symbol. Appears when the battery charge is <25 %.
Figure 3. 561 Thermometer Display
566/568 Display
The 566/568 thermometers display can show data in the following languages:
- English
- Spanish
- French
- German
- Portuguese
- Japanese
- Simplified Chinese
To change the displayed language, refer to “Setup”.
566/568 Menu Overview
There are many settings that can be easily changed by using the menu. Figure 4
shows the LCD and menu interface. Selecting the Menu softkey advances the menu
to the next level. Table 2 is a top-level description of the menu.
Table 2. Top-Level Menu Description
Level| Left Softkey| Description| Center Softkey|
Right Softkey| Description
---|---|---|---|---|---
1| Save| Save reading to memory| Menu| Light| Adjust
backlight brightness
2| Mem| Review/delete memories| Menu| | Set emissivity
3| MnMx| Enables Min/Max| Menu| Avg| Enable Avg/Diff
4| ° C/ ° F| Toggle between C and F| Menu| Alarm| Set and
enable alarms
5| (Lock)| Lock the thermometer on| Menu| Laser| Toggle the
laser on/off
6| Setup| – Turn off/on backlight
– Change Time/Date
– Change Language| Menu| –| –
Each menu item and function is explained in greater detail in the following sections.
Save
To save readings:
- Pull the trigger to take a measurement.
- Release the trigger to stop taking the measurement.
- Press the Save softkey to enter the Save menu.
- Press the Yes softkey to save the reading.
The reading is assigned a memory location and a time and date stamp.
The reading includes:
- IR temperature
- Thermocouple temperature (if connected)
- Emissivity
- Min/Max/Avg/Dif (if either Min/Max or Avg/Dif is enabled)
- Date/Time
You may also press the Cancel softkey to abort saving the reading.
Light
The thermometers are equipped with a backlit display with two brightness
levels.
The Light softkey is used to adjust the backlight brightness. The backlight is
on each time the trigger is pulled.
To toggle the backlight brightness, press the Light softkey.
The backlight can be disabled using the Setup menu. See “Setup” for more
information.
Memory
The thermometers can store measurement records including time, date,
emissivity, and measurement record numbers (see “Save” for more information).
The 566 can store 20 records and the 568 can store 99.
To access records stored in memory:
- Press the Menu softkey until Mem appears as the left softkey function.
- Press the Mem softkey to access the Memory menu. Saved readings can be read.
Emissivity Menu
The Emissivity menu includes a list of pre-defined materials and lists their
typical emissivity values, see Tables 3 and 4. For further information, see
“Emissivity”.
Note
Default emissivity is 0.95.
To access the Emissivity menu:
- Press the Menu softkey until (emissivity) appears as the right softkey function.
- Press the softkey.
The Emissivity list may be accessed by pressing the Table softkey or a
material’s typical emissivity may be entered manually by pressing the No.
softkey.
• If the Emissivity table is accessed, a listing of materials and their
suggested emissivity is shown.
- Use the down arrow to navigate through the list.
- Press the Enter softkey to choose the desired material.
• To enter an emissivity value manually:
- Press the No. softkey.
- Use the down or up arrow softkey to change the entry. Hold down the up or down arrow softkey to increase the rate of change.
- Press the Done softkey when finished to return to the main menu.
