LaMotte AR-02 Code 4783-03 Test Kit Instruction Manual

June 6, 2024
lamotte

LaMotte AR-02 Code 4783-03 Test Kit Instruction Manual
LaMotte AR-02 Code 4783-03 Test Kit

USE OF THE OCTA-SLIDE 2 VIEWER

The Octa-Slide 2 Viewer should be held so non-direct light enters through the back of the Viewer. Slide the Octa-Slide 2 Bar into the Viewer. Insert the reacted sample into the top of the Viewer. Match the color of the reaction to the color standards.

  1. Insert the OctaSlide 2 Bar into the Octa-Slide 2 Viewer (1101)
    Using Instructions

  2. Insert test tube into the Octa-Slide 2 Viewer (1101)
    Using Instructions

  3. Match sample color to a color standard. Record results.
    Using Instructions

COMBINATION WATER OUTFIT

CODE 4783-03
CODE 3590-03

QUANTITY CONTENTS CODE
15 mL *Hardness Reagent #5 *4483-E
50 Hardness Reagent #6 Tablets 4484A-H
60 mL Hardness Reagent #7 4487WT-H
30 mL *Iron Reagent #1 *4450-G
4.5 g *Iron Reagent #2 Powder *4451-S
30 mL *Wide Range Indicator *2218-G
30 mL† *Sulfide Reagent A *4458-G
15 mL† *Sulfide Reagent B *4459-E
60 mL† Sulfide Reagent C 4460-H
1 Test Tube, w/cap 4488
1 Spoon, 0.05 g, plastic 0696
4 (6†) Test Tubes, plastic, 2.5-5-10 mL, w/caps 0106
1† Pipet, 1.0 mL, plastic 0354
1 Iron Octa-Slide 2 Bar, 0.5-10.0 ppm 4448-01
1 Wide Range pH Octa-Slide 2 Bar, 5.0-10.0 ppm 3483-01
1† Sulfide Octa-Slide 2 Bar, 0.2-20 ppm 4457-01
1 Octa Slide 2 Viewer 1101

WARNING: Reagents marked with an are considered to be potential health hazards. To view or print a Safety Data Sheet (SDS) for these reagents go to www.lamotte.com. Search for the four digit reagent code number listed on the reagent label, in the contents list or in the test procedures. Omit any letter that follows or precedes the four digit code number. For example, if the code is 4450WT-H, search 4450. To obtain a printed copy, contact LaMotte by email, phone or fax.

Emergency information for all LaMotte reagents is available from Chem-Tel (US, 1-800-255-3924) (International, call collect, 813-248-0585).

To order individual reagents or test kit components, use the specified code number.

TOTAL HARDNESS TEST PROCEDURE

  1. Fill the test tube (4488) to the desired line with the sample water,
    upper line: 1 drop = 10 ppm CaCO3
    middle line: 1 drop = 1 gpg CaCO3
    lower line: 1 drop = 20 ppm CaCO3
    TEST PROCEDURE

  2. Add 5 drops of *Hardness Reagent #5 (4483).

  3. Swirl to mix.
    TEST PROCEDURE

  4. Add 1 Hardness Reagent #6 Tablet (4484A).
    TEST PROCEDURE

  5. Swirl until tablet has disintegrated. Solution will turn Red if hardness is present. Proceed to next step. If solution is Blue, there is no measurable amount of hardness.
    TEST PROCEDURE

  6. While gently swirling the tube, add Hardness Reagent #7 (4487WT) one drop at a time until the red color changes to blue. Count the number of drops added. Hold bottle vertically.
    TEST PROCEDURE

  7. Multiply the number of drops used in Step 6 as follows:
    upper line: each drop equals 10 ppm Hardness as CaCO3
    middle line: each drop equals 1 gpg Hardness as CaCO3
    lower line: each drop equals 20 ppm Hardness as CaCO

IRON TEST PROCEDURE

  1. Insert the Iron Octa-Slide 2 Bar (4448-01) into the Octa-Slide 2 Viewer (1101).
    IRON TEST

  2. Rinse test tube (0106) with sample water. Fill to 5 mL line.

  3. Add 5 drops of *Iron Reagent #1 (4450).
    IRON TEST

  4. Cap and mix
    IRON TEST

  5. Use the 0.05g spoon (0690) to add one level measure of *Iron Reagent #2 Powder (4451).
    IRON TEST

  6. Cap and shake until powder dissolves.
    IRON TEST

  7. Wait 3 minutes
    IRON TEST

  8. Insert test tube into the OctaSlide 2 Viewer (1101).
    IRON TEST

  9. Match sample color to a standard. Record results as ppm iron.
    IRON TEST

pH TEST PROCEDURE

  1. Insert Wide Range pH OctaSlide 2 Bar (3483-01) into the Octa-Slide 2 Viewer (1101)
    pH TEST

  2. Fill a test tube (0106) to the 10 mL line with sample water.
    pH TEST

  3. Add 8 drops of *Wide Range pH Indicator (2218).
    pH TEST

  4. Cap and mix.
    pH TEST

  5. Insert test tube into the Octa-Slide 2 Viewer (1101).
    pH TEST

  6. Match sample color to a color standard. Record results as pH.

TOTAL SULFIDE TEST PROCEDURE

†Supplied only with Code 3590-03
Note: Collect sample with a minimum of aeration and analyze promptly.

  1. Insert the Sulfide Octa-Slide 2 Bar (4457-01) into the Octa-Slide 2 Viewer (1101)
    TOTAL SULFIDE

  2. Fill a test tube (0106) to the 5 mL line with the sample water.
    TOTAL SULFIDE

  3. Add 15 drops of *Sulfide Reagent A (4458)
    TOTAL SULFIDE

  4. Cap and gently invert to mix.
    TOTAL SULFIDE

  5. Add 3 drops of *Sulfide Reagent B (4459). Cap and mix.
    TOTAL SULFIDE

  6. Wait 1 minute.
    TOTAL SULFIDE

  7. Use the pipet (0354) to add 1.0 mL of Sulfide Reagent C (4460). Cap and mix.

  8. Insert test tube into the OctaSlide 2 Viewer (1101).
    TOTAL SULFIDE

  9. Match sample color to a color standard. Record results as ppm Sulfide.
    TOTAL SULFIDE

EPA accepted procedure

A known concentration of sulfide in the range of 0.2-20.0 ppm S= may be prepared as a check standard by referring to APHA Standard Methods 4500-S2-D, 17th Ed., 1989. Run this test on the check standard. If the result is incorrect, discard the old reagents and order the reagent refill package.

LaMOTTE COMPANY
Helping People Solve Analytical Challenges
802 Washington Ave
Chestertown
Maryland
21620
USA
800-344-3100
410-778-3100 (Outside USA)
Fax 410-778-6394
www.lamotte.com

©2020 LaMotte Company | 63590-03 | 1.21

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals