Product Usage Instructions
Free Chlorine (FCL) Testing
- Fill the sample bottle (0688) with a water sample and replace the cap.
- Fill a clean tube (0205) to the 5 mL line with a water sample and cap the tube with a blue cap.
- Insert the tube into the ColorQ 2x.
- Press the button to turn the meter on. When “Blank” appears, press the button to blank the meter, and proceed to Free Chlorine (FCL). Remove the tube.
Total Chlorine (TCL) Testing
- Remove the cap from the reacted FCL tube.
- Add 5 drops of DPD 3 to the tube.
- Cap the tube with a pink cap and invert it 3 times to mix.
- Insert the tube into the ColorQ 2x.
- Press the button to read TCL Total Chlorine. Press past Br to pH. Remove the tube.
pH Testing
- Fill a clean tube (0205) to the 5 mL line with a water sample.
- Add 5 drops of pH reagent.
- Cap the tube with a red cap and invert 3 times to mix.
- Insert the tube into the ColorQ 2x and press the button to read pH. Remove the tube.
Alkalinity Testing
- Press the button to go to Alk Total Alkalinity.
- Fill a clean tube (0205) to the 5 mL line with a water sample.
- Add 1 ALK 2X IG Tablet and crush it with a tablet crusher.
- Cap the tube with a green cap and invert 10 times to mix.
- Insert the tube into the ColorQ 2x and press the button to read Alk Total Alkalinity. Remove the tube.
Calcium Hardness Testing
- Press the button to go to Ca Hard Calcium Hardness.
- Fill a clean tube (0205) to the 5 mL line with a water sample.
- Add 1 CA HARD 2X IG tablet and crush it with a tablet crusher.
- Cap the tube with a purple cap and invert it 3 times to mix.
- Insert the tube into the ColorQ 2x and press the button to read Ca Hard Calcium Hardness. Remove the tube.
Cyanuric Acid (CYA) Testing
- Press the button to go to CYA Cyanuric Acid.
- Fill a clean tube (0206) to the line with the water sample.
- Add 1 CYANURIC ACID IG tablet and crush thoroughly with a tablet crusher.
- Cap the tube with a grey cap, invert it 3 times to mix, and allow it to sit for two minutes without mixing.
- Insert the tube into the ColorQ 2x and press the button to read CYA Cyanuric Acid. Remove the tube.
Frequently Asked Questions
Q: Is this product safe for children?
A: This product contains chemicals that may be harmful if misused, so it should not be used by children except under adult supervision.
INTRODUCTION
- Read the instructions before testing.
- Press the button to turn the meter ON and OFF
- Quickly press the button twice to skip a test.
- Cap the tubes securely to prevent leaks.
- Do not interchange color-coded caps.
- Dry the tubes before placing them in the chamber.
- Rinse the tablet crusher after each use. Brush and rinse tubes, caps, and tablet crusher after testing.
- Allow the tubes to dry before returning them to the case.
- The tube specified in the test procedure must be used.
- Use only LaMotte reagents. Cap securely.
- If the unit should auto-off during testing, rinse and fill a tube. Repeat Blank and return to testing.
- Cap tube to eliminate stray light.
- WARNING! This set contains chemicals that may be harmful if misused. Read cautions on individual containers carefully.
- Not to be used by children except under adult supervision.
BLANK CALIBRATION
Blank
- Fill sample bottle (0688) with water sample. Replace cap.
- Fill clean tube (0205) to the 5 mL line with water sample. Cap tube with blue cap.
- Insert tube into the ColorQ 2x.
- Press the button to turn the meter on . When “Blank” appears press the button to “Blank” the meter, and go to Free Chlorine (FCL). Remove tube.
- Reagent is a potential health hazard.
- READ SDS: lamotte.com Emergency information.
- Chem-Tel USA 1-800-255-3924 Int’l, call collect, 813-248-0585
FREE CHLORINE/BROMINE
FCL/Br
- Add 5 drops of DPD 1A and 5 drops of DPD 1B to the same tube.
- Cap tube with blue cap. Invert 3 times to mix. Insert tube into the ColorQ 2x.
- Press the button to read “FCL” Free Chlorine.
- NOTE: To read as Bromine, press the button past “FCL” and “TCL” to “Br”.