Table 3. Surface Emissivity (561)
Materials | Switch Setting | Materials | Switch Setting |
---|---|---|---|
Aluminum | Iron, Cast | ||
Oxidized | Low | Oxidized | High, Medium |
Alloy A3003 | Unoxidized | Low | |
Oxidized | Low | Molten | Low |
Roughened | Low | Iron, Wrought | |
Brass | Dull | High | |
Burnished | Low | Lead | |
Oxidized | Low | Rough | Low |
Copper | Oxidized | Low, Medium | |
Oxidized | Medium | Molybdenum | |
Electrical Terminal Blocks | Medium | Oxidized | Low, Medium |
Haynes | Nickel | ||
Alloy | Medium | Oxidized | Low |
Inconel | Platinum | ||
Oxidized | High, Medium | Black | High |
Sandblasted | Medium | Steel | |
Electoropolished | Low | Cold-Rolled | High |
Iron | Ground Sheet | Medium | |
Oxidized | High, Medium | Polished Sheet | Low |
Rusted | Medium | Zinc | |
Oxidized | Low |
Table 4. Nominal Surface Emissivity (566/568)
Material | Value | Material | Value | |
---|---|---|---|---|
Default**** | 0.95 | Glass (plate) | 0.85 | |
Aluminum* | 0.30 | Iron* | 0.70 | |
Asbestos | 0.95 | Lead* | 0.50 | |
Asphalt | 0.95 | Oil | 0.94 | |
Brass* | 0.50 | Paint | 0.93 | |
Ceramic | 0.95 | Plastic** | 0.95 | |
Concrete | 0.95 | Rubber | 0.95 | |
Copper* | 0.60 | Sand | 0.90 | |
Food – Frozen | 0.90 | Steel* | 0.80 | |
Food – Hot | 0.93 | Water | 0.93 | |
Wood *** | 0.94 |
- Oxidized
** Opaque, over 20 mils
* Natural
** Factory Setting
Highlighted items may also be found in the emissivity table built into the thermometer.
°C and °F
To toggle between °C (Celsius) and °F (Fahrenheit) measurements, press the
Menu softkey until °C or °F appears as the left softkey function. Press the
corresponding softkey to change between the measurements.
Min, Max, Avg, Differential
The thermometers can measure minimum (MIN), maximum (MAX), average
(AVG), or differential (Δ) temperatures each time a reading is taken. These
values are not shown if a thermocouple is plugged into the thermometer.
To turn on the Min Max mode:
- Press the Menu softkey until MnMx (Min Max) appears as the left softkey function.
- Press the MnMx softkey.
The display shows the present reading, maximum and minimum readings, and the
emissivity setting.
To turn on the Avg/Dif mode:
- Press the Menu softkey until Avg appears as the right softkey function.
- Press the Avg softkey.
The display shows the present reading, average reading, the differential reading between max and min (designated by Δ), and the emissivity setting.
Note
Min, Max, Avg, and Differential readings are saved as part of the saved data
when either Min/Max or Avg/Dif mode is enabled.
Alarm
The thermometers have a programmable high and low alarm to designate high or
low readings depending on the thresholds entered. When the alarm level is
reached, an alarm will sound and the display will flash orange and white. To
set either the high or low alarm:
- Press the Menu softkey until Alarm appears as the right softkey function.
- Press the Alarm softkey to access the Alarm menu.
- Press either the Hi or Lo softkey, depending on which alarm will be set.
- Press the ON softkey to turn the alarm on.
- Press the OFF softkey to turn the alarm off.
- Use the Set softkey to access the Hi or Lo Alarm Set menu.
- Use the down or up softkeys to change the alarm setting.
- Once the desired settings have been entered, press the Done softkey.
Trigger Lock
The thermometer trigger can be locked on for continuous measurement. To lock
the trigger:
- Press the Menu softkey until the lock symbol ( ) appears as the left softkey function.
- Press the softkey to lock the trigger. The lock symbol appears on the display to designate a locked trigger. When the trigger is locked, the softkey changes to . Press this softkey to unlock the trigger.
Laser
Warning
To prevent eye damage and personal injury:
- Do not look into the laser. Do not point laser directly at persons or animals or indirectly off reflective surfaces.
The thermometer is equipped with a laser used for aiming purposes only. The laser turns off when the trigger is released.
To enable or disable the laser:
- Press the Menu softkey until Laser appears as the right softkey function.
- Press the Laser softkey to enable or disable the laser. * appears on the display when the laser is enabled.
Setup
From the Setup menu, the backlight, time and date, and display language can be
changed.
Backlight
In normal use, the backlight is always on. Use this menu to change the
backlight setting from on to off. Turning the backlight off conserves battery
power.
