Lamotte 1705 Waterlink Spin Touch Instructions
- August 6, 2024
- lamotte
Table of Contents
Lamotte 1705 Waterlink Spin Touch Instructions
KEEP THESE INSTRUCTIONS
METER CHECK DISK | Code 1705
Do not attempt to separate the components of
the Meter Check Disk (Code 1705). The Meter
Check Disk consists of a disk with a
permanently attached cover.
Do not fill the Meter Check Disk with
water. Water is not used during the meter
check procedure.
CALIBRATION OPERATIONS The Meter Check
Disk (MCD) is used to perform two calibration operations – Check Calibration and Start Calibration. The Start Calibration procedure should only be performed if the meter fails the Check Calibration procedure.
Check Calibration Meters are calibrated at the time of manufacture. However, it is possible for the calibration settings to be lost due to power anomalies or other circumstances. The Meter Check Disk is used in the Check Calibration procedure to determine whether the alignment of the hub and disk are correct. For some meters, it also evaluates the brightness of the individual LEDs.
- Follow the CLEANING procedure to clean the light chamber and optic
- From the main test page, tap to select
- Tap Calibration.
- Tap Check Calibration.
- Remove the Meter Check Disk (Code 1705) from the foil DO NOT remove the black cover from the disk. Insert the Meter Check Disk (MCD) into the meter and close the lid.
- Tap Start.
- The meter will briefly When it is complete, six channel values will be displayed. Compare the displayed channel values to those printed on the MCD pouch. If the displayed channel values are within the ranges provided on
the Meter Check Disk pouch, the meter is calibrated and performing normally.
If the displayed channel values are not within the ranges provided on the
Meter Check Disk pouch, perform the Start Calibration procedure.
Note: Range specifications are specific to the disk identified by the serial
number on the pouch. The range specifications will vary from disk to disk. The
exact readings from a specific disk may vary from meter to meter.
- Tap to return to the test
- Remove the MCD from the meter and return to the foil pouch for
Start Calibration The Start Calibration procedure calibrates the alignment of the hub and disk
for all meters and sets the brightness of the individual LEDs for compatible
meters. Before performing this calibration procedure, run the Check
Calibration procedure to determine
whether the meter is calibrated and operating normally. The Start Calibration
procedure should only be completed if the meter fails the Check Calibration
procedure.
- Follow the CLEANING procedure to clean the light chamber and optic
- Fromthe main test screen, tap to select
- Tap Calibration.
- Tap Start Calibration.
- Remove the Meter Check Disk (Code 1705) from the foil DO NOT remove the black cover from the disk. Insert the Meter Check Disk. Close the lid.
- Tap Start.
- When the calibration is complete the message “Angle Calibration Successful” will Meters that can also perform an LED calibration will display “LED Calibration Successful”.
- Tap to return to the test
Angle Calibration checks the alignment of the hub and disk. The LED
Calibration sets the brightness of the individual LEDs. The results are
reported as pass or fail. If the measurements pass, the settings will be
saved, and the meteris calibrated. If the analysis fails, contact Support.
Storage Store the disk in a cool, dry place in the foil pouch with the
desiccant. The disk is sensitive to moisture. Exposure to excessive heat or
moisture may cause degradation of the disk. The Meter Check Disk contains
delicate materials. Do not drop, jar, or twist the disk. Do not get the disk
wet. Keep the Meter Check Disk clean and dry.