PB-40002-100B Pressure Injectable Arrowg plus ard Owner’s Manual
- August 16, 2024
- Arrow
Table of Contents
PB-40002-100B
Pressure Injectable Arrowgtard Blue Advance® Three-Lumen PICC with 130
cm Hydrophilic-Coated Nitinol Guidewire
Contents:
1: Three Lumen TaperFree® Cathetert}+ with Arrowgtard Blue Advance®
Antimicrobial/Antithrombogenic Protection”: 6 Fr. (2.11 mm OD) x 55 cm
Pressure Injectable and Contamination Guard
1: GlideThru™ Peel-Away Sheath: 6 Fr. x 4″ (10 cm) Radiopaque over 6 Fr.
Dilator
1: Spring-Wire Guide, Nitinol, Marked: .018″ (0.46 mm) dia. x 17-3/4″ (45 cm)
(Straight Soft Tip on One End – Straight Stiff Tip on Other)
1: Guidewire, Marked: .018″ (0.46 mm) dia. x 130 cm Hydrophilic-Coated Nitinol
with Soft Tip Tungsten Coil
1: Introducer Needle: Echogenic 21 Ga. x 2-3/4″ (7 cm) TW
1: Syringe: 10 mL Luer-Lock
3: Dust Cap: Non-Vented
1: Catheter Trimmer
1: Safety Scalpel: #11
1: Patient ID Card
1: Patient Information Booklet
1: Paper Tape Measure
1: Dressing: STATLOCK®2 PICC Plus Catheter Stabilization Device
‘Licensed under US Patent No. 7,329,412 and 11,219,706.
2A registered trademark of C. R. Bard, Inc.
All components are CE 2797 unless otherwise noted.
Rx only
Warning: Read all package insert warnings, precautions, and instructions
prior to use. Failure to do so may result in severe patient injury or death.
www.teleflex.com/IFU California Prop. 65
WARNING: Cancer and Reproductive Harm.
www.P65Warnings.ca.gov
+++MR Safe
Not made with natural rubber latex.
Store below 25°C (77°F). Avoid excessive heat above 30°C (86°F).
Fluid path components are non-pyrogenic.
Contraindications: The Pressure Injectable Arrowgtard Blue Advance
antimicrobial/antithrombogenic catheter is contraindicated:
- for patients with known hypersensitivity to chlorhexidine
- in the presence of device related infection in the intended insertion vessel or catheter pathway
- in the presence of thrombosis in the intended insertion vessel or catheter pathway
Lumen| Priming Volume*
(mL)| Gravity Flow Rate† (mL/hr)| Pump Flow Rate†† (mL/hr)| MAX Pressure
Injection Flow Rate** (mL/sec)
---|---|---|---|---
Pink (18 Ga.)
Blue (19 Ga.)
White (19 Ga.)| 0.70
0.52
0.49| 990
240
270| 5440
1930
2070| 6
not rated
not rated
- Priming volumes are approximate and are done without accessories.
† Flow rate values are approximate and are determined using deionized water at 100 cm head eight.
†† Pump flow rates are determined at maximum pump pressure of 10 psig and represent approximate flow capabilities.
** Pressure injection flow rates are determined at the injector pressure setting of 300 psi maximum using media of 11.8 centipoise viscosity, with 152 cm pressure tubing.
EU Authorized
Representative and Importer:
Teleflex Medical
IDA Business and
Technology Park
Dublin Road, Athlone
Co. Westmeath, Ireland
All components are CE 2797 unless otherwise noted.
Arrow International LLC
Subsidiary of Teleflex Incorporated
3015 Carrington Mill Blvd.
Morrisville, NC 27560 USA
Product of XXX
Packaged in XXX
www.teleflex.com
LBL070055 R01 (2024-05)