- Press the Menu softkey until Setup appears as the left softkey function.
- Press the Setup softkey.
- Backlight has an indicator next to it. To enter the backlight menu, press the Enter softkey.
- Press the OFF softkey to turn the backlight off, press the ON softkey to turn it on.
- Press the Back softkey to return to the Setup Menu.
Time/Date
To change the time on the thermometer:
- Press the Menu softkey until Setup appears as the left softkey function.
- Press the Setup softkey to enter the Setup menu.
- Press the down arrow softkey to select Time/Date.
- Press the Enter softkey.
- Select the Time softkey.
- Select the desired time format softkey (24hr or 12hr).
- Use the up and down softkeys to change the selected hour to the desired hour.
- Press the Next softkey to select the minutes.
- Use the up and down softkeys to change the selection to the desired minute.
- When in 12 hour mode, press the Next softkey to highlight the am/pm parameter.
- Use the up and down softkey to change to am or pm.
- Press the Done softkey when finished. The display reverts to the first page of the Time/Date menu.
To change the date on the thermometer:
- From the main menu, press the Menu softkey until Setup appears as the left softkey function.
- Press the Setup softkey.
- Press the down arrow softkey to select Time/Date.
- Press the Enter softkey.
- Press the Date softkey.
- Select the date format: Day/Month/Year (dmy) or Month/Day/Year (mdy).
- Use the up and down softkeys to change the selected parameter.
- Press the Next softkey and the arrow softkeys to select the month, date, or year parameters.
- Use the up and down softkeys to change the selected parameter.
- Press the Next softkey to move through each parameter.
- Press the Done softkey when finished. The display reverts to the beginning of the Time/Date menu.
Language
To change the displayed language:
- From the main menu, press the Menu softkey until Setup appears as the left softkey function.
- Press the Setup softkey.
- Use the down arrow softkey to move the indicator to Language.
- Press the Enter softkey.
- Use the down arrow to move the indicator to the desired language.
- Press the Enter softkey to complete the language selection or press the Back softkey to return to the Setup menu.
Deleting Data
To delete stored data from the thermometer, from the main menu, press the menu
softkey until Mem shows as the left softkey function. The last memory location
appears on the display. To access the Delete menu, press the Delete softkey.
From here, either all of the thermometer memory can be deleted or
de3d3dw3de3de33de3de23d23d23d3d3edeceecececececececcececefececececdecdecececececefcecfeindividual
records.
Deleting All Data
To delete all of the records, press the All softkey. At the confirmation
screen, press the Yes softkey.
Deleting Individual Data Records
To delete individual records, press the View softkey and then use the down and
up arrow softkeys to access the desired record. Once the desired record is
shown, press the Yes softkey to delete the record. To cancel data deletion,
pull the trigger.
Buttons and Connector
Table 5 shows the buttons and connector used for the thermometers.
Table 5. Buttons and Connector
Button/ Connector | Description |
---|---|
(561) | Press and then press to toggle between the MIN, MAX, and DIF options. |
(561) | Selects the emissivity setting. You can toggle between LO (0.3), MED |
(0.7), or HI (0.95) using .
| The button is used to show the MIN, MAX, and DIF functions in the secondary
display, whichever was pressed most recently.
****| K-type thermocouple probe used to make contact temperature measurement.
How the Thermometers Work
Infrared thermometers measure the surface temperature of an object. The thermometer’s optics sense emitted, reflected, and transmitted energy, which is collected and focused onto a detector. The unit’s electronics translate the signal into a temperature reading which the unit displays (see Figure 5).
Operating the Thermometer
Temperature Measurement
To measure temperature, point the thermometer at an object and pull the
trigger.
You can use the laser pointer to help aim the thermometer. You may also insert
the K-type thermocouple probe for contact measurement. Be sure to consider
distance-to-spot size ratio and field of view (see “Distance and Spot Size”
and “Field of View”). The temperature appears on the display.
Note
The laser is used for aiming purposes only and is not related to temperature
measurement.