- Press the button to go to “TCL” Total Chlorine. Remove tube.
TOTAL CHLORINE
TCL
- Remove the cap from the reacted FCL (Free Chlorine) tube.
- Add 5 drops of DPD 3 to the tube.
- Cap tube with pink cap. Invert 3 times to mix. Insert tube into the ColorQ 2x.
- Press the button to read “TCL” Total Chlorine.
- Press the button past “Br” to “pH”. Remove tube.
pH
- Fill a clean tube (0205) to the 5 mL line with a water sample.
- Add 5 drops of pH.
- Cap tube with the red cap. Invert 3 times to mix. Insert tube into the ColorQ 2x.
- Press the button to read pH. Remove the tube.
TOTAL ALKALINITY
Alk
- Press the button to go to “Alk” Total Alkalinity.
- Fill a clean tube (0205) to the 5 mL line with a water sample.
- Add 1 Tablet. C ALK 2X IG rush with tablet crusher.
- Cap tube with a green cap. Invert 10 times to mix. Insert tube into the ColorQ 2x.
- Press button to read “Alk” Total Alkalinity. Remove tube.
CALCIUM HARDNESS
Ca Hard
- Press the button to go to “Ca Hard” Calcium Hardness.
- Fill a clean tube (0205) to the 5 mL line with a water sample.
- Add 1 CA HARD 2X IG tablet. Crush with a tablet crusher. Cap tube with a purple cap.
- Invert 3 times to mix. Insert tube into the ColorQ 2x.
- Press the button to read “Ca Hard” Calcium Hardness. Remove tube.
CYANURIC ACID
CYA
- Press the button to go to “CYA” Cyanuric Acid.
- Fill a clean tube (0206) to the line with the water sample.
- Add 1 CYANURIC ACID IG tablet. Crush thoroughly with a tablet crusher. Cap tube with grey cap.
- Invert 3 times to mix. Allow the tube to sit for TWO minutes.
- Do not mix. Insert tube into the ColorQ 2x.
- Press the button to read “CYA” Cyanuric Acid. Remove tube.
COPPER
Cop
- Press the button to go to “Cop” Copper.
- Fill a clean tube (0205) to the 5 mL line with a water sample.
- Add 1 COPPER IG tablet. Crush with a tablet crusher. Cap tube with blue cap.
- Invert 3 times to mix. Allow the tube to sit for TWO minutes.
- Invert 1 time to mix. Insert into the ColorQ 2x.
- Press the button to read “Cop” Copper. Remove tube.
IRON
Ir
- Press the button to go to “Ir” Iron.
- Fill a clean tube (0205) to the 5 mL line with a water sample.
- Add 1 IRON IG Tablet. Crush with a tablet crusher. Cap tube with blue cap.
- Invert 3 times to mix. Insert into the ColorQ 2x.
- Press the button to read “Ir” Iron. Remove tube.
BIGUANIDE
big
- Press the button to go to “big” Biguanide.
- Fill a clean tube (0205) to the 5 mL line with a water sample.
- Add 5 drops of BIGUANIDE TEST SOLUTION. Cap tube with blue cap.
- Invert 3 times to mix. Insert tube into the ColorQ 2x.
- Press the button to read “big” Biguanide. Remove tube.
BIGUANIDE SHOCK PEROXIDE
bsh
- Press the button to go to “bsh” Biguanide Shock.
- Fill a clean tube (0205) to the 5 mL line with a water sample.
- Add 1 SHOCK PEROXIDE IG Tablet. Crush with a tablet crusher. Cap tube with blue cap.
- Invert 3 times to mix. Allow the tube to sit for TWO minutes.
- Invert once to mix. Insert into the ColorQ 2x.
- Press the button to read “bsh” Biguanide Shock. Remove tube.
- Press the button to return to a blank meter or hold the button until is displayed.
- Send pool or spa test results from the ColorQ 2x through Bluetooth to the ® WaterLink Solutions App for chemical recommendations and dosage amounts.
- Go to learn more a www.lamotte.com/pool bout the Homeowner and Pool ® Professional WaterLink Solutions water analysis programs.
- 802 Washington Ave Chestertown Maryland 21620 USA
- 800-344-3100 410-778-3100 410-778-6394 lamotte.com