The thermometer features an auto off function that automatically powers down
the thermometer after 20 seconds of inactivity. To turn the thermometer on,
pull the trigger.
Locating a Hot or Cold Spot
To find a hot or cold spot, aim the thermometer outside the desired area.
Then, slowly scan across the area with an up and down motion until you locate
the hot or cold spot (see Figure 6).
Distance and Spot Size
As the distance (D) from the object being measured increases, the spot size
(S) of the area measured by the unit becomes larger. The relationship between
distance and spot size (D:S) for each unit is shown in Figure 7. The spot
sizes indicate 90 % encircled energy.
Field of View
For accurate measurements, make sure that the target is larger than the unit’s
spot size. The smaller the target, the closer you should be to it (see Figure
8).
Emissivity
Emissivity describes the energy-emitting characteristics of materials. Most
organic materials and painted or oxidized surfaces have an emissivity of
approximately 0.95, the default setting for the thermometer.
To compensate for inaccurate readings that may result from measuring shiny
metal surfaces, you can cover the surface to be measured with electrical tape
or flat black paint (<148 °C/300 °F) with emissivity set to 0.95. Allow time
for the tape or paint to reach the same temperature as the surface beneath it.
Measure the temperature of the tape or painted surface.
If you cannot paint or use tape, then you can improve the accuracy of your
measurements by either numerically adjusting emissivity or by using the
Emissivity Menu to access a table of some common materials. The thermometers
have numerically adjustable emissivity from 0.10 to 1.00 that can be used in
conjunction with the emissivities in Tables 3 and 4. Also, the thermometers
have a built-in table of the most common materials that can be set from the
Emissivity Menu. The built-in values are indicated in Tables 3 and 4.
HOLD
The display retains its last infrared measurement for twenty seconds when the
trigger is released, with HOLD appearing on the display. With the probe
inserted, the contact thermometer remains on. To freeze the infrared
temperature when a probe is not inserted, release the trigger until HOLD
appears on the display.
Storing Data
The 566 thermometer can store up to 20 data records. The 568 thermometer can
store up to 99 data records. The following information is stored in each
record:
- Record number
- IR and probe temperature in °F or °C
- Date/Time
- Emissivity
- Max/Min/Avg/Dif Temperature (if enabled)
For more information, see “Save”.
Downloading Data
The stored data of the 568 can be downloaded to a personal computer (PC) via
the included USB cable and the included FlukeView® Forms Documenting Software.
See the FlukeView® Forms documentation for details. The USB input 2 is located
on the top of the thermometer next to the external probe input 1 (see Figure
9).
Note
To prevent incorrect readings, do not perform a temperature measurement of an
earthed conductor while the 568 is connected to a PC that is earthed by a
three-phase grounding plug.
External Contact Probe
Warning
To avoid electrical shock or personal injury, do not connect the external
contact probe to live electrical circuits.
The thermometers come with a bead K-type thermocouple probe. The probes attach
to the thermometers via the probe input located on the top of the thermometer
(see Figure 9).
With the probe installed, the probe symbol ( ) appears on the display. The
probe can be used simultaneously while the thermometer is taking non-contact
measurements. The probe readings are shown below the non-contact measurements.
The thermometer will stay on when there is a probe inserted.
When used with a personal computer and FlukeView Forms (568 only), the auto
off feature is disabled.
Troubleshooting
See Table 6 for remedies to possible problems when using the thermometer.
Table 6. Troubleshooting
Symptom | Cause | Action |
---|---|---|
— (on display) | Target temperature is over or under range. | Select target |
within specifications
| Low batteries| Replace Batteries
Blank display| Thermometer is asleep Possible dead batteries| Pull trigger
Replace batteries
Laser does not work| Low or dead batteries Ambient temperature is above 40 °C
(104 °F)| Replace batteries
Use in area with lower ambient temperature
Failed communication via USB| Unit is off FlukeView Forms is not running| Pull
the trigger
Start FlukeView Forms
Inaccuracy| Possible incorrect emissivity setting, field of view, or spot
size| See “Emissivity”, “Field of View” and “Distance and Spot Size” sections.
Settings such as emissivity, date/time, F/C, and saved data lost| Battery dead
or not replaced in <1 minute of removal| Reset settings. Replace batteries as
soon as low battery indicated; Exchange the batteries within one minute of
removal. Refer to “Changing the Batteries”.
Maintenance
Battery Charge
Use the battery charge symbols to gauge the approximate level of charge left
on the batteries.
Notes
When in low battery mode, the thermometer does not store values.
“Err” appears on the display if attempted.
Batteries are at 5 %. Before taking further readings, the batteries must be
changed.
Changing the Batteries
Warning
Batteries contain hazardous chemicals that can cause burns or explode. If
exposure to chemicals occurs, clean with water and get medical aid. To prevent
injury and for safe operation and maintenance:
- Remove all probes, test leads, and accessories before the battery door is opened.
- Do not disassemble the battery.
- Repair the Product before use if the battery leaks.
- Remove batteries to prevent battery leakage and damage to the Product if it is not used for an extended period.
- Be sure that the battery polarity is correct to prevent battery leakage.
- Do not short the battery terminals together.
- Do not disassemble or crush battery cells and battery packs.
- Do not keep cells or batteries in a container where the terminals can be shorted.
- Do not put battery cells and battery packs near heat or fire.
Do not put in sunlight.
Note
The batteries should be replaced in less than one minute after removal to
avoid manually re-initializing the thermometer clock and date.
To install or change the two AA batteries (see Figure 10):
- Slide the locking tab on the battery door to the “unlock” position and then open the handle.
- Insert the batteries noting their correct polarity.
- Close and lock the handle.
Cleaning the Lens
Blow off loose particles using clean compressed air. Carefully wipe the
surface with a water-moistened cotton swab.
Cleaning the Housing
Use soap and water on a damp sponge or soft cloth.
Caution
To avoid damaging the thermometers, do NOT submerge them in water. Do not use
abrasive cleaners, they will damage the case.
User Replaceable Parts and Accessories
User Replaceable Parts
See Table 7 for a list of replaceable parts.
Table 7. Replaceable Parts
Part | Part Number |
---|---|
56x Manuals CD | 3833037 |
56x Getting Started
(English, French, German, Spanish, Portuguese, Japanese, Simplified Chinese,
Italian, Russian, Polish, Czech)| 3833028
Accessories
Optional accessories for the thermometers are:
- Soft Carrying Case (H6)
- Calibration Certification
- All Type-K Temperature Probes with standard mini-connector. See Table 8 for more information.
Recommended Temperature Probes
See Table 8 for a list of recommended temperature probes.
Table 8. Recommended Temperature Probes
Probe | Usage |
---|---|
80PK- 1 | The general purpose bead probe is an alternative, for |
quick, accurate surface temperatures and air temperatures within ducts, vent
temperatures.
80PK- 8| Pipe clamp probes (2) are essential for tracking
continuously changing temperature differentials on hydronic tubing and pipe
loops, and good for quick, accurate refrigerant temperatures.
80PK- 9| The insulation piercing probe provides a sharp tip to
pierce pipe insulation and flat probe tip for good surface thermal contact,
air temperatures within ducts, and vent temperatures.
80PK- 11| Flexible Cuff Thermocouple Temperature Probe is a
convenient way to attach a thermocouple to a pipe while keeping hands free.
80PK- 25| The piercing probe is the most versatile option. Good for
checking air temperature in ducts, surface temperature under carpets/pads,
liquids, thermometer wells, vent temperatures, and for penetrating pipe
insulation.
80PK- 26| The tapered probe is a good general-purpose gas and
surface probe, with a good length and low mass tip casing for faster reaction
to surface and air temperatures.
Specifications
Feature | 561 | 566 | 568 |
---|---|---|---|
IR Temperature Range | -40 °C to 550 °C | ||
(-40 °F to 1022 °F) | -40 °C to 650 °C | ||
(-40 °F to 1202 °F) | -40 °C to 800 °C |
(-40 °F to 1472 °F)
Accuracy| <0 °C: ±(1.0 °C + 0.1 °/1 °C)
0 °C: ±1 % or ± 1.0 °C, whichever is greater
(<32 F±2 °F ±0.1 °/1 °F)
32 °F: ±1 % or ±2 °F), whichever is greater
Repeatability| ±0.5 % of reading or ±1 °C (2 °F),
whichever is greater.| ±0.5 % of reading or ±0.5 °C (1 °F), whichever is greater.
Display Resolution| 0.1 °C / 0.1 °F
Spectral Response| 8 to 14 mm
Response Time (95 %)| <500 ms
K-Type Thermocouple Input Temperature Range| 0 °C to 100 °C (32 °F to 212 °F)| -270 °C to 1372 °C (-454 °F to 2501 °F)
__**K-Type Thermocouple Input Accuracy| Input accuracy ±2.2 °C (±4 °F)| -270 °C to -40 °C: ±(1 °C + 0.2 °/1 °C) (-454 °F to -40 °F: ±(2 °F + 0.2 °/1 °F))
-40 °C to 1372 °C: ±1 % or 1 °C (-40 °F to 2501 °F: ±1 % or 2 °F), whichever is greater
K-Type Thermocouple Resolution| 0.1 °C (0.1 °F)| 0.1 °C/0.1 °F
Distance:Spot (90 % energy)| 12:1| 30:1| 50:1
Laser sighting| Single laser, output <1 mW Class II, wavelength 630 to 670 nm
Emissivity| Lo, Med, Hi| Digitally adjustable from 0.10 to 1.00 by 0.01 or via built-in table of common materials
Data storage| –| 20 points| 99 points
Communication| none| USB 2.0
Operating Altitude| 3000 meters above mean sea level
Storage Altitude| 12,000 meters above mean sea level
Relative Humidity| 10 % to 90 % RH non-condensing up to 30 °C (86 °F)
Operating Temperature| 0 °C to 50 °C (32 °F to 122 °F)| 0 °C to 50 °C (32 °F to 122 °F)
Storage Temperature| -20 °C to 65 °C (-4 °F to 149 °F)| -20 °C to 60 °C (-4 °F to 149 °F)
Vibration| 2.5 G, IEC 68-2-6
Weight| 0.322 kg (0.7099 lb)
Dimensions| 17.69 cm (6.965 in) H x 16.36 cm (6.441 in) L x 5.18 cm (2.039 in) W
Power| 2 AA /LR6 Batteries (alkaline or NiCD)| 2 AA /LR6 Batteries or USB connection when used with a PC
Battery Life| 12 hours with laser and backlight on; 100 hours with laser and backlight off, at 100 % duty cycle (thermometer continuously on)
CE Certification**| EN/IEC 61326-1:2006, Class B, Criteria A EN/IEC 61010-1:2001
EN/IEC 60825-1:2007
561 Accessory
Feature | K-Type Thermocouple Probe (Wrap Type) |
---|---|
Measurement Range | 0 °C to 100 °C (32 °F to 212 °F) |
Accuracy | ±2.2 °C (4.0 °F) |
Cable Length | 505 mm (20 in) cable terminated with a Type K thermocouple |
inside a 495 mm (19.5 in) nylon cuff
566/568 Accessory
Feature | K-Type Thermocouple Probe (Bead Type) |
---|---|
Measurement Range | -40 °C to 260 °C (-40 °F to 500 °F) |
Accuracy | ±1.1 °C (±2.0 °F) from 0 °C to 260 °C (32 °F to 500 °F). |
Typically
within 1.1 °C (2.0 °F) from -40 °C to 0 °C (-40 °F to 32 °F)
Cable Length| 1 m (40 in) K-type thermocouple cable with standard
miniature thermocouple connector and bead termination